XPER INFORMATION MANAGEMENT AND FLEX CARDIO PHYSIOMONITORING SYSTEMS
Applicant
Witt Biomedical Corp
Product Code
MWI · Cardiovascular
Decision Date
Oct 26, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The physiomonitoring system is intended for use by professional healthcare providers for complete physiologic/hemodynamic monitoring, clinical data acquisition, medical image/data processing. and analytical assessment. The system is indicated for use in the cardiac catheterization, following areas: cardiology, electrophysiology, radiology, invasive radiology and other areas where cardiac monitoring may be required. The system is not intended to be used in the proximity of magnetic resonance imaging. The data may also be acquired from and/or sent to other devices, such as physiological monitoring systems, information management systems, image acquisition/storage devices, and other medical devices. User-adjustable alarms (both visual and audible) available in the system alert the operator to anomalous occurrences and facilitate timely responses. Use of the system is not intended where unattended patient monitoring is desired, or in situations where arrhythmia detection is required. The system provides the ability to transmit patient data files for storage, viewing and analysis at distributed locations via the intranet or internet, or may function as a stand-alone device.
Device Story
Xper Flex Cardio Physiomonitoring System and Xper Information Management System are enhancements to the cleared Xper Information Management System (K063840). The system acquires, displays, trends, stores, and transmits physiologic/hemodynamic data, including surface ECG (up to 16-lead), respiration, invasive/non-invasive pressure, SpO2, ETCO2, temperature, and thermal cardiac output. It processes multi-channel ECG signals and generates graphical ST segment maps. Used in clinical settings like cardiac catheterization labs and cardiology departments by healthcare professionals. The system features redesigned front-end/patient care consoles (DC power), new software modules (hemodynamic, EP logging/reporting, web portal, patient status viewer), and connectivity to other cleared devices. It supports server functions in virtual environments and includes alarm silencing. Outputs assist clinicians in evaluating cardiopulmonary performance and vascular status. The system functions as a standalone device or networked via intranet/internet for distributed data access.
Clinical Evidence
Bench testing only. Performance testing, including master test plans and hazard analysis, demonstrated substantial equivalence to predicate devices. Compliance with voluntary standards (IEC 60601-1, IEC 60601-1-1, EN 60601-1-2, IEC 60601-2-30, IEC 60601-2-49) was verified.
Technological Characteristics
Physiomonitoring and information management system. Hardware includes front-end and patient care consoles (DC power). Supports up to 16-lead ECG. Connectivity via intranet/internet. Software includes hemodynamic modules, EP logging/reporting, and patient status viewer. Complies with IEC 60601-1, 60601-1-1, 60601-2-30, 60601-2-49, and EN 60601-1-2. Integrates Philips ECG algorithm (K073376).
Indications for Use
Indicated for professional healthcare use in cardiology, cardiac catheterization, electrophysiology, and radiology for physiologic/hemodynamic monitoring and data management. Patient population includes those requiring cardiac monitoring. Contraindicated for use in proximity to MRI, unattended patient monitoring, or situations requiring arrhythmia detection.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
(015-7)
.
: :
0CT 26 2010
# DHILIPS
| 510(K) SUMMARY | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by: | Witt Biomedical Corporation (a wholly owned subsidiary of Philips<br>Holding USA, Inc.)<br>305 North Drive, Melbourne Florida 32934 |
| Contact Person: | James Luker<br>Phone: (321) 253-5693 ext 1161<br>Fax: (321) 253-0372 |
| Date Prepared: | May 28, 2010 |
| Proprietary Name: | Xper Flex Cardio physiomonitoring system ,<br>Xper Information Management System |
| Common Name: | Physiomonitoring System<br>Information System |
| Classification Name: | 21 CFR § 870.2300 74 MWI (for Physiomonitoring System)<br>Monitor, Physiological, Patient (Without Arrhythmia Detection Or<br>Alarms)<br>Class II |
| | 21 CFR § 870.1425 74 DQK (for Information system)<br>Computer, Diagnostic, Programmable<br>Class II |
| Predicate Device(s): | Xper Information Management System- K063840 |
| | GE MacLab/CardioLab EP/ComboLab System- K050093 |
| | Poise DataEngine- K040969 |
| | Siemens AXIOM Sensis- K020440<br>Alliance Instruments Integrity Patient Monitor- K980688 |
Philips Xper Flex Cardio Physiomonitoring System / Xper Information Management System Traditional 510(k) . . ・ . '
. .
:
・
·
Page 5-2
·
{1}------------------------------------------------
2/5-
#### Device Description:
The Xper Flex Cardio physiomonitoring system and Xper Information Management systems are enhancements of the cleared Xper Information Management system (K063840).
