The Connect App for iHealth Next is intended for use in home settings as an aid for people to view vital signs which are measured by iHealth devices (noninvasive blood pressure monitor and pulse oximeter). The Connect App for iHealth Next should not be used to either self-diagnose a disease state or exclude it. Please consult your physician if you find the reading is above a normal threshold.
Device Story
Connect App for iHealth Next is mobile software (iOS/Android) for home use; interfaces with iHealth blood pressure monitors and pulse oximeters via Bluetooth/BLE/Audio. App initiates/stops measurements on connected devices; receives, displays, and stores vital sign data. Users view historical data and trends. Intended as an aid for viewing measurements; not for diagnostic purposes. Healthcare providers do not interact with the app directly; patients use it to monitor their own health data. Output affects clinical decision-making only if patient shares data with physician. Benefits include convenient data tracking and visualization of vital signs.
Clinical Evidence
No clinical data. Evidence consists of software validation per FDA guidance, usability testing with 40 participants (20 lay users, 20 professionals) in home/clinical environments, and wireless coexistence testing.
Technological Characteristics
Software-based data management system. Operates on iOS and Android mobile platforms. Connectivity via Bluetooth, Bluetooth Low Energy, and Audio. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC) as inherited from connected peripheral devices. Software validation performed per FDA guidance.
Indications for Use
Indicated for home users to view vital signs (blood pressure, pulse oximetry) measured by iHealth devices. Not for self-diagnosis or disease exclusion.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG" in blue, and below that is the text "ADMINISTRATION" in blue.
April 16, 2019
Andon health Co., Ltd Yi Liu President No.3 Jinping Street, Ya An Road, Nankai District Tianjin, P.R. China Tianjin, 300190 CN
Re: K181541
Trade/Device Name: Connect App for iHealth Next Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, MNW, DOA Dated: March 15, 2019 Received: March 18, 2019
Dear Yi Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
### 1.0 submitter information
| Name: | Andon Health Co., Ltd. |
|----------------------|--------------------------------------------------------------------------|
| Address: | No.3 Jinping Street, Ya An Road, Nankai District, Tianjin,<br>P.R. China |
| Phone number: | 86-22-87611660 |
| Fax number: | 86-22-6052 6162 |
| Contact: | Yi Liu |
| Date of Preparation: | 6/6/2018 |
### 2.0 Device information
| Trade name: | Connect App for iHealth Next |
|----------------------|------------------------------|
| Common name: | Data management software |
| Classification name: | Data management software |
# 3.0 Classification
Production code: DXN, MNW, DQA Regulation number: 21 CFR 870.2770, 21 CFR 870.1130, 21 CFR 870.2700, Classification: II Panel: 870 Cardiovascular
### 4.0 Predicate device information
Manufacturer: Andon Health Co., Ltd.
Device: Apps-Health01
510(k) number: K122098
### 5.0 Device description
The Connect App for iHealth Next is a software for using it with health devices,
{3}------------------------------------------------
including iHealth medical devices such as Blood Pressure Monitor and pulse oximeter, etc. It can also be used with non-medical devices, such as active monitor, weight scale, etc.
The Connect App for iHealth Next can:
(1) Connect to blood pressure monitor and control the meter to complete measurement, and the data can be transferred to the Connect App for iHealth Next;
(2) Connect to pulse oximeter and transfer data from the pulse oximeter to the Connect App for iHealth Next.
| Device type | Model (510K number) | Control description | Control way | Electrical safety and EMC test evidence |
|------------------------------|-----------------------|---------------------|-------------|-----------------------------------------|
| Blood<br>pressure<br>monitor | KD-936<br>(K120672) | a* | b** | c* d* |
| Blood<br>pressure<br>monitor | KD-972<br>(K121470) | a* | b** | c* d* |
| Blood<br>pressure<br>monitor | BP3L<br>(K152003) | a* | b* | c* d* |
| Blood<br>pressure<br>monitor | KN-550BT<br>(K160014) | a** | b* | c* d* |
| Blood<br>pressure<br>monitor | BP7S<br>(K152379) | a** | b* | c* d* |
| Pulse<br>oximeter | PO3M<br>(K131111) | a** | b* | c* d* |
Below is detailed control information:
Control description:
the App can control medical devices to start/stop the measurement and receive a* data from blood pressure monitor.
a** the App only receives data from medical devices
b* the medical device can only be controlled by App
- b** the medical device can be controlled both by medical device and by App
#### Electrical safety test performed in the previous 510Ks
- C* Electrical safety test according to IEC 60601-1:2005+C1:2006,
#### EMC test performed in the previous 510Ks
- d* EMC test according to IEC 60601-1-2:2007
# 6.0 Intended use
The Connect App for iHealth Next is intended for use in home settings as an aid for people to view vital signs which are measured by iHealth devices (noninvasive blood pressure monitor and pulse oximeter).
