EWC: ECG CABLE AND LEADWIRE SYSTEMS

K982817 · Ewc · DSA · Sep 23, 1998 · Cardiovascular

Device Facts

Record IDK982817
Device NameEWC: ECG CABLE AND LEADWIRE SYSTEMS
ApplicantEwc
Product CodeDSA · Cardiovascular
Decision DateSep 23, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2900
Device ClassClass 2

Intended Use

The EWC: ECG Cable and Leadwire Systems are reusable electrode cable systems used to transmit signals from patient electrodes (not supplied by EWC) to various electrocardiograph recorders/monitors (not supplied by EWC) for both diagnostic and monitoring purposes. The EWC: ECG Cable and Leadwire Systems are limited by the indications for use of the connected monitoring or diagnostic equipment. Such equipment is commonly located in hospitals, doctor's offices, emergency vehicles, as well as in home use.

Device Story

Reusable electrode cable and leadwire system; transmits electrical signals from patient-applied electrodes to external ECG recorders/monitors. Configurations include 3, 5, 7, and 10-lead versions with various connectors (snaps, grippers, pins). Used in hospitals, clinics, emergency vehicles, and home settings. Operated by healthcare professionals or patients (depending on host monitor). Device acts as passive conduit for physiological signals; does not process data. Benefits include enabling signal acquisition for cardiac monitoring/diagnosis.

Clinical Evidence

Bench testing only. Device tested for compliance with 21 CFR 898 performance standards for electrode lead wires and patient cables. Comparative simulated use testing performed against predicate device; no significant performance differences observed.

Technological Characteristics

Reusable ECG cable and leadwire system. Configurations: 3, 5, 7, 10-lead. Connectors: snaps, grippers, pins. Passive signal transmission. Complies with 21 CFR 898. Partial evaluation against ANSI/AAMI EC53-1995.

Indications for Use

Indicated for patients requiring ECG diagnostic or monitoring procedures. Use is limited by the indications of the connected ECG recording or monitoring equipment in clinical, emergency, or home settings.

Regulatory Classification

Identification

A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the date SEP 23 1998 at the top. Below the date is the logo for EWC, which is in a bold, sans-serif font. To the right of the letters is a symbol that consists of five horizontal lines stacked on top of each other. The logo and date are in black and white. K982817- Image /page/0/Picture/2 description: The image shows a BSI Registered Firm logo. The logo is circular with the letters "BSI" at the top and the words "REGISTERED FIRM" around the circle. Inside the circle is a white triangle pointing downwards. Below the logo, the text "FM 34863" and "ISO 9001" are printed. ## Summary and Certification 510(k) Summary Date: August 6, 1998 ## 1. Establishment Information: | Submitter: | EWC (Electri-Wire Corporation)<br>N26 W23315 Paul Road<br>Pewaukee, WI 53072-4061 | |------------------|-----------------------------------------------------------------------------------| | Registration #: | 9921058 (Owner/Operator) | | Contact Name: | Timothy Davis | | Contact Phone #: | (414) 548-3700 or 800-786-3707 | | Manufacturing: | EWC (Electri-Wire Corporation)<br>208 West Street<br>Waupun, WI 53963 | | Registration #: | 2183764 | 2. General Device Information: | Common Name: | ECG Cable and Leadwire Systems | |------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name: | EWC: ECG Cable and Leadwire Systems | | Classification Name: | Patient transducer and electrode cable (including connector) | | Device Classification: | Class II (21 CFR: part 870.2900) | | Performance Std.: | 21 CFR Part 898: Performance Standard for Electrode Lead<br>wire and Patient Cables. This Standard is effective on May<br>11, 1998 and was published in the Friday, May 9, 1997<br>Federal Register. | 3. Substantial Equivalence: The EWC: ECG Cable and Leadwire Systems is substantially equivalent to the Tronomed Patient Cable and Lead Wire Systems which were marketed under 510(k) numbers; K771645, K771027 and K770884. 4. Device Description: The EWC: ECG Cable and Leadwire Systems are reuseable electrode cables designed to transmit signals from the patient electrode (not supplied by EWC) to various electrocardiograph recorders/monitors (not supplied by EWC) for both diagnostic and monitoring purposes. This device is common to both the industry and to most medical establishments. The EWC Cable and Leadwire Systems is offered in various configurations of cable types (3, 5, 7 and 10 lead versions), various lead styles (length, color, etc.) and various electrode connector configurations (snaps, grippers and pins). {1}------------------------------------------------ 5. Intended Use: The EWC: ECG Cable and Leadwire Systems are reusable electrode cable systems used to transmit signals from patient electrodes (not supplied by EWC) to various electrocardiograph recorders/monitors (not supplied by EWC) for both diagnostic and monitoring purposes. The EWC: ECG Cable and Leadwire Systems are limited by the indications for use of the connected monitoring or diagnostic equipment. Such equipment is commonly located in hospitals, doctor's offices, emergency vehicles, as well as in home use. 6. Technological comparison to legally marketed predicate device: The EWC: ECG Cable and Leadwire Systems technological characteristics are similar in comparison to the Tronomed Patient Cable and Lead Wire Systems as identified in our response to 3 above. 7. Test Summary and Conclusion: The EWC: ECG Cable and Leadwire Systems was tested to the requirements of 21 CFR 898 and shown to comply with the performance standard. Although the EWC: ECG Cable and Leadwire System was not designed to comply, in total, with the ANSI/AAMI EC53-1995, voluntary standard covering ECG Cables and Leadwires, a number of the test identified in the standard were performed, as necessary to evaluate the overall cable design and utility. Simulated use testing was also completed on both the predicate device cable and leadwire system and compared against the same tests conducted on the EWC: ECG Cable and Leadwire Systems and no significant difference in performance could be observed. Based on the results of the engineering/design testing, along with the simulated use performance testing (comparison testing), it is felt that the EWC: Cable and Leadwire Systems performs as expected and compares well, in terms of overall performance to the selected Tronomed device (predicate device). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized human profiles. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 3 1998 Mr. Timothy M. Davis Corporate Director RA/QA EWC Corporate Center N26 W23315 Paul Road Pewaukee, WI 53072-4061 K982817 Re: ECG Cable and Leadwire Systems Regulatory Class: II (two) Product Code: DSA Dated: August 6, 1998 Received: August 11, 1998 Dear Mr. Davis: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any {3}------------------------------------------------ obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, 2Misbranding by reference to premarket notification? (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## "Indication for Use Statement" ## INDICATION FOR USE PAGE 510(k) Number (if known): Unknown Device Name: EWC: ECG Cable and Leadwire Systems Indication for Use: The EWC: ECG Cable and Leadwire Systems are reusable electrode cable systems used to transmit signals from patient electrodes (not supplied by EWC) to various electrocardiograph recorders/monitors (not supplied by EWC) for both diagnostic and monitoring purposes. The EWC: ECG Cable and Leadwire Systems are limited by the indications for use of the connected monitoring or diagnostic equipment. Such equipment is commonly located in hospitals, doctor's offices, emergency vehicles, as well as in home use. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF REQUIRED) Prescription Use: (Per 21 CFR 801.109) or Over- Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use Dalk T. Wh Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%