CareEvent inclusive of the CareEvent Mobile Application accessory
K161164 · Philips Medical Systems · MSX · Jun 24, 2016 · Cardiovascular
Device Facts
Record ID
K161164
Device Name
CareEvent inclusive of the CareEvent Mobile Application accessory
Applicant
Philips Medical Systems
Product Code
MSX · Cardiovascular
Decision Date
Jun 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The intended use of Philips CareEvent is to deliver supplemental medical device data associated with physiological alarms, technical alarms, clinical notifications, nurse call alarms and informational messages to a healthcare professional's end device. The user may receive visual or audible notifications, and/or other message notification types based on the communicator in use. Philips CareEvent is a component of a distributed alarming system. It does not generate the alarm, alter the behavior of the alarm-generating system, replace the alarming system that generates the alarms, nor is it intended to provide real-time information, therefore the device producing the alarm or event remains the primary notification system. Philips CareEvent provides confirmed delivery features when it is used with the Philips CareEvent mobile application on Philips-approved devices. The Philips CareEvent mobile application either communicates alarm information that is sent from the Philips CareEvent server, or the user is notified that communication is not possible. Receipt of alarm messages or events by all other external devices is not confirmed and delivery to the end device is not guaranteed.
Device Story
CareEvent is a software-based distributed alarming system component; it routes supplemental alarm data (physiological, technical, nurse call) from primary alarm-generating systems to healthcare professionals' mobile devices. It does not generate alarms, replace primary monitoring systems, or provide real-time clinical data. The system operates via a server that pushes notifications to the CareEvent mobile application on approved devices; it supports confirmed delivery features. Users receive visual/audible alerts and can perform team communication, accept/reject delegation requests, or launch the MobileCaregiver application for patient data viewing. The device benefits patients by facilitating timely notification of clinical events to caregivers, though it is not a primary life-safety alarm source.
Clinical Evidence
No clinical testing was performed. Substantial equivalence was supported by non-clinical performance testing, including risk analysis, product specifications, design reviews, and system-level verification and validation.
Technological Characteristics
Software-only device; operates on off-the-shelf PC and ITE hardware. Functions as a distributed alarm system component. Connectivity via network to primary alarm-generating systems and mobile end-devices. No specific materials or sterilization required as it is a software application.
Indications for Use
Indicated for delivery of supplemental medical device data (physiological/technical alarms, clinical notifications, nurse call alarms) to healthcare professionals. Patient population determined by the Philips patient monitors in use at the facility. Rx only.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and recognizable, representing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 24, 2016
Philips Medical Systems Theresa Poole Regulatory Specialist 3000 Minuteman Road Andover, Massachusetts 01810
Re: K161164
Trade/Device Name: CareEvent Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MSX, OUG Dated: April 14, 2016 Received: April 25, 2016
Dear Theresa Poole:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161164
Page 1 of 1
Device Name CareEvent
#### Indications for Use (Describe)
The intended use of Philips CareEvent is to deliver supplemental medical device data associated with physiological alarms, technical alarms, clinical notifications, nurse call alarms and informational messages to a healthcare professional's end device. The user may receive visual or audible notifications, and/or other message notification types based on the communicator in use.
Philips CareEvent is a component of a distributed alarming system. It does not generate the alarm, alter the behavior of the alarm-generating system, replace the alarming system that generates the alarms, nor is it intended to provide real-time information, therefore the device producing the alarm or event remains the primary notification system. Philips CareEvent provides confirmed delivery features when it is used with the Philips CareEvent mobile application on Philips-approved devices. The Philips CareEvent mobile application either communicates alarm information that is sent from the Philips CareEvent server, or the user is notified that communication is not possible. Receipt of alarm messages or events by all other external devices is not confirmed and delivery to the end device is not guaranteed.
Rx Only
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Philips logo in blue, followed by the words "CareEvent" and "Traditional 510(k)" in black. The Philips logo is a stylized version of the company's name. The text below the logo appears to be related to a product or service offered by Philips, possibly referencing a regulatory pathway for medical devices.
# 510(k) Summary Philips CareEvent release B.01
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92(c).