{4}------------------------------------------------
The Connect App for iHealth Next should not be used to either self-diagnose a disease state or exclude it. Please consult your physician if you find the reading is above a normal threshold.
# 7.0 Summary comparing technological characteristics with predicate device
| | Predicate device | Proposed device |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Apps-Health01 | Connect App for iHealth Next |
| Product type | software | software |
| Indications for<br>use | The Apps-Health01 is intended for<br>use in home settings as an aid for<br>people to review, analyze, and<br>evaluate test results which are<br>measured from a blood pressure<br>monitor or scale. | The Connect App for iHealth Next<br>App is intended for use in home<br>settings as an aid for people to<br>view vital signs which are<br>measured by iHealth devices. |
| Software<br>platform | iOS | iOS and Android |
| Hardware<br>requirements | iPhone, iTouch, iPad | iPhone, iTouch, iPad, mobile<br>device with Android system |
| Measurement<br>control | Measurement module with Native<br>SDK as transfer module | Measurement module |
| Function-basic | Blood pressure: Test or manually<br>input blood pressure data.<br>Weight: Test or manually input<br>body composition data. | Connect to blood pressure:<br>complete measurement and<br>transfer test data<br>Connect to pulse oximeter and<br>transfer test data |
### Specification Comparison table
{5}------------------------------------------------
| Function-other | My calorie Trends: View your calories consumed vs. calories burned.<br>Reminders: Set reminders<br>(Medicine/Measure blood pressure /Measure weight/Sport) to stay focused on your goals.<br>My plan: Add and modify your weight goals.<br>My BP trends: View your blood pressure trends.<br>My weight Trends: View your weight data.<br>History: view of your historical data.<br>My Diary: Log food and activities to stay on track. | Check FAQ |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Meter<br>Compatibility | Scales, Blood Pressure Monitor | Blood Pressure Monitor,<br>Pulse Oximeter,<br>Weight scale<br>Active monitor |
| Connection<br>with peripheral<br>device by | Bluetooth (HS5 with WIFI additionally, KD-931 by apple 30PIN) | Bluetooth<br>Bluetooth Low Energy<br>Audio |
| Performance<br>test provided | Software validation | Software validation |
### 8.0 Performance summary
### Non-clinical test
### Software validation
Software validation has been performed according to FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". And the result conforms that the proposed device is as safe and effective as the predicate device.
### Usability study
Usability test has been performed according to FDA guidance "Apply for applying human factors and usability engineering to medical devices".
20 lay users and 20 professionals participated in this study on their actual use environment such as home, hospital, clinic and office etc.,
{6}------------------------------------------------
in order to verify that:
- user can choose the device correctly, >
- > user can operate the device in proper environment as intended,
- user can get all the warning and precaution as intended, >
- user can download and run the app correctly, and >
- > user can operate the device correctly,
The following main critical tasks were identified in this study,
- & Read App operation guide and medical device labeling
- < Download and run the App
- � Connect medical device to App
- � Complete measurement
The study results demonstrate that the Connect App for iHealth Next are understood by a variety of users, and provide sufficient information for the safe and effective use of the device.
#### Wireless Coexistence test
Wireless coexistence test has been performed to verify that the proposed device can be used in intended environments.
### Clinical test
This section is not applicable.
### 9.0 Comparison to the predicate device and the conclusion
The Connect App for iHealth Next is similar with the predicate device Apps-Health01. The new device can be used with more medical devices such as the pulse oximeter. Additionally, the proposed device can also be installed in Android mobile devices.
The tests and data in this submission demonstrate that the proposed device is as safe and effective as the predicate device.
{7}------------------------------------------------
è-¦å'Šï¼šå½"剺版本的 Acrobat 或 Adobe Reader丕æ" 挤本表â‰ã€' 请啇级至最新版本以获得完全支挕。
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.