The iSE series electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. The electrocardiograph is capable of network communications and supports the informatized management of workflows in hospital and healthcare facilities.
Device Story
The iSE series (iSE-1210, iSE-1810) are electrocardiographs acquiring ECG signals via body surface electrodes (up to 12 or 18 leads). Used in hospitals/clinics by physicians and trained professionals; device performs signal processing to generate measurements, graphical presentations, and interpretive statements. Output is provided to clinicians on an advisory basis to assist in heart disease diagnosis. Device supports network communications (WiFi, Bluetooth, 4G, NFC) for workflow management. Benefits include automated ECG analysis and streamlined clinical data management.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, electrical safety (ANSI AAMI ES 60601-1), electromagnetic compatibility (IEC 60601-1-2), performance testing (IEC 60601-2-25), human factors/usability engineering (IEC 62366-1), and software verification and validation.
Technological Characteristics
Electrocardiograph supporting 12-lead (iSE-1210) or 18-lead (iSE-1810) acquisition. Features 24-bit A/D conversion, 64,000 Hz sampling frequency, and 0.01-350 Hz frequency response. Includes thermal dot-matrix recorder, lithium battery, and wireless connectivity (WiFi, Bluetooth, 4G, NFC). Software utilizes SEMIP V1.92 algorithm. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-25.
Indications for Use
Indicated for adult and pediatric patients requiring ECG signal acquisition via body surface electrodes for the analysis and diagnosis of heart disease in hospital or healthcare facility settings.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 7, 2022
Edan Instruments, Inc. Ying Dai Regulatory Engineer #15 Jinhui Road, Jinsha Community, Kengzi Sub-District Pingshan District Shenzhen, Guangdong 518122 China
Re: K212278
Trade/Device Name: Electrocardiograph Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: February 28, 2022 Received: March 7, 2022
Dear Ying Dai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Electrocardiograph
Indications for Use (Describe)
The iSE series electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. The electrocardiograph is capable of network communications and supports the informatized management of workflows in hospital and healthcare facilities.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc.<br>#15 Jinhui Road, Jinsha Community, Kengzi Sub-District, Pingshan District,<br>Shenzhen, 518122 P.R.China.<br>Tel: +86(0755) 84513592 |
|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person:<br>Preparing date: | Joan. Wei<br>June 25th, 2021 |
| 2. Device name and<br>classification: | Electrocardiograph<br>Classification Name/ Product code: |
| | 21 CFR 870.2340 Electrocardiograph, DPS<br>Regulatory Class: Class II |
| 3.Premarket Notification<br>Class III Certification<br>and Summary | Not applicable, the subject device is Class II. |
| 4. Predicate Device(s): | 1) Edan Instruments, SE-18 Electrocardiograph, cleared under K170995<br>(Primary)<br>2) Edan Instruments, SE-12, SE-12 Express, SE-1200, and SE-1200 Express<br>Electrocardiograph, cleared under K171942(Reference) |
| 5. Reason for<br>Submission | By submission of the Traditional 510(k), Edan Instruments is requesting<br>clearance for new devices iSE-1210 and iSE-1810 Electrocardiograph. |
| 6.Pre-Submission,<br>IDE | Not applicable, there is no pre-submission. |
| 7. Device Description: | iSE-1210 and iSE-1810 Electrocardiograph can support ECG Data Sampling of<br>maximum 12 leads and maximum 18 leads separately. iSE series<br>electrocardiograph uses algorithm to generate measurements, data presentations,<br>graphical presentations and interpretative statements. The electrocardiograph is<br>only intended to be used in hospitals or healthcare facilities by doctors and<br>trained healthcare professionals. |
{4}------------------------------------------------
- 8. Indication for Use The iSE series electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. The electrocardiograph is capable of network communications and supports the informatized management of workflows in hospital and healthcare facilities.