The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular disease when non-invasive indicators warrant such. The Xper Flex Cardio physiomonitoring system may be used to display and analyze surface ECG (electrocardiogram), respiration, invasive pressure, SpO2 (pulse oximetry), end tidal CO2 (ETCO2) and non-invasive pressure waveforms; surface body temperature and thermal cardiac output curves. The system is capable of processing/analyzing information such as multi-channel ECG signals and displaying a graphical ST segment map. The Xper Flex Cardio physiomonitoring system makes measurements that assist physicians and other clinicians to evaluate a patient's overall cardiopulmonary performance, conductive system and general vascular status. The Xper Flex Cardio physiomonitoring system is not intended for use in the proximity of magnetic resonance imaging.
The Xper Information Management System is used for acquiring, displaying, trending, storing and transmitting various types of data such as physiologic/hemodynamic, clinical, medical image and other related data. The system is capable of processing/analyzing information such as multi-channel ECG signals, displaying a graphical ST segment map and performing other data management functions such as creating reports. Data may be acquired from and/or sent to other devices such as physiological monitoring systems, information management systems, image acquisition/storage devices and other medical devices.
Philips Xper Flex Cardio Physiomonitoring System / Xper Information Management System Traditional 510(k)
Page 5-3
{2}------------------------------------------------
Intended Use Statement(s): Xper Flex Cardio Physiomonitoring System-The physiomonitoring system is intended for use by professional healthcare providers for complete physiologic/hemodynamic monitoring, clinical data acquisition, medical image/data processing. and analytical assessment. The system is indicated for use in the cardiac catheterization, following areas: cardiology, electrophysiology, radiology, invasive radiology and other areas where cardiac monitoring may be required. The system is not intended to be used in the proximity of magnetic resonance imaging. The data may also be acquired from and/or sent to other devices, such as physiological monitoring systems, information management systems, image acquisition/storage devices, and other medical devices.
User-adjustable alarms (both visual and audible) available in the system alert the operator to anomalous occurrences and facilitate timely responses. Use of the system is not intended where unattended patient monitoring is desired, or in situations where arrhythmia detection is required.
The system provides the ability to transmit patient data files for storage, viewing and analysis at distributed locations via the intranet or internet, or may function as a stand-alone device.
Xper Information Management- The information management system is intended for use under the direct supervision of a healthcare practitioner for acquiring, displaying, trending, storing and transmitting various types of data. such as physiologic/ hemodynamic, clinical, medical image and other related data. The system is capable of processing/analyzing information, such as multichannel ECG signals, and performing other data management functions, such as creating reports. Data may be acquired from and/or sent to other devices, such as physiological monitoring systems, information management systems, image acquisition/storage devices, and other medical devices.
The system is indicated for use in the following areas: cardiology, cardiac catheterization. electrophysiology, radiology, invasive radiology, and surrounding areas where access to the information is needed.
The system consists of modules and may be entirely a software offering or a hardware/software offering. It is intended for use on standard computer systems and does not require proprietary hardware. The solution is available as a single module or combination of modules, or may function as a standalone system.
The system is capable of receiving and displaying user-adjustable alarms (both visual and audible) available in the system, which alert the operator to anomalous occurrences and facilitate timely responses. Use of the system is not intended where unattended patient
Philips Xper Flex Cardio Physiomonitoring System / Xper Information Management System Traditional 510(k)
Page 5-4
75-
{3}------------------------------------------------
monitoring is desired, or in situations where arrhythmia detection is required.
The system provides the ability to transmit patient data files for storage, viewing and analysis at distributed locations via the intranet or internet, or may function as a stand-alone device.
The enhanced device has the same technological characteristics as the legally marketed predicate device(s). The enhancements consist of these primary changes:
- > Introduction of trade name Xper Flex Cardio Physiomonitoring System as well as names for devices and software components.
- A Introduction of "re-designed" Front-End & Patient Care Console devices which include a change from AC power to DC power.
- A Introduction of separate Intended Use statements for the monitoring aspects (Xper Flex Cardio Physiomonitoring System) and the information management aspects (Xper Information Management System)
- > ' Change from CAS NE Non-Invasive Blood Pressure (NIBP) module to CAS ND+ Non-Invasive Blood Pressure (NIBP) module.
- A Change from Respironics ETCO2 sensors to Philips branded ETCO2 sensors.
- > Introduction of new software modules:
- Hemodynamic Software Module O
- Hemodynamic Control Software O
- Electrophysiology Logging 0 દિદ Reporting (EP Logging & Reporting)
- Web Portal O
- Patient Status Viewer (Whiteboard) O
- o Gonnectivity to cleared devices 9
- > Increase from 12 lead ECG (maximum) to 16 Lead (maximum)
- > Introduction of Philips ECG algorithm (individually cleared in K073376)
- A Upgrades in hardware and software (custom and off the shelf) to facilitate integration of the latest technology.
- A Introduction of connectivity to other cleared devices.
- A Introduction of ability to run server functions in a virtual! software environment.
- A Introduction of alarm silencing functionality
Philips Xper Flex Cardio Physiomonitoring System / Xper Information Management System Traditional 510(k)
Technological Characteristics:
{4}------------------------------------------------
Performance Data:
The Performance Testing as well as the hazard analysis for the Xper Flex Cardio physiomonitoring system and Xper Information Management System provides objective evidence that it is substantially equivalent to the predicate Xper Information Management system and other listed predicate devices.
