The Elfor L is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is intended to be used for clinical research applications and pre clinical research applications.
Device Story
Elfor-L uses Miniaturized Dynamic Light Scattering (mDLS) to non-invasively measure micro-vascular perfusion. System components include a sensor with a Vertical-Cavity Surface-Emitting Laser (VCSEL) light source (850nm) and two photodetectors, a control unit with a microcontroller, and dedicated software for data processing and recording on a PC. The device is operated by professionals in hospital, laboratory, or intensive care environments. The sensor captures analog signals from skin/muscle tissue, which are processed by the control unit and transmitted to a computer. The output provides perfusion measurements (0-2000 PU) to clinicians for research purposes, aiding in the assessment of micro-vascular blood flow.
Clinical Evidence
Bench testing only. Performance evaluation included comparison testing against the predicate device, accuracy testing using a standardized Motility Standard solution, and software validation. Biocompatibility testing (cytotoxicity, sensitization, intracutaneous irritation) per ISO 10993-1 confirmed safety. Electrical safety and EMC testing per IEC 60601-1, 60601-1-2, 60601-4-2, and 60825-1 were successfully completed.
Technological Characteristics
mDLS technology; VCSEL light source (850nm); two photodetectors; USB-powered (5V, 0.5A); signal bandwidth 5Hz-24KHz; perfusion range 0-2000 PU. Complies with IEC 60601-1 (Electrical Safety), IEC 60601-1-2 (EMC), and IEC 60825-1 (Laser Class I).
Indications for Use
Indicated for measuring micro-vascular perfusion in skin and muscle in humans of all ages for clinical and pre-clinical research applications. Intended for use on intact skin.
Regulatory Classification
Identification
A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 14, 2025
Elfi-Tech Ltd.
Orly Maor
Regulatory consultant
150 Derech Menachem Begin
Tel Aviv, 6492105
Israel
Re: K243852
Trade/Device Name: Elfor-L
Regulation Number: 21 CFR 870.2100
Regulation Name: Cardiovascular Blood Flowmeter
Regulatory Class: Class II
Product Code: DPW
Dated: December 12, 2024
Received: December 16, 2024
Dear Orly Maor:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K243852 - Orly Maor
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K243852 - Orly Maor
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.04.14
17:03:18 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243852 | |
| Device Name Elfor-L | |
| Indications for Use (Describe) The Elfor L is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is intended to be used for clinical research applications and pre clinical research applications. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
Traditional Premarket Notification Submission – 510(k)
Elfor-L
510(k) Number __________
Date Prepared: December 12, 2024
I. SUBMITTER
ELFI-TECH Ltd.
150 Derech Menachem Begin Tel-Aviv
Zip code: 6492105, Israel
Tel: +972-52-336-2625
E-mail: erez.herman@elfi-tech.com
Regulatory Correspondent:
Orly Maor
25A Sirkin Street
Kfar Saba 4442157, Israel
Tel: +972-9-7453607
Fax: +972-153-9-7453607
oram.ma@gmail.com
II. DEVICE
Name of Device: Elfor-L
Common or Usual Name: Elfor-L
Classification Name: 21CFR 870.2100- Cardiovascular blood flowmeter
Regulatory Class: II
Product Code: DPW
III. PREDICATE DEVICE
ELFI-TECH Ltd. believes that Elfor-L is substantially equivalent to the following predicate device:
- Perimed Inc. PeriFlux 5000 cleared under K974285, Product code DPW, Regulation number 21 CFR 870.2100 and in subsequent submission under K152930, PeriFlux 6000, Product code DPW, Regulation number 21 CFR 870.2100.
IV. DEVICE DESCRIPTION
The Elfor-L device is making use of Miniaturized Dynamic Light Scattering (mDLS) technology to non-invasively measure micro-vascular perfusion in the skin. The device is comprised of sensor and signal processing station. The device uses Vertical-Cavity Surface-Emitting Lasers (VCSEL) as the light source and two photodetectors for detection.
The Elfor-L design incorporates three major components.
Page 1 of 4
{6}
1. Sensor- the sensor as its predicate includes a laser light source and two photo sensitive elements. The sensor transfers analog signal to the control unit.
2. Control unit - the control unit is connected to the sensor by cable and includes the control circuit for the sensor, microcontroller and communication interface for the PC.
3. Dedicated software that processes, displays and record data on the computer. The computer is not part of the system.
# V. INDICATIONS FOR USE
The Elfor-L is intended for measuring microvascular perfusion in skin and muscle in humans. It is intended to be used for clinical research applications and pre-clinical research applications.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Elfor-L has the same intended use as the predicate device. Its indications for use are identical to that of the predicate device.
