Transonic Systems' BLF22 and BLF22A tissue perfusion monitors are intended for use in the measurement of microvascular perfusion.
Device Story
Transonic BLF22/BLF22A are laser Doppler monitors measuring microvascular perfusion; input via tissue perfusion probes; principle of operation based on laser Doppler flowmetry; device displays flow, mass, or velocity on front panel; BLF22A includes adjustable low-flow audio indicator tone; features analog outputs and USB interface; used in clinical settings; provides real-time perfusion data to clinicians to assist in monitoring tissue health; benefits include objective assessment of microvascular blood flow.
Clinical Evidence
Bench testing only. Studies verified functional requirement specifications, flow value accuracy, and message output. Electrical safety testing confirmed compliance with IEC 60601-1. No animal or clinical testing required.
Technological Characteristics
Laser Doppler flowmeter; metal case; analog outputs (0-10V DC) and USB interface; BLF22A includes low-flow indicator tone circuitry; complies with IEC 60601-1 electrical safety standards.
Indications for Use
Indicated for the measurement of microvascular perfusion in patients requiring such monitoring. Prescription use only.
Regulatory Classification
Identification
A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Transonic Systems Incorporated Ms. Leah Van De Water Senior Regulatory Affairs Manager 34 Dutch Mill Road Ithaca, New York 14850
July 23, 2015
Re: K140740
Trade/Device Name: Transonic Tissue Perfusion Monitors Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular blood flowmeter Regulatory Class: Class II Product Code: DPW Dated: June 8, 2015 Received: June 9, 2015
Dear Mr. Klementowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
For
Sincerelv vours.
Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140740
#### Device Name
Transonic Tissue Perfusion Monitors Model numbers: BLF22 and BLF22A.
Indications for Use (Describe)
Transonic Systems' BLF22 and BLF22A tissue perfusion monitors are intended for use in the measurement of microvascular perfusion.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K140740
## 510(k) SUMMARY
## Summary of Safety & Effectiveness
| Submitter's Name & Address: | Transonic Systems, Inc.<br>34 Dutch Mill Road<br>Ithaca, NY 14850 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person & Telephone: | Leah Van De Water<br>607-257-5300 |
| Date Summary Prepared: | March 24, 2014 (Revised July 17, 2015) |
| Device Name: | Proprietary Name – Transonic Tissue Perfusion Monitor<br>Common/Usual Name – Laser Doppler Meter<br>Classification Name – Cardiovascular Blood Flowmeter<br>(21 CFR 870.2100)<br>Product Code: DPW<br>Regulatory Class: II |
| Predicate Device: | K903633 ALF21 Laser Flow Meter |
### Device Description:
The BLF22 (single-channel) Tissue Perfusion Monitor reports Flow, Mass or Velocity on its front panel digital display. Housed in a small metal case, the Monitor is equipped with a convenient multi-position handle/stand. It also features: rear panel analog outputs for Flow/Mass/Velocity in 0 - 10 Volts DC; and a USB computer interface.
The BLF22A (single-channel + Indicator Tone) Tissue Perfusion Monitor incorporates all the functionality of the BLF22, plus adjustable Low Flow Indicator Tone circuitry. This will sound an audio signal when tissue perfusion flow drops below a preset level.
### Models:
Transonic Systems Laser Doppler BLF22 Series monitors consist of the following model numbers: BLF22 and BLF22A.
### Indications for Use (Prescription Device):
Transonic Systems' BLF22 and BLF22A tissue perfusion monitors are intended for use in the measurement of microvascular perfusion.
### Substantial Equivalence:
The Transonic BLF22 and BLF22A Monitors are similar in materials, form, and operating principle to the ALF21 Laser Flowmeter in K903633. The BLF22 and BLF22A monitors have
{4}------------------------------------------------
# K140740
the same principle of operation and electronics design. The case and circuit has been updated to utilize alarm electronics, clinical power supply, and poser entry module inside the case.
### Performance Testing:
Bench studies demonstrated the functional requirement specifications of the BLF22A including flow values and messages to user.
Electrical safety testing was conducted and the BLF22/ BLF22A comply with IEC 60601-1.
The product is tested and must meet all required release specifications before distribution. Testing is completed for the tissue perfusion probe/monitor combination demonstrating the volume flow is measured within the specified full range and accuracy.
Animal and clinical testing was not required to support substantial equivalence.
The successful testing demonstrates the safety and effectiveness of the BLF22 and BLF22A Tissue Perfusion Monitors when used for the defined indications for use.
### Conclusion:
The sponsor believes that the BLF22 and BLF22A tissue perfusion monitoring systems utilize similar technology as the predicate K903633 device. The sponsor has provided technical performance specifications and testing data. This information demonstrates that the devices do not raise new types of questions regarding safety or effectiveness when compared to the K903633 device and when used for the indications for use provided above. The sponsor thus concludes that the BLF22 and BLF22A devices are substantially equivalent to the predicate K903633 ALF21 device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.