HEARTVIEW P12/8 ELECTROCARDIOGRAPH RECORDER AND TRANSMITTER
Applicant
Aerotel, Ltd.
Product Code
DXH · Cardiovascular
Decision Date
Jun 16, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Indications for Use
The HeartView P12/8 is a 12/8-lead Transtelephonic electrocardiographic recorder and transmitter which is intended to be used to record and transmit a patient's electrocardiogram (ECG) via the telephone to a diagnostic/emergency center. The Heart View P 12/8 can record either an eight lead or a twelve lead electrocardiogram, and has been designed so that the patient can attach the electrodes him/herself.
Device Story
HeartView P12/8 is a transtelephonic ECG recorder/transmitter. Input: patient ECG signals captured via disposable electrodes and integrated 'button' electrodes. Operation: patient self-attaches electrodes and positions device on chest to record 8-lead or 12-lead ECG; internal memory stores ECG event upon button press; stored data transmitted via telephone to diagnostic/emergency center. Output: transmitted ECG waveform. Context: home-use by patient; remote analysis by healthcare providers at receiving center. Benefit: enables remote cardiac event monitoring without physician presence for electrode placement.
Clinical Evidence
Clinical study compared ECGs acquired by patients using the HeartView P12/8 against ECGs acquired by professionals using the predicate HeartView 12-lead recorder. A cardiologist reviewed the recordings and determined the signal quality was equivalent.
Technological Characteristics
Transtelephonic ECG recorder; 8-lead and 12-lead configurations; utilizes disposable electrodes and integrated case-mounted 'button' electrodes; internal memory for ECG storage; telephone transmission interface; tested for compliance with IEC 60601-1/1990.
Indications for Use
Indicated for patients requiring transtelephonic ECG monitoring. Allows patient self-application of electrodes and self-recording of 8-lead or 12-lead ECGs for transmission to a diagnostic or emergency center.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
K980144
Page A 3
JUN 16 1998
Aerotel Ltd HeartView P 12/8 Event Recorder 510(k) Submission
510(k) Summary
## (1) Submitter Information
Name: Aerotel Ltd.
Address: 5 Hazoref Street 58858 Holon Israel
Telephone Number: 972-3-559-6111
Contact Person: Dr. George Myers (Official Correspondent) Medsys Inc. 377 Rt. 17 S Hasbrouck Heights, NJ 07604 201-727-1703
Date Prepared: Jan. 5, 1998
(2) Name of Device:
Trade Name: HeartView P 12/8 Event Recorder Common Name: Cardiac Event Recorder Classification Name: Transmitters and Receivers, Electrocardiograph, Telephone (74DXH)
(3) Equivalent legally-marketed devices:
1.Aerotel HeartView 12-lead event recorder, K950004
2. Meridian CardioBeeper, K965101
3. Aerotel H 1001, K931020
(4) Description
The HeartView P12/8 is a 12/8-lead Transtelephonic electrocardiographic recorder and transmitter which is intended to be used to record and transmit a patient's electrocardiogram (ECG) via the telephone to a diagnostic/emergency center. It is a modification to the Aerotel HeartView event recorder, K950004. For the original HeartView, the patient had 10 standard disposable electrodes attached to his chest by a physician, and the unit always recorded and over and electrocardiogram. The HeartView P 12/8 can record either an eight lead or a 12 clead electrocardiogram, and has been designed so that the patient can attach the electredes him/le rself.
{1}------------------------------------------------
#### Aerotel HeartView P12/8
#### Page A 4
To record an 8-lead electrocardiogram, three standard disposable electrodes are attached to the patient's chest, and in addition the patient places the recorder in a specific place on his chest and makes recordings using "button" electrodes embedded in the case. To record a 12-lead electrocardiogram, the disposable electrodes are in the same locations but the patient makes three recordings using the embedded "button" electrodes, each with the recorder in a different location on the chest (also explained later). The system is designed so that the patient can attach the electrodes and place the embedded electrodes him/herself with the recorder in the proper locations. Clinical tests have been done which show that the electrocardiograms produced when the patient attaches the electrodes and manipulates the recorder are of the same quality as in the original HeartView, K950004, in which disposable electrodes are attached by the physician.
Heart View P12/8 has internal memory, which enables it to record and store the ECG of a cardiac event when the Record Transmit button is pressed. The stored ECG can then be transmitted to the center by telephone at a later time. Transmitting the ECG does not erase it, so that an inadvertent or faulty transmission does not affect the recording.
The system is designed so that the patient can record an episode on the recorder and then transmit it in a simple, non-confusing way.
#### (5) Intended Use
The HeartView P12/8 is a 12/8-lead Transtelephonic electrocardiographic recorder and transmitter which is intended to be used to record and transmit a patient's electrocardiogram (ECG) via the telephone, to a diagnostic/emergency center. The Heart View P 12/8 can record either an eight lead or a twelve lead electrocardiogram, and has been designed so that the patient can attach the electrodes him/herself.
(b) Performance data
(1) Non-clinical tests
The HeartView P 12/8 has had to undergo safety and emissions tests. The HeartView P/12/8 has been tested for compliance with IEC 60601-1/1990 by Israel Testing Laboratories (ITL).
### (2) Clinical tests
In a clinical test, electrocardiograms taken with the HeartView P 12/8 with the electrodes attached by the patient were compared to electrocardiograms taken with the predicate device, the HeartView 12-lead event recorder, with electrodes professionally attached. The electrodes were given to a cardiologist for comparison, who determined that the quality of the electrocardiograms from the two devices was equivalent.
(3) Conclusions
{2}------------------------------------------------
Aerotel HeartView P12/8
(ille][imalle]dial aroma: 0 + puper = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
The HeartView P 12/8 is equivalent in safety and efficacy to the legally-marketed predicate devices.
ー・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, with three curved lines forming the body and wings.
1
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 6 1998
George H. Myers, Sc.D. Official Correspondent for Aerotel, Ltd. C/O Medsys, Inc. 377 Route 17 South Hasbrouck Heights, NJ 07604
K980144 Re: Aerotel HeartView P12/8 Cardiac Event Recorder Trade Name: Regulatory Class: II (two) Product Code: 74 DHX April 8, 1988 Dated: April 9, 1998 Received:
Dear Dr. Myers:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce.prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - George H. Myers, Sc.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
ੇ ਵਿ
{5}------------------------------------------------
Question 14. Indications for Use
The revised statement follows.
1 of of ______________________________________________________________________________________________________________________________________________________________________ Page _________________________________________________________________________________________________________________________________________________________________________ 1
K980144 510(k) Number (if known):
Device Name: HeartView Cardiac Event Recorder
Indications for Use:
ి శు
The HeartView P12/8 is a 12/8-lead Transtelephonic electrocardiographic recorder and transmitter which is intended to be used to record and transmit a patient's electrocardiogram (ECG) via the telephone to a diagnostic/emergency center. The HeartView P 12/8 can record either an eight lead or a twelve lead electrocardiogram, and has been designed so that the patient can attach the electrodes him/herself.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
| 510(k) Number | K980144 |
|---------------|---------|
|---------------|---------|
| Prescription Use Use (Per 21 CFR 810.109) | OR | Over-the-Counter |
|-------------------------------------------|----|------------------|
|-------------------------------------------|----|------------------|
(Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.