MAGLIFE C/MAGLIFE C PLUS

K023195 · Schiller Medical S.A. · MWI · Jun 3, 2003 · Cardiovascular

Device Facts

Record IDK023195
Device NameMAGLIFE C/MAGLIFE C PLUS
ApplicantSchiller Medical S.A.
Product CodeMWI · Cardiovascular
Decision DateJun 3, 2003
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.2300
Device ClassClass 2

Indications for Use

MAGLIFE C is intended for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N20), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) and temperature tracking in the immediate vicinity of Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations.

Device Story

Multi-parameter patient monitor for MRI environment; console placed outside magnet bore; uses non-metallic sensors and fiber optic signal transmission to prevent burns and minimize MRI interference; Faraday cage enclosure protects electronics. Inputs: ECG, SpO2, pulse rate, EtCO2, N2O, inspired CO2, respiratory rate, NIBP, IBP, anesthetic gases, FiO2, temperature. Outputs: real-time physiological data displayed on console for clinician surveillance during MRI. Benefits: enables continuous patient monitoring within MRI suite without compromising image quality or patient safety.

Clinical Evidence

Bench testing only. Validation included unit and system-level testing for compliance with IEC 601-1 and IEC 601-1-2. Testing in 1.5 Tesla MRI unit verified performance stability, lack of ferrous material movement, and no degradation of MRI image homogeneity at distances as close as 60 cm from the magnet bore.

Technological Characteristics

Multi-parameter monitor; line-powered console with Faraday cage shielding; non-metallic sensors; fiber optic signal transmission. Dimensions/form factor: console-based. Connectivity: standalone. Sterilization: N/A. Software: embedded firmware for physiological signal processing.

Indications for Use

Indicated for monitoring ECG, SpO2, pulse rate, EtCO2, N2O, inspired CO2, respiratory rate, NIBP, IBP, anesthetic gas concentrations, FiO2, and temperature in patients undergoing MRI examinations.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K023195 page 1 of 3 # 510(k) Summary for MAGLIFE C/MAGLIFE C Plus ## 1. Date this summary was prepared: February 5, 2003 #### 2. Submitter's Name and Address SCHILLER MEDICAL 4, rue Louis Pasteur ZAE Sud BP 90050 67162 WISSEMBOURG CEDEX FRANCE #### 3. Contact person M. Court GOEHRY Regulatory Affairs & Vigilance Manager SCHILLER MEDICAL 4. rue Louis Pasteur Z.A.E Sud BP 90050 67162 WISSEMBOURG CEDEX France Tel: + 33 3 88 63 36 00 Fax: + 33 3 88 94 12 82 E-mail: court.goehry@schiller.fr ## 4. Device Name Maglife C/Maglife C Plus Trade/Proprietary Name: MR Safe Patient Monitor Common Name: Clasification Names: Patient physiological monitor ## 5. . Predicate Devices The legally marketed devices to which equivalence is being claimed are: - MAGLIFE, marketed by O.D.A.M. (K950264) . - MRI-Compatible Model 9500 manufactured by Magnetic Resonance Corp. (K954120) - Omni Trak 3100 MRI manufactured by In Vivo Research (K864889) ● - . Model 3100 LUXTRON CK {1}------------------------------------------------ K023195 page 2 of 3 ## 6. Device Description The Maglife C is a multi-parameter patient monitor which is indicated for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO-), nitrous oxide concentration (%N2O), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, halothane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) and temperature tracking. The monitor consists of line-powered console (with optional back-up battery) which is placed outside the magnet boren, and appropriate electrodes, transducers, cables and tubes to allow the patient to be monitored during the examination. The console enclosure is a Faraday cage for the protection of sensitive electronic circuits. ## 7. Intended Use MAGLIFE C is intended for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N20), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) and temperature tracking in the immediate vicinity of Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations. ## 8. Comparison of Technological Characteristics All four MRI compatible patient monitors employ established measurement methods combined with non-metallic sensors and fiber optic signal transmission to minimize the risk of burns and special shielding and filtering to minimize interference to and from the MRI unit. ## 9. Nonclinical Tests Used in Determination of Substancial Equivalence The design of the Madife C has been thoroughly validated at the unit and system level. Non-clinical test were conducted to demonstrate compliance with the following standards: - IEC 601-1 ● - IEC 601-1-2 ● Test were conducted in an MRI unit to determine the maximum field strength that Maglife C can withstand without degradation of performance and to verify that the amount of ferrous material in the Maglife C is sufficiently small so that movement of the unit due to magnetic attraction is not possible. {2}------------------------------------------------ Tests were conducted to assess the effect of the Maglife C on the Homogeneity of the magnetic field inside the magnet bore. These tests, using a 1.5 Tesla MR unit, showed that the Maglife could be placed as close as 60 cm from the front of the opening from the magnet without causing any visible difference in the images. ## 10. Conclusions From Nonclinical Testing The testing of the Maglife C demonstrates that the performance is substantially equivalent to predicate devices cited above. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, formed by a series of curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN - 3 2003 Schiller Medical SA c/o Mr. Court Goehry Regulatory Affairs Specialist 4, rue Louis Pasteur Z.A.E Sud BP 90050 67162 Wissembourg Cedex France Re: K023195 Trade Name: Maglife C/Maglife C Plus Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm). Regulatory Class: Class II (two) Product Code: MWI Dated: May 2, 2003 Received: May 8, 2003 Dear Mr. Goehry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. • {4}------------------------------------------------ Page 2 - Mr. Court Goehry Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bram D. Zuckerman, M.D. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): __________ Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications For Use: The Maglife C is a multi-parameter patient monitor which is indicated for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N₂O), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired (isoflurane, halothane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) and temperature tracking in the immediate vicinity of a Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|---------| | (Division Sign-Off) | | | Division of Cardiovascular Devices | | | 510(k) Number | K023195 | | Prescription Use | X | OR | Over-The-Counter Use | |----------------------|---|----|----------------------| | (Per 21 CFR 801.109) | | | | I / I
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...