The Elfor-L has similar technological characteristics as the predicate device as demonstrated in the table below:
| | ELFOR-L | PERIFLUX 5000 | PERIFLUX 6000 | SE JUSTIFICATION |
| --- | --- | --- | --- | --- |
| 510(k) Number | TBD | K974285 | K152930 | — |
| Manufacturer | ELFI-TECH | Perimed AB | Perimed AB | — |
| Product Code | DPW | DPW | DPW | Same |
| CFR | 870.2100 | 870.2100 | 870.2100 | Same |
| Intended Use | The Elfor L is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is intended to be used for clinical research applications and pre-clinical research applications. | The PF 5010 LDPM Unit is indicated for use in measuring micro-vascular perfusion in skin and muscle in humans. It is also intended for use in measuring micro-vascular perfusion in all tissues in animals. | PeriFlux 6000, equipped with PF 6010 is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is also intended for measuring micro-vascular perfusion in all tissues in animals for research purposes. | Same |
{7}
| | ELFOR-L | PERIFLUX 5000 | PERIFLUX 6000 | SE JUSTIFICATION |
| --- | --- | --- | --- | --- |
| Intended User | The Elfor-L is intended for professional use in a hospital environment, laboratory environment and in a hospital intensive care unit environment. | The PF 5010 LDPM Unit is intended for professional use in a hospital environment, laboratory environment and in a hospital intensive care unit environment. | The PeriFlux 6000 instrument is intended for professional use in a hospital environment, laboratory environment and in a hospital intensive care unit environment. | Same |
| Patient Population | The Elfor-L can be used for measurements on patients of all ages. | The PF 5010 LDPM Unit can be used for measurements on patients of all ages. | The PF 6010 LDPM/Temp Unit can be used for measurements on patients of all ages. | Same |
| Body contact | The Elfor-L is intended for measurements on intact skin. | The PeriFlux 5000 is intended for measurements on intact skin. | The PeriFlux 6000 is intended for measurements on intact skin. | Same |
| Maximum laser current | 3-5mA | Not specified. | 5mA | Within the predicate range |
| Laser product classification | Class I laser product per IEC 60825-1. | Class I laser product per IEC 60825-1. | Class I laser product per IEC 60825-1. | Same |
| Laser Type | Invisible, near-infrared, 850nm | Invisible, near-infrared, 780 nm | Invisible, near-infrared, 785nm | Similar |
| Perfusion measurement | 0-2000 PU | Upper limit not specified. | 0-2000 PU | Same as PeriFlux 6000 |
| Perfusion Accuracy | 250±10 | Motility standard: 250±20 PU Zeroing disc: 0.0±1.0 PU | ± 2 PU at zero, ±4 % in motility standard | Same |
| Signal Bandwidth | 5Hz -24KHz | 20 Hz to 15 kHz | 5 Hz - 25 kHz | Substantially equivalent |
| Calibration | Required | Required. | Required | Same |
| List of main components | 1. VCSEL laser source2. Two photodetectors3. Amplifier4. Current driver5. A2D convertor6. Microprocessor Adhesive Tape and Disposable patches | 1. Laser source2. Fiber optic probe3. Photodetector4. Amplifier5. A2D convertor6. Microprocessor Adhesive Tape and Probe holder | 1. Laser source2. Fiber optic probe3. Photodetector4. Amplifier5. A2D convertor6. Microprocessor Adhesive Tape and Probe holder | Similar |
{8}
| | ELFOR-L | PERIFLUX 5000 | PERIFLUX 6000 | SE JUSTIFICATION |
| --- | --- | --- | --- | --- |
| Power source | The Elfor-L receives its power from a USB port (5V; 0.5A) | 100–240 Volts, 50 or 60 Hz. | 100–240 Volts, 50 or 60 Hz. | Different however, no safety issue |
| Standards | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | Same |
| | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | |
| | IEC 60825-1 | IEC 60825-1 | IEC 60825-1 | |
VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
- Biocompatibility
Biocompatibility evaluation in compliance with ISO 10993-1 was performed.
The following tests were conducted:
Cytotoxicity Study Using the ISO Elution Method
ISO Guinea Pig Maximization Sensitization Test
ISO Intracutaneous -irritation Study in Rabbits
The device was found biocompatible.
- Performance Testing
Performance testing included comparison testing of the Elfor-L to its predicate device. The main purpose of this test was to verify the Elfor-L performance is similar to that of its predicate.
The test passed and met the predefined acceptance criteria.
Accuracy of the Elfor-L device- the purpose of this test was to evaluate the accuracy of the Elfor-L by repeated measurements using a standardized Motility Standard solution. The test passed.
- Software Validation
Software verification and validation testing of the Elfor-L were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions”.
- Electrical Safely and EMC
Electrical Safety per IEC 60601-1, Electromagnetic compatibility (EMC) per IEC 60601-1-2, IEC 60601-4-2 and IEC 60825-1 were conducted on the Elfor-L. The tests passed.
VIII. CONCLUSION
The Elfor-L was determined to be substantially equivalent to the predicate device.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.