The Datex-Ohmeda S/5 Device Interfacing Solution, DIS, is intended to be used with a Datex-Ohmeda monitoring systems for transferring data between external devices and a monitor.
Device Story
Device Interfacing Solution (DIS) consists of plug-in converter modules connecting external medical devices (ventilators, anesthesia machines, stand-alone monitors, blood gas analyzers, heart-lung machines) to Datex-Ohmeda S/5 monitoring systems. DIS modules convert external device data into a format compatible with S/5 monitors; data is transferred via a bus cable system (up to 10 modules). DIS transfers numerical, waveform, and event data for display, trending, and record-keeping on the bedside monitor or central station. It does not perform physiological measurements or transfer alarms. A specific module (N-DISVENT) enables time synchronization between the monitor and the Datex-Ohmeda AiSys Carestation. Used in clinical settings by qualified medical personnel; facilitates centralized patient data management and clinical decision-making by integrating disparate device outputs into a single monitoring interface.
Clinical Evidence
Bench testing only. Device verified against EN 60601-1, IEC 60601-1-2, EN 980, EN 1041, and EN 60529 standards for safety, electromagnetic compatibility, and ingress protection.
Technological Characteristics
Modular data converter system; connects via bus cables. Supports 15Vdc and 8Vdc power. Connectivity: proprietary bus interface to S/5 monitors. Software: embedded firmware for data conversion and time synchronization. Standards: EN 60601-1, IEC 60601-1-2, EN 60529 (IPX1).
Indications for Use
Indicated for data transfer between stand-alone monitors, ventilators/anesthesia machines, blood gas analyzers, and heart-lung machines and Datex-Ohmeda bedside monitors for displaying and patient care information purposes. For use by qualified medical personnel only.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
JUL 1 9 2005
## Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01
## GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
#### COMPANY NAME/ADDRESS/PHONE/FAX:
GE Healthcare 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
#### NAME OF CONTACT:
Mr. Joel Kent
#### DATE:
June 17, 2005
## DEVICE NAME as required by 807.92(a)(2)
#### TRADE NAME:
Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01.
## COMMON NAME:
Medical device data converter
## CLASSIFICATION NAME:
## The following Class II classification appears applicable:
| Product Code | Classification Name | CFR Section |
|--------------|--------------------------------------------------------------|-------------|
| MSX | System, Network And Communication, Physiological<br>Monitors | 870.2300 |
## NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The S/5TM Device Interfacing Solution, N-DISxxxx..01 is substantially equivalent in safety and effectiveness to the legally marketed (predicate) S/5TM Device Interfacing Solution, N-DISxxxx..00 (K012531).
{1}------------------------------------------------
#### DEVICE DESCRIPTION as required by 807.92(a)(4)
External devices are connected to the monitoring system by using small plug-in converter modules that handle the communication between the device and S/5 monitoring system. These DIS converter modules convert the data coming from the connected device to a format that can be utilized in the Datex-Ohmeda S/5 Anesthesia Monitor, Datex-Ohmeda S/5 Critical Care Monitor, Datex-Ohmeda S/5 Compact Anesthesia Monitor. Datex-Ohmeda S/5 Compact Critical Care Monitor, or Datex-Ohmeda S/5 FM. The use of a DIS system consists of making the physical connections connecting external devices to DIS and linking DIS modules together to make a complete bus. Then the DIS transfers data between a device and the S/5 monitoring system. The user can then select the source of measurement data for physiologic parameters displayed on the Datex-Ohmeda monitor. The first DIS converter module is connected to the socket at the Datex-Ohmeda S/5 monitor. In S/5 Anesthesia Monitor, and S/5 Critical Care Monitor the DIS socket is located at the rear of the monitor and in \$/5 Compact Anesthesia Monitor, \$/5 Compact Critical Care Monitor, and S/5 FM monitor it is located in left hand side of the monitor. Additional DIS converter modules in a system are connected to each other with the bus cable. The external device is connected to the DIS converter module with a device specific cable. Up to ten DIS converter modules can be connected in the system. The maximum number of simultaneous interfaces is ten. The maximum length of interface cable is 10 meters (33ft). The number of DIS interfaces that can be used depend on the length of the interface cables and the particular monitor used. The Device Interfacing Solution supports interfacing of the following device categories: ventilators/anesthesia machines, stand-alone monitors, blood gas analyzers and heart-lung machines. The Device Interfacing Solution can interface numerical, waveform and event type of data from the external device. Alarms are not transferred. Interfaced data can be displayed on the monitor screen, trended, printed and used for record keeping purposes. Also, interfaced physiologic data is sent to the network to be viewed at the Central station monitor.
