K150992 · Vios Medical, Inc. · DRT · Dec 16, 2015 · Cardiovascular
Device Facts
Record ID
K150992
Device Name
Vios Monitoring System
Applicant
Vios Medical, Inc.
Product Code
DRT · Cardiovascular
Decision Date
Dec 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Vios Monitoring System (VMS) is intended for use by medically qualified personnel for physiological and vital signs monitoring of adult (18+) patients in healthcare facilities. It is indicated for use in monitoring of ECG, heart rate, pulse rate, functional oxygen saturation of arterial hemoglobin, and axillary temperature. VMS allows for the input of non-invasive blood pressure and can display data from peripheral devices. VMS can generate alerts when rate-based cardiac arrhythmias are detected and when physiological vital signs fall outside of selected parameters.
Device Story
Vios Monitoring System (VMS) monitors adult patients in healthcare facilities; operated by medically qualified personnel. System receives physiological data (ECG, heart rate, pulse rate, SpO2, axillary temperature, non-invasive blood pressure) from sensors and peripheral devices. Software processes and displays vitals; generates alerts for rate-based cardiac arrhythmias or out-of-range physiological parameters. Operates in stand-alone or networked modes using standard IT equipment/protocols. Enables clinician oversight of patient status; facilitates timely clinical intervention based on real-time monitoring and alarm notifications.
Clinical Evidence
Bench testing only. Verification and validation included standards compliance (IEC 60601-1, 60601-1-2, 60601-1-8, 60601-2-27, 60601-2-49), formative and summative usability testing, reliability/performance testing, and transportation simulation.
Technological Characteristics
System components include bedside monitor software running on commercial IT equipment. Supports stand-alone or networked operation via standard protocols. Conforms to IEC 60601 series (safety/EMC), ISO 14971 (risk), IEC 62304 (software lifecycle), IEC 62366 (usability), and ISO 10993 (biocompatibility).
Indications for Use
Indicated for physiological and vital signs monitoring of adult (18+) patients in healthcare facilities. Monitors ECG, heart rate, pulse rate, SpO2, and axillary temperature; allows input of non-invasive blood pressure and peripheral device data.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 16, 2015
Vios Medical, Inc. Megan Graham Quality And Regulatory Advisor 7300 Hudson Blvd N. Ste 140 St. Paul, Minnesota 55128
Re: K150992
Trade/Device Name: Vios Monitoring System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: DRT, DRG, FLL Dated: November 16, 2015 Received: November 17, 2015
Dear Megan Graham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Page 1 of 1
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K150992
Device Name Vios Monitoring System
### Indications for Use (Describe)
The Vios Monitoring System (VMS) is intended for use by medically qualified personnel for physiological and vital signs monitoring of adult (18+) patients in healthcare facilities. It is indicated for use in monitoring of ECG, hear rals, pull si rate, functional oxygen saturation of arterial hemoglobin, and axillary temperature. VMS allews for your and invasive blood pressure and can display data from peripheral devices. VMS can generate alserts when on notiac arrhythmias are detected and when physiological vital signs fall outside of selected paramaters.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 7. 510(k) Summary
| Submitter: | Vios Medical, Inc.<br>7300 Hudson Rd Blvd N<br>Suite 140<br>St. Paul, MN 55128 |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Megan Graham<br>Quality and Regulatory Advisor<br>megan@viosmedical.com<br>Office: 651-764-8467<br>Fax: 651-237-7003 |
| Date Prepared: | April 28, 2015 |
| Trade Name: | Vios Monitoring System™ |
| Common Name: | Vitals signs monitor |
| Regulation<br>Classification: | 21 CFR 870.2300 Cardiac Monitor including Cardiotachometer<br>& Rate Alarm (Class II, Cardiovascular) 21 CFR 890.2910 Transmitters And Receivers, Physiological<br>Signal, Radiofrequency (Class II, Cardiovascular) 21 CFR 880.2910 Clinical Electronic Thermometer (Class II<br>exempt, General Hospital) |
| Review Panel: | Cardiovascular |
| Product Codes: | DRT, DRG, FLL |
| Predicate<br>Devices: | ViSi Mobile Monitoring System (K143751) for vitals monitoring<br>and temperature VitalConnect Platform (K141167) for physiological data<br>transmission and receipt |
| Device<br>Description: | The Vios Monitoring System includes the following components:<br>Bedside Monitor Software (BSM SW): Medical device software<br>that analyzes and/or displays vitals data received via standard<br>communication protocols from one or more compatible sensors<br>or peripheral devices. The software runs on standard<br>commercial IT equipment and can operate in stand-alone mode<br>or networked mode using standard networking protocols. In<br>networked mode, it can share its display with one or more |
| Indications for<br>Use: | The Vios Monitoring System (VMS) is intended for use by<br>medically qualified personnel for physiological and vital signs<br>monitoring of adult (18+) patients in healthcare facilities. It is<br>indicated for use in monitoring of ECG, heart rate, pulse rate,<br>functional oxygen saturation of arterial hemoglobin, and axillary<br>temperature. VMS allows for the input of non-invasive blood<br>pressure and can display data from peripheral devices. VMS can<br>generate alerts when rate-based cardiac arrhythmias are detected<br>and when physiological vital signs fall outside of selected<br>parameters. |
| Summary of<br>Substantial<br>Equivalence: | The Vios Monitoring System is substantially equivalent to the<br>predicate devices with respect to vitals signs monitoring in a<br>healthcare facility (product codes DRT, DRG, and FLL). Both the<br>VMS and its predicates are wireless vitals monitoring systems that<br>can operate in stand-alone or networked mode. Additional<br>functionality of the predicate devices is outside the scope of the<br>Vios Monitoring System.<br><br>The safety and effectiveness of the Vios Monitoring System design<br>have been confirmed through non-clinical testing and conformance<br>to vital signs monitoring, electrical safety (60601), electromagnetic<br>compatibility (60601), usability (62366), risk management (14971),<br>software development lifecycle (62304), and biocompatibility<br>standards (10993). |
| Non-Clinical<br>Testing: | The following bench testing was conducted to demonstrate safety<br>and efficacy of the Vios Monitoring System:<br>Design Verification and Validation Standards Compliance Testing (60601-1, 60601-1-2,<br>60601-1-8, 60601-2-27, and 60601-2-49) Usability Testing (Formative and Summative) Reliability and Performance Testing Transportation Simulation Testing |
| Conclusion: | Vios Medical considers the Vios Monitoring System to be safe,<br>effective, and substantially equivalent to the legally marketed<br>predicate devices. |
{4}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.