K250882 · Huxley Medical · MNR · Oct 29, 2025 · Anesthesiology
Device Facts
Record ID
K250882
Device Name
SANSA HSAT
Applicant
Huxley Medical
Product Code
MNR · Anesthesiology
Decision Date
Oct 29, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
AI/ML
Intended Use
The Huxley Home Sleep Apnea Test (SANSA) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing and cardiac disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only. The SANSA device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.
Device Story
Wearable chest patch; collects PPG, single-lead ECG, and 3-axis accelerometer data. Records up to 10 hours of physiological signals (oximetry, heart rate, chest movement, snoring, body position, respiratory effort, actigraphy, sleep staging, ECG). Data uploaded to cloud-based portal via cellular (LTE-M) or USB. Cloud algorithm uses signal processing and AI/ML to compute time-series data and summary metrics (sAHI, TST). Healthcare professionals review tracings and perform manual event editing. No automated ECG analysis; no pacemaker analysis. Intended to aid diagnosis of sleep-related breathing and cardiac disorders; provides clinicians with objective data to support clinical decision-making.
Clinical Evidence
Clinical validation performed via comparison to a simultaneously collected reference standard Holter monitor. Qualitative and quantitative analysis of the ECG signal demonstrated clinically acceptable performance. Additional performance metrics for OSA diagnosis (AHI≥15) include 88.2% sensitivity and 87.3% specificity.
Technological Characteristics
Wearable chest patch; sensors: reflective PPG, single-lead ECG (modified Lead II), 3-axis accelerometer. ECG sampling: 250 Hz, 15.5-bit resolution. Frequency response: 0.67-40 Hz. Input impedance: >10 MΩ. Connectivity: Cellular (LTE-M) or USB. Power: Rechargeable Lithium Polymer battery. Software: Cloud-based portal for manual annotation. Standards: IEC 60601-2-47, ANSI AAMI EC12.
Indications for Use
Indicated for adults (22+ years) suspected of sleep apnea to aid in evaluation of sleep-related breathing and cardiac disorders. Prescription use only in clinical or home settings.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 29, 2025
Huxley Medical
% Grace Powers
Founder / Principal Consultant
Powers Regulatory Consulting
2451 Cumberland Blvd SE
Suite 3740
Atlanta, Georgia 30339
Re: K250882
Trade/Device Name: SANSA HSAT
Regulation Number: 21 CFR 868.2375
Regulation Name: Breathing Frequency Monitor
Regulatory Class: Class II
Product Code: MNR, BZQ, MWJ
Dated: September 26, 2025
Received: September 30, 2025
Dear Grace Powers:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250882 - Grace Powers
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250882 - Grace Powers
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
SANSA HSAT
Page 8 of 61
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250882 | ? |
| Please provide the device trade name(s). | | ? |
| SANSA HSAT | | |
| Please provide your Indications for Use below. | | ? |
| The Huxley Home Sleep Apnea Test (SANSA) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing and cardiac disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only.
The SANSA device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
K250882
Traditional
510(k) - Huxley Medical
SANSA HSAT
# HUXLEY™
# MEDICAL
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Huxley Medical Traditional 510(k) premarket notification.
Sponsor: Huxley Medical, Inc.
1465 Northside Dr NW Ste 217
Atlanta, GA, 30318
info@huxleymed.com
Submission Contact: Grace Powers, MS, MBA, RAC
Founder/Principal Consultant
Powers Regulatory Consulting
grace@powersregulatory.com
404-931-8730
Submission Date: March 24, 2025
Subject Device: Trade Name: SANSA™ HSAT
Common Name: Ventilatory Effort Recorder
Classification Name: Breathing Frequency
Monitor Regulation: 21 CFR §868.2375
Regulatory Classification: Class 2
Product Code: MNR, BZQ, MWJ
Predicate Device: iRhythm Zio Monitor (K202359)
Reference Device: Huxley Medical SANSA HSAT (K244027)
# Device Description
The Huxley Home Sleep Apnea Test (SANSA™) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only. The SANSA™ device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA™ does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.
The SANSA HSAT collects multiple physiological signals using a single wearable patch worn on the chest. The SANSA device contains a reflective PPG sensor, a single-lead ECG sensor, and a 3-axis accelerometer. The signals from these sensors are passed into a cloud-based algorithm which utilizes a combination of signal processing and AI/ML components to compute time-series data for clinician review and summary metrics for report output. The device outputs the following time-series channels: Oximetry, Heart Rate,
{5}
510(k) Summary
K250882
Traditional
510(k) - Huxley Medical
SANSA HSAT
Chest Movement, Snoring, Body Position, Respiratory Effort, Actigraphy, Sleep staging (Sleep/Wake), and ECG. The following summary metrics are calculated: sansa-Apnea Hypopnea Index (sAHI) and Total Sleep Time (TST).
Recorded data are uploaded to a software portal where physiological tracings are made available for review and event editing by a qualified healthcare professional. The device is intended to be worn for 10 hours per study.