#### INTENDED USE as required by 807.92(a)(5)
#### Intended Use:
. .
The Datex-Ohmeda S/5 Device Interfacing Solution, DIS, is intended to be used with a Datex-Ohmeda monitoring systems for transferring data between external devices and a monitor.
### Indications for use:
The Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01, is indicated for data transfer between stand-alone monitors, ventilators/anesthesia machines, blood gas analyzers, and heartlung machines and Datex-Ohmeda bedside monitors for displaying and patient care information purposes. The device is indicated for use by qualified medical personnel only.
{2}------------------------------------------------
## SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The S/5™ Device Interfacing Solution, N-DISxxxx..01 is substantially equivalent in safety and effectiveness to the legally marketed (predicate) S/5TM Device Interfacing Solution, N-DISxxxx..00 (K012531).
The S/5TM Device Interfacing Solution (later referred to as DIS) is an additional part of a Datex-Ohmeda modular S/5TM monitoring system. The Device Interfacing Solution enables connection of external medical bedside devices such as stand-alone monitors, ventilators/anesthesia machines, blood gas analyzers and heart-lung machines to the monitoring systems.
The S/5™ Device Interfacing Solution is a module that can be placed close to the interfaced external device. The module is device specific: a separate DIS module is needed to interface each external device. The DIS is able to transfer numerical, waveform and event type of data such as ventilator settings, laboratory results and heart-lung machine settings.
The N-DISxxxx..01 like the predicate N-DISxxxx..00 (K12531) does not perform any physiological measurements itself. It transfers data between connected external devices and S/5™ monitoring system. The Datex-Ohmeda S/5™ monitor displays, trends and uses data for calculations and transfers it to the record keeping system and network accordingly. The N-DISxxxx..01, in terms of general function is identical to its S/5™ Device Interfacing Solution, N-DISxxxx..00 (K012531). The N-DISxxxx..01 simply extends capability of Interfacing Solution to the Datex-Ohmeda S/5TM monitor. Neither the Device Interfacing Solution, N-DISxxxx..01 nor the predicate N-DISxxxx..00 transfer alarms. A new feature of the N-DISVENT module is that it can transfer a current clock time from the monitor to the Datex-Ohmeda AiSys Carestation (only this ventilator). This synchronization is done to have same clock time both in the ventilator and monitor. The same clock time in both devices is important for the record keeping purposes. This clock time transfer (time synchronization) is done only when there is no case open in the the monitor/ventilator. The comparison above as well as supporting data and analysis shows that there are no new questions of safety and effectiveness for the S/5™ Device Interfacing Solution or accessories.
The N-DISxxxx..01 has following similarities compared to the predicate N-DISxxxx..00 (K012531).
- ·Identical intended use and nearly identical indications for use .
- . ·The same (improved) electronic board
- . ·Essentially the same software.
. 1
- ·Uses same operating principle .
- . ·Have same user interface at the monitor
- ·Have the same safety and effectiveness
- · Are manufactured using the same processes .
The main differences between the new N-DISxxxx..01 and the predicate N-DISAxxxx..00 (K0124531) are as follows:
- · Electronic board uses various voltage comparing the old board. With FM monitor DIS . uses a 15Vdc power supply instead of old DIS modules which use 8Vdc. New board supports both voltages and automatically selects the correct voltage.