### 9. Predicate Device Comparison
Comparison to the predicate and reference devices, the subject device has the same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following tables:
| Item | Predicate device<br>(SE-18) | Proposed device<br>(iSE Series) | Comparison<br>Result | |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-----------|
| K# | K170995 | K212278 | — | |
| Indications for Use/<br>Intended Use | The SE-18 18-lead electrocardiograph is<br>intended to acquire ECG signals from<br>adult and pediatric patients through body<br>surface ECG electrodes. The<br>electrocardiograph is only intended to be<br>used in hospitals or healthcare facilities<br>by doctors and trained healthcare<br>professionals. The cardiogram recorded<br>by the electrocardiograph can help users<br>to analyze and diagnose heart disease.<br>However, the interpreted ECG with<br>measurements and interpretive statements<br>is offered to clinicians on an advisory<br>basis only. | The iSE Series electrocardiograph is<br>intended to acquire ECG signals from<br>adult and pediatric patients through<br>body surface ECG electrodes. The<br>electrocardiograph is only intended to<br>be used in hospitals or healthcare<br>facilities by doctors and trained<br>healthcare professionals. The<br>cardiogram recorded by the<br>electrocardiograph can help users to<br>analyze and diagnose heart disease.<br>However, the interpreted ECG with<br>measurements and interpretive<br>statements is offered to clinicians on an<br>advisory basis only. The<br>electrocardiograph is capable of<br>network communications and supports<br>the informatized management of | Different | |
| | | workflows in hospital and healthcare<br>facilities. | | |
| Algorithm | SEMIP V1.8 | SEMIP V1.92 | Different | |
| The number of<br>electrodes | 16 | iSE-1810: 16<br>iSE-1210: 10 | Different | |
| Power Supply Specifications | | | | |
| Mains Supply: | Operating Voltage = 100V-240V~<br>Operating Frequency = 50Hz/60Hz<br>Input Current = 0.9A ~ 0.4A | Operating Voltage = 100V-240V~<br>Operating frequency=50Hz/60Hz<br>Input current=1.1 A | Same<br>Same<br>Different | |
| Lithium Battery Pack: | Rated voltage = 14.8 V<br>Rated capacity = 5000mAh<br>-- | Rated voltage = 15.2 V<br>Rated capacity = 3550mAh<br>When the battery is fully charged, iSE can work (without printing) continuously at least 8 hours.<br>100% recharge time: ≤ 5 hours<br>90% recharge time: < 4 hours | Different | |
| Performance Specifications | | | | |
| Recording (Optional for iSE with thermal printer) | | | | |
| Recorder: | Thermal dot-matrix recorder | Thermal dot-matrix recorder | Same | |
| HR Recognition | HR Range | 30 BPM ~ 300 BPM | 30 BPM ~ 300 BPM | Same |
| ECG Unit | Leads: | iSE-1810: 9, 12, 15, 16 or 18 standard leads<br>iSE-1210: 9 or 12 standard leads | Different | |
| | A/D | 24 bits | 24 bits | Same |
| | Sampling Frequency | 16, 000 Hz | 64,000 Hz | Different |
| | Frequency<br>Response: | 0.01Hz~300Hz (-3 dB) | 0.01 Hz ~ 350 Hz (-3 dB) | Different |
| Filter | AC Filter | AC Filter | Same | |
| | DFT Filter | DFT Filter | Same | |
| | EMG Filter | EMG Filter | Same | |
| | LOWPASS Filter:<br>300Hz/270Hz/150Hz/100Hz/75Hz | LOWPASS Filter:<br>350Hz/300Hz/270Hz/150Hz/100Hz/75 Hz | Different | |
| Pacemaker Detection | | | | |
| | Sampling Frequency | iSE-1810(DE18): 16,000Hz, Rhythm Lead<br>iSE-1210(DE12): 80,000Hz, Rhythm Lead | Different | |
| Connection | | | | |
| Wireless connection | WiFi | WiFi, Bluetooth, 4G, NFC | Different | |
| Item | Reference device<br>(SE-12, SE-12 Express, SE-1200, and<br>SE-1200 Express) | Proposed device<br>(iSE Series) | Comparison<br>Result | |
| K# | K171942 | K212278 | — | |
| Indications for Use/<br>Intended Use | The intended use of SE-12 series<br>electrocardiograph is to acquire ECG<br>signals from adult and pediatric patients<br>through body surface ECG electrodes.<br>The electrocardiograph is only intended to<br>be used in hospitals or healthcare facilities<br>by doctors and trained healthcare<br>professionals. The cardiogram recorded<br>by the electrocardiograph can help users<br>to analyze and diagnose heart disease.<br>However, the interpreted ECG with<br>measurements and interpretive statements<br>is offered to clinicians on an advisory<br>basis only. | The iSE Series electrocardiograph is<br>intended to acquire ECG signals from<br>adult and pediatric patients through<br>body surface ECG electrodes. The<br>electrocardiograph is only intended to<br>be used in hospitals or healthcare<br>facilities by doctors and trained<br>healthcare professionals. The<br>cardiogram recorded by the<br>electrocardiograph can help users to<br>analyze and diagnose heart disease.<br>However, the interpreted ECG with<br>measurements and interpretive<br>statements is offered to clinicians on an<br>advisory basis only. The<br>electrocardiograph is capable of<br>network communications and supports<br>the informatized management of<br>workflows in hospital and healthcare<br>facilities. | Different | |
| The number of<br>electrodes | 10 | iSE-1210: 10 | Same | |
| Leads: | Standard 12 leads | iSE-1210: 9 or 12 standard leads | Different | |
### Table 1: Comparison between iSE Series and SE-18
{5}------------------------------------------------
# Table 2: Comparison between iSE Series and SE-12 Series
{6}------------------------------------------------
As seen in the comparison tables, the subject and predicate devices have similar intended use, design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness.
## 10. Performance Data:
### Non-clinical data:
### Electrical safety and electromagnetic compatibility (EMC)
The iSE series electrocardiograph were assessed for conformity with the relevant requirements of the following standards:
- ANSI AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R) 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014 (Fourth Edition) Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests.
{7}------------------------------------------------
### Performance testing-Bench
Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.
- IEC 60601-2-25 Edition 2.0 2011-10 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs
### Human Factors Engineering Testing
Edan has conducted usability testing to validate the usability of the devices. The results of usability testing show that the subject device safe and effective for the intended users, uses, and use environments and meet the requirement of following FDA s are Guidance and consensus standard:
- Applying Human Factors and Usability Engineering to Medical Devices, issued on February 3, 2016
- IEC 62366-1:2020 Medical Devices-Part 1:Application of usability engineering to medical devices
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
#### Clinical data: Not applicable.
#### Summary
The non-clinical performance testing showed that the subject devices are as safe and as effective as predicate devices.
#### 11. Conclusion
The non-clinical data, bench testing data, and software verification testing demonstrate that iSE series electrocardiograph are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.