PRUCKA ENGINEERING INC. CARDIOLAB EP SYSTEM,VERSION 4.2
Applicant
Prucka Engineering, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Apr 7, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The intended use of the Prucka Engineering, Inc. CardioLab EP 4.2 System is to acquire, filter, digitize, amplify, display, and record electrical signals obtained during electrophysiological studies and related procedures conducted in an electrophysiological laboratory. Signal types acquired include ECG signals, direct cardiac signals, and pressure recordings. Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, store the data, perform elementary caliper-type measurements of the data, and generate reports on the data. Additionally, the system may acquire, amplify, display, and record data received from other medical devices typically used during these procedures, such as imaging devices and RF generators. The CardioLab EP system does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze physiological data or other data acquired during an EP procedure. It does not transmit alarms or arrhythmias, and does not have arrhythmia detection capabilities.
Device Story
Microprocessor-based data acquisition system for electrophysiology (EP) labs; acquires ECG, intracardiac, and pressure signals via third-party leadwires/catheters; accepts video input from fluoroscopy and serial data from RF generators. Amplifier filters, digitizes, and transmits data via fiber optic cable to a PC. Software functions: data acquisition, display, storage, and reporting. User performs caliper-type measurements and creates isopotential/isochronal activation maps. Used by clinicians in EP labs to monitor procedures and generate reports. Does not control energy delivery, administer drugs, perform life-support, or detect arrhythmias. Benefits include centralized visualization and recording of multi-modal procedural data.
Clinical Evidence
Bench testing only. System certified to IEC 601-1, IEC 601-1-1, IEC 601-1-2, IEC 601-2-27, and IEC 601-2-34 standards. In-house validation performed to confirm safety and efficacy of functional aspects.
Technological Characteristics
Microprocessor-based data acquisition system; fiber optic cable for electrical isolation between amplifier and PC; standard PC architecture; read/write optical media storage; video frame grabber board; RS-232 serial interface for RF generator connectivity. Complies with IEC 601-1 series safety standards.
Indications for Use
Indicated for use in electrophysiological laboratories to acquire, filter, digitize, amplify, display, and record ECG, intracardiac, and pressure signals during electrophysiological studies and related procedures.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
APR 7 2000
K993414
Image /page/0/Picture/2 description: The image shows the logo for Prucka Engineering, Inc. The logo consists of a black geometric shape on the left, resembling a stylized building or structure. To the right of the shape is the text "Prucka Engineering, Inc." in a slanted, sans-serif font. The word "Engineering" is slightly smaller than "Prucka", and "Inc." is even smaller and set off by a comma.
Advanced Technology for Cardiology
# Section C - 510(k) Summary
October 8, 1999
#### A. Contact Information:
Matthew W. Prucka Prucka Engineering, Inc. 13000 Executive Drive Sugar Land, TX 77478
Phone: (281) 275-5011 Fax: (281) 275-5001
#### B. Device names:
Trade Name: Prucka Engineering, Inc. CardioLab EP System Common Name: Electrophysiology lab system Classification Names: Electrocardiograph Biopotential Amplifier and Signal Conditioner Programmable Diagnostic Computer Blood Pressure Computer
#### C. Substantial equivalence is claimed to the following devices:
Quinton EP System (K971570) EP Medsystem EP System (K935186) Biosense Carto System (K954395)
#### D. Description of device:
The CardioLab EP System is a microprocessor based data acquisition system used during electrophysiology procedures to accquire ECG, intracardiac, pressure, and digital data from other devices such as RF generators and fluoro video systems. The ECG, intracardiac and pressure data are acquired by an amplifier that is connected to the patient by third-party
{1}------------------------------------------------
devices such as ECG leadwires and catheters. The amplifier filters, amplifies, digitizes and transmits the data to the computer. The computer stores the data on optical disks, displays the data on the video monitors, allows the user to perform basic signal measurements, and prints out waveforms on a laser printer or continuous paper recorder. The software has three major functions: data acquisition and display, data storage, and reporting of data. The system is able to acquire signal data in the presence of pacing pulses. In addition to signal waveform display and basic caliper-type measurement, the system allows the user to create isopotential and isochronal activation maps of waveform timing information. The system can also be used to accept video input from a fluoroscopy system for either single frame capture or full motion video display. The system can also acquire data from an RF generator via standard computer RS-232 serial port. The CardioLab EP system does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze physiological data or other data acquired during an EP procedure. It does not transmit alarms or arrhythmias, and does not have arrhythmia detection capabilities.
### E. Intended use of device:
The intended use of the Prucka Engineering, Inc. CardioLab EP 4.2 System is to acquire, filter, digitize, amplify, display, and record electrical signals obtained during electrophysiological studies and related procedures conducted in an electrophysiological laboratory. Signal types acquired include ECG signals, direct cardiac signals, and pressure recordings. Physiological parameters such as diastolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, store the data, perform elementary caliper-type measurements of the data, and generate reports on the data. Additionally, the system may acquire, amplify, display, and record data received from other medical devices typically used during these procedures, such as imaging devices and RF generators.