- . · A new N-DISVENT module can communicate also with other Datex-Ohmeda ventilators and it will replace the N-DISAEST module,
- . ·N-DISVENT can transfer time from monitor to ventilator (only with Datex-Ohmeda AiSys Carestation). This is used for time synchronization.
Based on the analysis and other documentation included in this 510(k) notification and attachments it is evident that the main features and indications for use of the S/5™ Device Interfacing Solution, N-DISxxxx.01 is substantially equivalent to the predicate S/5TM Device Interfacing Solution, N-DISxxxx..00 (K012531).
{3}------------------------------------------------
## SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01. has been assessed against the standards below. The device has been thoroughly tested through validation and verification of specifications.
- EN 60601-1:1990 + Amdt 1:1993 + Amdt 2:1995 + Amdt 3:1996 Medical electrical equipment Part . 1: General requirements for safety
- IEC 60601-1:1988 +Amdt 1:1991 + Amdt 2:1995 .
- . CAN/CSA C22.2 No. 601.1-M90 + S1:1994 + Amdt 2:1998
- . UL 2601-1,October 24, 1997
- . IEC 60601-1-2:2001
- Electromagnetic compatibility Requirements and tests .
- EN 980: 1996 Terminology, symbols and information provided with medical devices- Graphical . symbols for use in the labeling of medical devices
- EN 1041 1998 Terminology, symbols and information provided with medical devices; information . supplied by the manufacturer with medical devices.
- . Protection against ingress of liquid: EN 60529 (IPX1):1992
#### CONCLUSION:
. .
The summary above shows that there are no new questions of safety and effectiveness for the S/5TM Device Interfacing Solution, N-DISxxxx..01 and it is substantially equivalent in safety and effectiveness to the legally marketed (predicate) S/5TM Device Interfacing Solution, N-DISxxxx.00 (K012531).
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a stylized eagle emblem. The eagle is depicted with three lines forming its body and wings, and two curved lines representing its legs.
JUL 1 9 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
GE Healthcare c/o Mr. Joel C. Kent Manager, Quality and Regulatory Affairs 86 Pilgrim Road Needham, MA 02492
Re: K051634
Trade Name: Datex-Ohmeda S/5 Device Interfacing Solution N-DISxxxx..01 Regulation Number: 21 CFR 870.2300 Regulation Name: Physiological Network and Communication System Regulatory Class: Class II (two) Product Code: MSX Dated: June 17, 2005 Received: June 20, 2005
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Mr. Joel C. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loast be advisou that I Dr wintion that your device complies with other requirements of the Act that I Dri has intacts and regulations administered by other Federal agencies. You must or any I cacal statuted and registements, including, but not limited to: registration and listing (21 Comply with an the 110 - 510 and manufacturing practice requirements as set CFR Part 8077, adoning (21 CFR Part 820); and if applicable, the electronic form in the qualis (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation come of pro herem (evice as described in your Section 510(k) rins letter wilf anow your to organization of substantial equivalence of your device to a legally prematice notification: "The President on for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acreoliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Whoording of reservir on your responsibilities under the Act may be obtained from the Other general mionmation on Journational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bfimmermon for
S. R. Zimmerman, M.D.
Bram-D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): __ K05 | 634
Device Name: Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01
Indications for Use:
The Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx..01, is indicated for data transfer between stand-alone monitors, ventilators/anesthesia machines, blood gas analyzers, and heart-lung machines and Datex-Ohmeda bedside monitors for displaying and patient care information purposes.
The device is indicated for use by qualified medical personnel only.
Over-The-Counter Use _ X Prescription Use __ AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (Please do not write below this line-continue on another page of neëoed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ of ___________________________________________________________________________________________________________________________________________________________________
B.H. Harrison
(Division Sign-OHJ
Division of Cardiovascular Devices
510(k) Number K05/454