DATEX-OHMEDA S/5 DEVICE INTERFACING SOLUTION, N-DISXXXX..01 WITH N-DISVENT..02
K070421 · GE Healthcare · MSX · Mar 13, 2007 · Cardiovascular
Device Facts
Record ID
K070421
Device Name
DATEX-OHMEDA S/5 DEVICE INTERFACING SOLUTION, N-DISXXXX..01 WITH N-DISVENT..02
Applicant
GE Healthcare
Product Code
MSX · Cardiovascular
Decision Date
Mar 13, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Datex-Ohmeda S/5 Device Interfacing Solution, DIS, is intended to be used with a Datex-Ohmeda monitoring systems for transferring data between external devices and a monitor. The Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx.01 with N-DISVENT..02, is indicated for data transfer between stand-alone monitors, ventilators/anesthesia machines, blood gas analyzers, and heart-lung machines and Datex-Ohmeda bedside monitors for displaying and patient care information purposes. The device is indicated for use by qualified medical personnel only.
Device Story
Module transfers data between external medical devices (ventilators, anesthesia machines, blood gas analyzers, stand-alone monitors) and Datex-Ohmeda S/5 bedside monitors. System uses plug-in converter modules connected via bus cables; supports up to ten simultaneous interfaces. Inputs include numerical, waveform, and event data; outputs include display on monitor, trending, printing, and network transmission to central stations. Does not perform physiological measurements; does not transfer alarms, but transmits ventilator alarm status events, patient demographics, and gas consumption data. Used in clinical settings by qualified medical personnel. Facilitates centralized patient data management and record keeping.
Clinical Evidence
Bench testing only. Device verified through validation and verification of specifications against standards including EN 60601-1, IEC 60601-1-2, and EN 60529.
Technological Characteristics
Module-based data converter system. Connects via bus cables (max 10m). Supports up to 10 modules. Interfaces with Datex-Ohmeda S/5 monitor frames (F-CU8, F-CU5(P)). Standards: EN 60601-1 (safety), IEC 60601-1-2 (EMC), EN 60529 (IPX1 ingress protection).
Indications for Use
Indicated for data transfer between stand-alone monitors, ventilators/anesthesia machines, blood gas analyzers, heart-lung machines, and Datex-Ohmeda bedside monitors for patient care information. For use by qualified medical personnel only.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Page 1 of 4
K070421
MAR 1 3 2007
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01 with N-DISVENT.02
# GENERAL COMPANY INFORMATION as required by 807.92(a){1}
## COMPANY NAME/ADDRESS/PHONE/FAX:
GE Healthcare 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
# NAME OF CONTACT:
Mr. Joel Kent
DATE:
February 3, 2007
## DEVICE NAME as required by 807.92(a)(2)
# TRADE NAME:
Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01 with N-DISVENT..02
## COMMON NAME:
Medical device data converter
### CLASSIFICATION NAME:
### The following Class II classification appears applicable:
Product Code Classification Name System, Network And Communication, MSX Physiological Monitors
CFR Section 870.2300
# NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The S/5TM Device Interfacing Solution, N-DISxxxx.01 with N-DISVENT..02 is substantially equivalent in safety and effectiveness to the legally marketed (predicate) S/5™ Device Interfacing Solution, N-DISxxxx..01 (K051634).
{1}------------------------------------------------
# DEVICE DESCRIPTION as required by 807.92(a)(4)
S/5 Device Interfacing Solution (later referred to as DIS) is a module that transfers data between an external device and S/5 monitor.
External devices are connected to the monitoring system by using small plug-in converter modules that handle the communication between the device and S/5 monitoring system. These DIS converter modules convert the data coming from the connected device to a format that can be utilized in the Datex-Ohmeda S/5 Anesthesia Monitor, Datex-Ohmeda S/5 Critical Care Monitor, Datex-Ohmeda S/5 Compact Anesthesia Monitor, Datex-Ohmeda S/5 Compact Critical Care Monitor, or Datex-Ohmeda S/5 FM.
The use of a DIS system consists of making the physical connections connecting external devices to DIS and linking DIS modules together to make a complete bus. Then the DIS transfers data between a device and the S/5 monitoring system. The user can then select the source of measurement data for physiologic parameters displayed on the Datex-Ohmeda monitor.
The first DIS converter module is connected to the socket at the Datex-Ohmeda S/5 monitor. In S/5 Anesthesia Monitor and S/5 Critical Care Monitor the DIS socket is located at the rear of the monitor in the F-CU8 monitor frame option and at the front of the monitor in the F-CU5(P) monitor frame option. In S/5 Compact Anesthesia Monitor, S/5 Compact Critical Care Monitor, and S/5 FM monitor the DIS socket is located in left hand side of the monitor. Additional DIS converter modules in a system are connected to each other with the bus cable. The external device is connected to the DIS converter module with a device specific cable.
Up to ten DIS converter modules can be connected in the system. The maximum number of simultaneous interfaces is ten. The maximum length of interface cable is 10 meters (33ft). The number of DIS interfaces that can be used depend on the length of the interface cables and the particular monitor used.
The Device Interfacing Solution supports interfacing of the following device categories: ventilators/anesthesia machines, stand-alone monitors, blood gas analyzers and heart-lung machines. The Device Interfacing Solution can interface numerical, waveform and event type of data from the external device. Alarms are not transferred but alarm status events are transferred.
Interfaced data can be displayed on the monitor screen, trended, printed and used for record keeping purposes. Also, interfaced physiologic data is sent to the network to be viewed at the Central station monitor.
#### INTENDED USE as required by 807.92(a)(5)
#### Intended Use:
Intended use:
The Datex-Ohmeda S/5 Device Interfacing Solution, DIS, is intended to be used with a Datex-Ohmeda monitoring systems for transferring data between external devices and a monitor.
