PC ECG 1200 ECG AND STRESS ELECTROCARDIOGRAPHY TESTING SYSTEM(VERSION 2.90)

K960449 · Biosensor Corp. · DPS · May 30, 1996 · Cardiovascular

Device Facts

Record IDK960449
Device NamePC ECG 1200 ECG AND STRESS ELECTROCARDIOGRAPHY TESTING SYSTEM(VERSION 2.90)
ApplicantBiosensor Corp.
Product CodeDPS · Cardiovascular
Decision DateMay 30, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2340
Device ClassClass 2

Device Story

PC ECG 1200 ECG and Stress Electrocardiography System captures ECG signals via electrodes; processes data using digital/analog hardware and firmware to measure heart rate, ST levels, and ECG parameters. System employs averaging of multiple beats to mitigate noise from myopotential, baseline wander, and line interference. Features include 50/60 Hz notch filters, low-pass filters for EMG interference, and high-pass filters for baseline wander, with on-screen user notifications when filters are active. Output provided to physicians for review of trends and diagnostic measurements. Used in clinical settings to assist in cardiac assessment and stress testing.

Clinical Evidence

No clinical data provided. Document references adherence to ANSI/AAMI/ES-1, IEC 601-1/2/3, and ANSI/AAMI EC11-1982 standards for safety and performance testing.

Technological Characteristics

System comprises digital and analog hardware, firmware, and software. Includes 50/60 Hz notch filters, low-pass filters for EMG, and high-pass filters for baseline wander. Complies with ANSI/AAMI/ES-1, IEC 601-1, IEC 601-2, IEC 601-3, and ANSI/AAMI EC11-1982 standards.

Indications for Use

Indicated for patients requiring ECG and stress electrocardiography monitoring to measure heart rate, ST values, and other ECG parameters.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K960449 MAY 30 1996 # Section 2 - Safety and Effectiveness Information ## 2.1 Summary of Safety and Effectiveness Information ### Safety Stress and ECG systems can be sources for macroshock and microshock hazards for the patient and the user. Details and protection requirements are explained in 'Medical Instrumentation: Theory and Application:, J.G. Webster, Ch. 13, 1978. These protection requirements are compiled into National and International Safety Standards. The most comprehensive ones are ANSI/AAMI/ES-1 and IEC 601-1 and 2. For safe use of Stress and ECG systems, labeling and documentation must be complete. Standards for these are also covered in the above mentioned standards and also in ANSI/AAMI EC11-1982. ### Effectiveness Stress and ECG systems contain digital and analog hardware, associated firmware and software, which must perform effectively, accurately, and reliably. Standards and test methods for the performance of these systems are compiled in detail in ANSI/AAMI EC11-1982 and IEC 601-2 and 3. In addition to these, because of lack of proper user training, problems may arise regarding the following: (a) Lead placement (Correct positions for electrodes are explained in ECG textbooks such as "Harrison's Principles of Internal Medicine", Eds. R.G. Petersdorf, R.D. Adams, E. Braunwald, K.J. Isselbacher, J.B. Martin and J.D. Wilson, 10th edition, pp: 1320-21. McGraw Hill, 1983). (b) Line Interference (A comprehensive reference list for causes and reduction methods for line interference is given in "A new technique for line interference monitoring and reduction in biopotential amplifiers", Y.Z. Ider and H. Koymen, IEEE Trans. Biomedical Engineering, Vol. 37, pp. 624-31, 1990.) 50/60 Hz Notch filters can reduce partially or substantially an problems with line interference. (c) EMG (myopotential) interference (A low pass filter may be provided as an option as too much muscle interference may be encountered during an ECG or stress recording. Users should be notified continuously when this filter is in use, since low pass filters may affect the diagnostic value of the ECG recording information. The PC ECG 1200 ECG and Stress Electrocardiography System provides such notification on screen.) (d) Baseline wander (High pass correction filters may be provided as an option as baseline wander may occur during ECG or stress recording. Users should be notified continuously when this filter is used, since baseline wander filters may affect the diagnostic value of the recording. The PC ECG 1200 ECG and Stress Electrocardiography System provides such notification on screen.) (e) Averaging (In Stress and ECG systems, ST level changes must be accurately measured for each lead and at each exercise step. Due to myopotential, baseline wander and other noise interference, such measurements cannot be reliably made using a single beat. Therefore, {1} beats are averaged and measurements are made from average beats to minimize errors. These average beats must be recorded for every exercise step level during stress testing.) (f) Diagnostic Accuracy (In Stress and ECG systems that measure heart rate, ST values and other ECG parameters, the algorithms used to make such measurements and report trends to the physician influence the quality of information provided to the physician. Testing of the diagnostic accuracy of the ST measurements, QRS detector and VE detector are necessary to evaluate the value of the measurements and trends supplied to the physician for review.) ## 2.2 Certification: I certify that we have conducted a reasonable search of all information known or otherwise available to me about the types and causes of safety and/or effectiveness problems that have been reported for the PC ECG 1200 ECG and Stress Electrocardiography System. I further certify that I am aware of the types of problems to which the PC ECG 1200 ECG and Stress Electrocardiography System is susceptible and that the above summary of the types and causes of safety and/or effectiveness problems about the PC ECG 1200 ECG and Stress Electrocardiography System is complete and accurate ![img-0.jpeg](img-0.jpeg) Steve Springrose President Biosensor Corporation January 23, 1996
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...