MULTI-LINK CABLE AND LEAD WIRE SYSTEMS

K980582 · Marquette Medical Systems, Inc. · DSA · Mar 16, 1998 · Cardiovascular

Device Facts

Record IDK980582
Device NameMULTI-LINK CABLE AND LEAD WIRE SYSTEMS
ApplicantMarquette Medical Systems, Inc.
Product CodeDSA · Cardiovascular
Decision DateMar 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2900
Device ClassClass 2

Indications for Use

Multi-Link Cable and Lead Wire Systems are reusable electrode cable systems used to transmit signals from patient electrodes to various electrocardiograph recorders / monitors for both diagnostic and monitoring purposes. Multi-Link Cable and Lead Wire Systems are limited by the indications for use of the connected monitoring or diagnostic equipment is commonly located in hospitals, doctor's offices, and emergency vehicles, as well as in home use.

Device Story

Multi-Link Cable and Lead Wire Systems are reusable electrode cables; transmit electrical ECG signals from patient surface electrodes to electrocardiograph recorders/monitors. Used in hospitals, clinics, emergency vehicles, and home settings; operated by healthcare professionals or patients. Device acts as a passive conduit for physiological signals; does not perform signal processing or analysis itself. Output is raw ECG signal provided to host monitor for clinical interpretation. Benefits include reliable signal transmission for diagnostic and monitoring purposes.

Clinical Evidence

Bench testing only. Comparison of electrical and mechanical performance against AAMI ECGC-5/83 standards and 21 CFR Part 898 requirements demonstrated safety and effectiveness equivalent to the predicate device.

Technological Characteristics

Reusable electrode cable and lead wire system. Passive signal transmission. Compliant with 21 CFR Part 898 performance standards. Tested per AAMI ECGC-5/83 for electrical and mechanical integrity.

Indications for Use

Indicated for patients requiring ECG signal transmission from surface electrodes to diagnostic or monitoring equipment. Use is restricted by the indications of the connected host device. Suitable for clinical settings including hospitals, doctor's offices, emergency vehicles, and home environments.

Regulatory Classification

Identification

A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K980582 ## Section 2 Summary and Certification 510(k) Summary Date: February 13, 1998 | 1. Establishment Information | Manufacturer/Submitter: Marquette Medical Systems<br>100 Marquette Drive<br>Jupiter, FL 33468 | |------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Registration Number: 1051778 | | | Contact Name/Phone #: Michael Laughran/ (561) 575-5000 | | 2. | General Device Information | | | Common/Usual Name: ECG Lead Wire and Cable System | | | Trade/Proprietary Name: Multi-Link Cable and Lead Wire Systems | | | Classification Name: Patient transducer and electrode cable (including connector) | | | Device Classification: Class II. | | | Performance Standards: 21CFR Part 898: Performance Standard for Electrode Lead Wire an<br>Patient Cables. This Standard is effective on May 11, 1998 and was | - Substantial Equivalence: Multi-Link Cable and Lead Wire Systems is substantially equivalent to ConMed 3. Corp. ECG Patient Cables and Leadwires which are currently legally marketed under 510(k) 933649. published in the Friday, May 9, 1997 Federal Register. - Device Description: Multi-Link Cable and Lead Wire Systems are a reusable electrode cable systems used 4. to transmit signals from patient electrodes to various electrocardiograph recorders / monitors for both diagnostic and monitoring purposes. This type of device is common to both the industry and to most medical establishments. - Intended Use: Multi-Link Cable and Lead Wire Systems are reusable electrode cable systems used to transmit ട. signals from patient electrodes to various electrocardiograph recorders / monitors for both diagnostic and monitoring purposes. Multi-Link Cable and Lead Wire Systems are limited by the indications for use of the connected monitoring or diagnostic equipment is commonly located in hospitals, doctor's offices, and emergency vehicles, as well as in home use. - Technological comparison to legally marketed predicate device: the Multi-Link Cable and Lead Wire 6. Systems technological characteristics are similar in comparison to the ConMed Corp. ECG Patient Cables and Leadwires - 7. Test Summary and Conclusion: A comparison of the predicate device's production and both electrical and mechanical testing to AAMI ECGC-5/83 was made to the equivalent production, electrical and mechanical testing of the Multi-Link Cable and Lead Wire Systems. Testing also demonstrated that the Multi-Link Cable and Lead Wire Systems meet 21CFR Part 898: Performance Standard for Electrode Lead Wire and Patient Cables. Marquette Medical Systems has demonstrated that the use of the Multi-Link Cable and Lead Wire Systems is as safe and as effective, and performs substantially equivalent to its predicate device. MAR 1 6 1998 {1}------------------------------------------------ Image /page/1/Picture/2 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the department's name around the perimeter. In the center of the seal is an image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 MAR 16 1998 Mr. Michael Laughran Marquette Medical Systems 100 Marquette Drive Jupiter, FL 33458 Re: K980582 Multi-Link Cable and Lead Wire Systems Requlatory Class: II (two) Product Code: 74 DSA February 13, 1998 Dated: Received: February 17, 1998 Dear Mr. Laughran: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set - forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ #### Page 2 - Mr. Michael Laughran This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4649. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Section 13 Indication for Use Statement ### INDICATIONS FOR USE PAGE 510(K) Number (if known): Unknown - 510(k) filed February 13, 1998 Device Name: Multi-Link Cable and Lead Wire Systems Indications for use: Multi-Link Cable and Lead Wire Systems are electrode cable systems used to transmit signals from patient surface electrodes to various electrocardiograph recorders/monitors for both diagnostic and monitoring purposes. Use is limited by the indications for use of the connected monitoring or diagnostic equipment. Such equipment is commonly located in hospitals, doctors offices, emergency vehicles, as well as in home use. #### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ### Concurrence of CDRH, Office of device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over- The- Counter Use ![.](Image) (Division Sign-Off)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...