K142935 · Philips Medical Systems · MSX · Jan 15, 2015 · Cardiovascular
Device Facts
Record ID
K142935
Device Name
CareEvent
Applicant
Philips Medical Systems
Product Code
MSX · Cardiovascular
Decision Date
Jan 15, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The intended use of the CareEvent solution is to provide healthcare professionals with supplemental information about patient alarms, technical alarms, nurse call alarms and system information messages (events). The product can route all of subsets of this information to selective remote devices such as pagers, phones, or marquees. Receipt of alarm messages or events by the external device, is not confirmed and delivery to the end device is not guaranteed. The primary alarm notification is the device producing the alarm or event. This product line is not intended to provide real-time information. nor is it a source of patient alarms, nor is it a replacement for alarming devices. The CareEvent mobile application software provides healthcare professionals with supplemental information about patient alarms, technical alarms, nurse call alarms and system information messages (events). Receipt of alarm messages or events by the external mobile device is not guaranteed. The primary alarm notification is the device producing the alarm or event. This product line is not intended to provide real-time information, nor is it a source of patient alarms, nor is it a replacement for alarming devices.
Device Story
CareEvent is a notification management system for healthcare environments; it aggregates patient alarms, technical alarms, and nurse call events from monitoring systems. The device routes these events to remote secondary devices, including pagers, phones, marquees, and a mobile application. It functions as a supplemental information tool; it does not provide real-time data and is not a substitute for primary bedside alarm systems. Delivery of notifications to external devices is not guaranteed. Healthcare professionals use the system to receive remote alerts, allowing for awareness of patient status or system issues while away from the primary monitor. The device does not perform clinical analysis or diagnostic functions; it serves as a communication conduit for existing alarm infrastructure.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level tests, performance tests, and safety testing based on hazard analysis. Pass/Fail criteria were established based on predicate device specifications; results demonstrated that the device meets all defined reliability requirements and performance claims.
Technological Characteristics
Networked communication system for physiological monitors. Software-based event routing to remote devices (pagers, phones, marquees) and mobile applications. Operates as a secondary notification layer. No specific hardware materials or energy sources described beyond standard network infrastructure requirements.
Indications for Use
Indicated for healthcare professionals to receive supplemental information regarding patient alarms, technical alarms, nurse call alarms, and system events via remote devices (pagers, phones, marquees) or mobile applications. Not for real-time use; not a replacement for primary alarming devices.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2015
Philips Medical Systems Theresa Poole Regulatory Affairs Specialist 3000 Minuteman Road Andover, Massachusetts 01810
Re: K142935 Trade/Device Name: CareEvent Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX Dated: December 15, 2014 Received: December 16, 2014
Dear Theresa Poole,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known):
CareEvent Device Name:
Indications for Use:
The intended use of the CareEvent solution is to provide healthcare professionals with supplemental information about patient alarms, technical alarms and system information messages (events). The product can route all of subsets of this information to selective remote devices such as pagers, phones, or marquees. Receipt of alarm messages or events by the external device, is not confirmed and delivery to the end device is not guaranteed. The primary alarm notification is the device producing the alarm or event. This product line is not intended to provide real-time information, nor is it a source of patient alarms, nor is it a replacement for alarming devices.
The CareEvent mobile application software provides healthcare professionals with supplemental information about patient alarms, technical alarms and system information messages (events). Receipt of alarm messages or events by the external mobile device is not guaranteed. The primary alarm notification is the device producing the alarm or event. This product line is not intended to provide real-time information, nor is it a source of patient alarms, nor is it a replacement for alarming devices.
Prescription Use Yes (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use No (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR §807.92.
The submitter of this premarket notification is:
Theresa Poole Regulatory Affairs Specialist Patient Monitoring Philips Medical Systems 3000 Minuteman Road. MS0480 Andover, MA 01810-1099 Tel: 978 659 7621 Fax: 978 685 5624 Email: claire.arakaki@philips.com
This summary was prepared on 15 January 2015.
The name of this device is the CareEvent.
## Device Description
The CareEvent solution is designed to provide healthcare professionals with supplemental information about patient alarms, technical alarms, nurse call alarms and system information messages (events).
Classification names are as follows:
| Device Panel | Classification | ProCode | Description |
|----------------------------|----------------|---------|-----------------------------------------------------------------|
| Panel 74<br>Cardiovascular | \$870.2300, II | MSX | System, Network and<br>Communication, Physiological<br>Monitors |
The CareEvent solution is substantially equivalent to the previously cleared predicate IntelliSphere Event Management (K102974).
## Intended Use
The intended use of the CareEvent solution is to provide healthcare professionals with supplemental information about patient alarms, technical alarms, nurse call alarms and system information messages (events). The product can route all of subsets of this information to selective remote devices such as pagers, phones, or marquees. Receipt of alarm messages or events by the external device, is not confirmed and delivery to the end device is not guaranteed. The primary alarm notification is the device producing the alarm or event. This product line is not intended to provide real-time information. nor is it a source of patient alarms, nor is it a replacement for alarming devices.
The CareEvent mobile application software provides healthcare professionals with supplemental information about patient alarms, technical alarms, nurse call alarms and system information messages (events). Receipt of alarm messages or events by the external mobile device is not guaranteed. The primary alarm notification is the device producing the alarm or event. This
{4}------------------------------------------------
product line is not intended to provide real-time information, nor is it a source of patient alarms, nor is it a replacement for alarming devices.
The device has the same Indications for Use and Intended Use as the legally marketed predicate device.
The technological characteristics are the same as the legally marketed predicate device.
Software changes have been made to accommodate the communication to the CareEvent mobile application accessory.
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the device meets all defined reliability requirements and performance claims.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.