Vios Monitoring System™ Model 2050

K172586 · Vios Medical, Inc. · DRT · Jun 22, 2018 · Cardiovascular

Device Facts

Record IDK172586
Device NameVios Monitoring System™ Model 2050
ApplicantVios Medical, Inc.
Product CodeDRT · Cardiovascular
Decision DateJun 22, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2

Indications for Use

The Vios Monitoring System (VMS) is intended for use by medically qualified personnel for physiological vital signs monitoring of adult (18+) patients in healthcare facilities. It is indicated for use in monitoring of 7-lead ECG, heart rate, respiratory rate, pulse rate, functional oxygen saturation of arterial hemoglobin, non-invasive blood pressure, and patient posture and activity. VMS allows for the input of body temperature, and can display data from peripheral devices. VMS can generate alerts when rate-based carthythmias are detected and when physiological vital signs fall outside of selected parameters.

Device Story

Wireless mobile platform for monitoring adult patient vitals in healthcare facilities; operated by medically qualified personnel. System components: patient-worn Chest Sensor (CS2050), pulse oximetry adapters (L2050E/F), and BSM SW (B2050) running on commercial IT platform. Chest Sensor acquires 3-channel ECG, bioimpedance, 2-channel pulse oximetry, and tri-axial accelerometer data; transmits via Bluetooth to BSM SW. BSM SW processes data to generate 7-lead ECG, respiratory rate, SpO2, pulse rate, posture, activity, and non-invasive blood pressure (from external Bluetooth cuff). System provides real-time display and generates alerts for rate-based arrhythmias or out-of-range vital signs. Enables continuous patient surveillance; facilitates timely clinical intervention; improves patient safety through automated monitoring and alerting.

Clinical Evidence

Clinical testing performed for pulse oximetry (per IEC 80601-2-61) and respiratory rate. Non-clinical testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), alarm standards (IEC 60601-1-8), ECG monitoring (IEC 60601-2-27), and patient monitoring (IEC 60601-2-49). Biocompatibility per ISO 10993, usability per EN 62366, and risk management per ISO 14971.

Technological Characteristics

Wireless mobile platform; Bluetooth connectivity. Chest Sensor: rechargeable, tri-axial accelerometer, bioimpedance, ECG, pulse oximetry. Adapters: plastic, non-sterile. BSM SW: standalone software on commercial IT platform. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 60601-2-49, EC53, ISO 10993, EN 62366, ASTM D4169-16, EN 62304, ISO 14971, IEC 80601-2-61.