Date Prepared: 23 June 2016
#### l. Submitter's name and address
| Manufacturer: | Philips Medical Systems<br>3000 Minuteman Road<br>Andover, MA 01810 USA |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Theresa Poole<br>Regulatory Affairs Specialist<br>Philips Medical Systems<br>3000 Minuteman Road, MS0480<br>Andover, MA 01810-1099<br>Tel: 978 659 7621 |
Fax: 978 685 5624
### II. Device information
Device Name: CareEvent Common Name: Communication System Classification panel: 74 - Cardiovascular
Classification names are as follows:
| Classification | ProCode | Description |
|----------------|---------|--------------------------------------------------------------|
| 870.2300, II | MSX | System, Network and Communication,<br>Physiological Monitors |
| 880.6310, I | OUG | Medical Device Data System |
Email: theresa.poole@philips.com
#### III. Predicate device information
| Trade name: | CareEvent |
|----------------------------|-------------------------------------------------------------|
| Manufacturer: | Philips Medical Systems |
| 510(k) clearance: | K142935 |
| Classification name | Cardiac Monitor (Including Cardiotachometer And Rate Alarm) |
| Device class: | Class II |
| Classification regulation: | 21 CFR 892.2300 |
| Classification panel: | Cardiovascular |
| Product code: | MSX |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Philips logo in blue, followed by the text "CareEvent" and "Traditional 510(k)" in black. The Philips logo is a stylized wordmark with bold, sans-serif letters. The text "CareEvent" is smaller than the Philips logo and is positioned below and to the right of it. The text "Traditional 510(k)" is below the "CareEvent" text.
### IV.Device Description
The CareEvent solution is designed to provide healthcare professionals with supplemental information about patient alarms, technical alarms, nurse call alarms and system information messages (events).
### V. Intended use/ Indications for Use
The Philips CareEvent software application change in the intended use wording was initiated to clarify the previous predicate intended use statement. The Philips CareEvent software application release B.01 has the intended use statement listed below.
The intended use of Philips CareEvent is to deliver supplemental medical device data associated with physiological alarms, technical alarms, clinical notifications, nurse call alarms and informational messages to a healthcare professional's end device. The user may receive visual or audible notifications, and/or other message notification types based on the communicator in use.
Philips CareEvent is a component of a distributed alarming system. It does not generate the alarm, alter the behavior of the alarm-generating system, replace the alarming system that generates the alarms, nor is it intended to provide real-time information, therefore the device producing the alarm or event remains the primary notification system. Philips CareEvent provides confirmed delivery features when it is used with the Philips CareEvent mobile application on Philips-approved devices. The Philips CareEvent mobile application either communicates alarm information that is sent from the Philips CareEvent server, or the user is notified that communication is not possible. Receipt of alarm messages or events by all other external devices is not confirmed and delivery to the end device is not guaranteed.
Rx Only
| Key<br>Characteristic | Predicate Device | Subject Device | Comments |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CareEvent (K142935) | CareEvent B.01 | |
| Intended<br>Use/Indications<br>for Use<br>Sentence by<br>Sentence<br>comparison | S1: The intended use of<br>the CareEvent solution is<br>to provide healthcare<br>professionals with<br>supplemental information<br>about patient alarms,<br>technical alarms, nurse<br>call alarms and system<br>information messages<br>(events). | S1: The intended use of<br>Philips CareEvent is to<br>deliver supplemental<br>medical device data<br>associated with<br>physiological alarms,<br>technical alarms, clinical<br>notifications, nurse call<br>alarms and informational | S1: Clarification of<br>timing, what type of<br>information and to<br>where. |
| S2: The product can route<br>all of subsets of this<br>information to selective<br>remote devices such as<br>pagers, phones, or<br>marquees. | messages to a healthcare<br>professional's end device.<br>S2: The user may receive<br>visual or audible<br>notifications, and/or other<br>message notification types<br>based on the communicator<br>in use. | S2: Changed focus from<br>product function to<br>user experience.<br>Clarified types of<br>notifications. Stated it<br>is dependent on end<br>device in use<br>(communicator)<br>aligned with IEC 80001<br>terminology. | |