# Intended Use/Indications for Use
The Huxley Home Sleep Apnea Test (SANSA™) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing and cardiac disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only.
The SANSA device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.
An overview comparison of the SANSA HSAT (subject device) to the predicate device and reference device are presented in the table below.
Table 1: Device Comparison
| Device Comparison | Subject Device: SANSA HSAT | Predicate Device: Zio Monitor (K202359) | Reference Device: SANSA HSAT (K244027) | Comparison |
| --- | --- | --- | --- | --- |
| Manufacturer | Huxley Medical, Inc. | iRhythm Technologies, Inc | Huxley Medical, Inc. | Identical to reference device. |
| FDA Product Codes | MNR, BZQ, MWJ | DSH, MWJ | MNR, BZQ | Identical to reference device with the addition of MWJ from the predicate device. |
| Primary Regulation | 868.2375 | 870.2800 | 868.2375 | Identical to reference device. |
| Classification | II | II | II | Identical to reference device. |
| Classification Name | Ventilatory Effort Recorder | Medical magnetic tape recorder | Ventilatory Effort Recorder | Identical to reference device. |
| Advisory Committee | Anesthesiology | Cardiovascular | Anesthesiology | Identical to reference device. |
| Prescription Use? | Yes | Yes | Yes | Identical |
| Indications for Use | The Huxley Home Sleep Apnea Test (SANSA™) is a | The Zio Monitor is a prescription-only, single-patient-use, | The Huxley Home Sleep Apnea Test (SANSA™) is a | Equivalent - The addition of ECG monitoring |
{6}
510(k) Summary K250882
Traditional 510(k) - Huxley Medical SANSA HSAT
| Device Comparison | Subject Device: SANSA HSAT | Predicate Device: Zio Monitor (K202359) | Reference Device: SANSA HSAT (K244027) | Comparison |
| --- | --- | --- | --- | --- |
| | wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only.
The SANSA™ device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA™ does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis. | ECG monitor that continuously records data for up to 14 days. It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, lightheadedness, pre-syncope, syncope, fatigue or anxiety. | wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). | capability for manual annotation and analysis of the ECG signal with no automated ECG analysis is supported by the indication for use of the predicate device. |
| Target Population | Adults (22 years age and older) | 18 years and older | Adults (22 years age and older) | Identical to reference device. |
| Intended Use | Clinics and Home | Clinics and Home | Clinics and Home | Identical |
{7}
510(k) Summary
K250882
Traditional 510(k) – Huxley Medical
SANSA HSAT
| Device Comparison | Subject Device: SANSA HSAT | Predicate Device: Zio Monitor (K202359) | Reference Device: SANSA HSAT (K244027) | Comparison |
| --- | --- | --- | --- | --- |
| Environment | Use | Use | Use | |
| Device Placement for Data Collection | Chest sensor | Chest Sensor | Chest sensor | Identical |
| Device Sensors | Chest Sensor
Accelerometer
ECG
Reflectance Photo-plethysmography | ECG | Chest Sensor
Accelerometer
ECG
Reflectance Photo-plethysmography | Equivalent - Subject and Predicate contain ECG. |
| Channels | Oximetry
Heart rate
Chest movement
Snoring
Body position
Respiratory effort
Actigraphy
Sleep stage (Sleep/Wake)
ECG | ECG | Oximetry
Heart rate
Chest movement
Snoring
Body position
Respiratory effort
Actigraphy
Sleep stage (Sleep/Wake)
ECG (Reference channel only) | Equivalent – Official ECG channel (not reference only) is equivalent to the ECG channel of the reference device. |
| ECG Recording Characteristics | Number of Leads: Single Lead
ECG Lead Orientation: Modified Lead II
ECG A/D Sampling Rate: 250 Hz
ECG Resolution: | Number of Leads: Single Lead
ECG Lead Orientation: Modified Lead II
ECG A/D Sampling Rate: 200 Hz
ECG Resolution: | Number of Leads: Single Lead
ECG Lead Orientation: Modified Lead II
ECG A/D Sampling Rate: 250 Hz
ECG Resolution: | Identical to reference device. Equivalent to predicate device in performance in recording capability and lead orientation. Higher sample rate in subject device does not raise questions of safety or effectiveness with predicate device as the output ECG is |
4
{8}
510(k) Summary K250882
Traditional 510(k) - Huxley Medical SANSA HSAT
| Device Comparison | Subject Device: SANSA HSAT | Predicate Device: Zio Monitor (K202359) | Reference Device: SANSA HSAT (K244027) | Comparison |
| --- | --- | --- | --- | --- |
| | 15.5 bits | 15.5 bits | 15.5 bits | equivalent. |
| Analysis Outputs | sAHI
Body Position
Discrete States
Heart Rate
Total Sleep Time
SpO2 | No analysis provided by device. | sAHI
Body Position
Discrete States
Heart Rate
Total Sleep Time
SpO2 | Identical to reference device. Subject device and predicate device lack ECG analysis output. |
| Performance | Heart Rate: Arms ≤ 3 bpm (range 30-250 bpm)
SpO2 Arms ≤ 3% (range 70-100%)