Indications for Use

Indicated for physiological vital signs monitoring of adult (18+) patients in healthcare facilities.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure, while the FDA acronym and full name are written in blue, with the acronym in a blue square. June 22, 2018 Vios Medical, Inc. Megan Graham Quality/Regulatory Advisor 7300 Hudson Blvd N St. Paul, Minnesota 55128 Re: K172586 Trade/Device Name: Vios Monitoring System™ Model 2050 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: DRT, DQA, DPZ, DRG Dated: May 22, 2018 Received: May 23, 2018 Dear Megan Graham: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ # Page 2 - Megan Graham and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Dlorali Castillo **For** Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K172586 Device Name Vios Monitoring System™ Model 2050 #### Indications for Use (Describe) The Vios Monitoring System (VMS) is intended for use by medically qualified personnel for physiological vital signs monitoring of adult (18+) patients in healthcare facilities. It is indicated for use in monitoring of 7-lead ECG, heart rate, respiratory rate, pulse rate, functional oxygen saturation of arterial hemoglobin, non-invasive blood pressure, and patient posture and activity. VMS allows for the input of body temperature, and can display data from peripheral devices. VMS can generate alerts when rate-based carthythmias are detected and when physiological vital signs fall outside of selected parameters. Type of Use (Select one or both, as applicable) | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |-------------------------------------------------------------------------------------------------------| | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary | Submitter: | Vios Medical, Inc.<br>700 Commerce Dr.<br>Suite 190<br>Woodbury, MN 55125 | |-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Megan Graham<br>Director, Quality and Compliance<br>megan@viosmedical.com<br>Office: 651-764-8467<br>Fax: 651-237-7003 | | Date Prepared: | June 22, 2018 | | Trade Name: | Vios Monitoring System™ Model 2050 | | Common Name: | Vitals signs monitor | | Primary Regulation: | 21 CFR 870.2300 Cardiac Monitor including Cardiotachometer & Rate Alarm | | Other Regulation: | 21 CFR 890.2910 Transmitters And Receivers, Physiological Signal,<br>Radiofrequency<br>21 CFR 870.2710 Ear oximeter<br>21 CFR 870.2700 Oximeter | | Classification: | Class II | | Review Panels: | Cardiovascular, Anesthesiology | | Product Codes: | Primary: DRT<br>Other: DQA, DPZ, DRG | | Predicate Devices: | VMS Model 1000 (reference K150992)<br>VisiMobile Monitoring System (reference K152341)<br>Nonin Pulse Ox Model 6100C (reference K092101). | | Device Description: | The Vios Monitoring System (VMS) Model 2050 is a wireless mobile<br>medical device platform that allows caregivers in healthcare settings to<br>monitor patient vitals.<br><br>VMS includes Vios-proprietary monitoring software and a Vios-<br>proprietary vitals sensor with two Vios-proprietary adapters. It is<br>compatible with a medical grade, Bluetooth™-enabled NIBP cuff.<br><br>The VMS BSM SW Model B2050 is stand-alone software that can receive,<br>analyze, and display physiological vitals data from one or more patient-<br>worn sensors via standard communication protocols (Bluetooth™). It<br>runs on a commercial IT platform and is intended to be used in<br>conjunction with the Vios Chest Sensor and Vios Lead Adapters and can<br>support peripheral, medical grade, Bluetooth™-enabled devices. | | | The VMS Chest Sensor Model CS2050 is a small, patient-worn, non-sterile,<br>multiple use, and rechargeable sensor that acquires 3-channel ECG,<br>bioimpedance, 2-channel pulse oximetry, and tri-axial accelerometer data.<br>The sensor contains signal acquisition firmware (embedded software) and<br>wirelessly communicates acquired data via standard communication<br>protocols (Bluetooth™) to the BSM SW for analysis and display. The Chest<br>Sensor has a button that, when pressed, sends a patient call alert to the<br>BSM SW.<br>VMS Chest Sensor Adapter Models L2050E (Pulse Ox Ear Adapter) and<br>L2050F (Pulse Ox Finger Adapter) are plastic, non-sterile, patient-worn,<br>multiple use pulse oxygenation sensors that connect to the Vios Chest<br>Sensor and are secured to the patient via medical grade ECG electrodes. | | Indications for Use: | The Vios Monitoring System (VMS) is intended for use by medically<br>qualified personnel for physiological vital signs monitoring of adult (18+)<br>patients in healthcare facilities. It is indicated for use in monitoring of 7-<br>lead ECG, heart rate, respiratory rate, pulse rate, functional oxygen<br>saturation of arterial hemoglobin, non-invasive blood pressure, and<br>patient posture and activity. VMS allows for the input of body<br>temperature, and can display data from peripheral devices. VMS can<br>generate alerts when rate-based cardiac arrhythmias are detected and<br>when physiological vital signs fall outside of selected parameters. | | Summary of<br>Technology<br>Comparison | The Vios Monitoring System Model 2050 technology is based on the Vios<br>Monitoring System Model 1000 and shares a common architecture and<br>run on a commercial IT platform. The VMS Model 2050 Chest Sensor uses<br>a standard medical grade ECG electrode and an adapter plate to collect<br>additional ECG, impedance, accelerometer, and pulse oximetry data<br>(finger or ear). The VMS Model 2050 BSM SW has been updated to<br>generate and display 7-lead ECG data, respiratory rate, SpO2, pulse rate,<br>posture, activity, and non-latched physiological alarms. It can also display<br>non-invasive blood pressure data from a medical grade, Bluetooth™-<br>enabled NIBP cuff. | | Summary of Non-<br>Clinical, Clinical,<br>and Conformance<br>Testing | The safety, effectiveness, and substantial equivalency of the VMS Model<br>2050 have been confirmed through the following non-clinical, clinical, and<br>conformance testing:<br>Electrical safety, EMC, and vitals sign monitoring standards (IEC<br>60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC<br>60601-2-49, EC53)Biocompatibility standards (ISO 10993)Usability and human factors standards (EN 62366)Transportation Simulation testing (ASTM D4169-16)Software development life cycle (EN 62304)Risk Management (ISO 14971)Pulse oximetry clinical testing (IEC 80601-2-61)Respiratory Rate clinical testing | | Conclusion: | Vios Medical's evaluation of the substantial equivalence of the Vitals<br>Monitoring System Model 2050 to the predicate devices was based on a<br>comparison of device classification, intended use, indications for use and | | | | | contraindications, warnings, technical characteristics, and performance | | | characteristics. Vios Medical also performed conformance testing and | | | clinical testing to demonstrate substantial equivalency. Based on this | | | comparison, Vios Medical concludes that Vios Monitoring System Model | | | 2050 is substantially equivalent to the predicate devise and does not | | | introduce new safety or effectiveness issues. | | {4}------------------------------------------------ {5}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...