Philips Hemodynamic Application R1.0

K181311 · Philips Medical Systems Nederlands B.V. · MWI · Sep 7, 2018 · Cardiovascular

Device Facts

Record IDK181311
Device NamePhilips Hemodynamic Application R1.0
ApplicantPhilips Medical Systems Nederlands B.V.
Product CodeMWI · Cardiovascular
Decision DateSep 7, 2018
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Indications for Use

The Philips Hemodynamic Application is intended for use by professional healthcare providers for physiologic/ hemodynamic monitoring, medical data processing and analytical assessment. The software may be used to display analyze surface Electrocardiogram (ECG), Respiration. Invasive Blood Pressure (IBP), Pulse Oximetry (SpO2), End Tidal CO2 (ETCO2), Fractional Flow Reserve (FFR), Instant Wave-Free Ratio (iFR), Non-Invasive Blood Pressure (NIBP), surface body Temperature and thermal Cardiac Output. The software is intended for use with other devices, such as physiological monitoring systems, information management systems, image acquisition and other medical devices. Use of the software in combination with physiological monitoring system is not intended to be used where unattended patient monitoring is desired, or in situations where arrhythmia detection is required. The software in combination with an information management system provides the ability to transmit patient data files for storage, viewing and analysis at distributed locations via the intranet or internet. The software is indicated for use in the following areas; (interventional) cardiology, electrophysiology, The Philips Hemodynamic Application is indicated for use for all human patients of all ages.

Device Story

Philips Hemodynamic Application is an optional software accessory to the Xper Flex Cardio Physiomonitoring System. It acquires real-time physiological data (ECG, respiration, IBP, SpO2, ETCO2, NIBP, temperature) from the host monitor. It performs hemodynamic calculations including FFR, iFR, thermal cardiac output, valve area, and valve gradient. Used by professional healthcare providers in interventional cardiology, electrophysiology, and radiology. It interfaces with Xper Information Management (XperIM) for data storage, analysis, and reporting. The application provides a user interface for waveform visualization, alarm limit setting, and table-side operation via X-ray system integration. It benefits patients by providing real-time hemodynamic assessment and data management during cardiac interventions.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical performance testing, including software verification and validation (usability and simulated use).

Technological Characteristics

Software-based hemodynamic monitoring application. Features include waveform visualization, hemodynamic calculations (FFR, iFR, cardiac output), and alarm limit management. Integrates with X-ray systems and XperIM for data management. Implements iFR algorithm for invasive blood pressure measurement. Complies with IEC 62304, IEC 62366-1, and ISO 14971 standards.

