iHealth Track Blood Pressure Monitor

K160014 · Andon Health Co, Ltd. · DXN · Feb 2, 2016 · Cardiovascular

Device Facts

Record IDK160014
Device NameiHealth Track Blood Pressure Monitor
ApplicantAndon Health Co, Ltd.
Product CodeDXN · Cardiovascular
Decision DateFeb 2, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

iHealth Track Blood Pressure Monitor(KN-550BT) is for use by medical professionals or at home and are non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Device Story

The iHealth Track Blood Pressure Monitor (KN-550BT) is an oscillometric device that uses an inflatable upper-arm cuff and a silicon-integrated pressure sensor to measure systolic/diastolic blood pressure and pulse rate. Designed for home or clinical use by adults, the device processes pressure signals to calculate blood pressure, detect irregular heartbeats, and compute measurement averages. Results are displayed on an integrated LCD or transmitted to iOS/Android devices. The device assists healthcare providers and patients in monitoring cardiovascular health by providing automated, non-invasive readings.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1), and performance characteristics for automated non-invasive sphygmomanometers (IEC 80601-2-30).

Technological Characteristics

Oscillometric measurement principle using silicon-integrated pressure sensors. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 80601-2-30. Features include LCD display and mobile connectivity (iOS/Android).

Indications for Use

Indicated for adult individuals requiring non-invasive measurement of systolic/diastolic blood pressure and pulse rate via upper arm cuff (22cm-48cm circumference). Suitable for use by medical professionals or home users.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Image /page/0/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human figures, one behind the other, with their arms raised. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 2, 2016 Andon Health Co., Ltd Ms. Liu Yi President No. 3 Jin Ping Street, Ya An Road, Nankai District Tianjin, 300190 CN Re: K160014 Trade/Device Name: iHealth Track Blood Pressure Monitor (KN-550BT) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 30, 2015 Received: January 4, 2016 Dear Ms. Liu Yi. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Mitchell Stein for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K160014 Device Name iHealth Track Blood Pressure Monitor (KN-550BT) Indications for Use (Describe) iHealth Track Blood Pressure Monitor(KN-550BT) is for use by medical professionals or at home and are non-invasive blood pressure measurement system intended to measure the diastolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the usper arm. The cuff circumference is limited to 22cm-48cm. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input checked="checked" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K160014 ## Page 1 of 4 ## 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92. #### 1.0 submitter's information | Name: | Andon Health Co., Ltd. | |----------------------|------------------------------------------------------------| | Address: | No 3, Jinping Street Ya An Road, Nankai District, Tianjin, | | | P.R. China | | Phone number: | 86-22-6052 6161 | | Fax number: | 86-22-6052 6162 | | Contact: | Liu Yi | | Date of Preparation: | 12/15/2015 | #### 2.0 Device information Trade name: iHealth Track Blood Pressure Monitor (KN-550BT) iHealth Track Blood Pressure Monitor Device name: Classification name: Noninvasive blood pressure measurement system #### 3.0 Classification Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular #### 4.0 Predicate device information Manufacturer: Andon Health Co., Ltd. KD-927 Fully Automatic Electronic Blood Pressure Monitor Device: 510(k) number: K141984 {4}------------------------------------------------ K160014 Page 2 of 4 #### 5.0 Intended use iHealth Track Blood Pressure Monitor(KN-550BT) is for use by medical professionals or at home and are non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. The intended use and the indication for use of iHealth Track Blood Pressure Monitor(KN-550BT), as described in its labeling are the same as the predicate device KD-927 (K141984). #### 6.0 Device description iHealth Track Blood Pressure Monitor(KN-550BT) is and manufactured according to IEC 80601-2-30. The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, the measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user. More over, it also obtains the function of averaging the measurement results. The new devices achieves its function by LCD or iOS/Andriod devices. #### 7.0 Summary comparing technological characteristics with predicate device | Technological Characteristics | Comparison result | |-------------------------------|-------------------| | Design principle | Identical | | Appearance | Similar | | Patients contact Materials | Similar | | Performance | Similar | | Biocompatibility | Identical | | Mechanical safety | Identical | | Standards met | Identical | | Electrical safety | Identical | | EMC | Identical | | Function | Similar | {5}------------------------------------------------ K160014 Page 3 of 4 #### 8.0 Discussion of non-clinical and clinical test performed #### Non-clinical Tests have been done as follows: a. Electromagnetic compatibility test according to IEC 60601-1-2; b. Electrical safety according test to IEC 60601-1; c. Safety and performance characteristics of the test according to IEC 80601-2-30 None of the test demonstrates that KN-550BT Track Blood Pressure Monitor bring new questions of safety and effectiveness. #### 9.0 Performance summary iHealth Track Blood Pressure Monitor(KN-550BT) conforms to the following standards: - IEC 60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD) - · IEC 60601-1-2:2007/(R)2012, Medical Electrical Equipment Part 1-2: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Compatibility - Requirements And Tests (Edition 3) - · IEC 80601-2-30 Edition 1.1 2013-07, Medical Electrical Equipment Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers #### 10.0 Comparison to the predicate device and the conclusion iHealth Track Blood Pressure Monitor(KN-550BT) is substantially equivalent to Fully Automatic Electronic Blood Pressure Monitor KD-927 whose 510(k) the number is K141984. {6}------------------------------------------------ K160014 Page 4 of 4 The new device KN-550BT is very similar with its predicate device in the intended use, the design principle, the material, the performance and the applicable standards. Only their appearance, the memory capacity, electrical power, measure process, average measure function and the MCU are different, and a new software platform has been added. However, the test in this submission provides demonstrates that these small differences do not raise any new questions of safety and effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...