Performance Testing:
- . Master Test Plan(s)
#### Voluntary Standards
General Requirements for Safety .
- . IEC 60601-1
- . IEC 60601-1-1
- EN 60601-1-2 ●
- IEC 60601-2-30 .
- IEC 60601-2-49 .
| | Substantial Equivalence |
|---|-------------------------|
| 2 | CALL |
| | |
The Xper Flex Cardio physiomonitoring system and Xper Information Management System are substantially equivalent to the following predicate devices:
| Predicate Device | Manufacturer | 510(k) number | Clearance date: |
|-------------------------------------------|--------------------------------------------------------------------------------------------------|---------------|-----------------|
| Xper Information<br>Management System | Witt Biomedical<br>Corporation (a wholly<br>owned subsidiary of<br>Philips Holding USA,<br>Inc.) | K063840 | 02/06/2007 |
| MacLab/CardioLab<br>EP/ComboLab<br>System | GE Medical Systems | K050093 | 05/13/2005 |
| Poise DataEngine | Poise Technology<br>Corp. | K040969 | 06/08/2004 |
| AXIOM Sensis | Siemens Healthcare | K020440 | 12/03/2002 |
| Integriti Patient<br>Monitor | Alliance Instruments | K980688 | 03/16/1999 |
Conclusion The results of the safety and performance testing demonstrate that the Xper Flex Cardio physiomonitoring system and the Xper Information Management System are as safe and effective as the predicate device(s) and perform as well as the predicate device(s).
Philips Xper Flex Cardio Physiomonitoring System / Xper Information Management System Traditional 510(k)
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure, with three curved lines representing its wings or body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Witt Biomedical Corporation c/o Mr. James Luker Q & R Engineer 305 North Drive Melbourne, FL 32934
0CT: 2 6 2010
Re: K101571 -
Trade/Device Name: Xper Flex Cardio Physiomonitoring System and Xper Information Management device
Regulation Number: 21 CFR 870.2300
Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (two) Product Code: MWI, DQK Dated: September 28, 2010 Received: September 29, 2010
Dear Mr. Luker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. James Luker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number:
K101571
OCT 26 2010
1/2
Device Name: Xper Flex Cardio physiomonitoring system Indications for Use:
The physiomonitoring system is intended for use by professional healthcare providers for complete physiologic/hemodynamic monitoring. The system may be used to display and analyze surface ECG (electrocardiogram), respiration, invasive pressure, pulse oximetry (SpO2), end tidal CO2 (ETCO2), fractional flow reserve (FFR), non-invasive blood pressure (NIBP), surface body temperature and thermal cardiac output. The system also provides for clinical data acquisition, medical image/data processing, and analytical assessment. The system is indicated for use in the following areas: cardiology, cardiac catheterization, electrophysiology, radiology, invasive radiology and other areas where cardiac monitoring may be required. The system is not intended to be used in the proximity of magnetic resonance imaging. The data may also be acquired from and/or sent to other devices, such as physiological monitoring systems, information management systems, image acquisition/storage devices, and other medical devices.
User-adjustable alarms (both visual and audible) available in the system alert the operator to anomalous occurrences and facilitate timely responses. Use of the system is not intended where unattended patient monitoring is desired, or in situations where arrhythmia detection is required.
The system provides the ability to transmit patient data files for storage, viewing and analysis at distributed locations via the intranet or internet, or may function as a standalone device.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K101571
Page 1 of 1
{8}------------------------------------------------
## 510(k) Number: 《 / 0757 |
Indications for Use:
The information management system is intended for use under the direct supervision of a healthcare practitioner for acquiring, displaying, trending, storing and transmitting various types of data, such as physiologic/ hemodynamic, clinical, medical image and other related data. The system is capable of processing/analyzing information, such as multi-channel ECG signals, and performing other data management functions, such as creating reports. Data may be acquired from and/or sent to other devices, such as physiological monitoring systems, information management systems, image acquisition/storage devices, and other medical devices.
The system is indicated for use in the following areas: cardiology, cardiac catheterization, electrophysiology, radiology, invasive radiology, and surrounding areas where access to the information is needed.
The system consists of modules and may be entirely a software offering or a hardware/software offering. It is intended for use on standard computer systems and does not require proprietary hardware. The solution is available as a single module or combination of modules, or may function as a standalone system.
The system is capable of receiving and displaying user-adjustable alarms (both visual and audible) available in the system, which alert the operator to anomalous occurrences and facilitate timely responses. Use of the system is not intended where unattended patient monitoring is desired, or in situations where arrhythmia detection is required.
The system provides the ability to transmit patient data files for storage, viewing and analysis at distributed locations via the intranet or internet, or may function as a standalone device.
| Prescription Use<br><span style="text-decoration: overline;">✓</span> | AND/OR |
|-----------------------------------------------------------------------|--------|
| (Part 21 CFR 801 Subpart D) | |
Over-The-Counter Use (21 CFR 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page Lof 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.