Indications for Use

Indicated for all human patients of all ages requiring physiologic/hemodynamic monitoring in interventional cardiology, electrophysiology, and radiology. Not for unattended monitoring or arrhythmia detection.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 7, 2018 Philips Medical Systems Nederland B.V. Owen Callaghan Regulatory Affairs Manager Veenpluis 4-6 5684 PC Best, The Netherlands Re: K181311 Trade/Device Name: Philips Hemodynamic Application R1.0 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: August 6, 2018 Received: August 8, 2018 Dear Owen Callaghan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Stephen C. Browning -S5 for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K181311 Device Name Philips Hemodynamic Application R1.0 ### Indications for Use (Describe) The Philips Hemodynamic Application is intended for use by professional healthcare providers for physiologic/ hemodynamic monitoring, medical data processing and analytical assessment. The software may be used to display analyze surface Electrocardiogram (ECG), Respiration. Invasive Blood Pressure (IBP), Pulse Oximetry (SpO2), End Tidal CO2 (ETCO2), Fractional Flow Reserve (FFR), Instant Wave-Free Ratio (iFR), Non-Invasive Blood Pressure (NIBP), surface body Temperature and thermal Cardiac Output. The software is intended for use with other devices, such as physiological monitoring systems, information management systems, image acquisition and other medical devices. Use of the software in combination with physiological monitoring system is not intended to be used where unattended patient monitoring is desired, or in situations where arrhythmia detection is required. The software in combination with an information management system provides the ability to transmit patient data files for storage, viewing and analysis at distributed locations via the intranet or internet. The software is indicated for use in the following areas; (interventional) cardiology, electrophysiology, The Philips Hemodynamic Application is indicated for use for all human patients of all ages. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------| | | <span> <span style="font-size:100%;"> <span style="font-family:Arial;">☒</span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | | | <span> <span style="font-size:100%;"> <span style="font-family:Arial;">☐</span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. ## 510(k) Summary This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92. | Date Prepared: | May 14, 2018 | | |------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best<br>The Netherlands<br>Establishment Registration Number: 3003768277 | | | Primary Contact<br>Person: | Owen Callaghan<br>Regulatory Affairs Officer<br>Phone: +31 621394159<br>E-mail: owen.callaghan@philips.com | | | Secondary Contact<br>Person: | Marta Walker<br>Head of Regulatory Affairs IGT Systems<br>Phone: +31 631978546<br>E-mail: marta.walker@philips.com | | | Device: | Trade Name: Philips Hemodynamic Application R1.0<br>Classification Name: Cardiac monitor (including cardiotachometer and rate alarm)<br>Classification Regulation: 21 CFR 870.2300<br>Classification Panel: Cardiovascular<br>Device Class: Class II<br>Primary Product Code: MWI Monitor, Physiological, Patient (without Arrhythmia Detection or alarms)<br>Secondary Product Code: -- | | | Primary Predicate<br>Device: | Trade Name: Xper Flex Cardio Physiomonitoring system<br>Manufacturer: Witt Biomedical Corporation (a wholly owned subsidiary of Philips Holding USA, Inc.) latterly transferred to Invivo, a division of Philips Medica Systems<br>510(k) Clearance: K101571 (October 26, 2010)<br>Classification Name: Cardiac monitor (including cardiotachometer and rate alarm)<br>Classification Regulation: 21 CFR 870.2300<br>Classification Panel: Cardiovascular<br>Device Class: Class II | | | | Product Code: | MWI Monitor, Physiological, Patient (without<br>Arrhythmia Detection or alarms) | | Reference Device 1: | Trade Name:<br>Manufacturer:<br>510(k) Clearance:<br>Classification Name:<br>Classification Regulation:<br>Classifcation Panel:<br>Device Class:<br>Product Code: | Volcano iFR® Modality<br>Volcano Corporation, 3721 Valley Center Drive,<br>San Diego, CA 92130<br>K133323 (March 14, 2013)<br>Ultrasonic pulsed echo imaging system<br>21 CFR Part 892.1560<br>Radiology<br>Class II<br>IYO System, Imaging, Pulsed Echo, Ultrasonic | | Reference Device 2: | Trade Name:<br>Manufacturer:<br>510(k) Clearance:<br>Classification Name:<br>Classification Regulation:<br>Classifcation Panel:<br>Device Class:<br>Product Code: | Volcano iFR® ScoutTM<br>Volcano Corporation, 3721 Valley Center,<br>Drive,San Diego, CA 92130<br>K150441 (March 20, 2015)<br>Ultrasonic pulsed echo imaging system<br>21 CFR Part 892.1560<br>Radiology<br>Class II<br>IYO System, Imaging, Pulsed Echo, Ultrasonic | {4}------------------------------------------------ Device description: Philips Hemodynamic Application is a new software medical device that enables invasive investigation of cardiac and vascular disease. It will be offered as an optional accessory to the Xper Flex Cardio Physiomonitoring System, (K101571). Currently, the functionality offered by the Philips Hemodynamic Application is provided by "Hemodynamic Control Software" of the currently marketed and predicate Xper Flex Cardio Physiomonitoring System. > The software connects to the patient monitor (i.e. the Xper Flex Cardio Physiomonitoring System) and during the intervention continuously acquires realtime physiological data and alarms. In addition, Philips Hemodynamic Application