#### Indications for use:
The Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx.01 with N-DISVENT..02, is indicated for data transfer between stand-alone monitors, ventilators/anesthesia machines, blood gas analyzers, and heart-lung machines and Datex-Ohmeda bedside monitors for displaying and patient care information purposes. The device is indicated for use by qualified medical personnel only.
{2}------------------------------------------------
# SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The S/5™ Device Interfacing Solution, N-DISxxxx..01 with N-DISVENT..02 is substantially equivalent in safety and effectiveness to the legally marketed (predicate) S/5™ Device Interfacing Solution, N-DISxxxx... 01 (K051634).
The N-DISxxxx..01 with N-DISVENT..02 like the predicate N-DISxxxx..01 (K051634) does not perform any physiological measurements itself. It transfers data between connected external devices and S/5™ monitoring system. The Datex-Ohmeda S/5™ monitor displays, trends and uses data for calculations and transfers it to the record keeping system and network accordingly. The N-DISxxxx..01 with N-DISVENT..02, in terms of general function is identical to its S/5™ Device Interfacing Solution, N-DISxxxx..01 (K051634). The N-DISxxxx..01 with N-DISVENT..02 simply extends capability of Interfacing Solution to the Datex-Ohmeda S/5TM monitor by introducing an enhanced version of the N-DISVENT. Neither the Device Interfacing Solution, N-DISxxxx..01 with N-DISVENT..02 nor the predicate N-DISxxxx..01 (K051634) transfer alarms. A new feature of the N-DISVENT..02 module is that it can transfer ventilator alarm status events (e.g. "FiO2 low", "Replace O2 sensor", "Inspiration stopped", "MVexp low", "Circuit leak", "RR low", "FiCO2 low. Absorbent OK?", "FiSEV high", "Alternate O2, Aisys", etc.) to the monitor. The other new features of the N-DISVENT..02 are: patient demographics data transfer between the monitor and the ventilator, gas consumption event data transfer from the ventilator to the monitor.
The main differences between the new N-DISxxxx..01 with N-DISVENT..02 and the predicate N-DISxxxx..01 (K0516341) are as follows:
- Demographics data of patient (Patient ID, weight, height and BSA) is transferred between 1. the S/5 Monitor and the external device with N-DISVENT..02.
- Ventilator alarm status events are transferred from the ventilator through 2. N-DISVENT..02 and S/5 Monitor to the network.
- Ventilator gas consumption event data is transferred from the ventilator through N-3. DISVENT..02 and S/5 Monitor to the network after the patient case has been ended at the ventilator.
- 4. Modifications to Labeling:
Changes to the labeling include revised Installation Guides, and Brochure for S/5 Device Interfacing Solution. The N-DISA2000 Installation Guide is updated with compatibility information of BIS Monitoring System's software BIS Engine version 3.31 and BIS Vista Monitoring System's version 1.01. Uset's Guides for Anesthesia, Critical Care and Flexible Monitors are essentially same. The User's Guide for S/5 Anesthesia Monitor has been updated with additional safety precautions, warnings and corrections.
Based on the analysis and other documentation included in this 510(k) notification and attachments it is evident that the main features and indications for use of the S/5™ Device Interfacing Solution, N-DISxxxx..01 with N-DISVENT..02 is substantially equivalent to the predicate S/57M Device Interfacing Solution, N-DISxxxxx.01 (K051634).
{3}------------------------------------------------
# SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01 with N-DISVENT..02 has been assessed against the standards below and details of conformity are presented in the attached 510(k) notification. The device has been thoroughly tested through validation and verification of specifications.
- . EN 60601-1:1990 + Amdt 1:1993 + Amdt 2:1995 + Amdt 3:1996 Medical electrical equipment Part 1: General requirements for safety
- IEC 60601-1:1988 +Amdt 1:1991 + Amdt 2:1995 .
- CAN/CSA C22.2 No. 601.1-M90 + S1:1994 + Amdt 2:1998 .
- UL 2601-1,October 24, 1997 .
- IEC 60601-1-2:2001, Electromagnetic compatibility Requirements and tests .
- . EN 980: 1996, Terminology, symbols and information provided with medical devices- Graphical symbols for use in the labeling of medical devices
- . EN 1041 1998, Terminology, symbols and information provided with medical devices; information supplied by the manufacturer with medical devices.
- . Protection against ingress of liquid: EN 60529 (IPX1): 1992
### CONCLUSION:
The summary above shows that the S/5™ Device Interfacing Solution, N-DISxxxx..01 with N-DISVENT.. 02 is substantially equivalent in safety and effectiveness to the legally marketed (predicate) S/5TM Device Interfacing Solution, N-DISxxxx..01 (K051634).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an abstract eagle design with three stylized lines forming the body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" are arranged in a circular pattern around the top of the seal. The seal is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Joel C. Kent Manager, Quality and Regulatory Affairs GE Healthcare 86 Pilgrim Road Needham, Massachusetts 02492
MAR 1 3 2007
Re: K070421
Trade/Device Name: Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01 with N-DISENT..02 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: II Product Code: MSX Dated: February 4, 2007 Received: February 13, 2007
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
# Page 2 - Mr. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Clues
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): _____________________
Device Name: Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01 with N-DISVENT..02
Indications for Use:
The Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01 with N-DISVENT..02, is indicated for data transfer between stand-alone monitors, ventilators/anesthesia machines, blood gas analyzers, and heart-lung machines and Datex-Ohmeda bedside monitors for displaying and patient care information purposes.
The device is indicated for use by qualified medical personnel only.
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | |
|------------------------|--|
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| | Page |
| | |
| | K070421 |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.