Aid to Diagnosis of Moderate to Severe OSA (AHI≥15): Sensitivity 88.2%, Specificity 87.3%
ECG Recording Accuracy:
• Frequency Response: 0.67 Hz to 40 Hz
• Input Impedance: >10 MΩ
• Gain Accuracy: Maximum amplitude error ±10% | No Heart Rate Analysis
Does not collect SpO2
Does not diagnose OSA
ECG Recording Accuracy:
• Frequency Response: 0.67 Hz to 40 Hz
• Input Impedance: >10 MΩ
• Gain Accuracy: Maximum amplitude error ±10%
• Gain Stability: <3% over a 24-hour period | Heart Rate: Arms ≤ 3 bpm (range 30-250 bpm)
SpO2 Arms ≤ 3% (range 70-100%)
Aid to Diagnosis of Moderate to Severe OSA (AHI≥15): Sensitivity 88.2%, Specificity 87.3%
ECG Recording Accuracy:
• Frequency Response: 0.67 Hz to 40 Hz
• Input Impedance: >10 MΩ
• Gain Accuracy: Maximum amplitude error ±10% | Identical to reference device. Identical ECG recording accuracy to predicate device. |
{9}
510(k) Summary K250882
Traditional 510(k) - Huxley Medical SANSA HSAT
| Device Comparison | Subject Device: SANSA HSAT | Predicate Device: Zio Monitor (K202359) | Reference Device: SANSA HSAT (K244027) | Comparison |
| --- | --- | --- | --- | --- |
| | ·Gain Stability: <3% over a 24-hour period
·Timing Accuracy: <30 seconds | ·Timing Accuracy: <30 seconds | ·Gain Stability: <3% over a 24-hour period
·Timing Accuracy: <30 seconds | |
| Data Collection and Transfer | Patient data is wirelessly transferred via Cellular (LTE-M) at the conclusion of the study. In the event of no connectivity or failed transfer, patient data is physically transferred via USB after study conclusion. | Patient data is transferred after study conclusion | Patient data is wirelessly transferred via Cellular (LTE-M) at the conclusion of the study. In the event of no connectivity or failed transfer, patient data is physically transferred via USB after study conclusion. | Identical to reference device. |
| Recording Capacity | Approx. 10 hours per study. 2 nights of study maximum. | Continuous recording for 14 days. | Approx. 10 hours per study. 2 nights of study maximum. | Identical to reference device. Predicate device is focused on 14 days of data and subject and reference are focused on sleep. |
| Energy Source | Rechargeable Lithium Polymer Battery | Lithium Manganese Dioxide Coin Cell | Rechargeable Lithium Polymer Battery | Identical to reference device. |
| Analysis Software | Analysis performed off the recording device, on a compatible cloud-based software platform.
No automated ECG analysis conducted. | No ECG analysis provided by the device. | Analysis performed off the recording device, on a compatible cloud-based software platform.
No automated ECG analysis conducted. | Identical to reference device. Equivalent to predicate. |
{10}
510(k) Summary
K250882
Traditional 510(k) – Huxley Medical
SANSA HSAT
# Non-Clinical Performance Data
The following non-clinical testing was conducted to support device changes:
- Software Unit & Integration Testing to test portal functionality. This testing verified the additional features, and that the data integrity was not altered in the EDF or PDF export process.
- The following performance testing verifies the accuracy of the SANSA device’s ECG recording:
- Testing to FDA recognized consensus standard IEC 60601-2-47 Ed. 2.0 (2012-02) Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems
- ECG recorder performance testing (Retested with latest device firmware)
- Database Accuracy testing for beat detection and heart rate (testing leveraged from previous clearance due to lack of changes)
- Testing to FDA recognized consensus standard ANSI AAMI EC12:2000/(R)2015 Disposable ECG electrodes (testing leveraged from previous clearance due to lack of changes)
- ECG electrode performance testing
- On-body adhesive performance testing
- Human Factors Usability Testing conducted on patients and clinicians in accordance with FDA Guidance Document “Applying Human Factors and Usability Engineering to Medical Devices” Issued February 3, 2016. (Testing leveraged from previous clearance due to lack of changes)
No further system or performance testing was required, as all updates made were isolated in cloud software functionality.
# Clinical Performance Data
Sansa ECG clinical performance was validated through comparison to a simultaneously collected reference standard Holter monitor. Clinically acceptable performance was demonstrated through qualitative and quantitative analysis of the ECG signal.
# Conclusion
The SANSA System is substantially equivalent to the legally marketed predicate device as demonstrated by the similar intended use and similar technologies, with the reference device leveraged to address the sleep-disordered breathing indications for use. The addition to the indications for use to allow usage of the existing ECG waveform for diagnostic purposes is supported by the predicate device. The updates made to the subject device in this submission do not raise questions of safety and effectiveness compared to the predicate and reference devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.