provides the following functionality: - Visualize and analyze: surface ECG, Respiration rate (RR), Invasive . Blood Pressure (IBP), Pulse Oximetry (SpO2), End Tidal CO2 (etCO2), Noninvasive monitoring and recording of Non-Invasive Blood Pressure (NIBP), Body surface temperature (Tskin); - Provide Hemodynamic calculations: Fractional Flow Reserve (FFR), Instant Wave-Free Ratio (iFR), thermal cardiac output parameters, valve area and valve gradient. Furthermore, Philips Hemodynamic Application also interfaces with Xper Information Management (XperIM) System (K101571) which it can transfer data to for the purpose of data collection/display, processing and patient reporting. {5}------------------------------------------------ ## PHILIPS The Philips Hemodynamic Application, provided as an optional accessory to Indications for Use: the Philips Xper Flex Cardio Physiomonitoring System, has the following indications for use: > The Philips Hemodynamic Application is intended for use by professional healthcare providers for physiologic/hemodynamic monitoring, medical data processing and analytical assessment. > The software may be used to display and/or analyze surface Electrocardiogram (ECG), Respiration, Invasive Blood Pressure (IBP), Pulse Oximetry (SpO2), End Tidal CO2 (ETCO2), Fractional Flow Reserve (FFR), Instant Wave-Free Ratio (iFR), Non-Invasive Blood Pressure (NIBP), surface body Temperature and thermal Cardiac Output. > The software is intended for use with other devices, such as physiological monitoring systems, information management systems, image acquisition and other medical devices. > Use of the software in combination with physiological monitoring system is not intended to be used where unattended patient monitoring is desired, or in situations where arrhythmia detection is required. > The software in combination with an information management system provides the ability to transmit patient data files for storage, viewing and analysis at distributed locations via the intranet or internet. > The software is indicated for use in the following areas: (interventional) cardiology, electrophysiology and radiology. > The Philips Hemodynamic Application is indicated for use for all human patients of all ages. > The indications for use statement of the Philips Hemodynamic Application are similar to the indications for use of the currently marketed predicate Xper Flex Cardio Physiomonitoring System (K101571). Both devices are: - Indicated for use by professional healthcare providers for physiologic/ ● hemodynamic monitoring, medical data processing and analytical assessment. - . Indicated for use in the following areas: (interventional) cardiology, electrophysiology and radiology. - Not intended to be used where unattended patient monitoring is desired, or in ● situations where arrhythmia detection is required. - Indicated for use for all human patients of all ages. The Indications for Use of Philips Hemodynamic Application have been further detailed by adding a description of the iFR feature. Addition of this functionality does not raise any new question of safety and effectiveness since Philips Hemodynamic Application still has the same intended use as the predicate, namely for use by professional healthcare providers for complete physiologic/hemodynamic monitoring. Based on the information provided above, the Philips Hemodynamic Application is considered substantially equivalent to the primary currently {6}------------------------------------------------ # PHILIPS Technological characteristics: marketed and predicate device Xper Flex Cardio Physiomonitoring System in terms of Indications for Use. Philips Hemodynamic Application employs comparable technology as implemented in the Hemodynamic Control Software module which is part of the Xper Flex Cardio Physiomonitoring System software. The technical similarities are such that both devices: - Provide visualization of physiology waveforms and/or values (surface ECG, ● respiration rate, invasive blood pressure, non-invasive blood pressure, pulse oximetry, end tidal CO2, body surface temperature). - . Provide user interface to display acquired physiological waveforms and values. - . Implement algorithms for hemodynamic calculations: Fractional Flow Reserve (FFR), thermal cardiac output parameters, valve area and valve gradient. - . Allow for the setting of alarm limits (upper/lower) and show visual alarms on the host PC. The Philips Hemodynamic Application provides the capability to set limits and visualize alarms, however it still requires the predicate to produce the audible alarm. - Provides a feature such that samples of the live monitoring can be captured. - Have X-ray system integration which allows table side operation via the touch screen module. - . Interfaces with the cleared Xper Information Management (XperIM) System (K101571), enabling data to be transferred for the purpose of patient data management. The Philips Hemodynamic Application implements the iFR algorithm for invasive blood pressure measurement (iFR spot and pullback) which is not present in the currently marketed predicate. These measurements are identical to the reference devices Volcano iFR® Modality and Volcano iFR® Scout™. This extends the existing Fractional Flow Reserve (FFR) functionality which uses hyperemic agent. This difference and the ones outlined above do not raise any new questions regarding safety and effectiveness. Based on the information provided above, the Philips Hemodynamic Application is considered substantially equivalent to the primary currently marketed and predicate device Xper Flex Cardio Physiomonitoring System in terms of fundamental scientific technology. {7}------------------------------------------------ # PHILIPS Summary of Non-Data: Non-clinical performance testing has been performed on Philips Hemodynamic Clinical Performance Application and demonstrates compliance with the following FDA recognized consensus standards and FDA guidance document(s): - IEC 62304 Medical device software Software life cycle processes (Edition ● 1.1, 2015-06). FDA/CDRH recognition number 13-79. - IEC 62366-1 Medical devices Part 1: Application of usability engineering . to medical devices (Edition 1.0, 2015-02). FDA/CDRH recognition number 5-114. - ISO 14971 Medical devices Application of risk management to medical ● devices (Edition. 