| S3-5: Receipt of alarm<br>messages or events by the<br>external device, is not<br>confirmed and delivery to<br>the end device is not<br>guaranteed.<br>The primary alarm<br>notification is the device<br>producing the alarm or<br>event.<br>This product line is not<br>intended to provide real-<br>time information, nor is it<br>a source of patient alarms,<br>nor is it a replacement for<br>alarming devices. | S3-5: Philips CareEvent is a<br>component of a distributed<br>alarming system.<br>It does not generate the<br>alarm, alter the behavior of<br>the alarm-generating<br>system, replace the<br>alarming system that<br>generates the alarms, nor is<br>it intended to provide real-<br>time information, therefore<br>the device producing the<br>alarm or event remains the<br>primary notification system.<br>Philips CareEvent provides<br>confirmed delivery features<br>when it is used with the<br>Philips CareEvent mobile<br>application on Philips-<br>approved devices. | S3-5: Removed entirely<br>S3 from Predicate<br>Intended use.<br>Replaced with<br>clarification of role<br>CareEvent plays in a<br>Distributed alarm<br>system.<br>Removed term<br>secondary used in<br>Predicate, it is not a<br>standardized term.<br>S6: Removed<br>redundancy of the | |
| | S6: The CareEvent mobile | S6: The Philips CareEvent | predicate intended use |
| | application software | mobile application either | for the CareEvent |
| | provides healthcare | communicates alarm | mobile application. |
| | professionals with | information that is sent | |
| | supplemental information | from the Philips CareEvent | |
| | about patient alarms, | server, or the user is | Replaced it with clear |
| | technical alarms, nurse | notified that communication | definition of the role |
| | call alarms and system | is not possible. | the mobile application |
| | information messages | | plays in the distributed |
| | (events). Receipt of alarm | | alarm system and |
| | messages or events by the | | confirmed delivery. |
| | external mobile device is | | |
| | not guaranteed. The | | |
| | primary alarm notification | | |
| | is the device producing | | |
| | the alarm or event. This | | S7: Added from |
| | product line is not | S7: Receipt of alarm<br>messages or events by all | predicate device to |
| | intended to provide real- | other external devices is not | clarify that only Philips |
| | time information, nor is it | confirmed and delivery to | approved devices are<br>different from the |
| | a source of patient alarms, | the end device is not | predicate. |
| | nor is it a replacement for | guaranteed. | |
| | alarming devices. | | |
| | | | |
| | | | Same |
| | Rx Only | Rx Only | |
| | | | |
| Target Patient<br>Population | As determined by the<br>Philips patient monitors in<br>use at the facility. | Same | |
| Users | Trained health care<br>professionals | Same | |
| | | | |
| Device<br>Software | Unchanged, as previously<br>cleared | 1. Allow one-way context<br>sharing from CareEvent<br>to MobileCaregiver. | 1. When in CareEvent<br>Mobile Application,<br>allows user to open<br>the Mobile<br>CareGiver<br>Application, if |
| | | 2. Text messaging<br>capability within the<br>CareEvent system via<br>the CareEvent mobile<br>application.<br>3. Add user button on the<br>CareEvent mobile<br>application to<br>Accept/Reject a<br>delegation request<br>when received. | installed on the<br>same mobile<br>device. Mobile<br>CareGiver is a view<br>of data from the<br>PIIC iX central<br>station.<br>2. Provides users an<br>alternative for<br>team<br>communication.<br>3. Enables additional<br>user control, if a<br>user is not able to<br>accept ownership<br>of alerts from<br>another device,<br>they can<br>intentionally reject<br>the request. |
| Technological<br>Characteristics | Unchanged, as previously<br>cleared. | Same | |
| Device Hardware<br>is specified by<br>Philips | Off the Shelf PC and other<br>ITE equipment, specification<br>available or can be ordered<br>from Philips. | Same. | |
| Device<br>Accessories | CareEvent Mobile<br>Application | Same | |
| Care and<br>Cleaning | Not applicable, software<br>device. | Same | |
| Device safety<br>and<br>environmental<br>specifications | Not applicable, software | Same |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.