## F. Summary of the technological characteristics of the CardioLab EP System compared to the predicate devices:
Technological characteristics of the Prucka Engineering, Inc. CardioLab EP System are the same as predicate devices with the following exceptions:
| Comparison Table<br>Reference* | Summary of differences |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2 | Amplifier leakage current data not available for Biosense Carto System |
| 3 | Amplifier input channel types data not available for Biosense Carto System |
| 4 | Amplifier output channel types data not available for Biosense Carto System |
| 6 | The amplifier in the CardioLab EP System sends digital data to the computer via fiber optic cable which provides complete electrical isolation between the computer and amplifier unit. The predicate devices use an electrical connection between the amplifier and computer, and electrical isolation is provided by an opto isolator inside the amplifier. |
| 7 and 8 | The CardioLab EP System, the Quinton EP System and the EP Medsystems EP System use standard personal computers to control the amplifier and use it as the main data processing unit. The Biosense Carto System uses a Sun workstation computer for these functions. |
| 9 | The CardioLab EP System, the Quinton EP System and the EP Medsystems EP System have independently controlled monitors which display real-time data, review information, and images. The Biosense Carto System has only one computer monitor. |
| 10 | The CardioLab EP System, the EP Medsystems EP System, and the Biosense Carto System store data on read/write optical media. The Quinton EP System stores data on "write once" media. |
| 12 | The CardioLab EP System, the Quinton EP System and the EP Medsystems EP System have the ability to print data on a strip chart recorder. The Biosense Carto System does not have this feature. |
| 13 | The CardioLab EP System and the Quinton EP System have input for a third party cardiac stimulator which will connect the stimulator to the intracardiac signal inputs with no control over the function of the third party cardiac stimulator. The EP Medsystems EP System has the same properties, except it controls the function of the external stimulator. Third party stimulator input capability data not available for the Biosense Carto System. |
| 15 | The CardioLab EP System, the Quinton EP System and the EP Medsystems EP System have the ability to record from intracardiac signal at the same time a third party cardiac stimulator is pacing to the intracardiac signal. Biosense Carto System data not available with respect to this feature. |
| 16 | The CardioLab EP System, the Quinton EP System and the EP Medsystems EP System can connect to third party RF generators using a standard RS-232 serial port in order to acquire, display, store and print data from the third party RF generators. The Biosense Carto System does not have this feature. |
| 17 | The CardioLab EP System, the Quinton EP System and the EP Medsystems EP System have standard video frame grabber board that operates in the personal computer and digitizes standard video images for display on the computer monitor and storage with the patient information. The Biosense Carto System does not have this feature. |
| 18 | The CardioLab EP System and the Biosense Carto System have the ability to display waveform timing information in a two dimensional color map that indicates either the isocronal or isopotential timing information. The Quinton EP System and EP Medsystems EP System do not have this feature. |
{2}------------------------------------------------
* See corresponding number in comparison table in Section F.
{3}------------------------------------------------
Each technological characteristic of the CardioLab EP System is found in at least one predicate device used in the same type of procedure as the CardioLab EP System is intended to be used in. The technological differences listed above are minor and do not raise new issues of safety and efficacy. The technological characteristics of the CardioLab EP System are substantially equivalent to technological characteristics found in predicate devices.
## G. Brief discussion of the nonclinical tests and how their results support a determination of substantial equivalence:
The equipment has been tested and certified to meet the following national and international safety stadards by SEMKO which is an Inchcape Testing Services Company.
IEC 601-1 IEC 601-1-1 IEC 601-1-2 IEC 601-2-27 IEC 601-2-34
The equipment has also been tested by ETL Testing Laboratories with the applicable requirements of the FDA Reviewer Guidance for premarket Notification Submissions, November 1993.
In addition, an in-house validation has been performed on the system with results that meet acceptance criteria, confirming the safety and efficacy of each functional aspect of the system.
These tests conducted by outside laboratories together with system level validation testing conducted in-house provide complete confirmation that the system is safe and effective for its intended use. This demonstrates that the Prucka Engineering. Inc. CardioLab EP 4.2 System has a substantially equivalent level of safety and efficacy as predicate devices.
{4}------------------------------------------------
Public Health Service
APR 7 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Matthew Prucka President Prucka Engineering, Inc. 13000 Executive Drive Sugar Land, TX 77478
Re: K993414 Prucka Engineering, Inc. CardioLab EP System, Version 4.2 Regulatory Class: II (two) Product Code: DPS Dated: February 3, 2000 Received: February 4, 2000
Dear Mr. Prucka:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Paqe 2 - Mr. Matthew W. Prucka
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CTegarlmgneher
for
James E. Dillard III Director
Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
K993414 510(k) Number (if known): __
Device Name: CardioLab EP System, Version 4.2
Indications For Use:
The intended use of the Prucka Engineering, Inc. CardioLab EP 4.2 System is to acquire, filter, digitize, amplify, display, and record electrical signals obtained during electrophysiological studies and related procedures conducted in an electrophysiological laboratory. Signal types acquired include ECG signals, direct cardiac signals, and pressure recordings. Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, store the data, perform elementary caliper-type measurements of the data, and generate reports on the data. Additionally, the system may acquire, amplify, display, and record data received from other medical devices typically used during these procedures, such as imaging devices and RF generators.
The CardioLab EP system does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze physiological data or other data acquired during an EP procedure. It does not transmit alarms or arrhythmias, and does not have arrhythmia detection capabilities.
(Please Do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) scular, Respiratory,
(Optional Format 3-10-98)
000-55
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.