2.0, 2007). FDA/CDRH recognition number 5-40, - ISO15223-1 Medical devices Symbols to be used with medical device ● labels, labelling and information to be supplied - Part 1: General requirements (Edition 3, 2016-11). FDA/CDRH recognition number 5-117, - Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005 (document number 337), - Guidance for Industry and FDA Staff Applying Human Factors and ● Usability Engineering to Medical Devices, February 3, 2016 (document number 1757) - . "Guidance for Industry and FDA Staff – Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", October 2, 2014 (document number 1825), - . "Guidance for Industry, FDA Reviewers and Compliance on – Off-The-Shelf Software Use in Medical Devices", September 9, 1999 (document number 585). - "Guidance for Industry and FDA Staff The 510(k) Program: Evaluating . Substantial Equivalence in Premarket Notifications [510(k)]", July 28, 2014 (document number 1766). Non-clinical software verification testing has been performed to verify that all requirements of the System Requirements Specification, User Interaction Design as well as the identified safety risk control measures from the Detailed Risk Management Matrix and the Privacy and Security requirements for Philips Hemodynamic Application have been implemented. Algorithm verification was performed using calibrated simulator tools that confirmed the algorithm was correctly implemented in the product. Results demonstrated that all executed verification tests were passed. Software validation testing has been performed to validate that Philips Hemodynamic Application conforms to its intended use, claims and user needs. The validation consisted of the following activities: - Usability validation was performed with both cardiologists and monitoring . nurse/technicians in a simulated use environment in a simulated environment. Philips Hemodynamic Application found to be safe and effective for the intended use, users and use environment; - In-house simulated use design validation was performed with experienced . Clinical Marketing specialists that fulfill the intended user profile. The participants executed validation protocols in the form of a device workflow {8}------------------------------------------------ # PHI | | to validate user needs, intended use and effectiveness of the safety and<br>security related measures. As part of the validation, the implemented<br>algorithms were evaluated as part of the workflow. Results demonstrated that<br>all executed validation protocols were passed. | |----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | All these tests were used to support substantial equivalence of the subject device<br>and demonstrate that <b>Philips Hemodynamic Application</b> :<br>• complies with the aforementioned international and FDA-recognized<br>consensus standards and FDA guidance documents, and<br>• meets the acceptance criteria and is adequate for its intended use. | | | Based on the information provided above, <b>Philips Hemodynamic Application</b><br>is substantially equivalent to the currently marketed predicate device <i>Xper Flex<br/>Cardio Physiomonitoring System</i> in terms of safety and effectiveness. | | <b>Summary of Clinical<br/>Performance Data:</b> | <b>Philips Hemodynamic Application</b> did not require a clinical study since<br>substantial equivalence to the currently marketed predicate device <i>Xper Flex<br/>Cardio Physiomonitoring System</i> was demonstrated with the following attributes:<br>• Indication for use;<br>• Technological characteristics;<br>• Non-clinical performance testing, including safety and effectiveness. | | | The verification and validation test results of <b>Philips Hemodynamic Application</b><br>described above support the safety and effectiveness of the product. It conforms<br>to the intended use, the user needs and the claims and is therefore considered<br>substantially equivalent to the currently marketed predicate device <i>Xper Flex<br/>Cardio Physiomonitoring System</i> . | | <b>Substantial<br/>Equivalence<br/>Conclusion:</b> | <b>Philips Hemodynamic Application</b> is substantially equivalent to the currently<br>marketed predicate device <i>Xper Flex Cardio Physiomonitoring System</i> in terms<br>of indications for use, technological characteristics and safety and effectiveness. | | | Additionally, substantial equivalence was demonstrated by non-clinical<br>performance tests provided in this 510(k) premarket notification. These tests<br>demonstrate that <b>Philips Hemodynamic Application</b> complies with the user<br>need requirements as well as the requirements specified in the international and<br>FDA-recognized consensus standards and is as safe and effective as its predicate<br>device and does not raise any new safety and/or effectiveness concerns. |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...