K173952 · Vectracor, Inc. · DPS · Aug 28, 2018 · Cardiovascular
Device Facts
Record ID
K173952
Device Name
Universal SmartECG
Applicant
Vectracor, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Aug 28, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Indications for Use
The System is intended to derive, display and print a derived 12 lead ECG as well as the X, Y, Z leads from the acquisition of just 3 leads (5 electrodes).The System also has the capability to acquire the standard 12 lead ECG using the standard 10 electrodes and convert the ECG signal into a digital format. The 12 lead interpretive software is a windows-based program intended to interpret electrocardiograms. The software receives, displays and stores a single, three or standard 12 lead simultaneous ECG recording, which is tres- mitted either locally or transtelephonically from an ECG monitor using a proprietary digital data transmission protocol. The device contains proprietary software algorithms to receive, store, analyze and interpret the 12 lead ECG signal only. In 5 electrode mode, the System analyzes data from 3 leads (5 electrodes) and produces visual and audible alarms for ECG changes that may be consistent with 12-lead ECG signs of Acute Myocardial Ischemic Injury (AMII), including Acute Myocardial Infarction (AMI).This mode is not intended to be a sole means of diagnosis, but prompts, the user to acquire a standard 12 lead ECG (using 10 electrodes) for interpretation by a physician when the Cardian Electrical Biomarker (CEB®) is 95 or greater. Monitoring patients with a CEB® is only indicated for patients presentinewith chest pain or other presumed anginal equivalents. The interpretation software is only available for the standard 12 lead ECG utilizing the standard 10 electrodes. The System is intended to be used by healthcare professionals, or trained personnel where ECG monitoring /acquisition is indicated for hospitals and/or clinics. The System can be used within electro-surgical environments. Device is for Adult use.
Device Story
Universal SmartECG acquires ECG signals via 10-lead (standard 12-lead ECG) or 3-lead (5-electrode) configurations. Device transforms 3-lead inputs into derived 12-lead and Frank vector (X, Y, Z) leads using geometric principles. Proprietary algorithm calculates Cardiac Electrical Biomarker (CEB) to monitor for Acute Myocardial Ischemic Injury (AMII) and AMI. System provides visual/audible alarms when CEB ≥ 95, prompting clinicians to acquire standard 12-lead ECG for physician interpretation. Used in hospitals/clinics by healthcare professionals; compatible with electrosurgical environments. Output includes ECG waveforms, vector loops, and CEB index. Assists clinical decision-making by identifying potential ischemic events, facilitating timely diagnostic intervention.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including verification and validation of software and hardware performance against IEC 60601-1, 60601-1-2, and 60601-2-25 standards.
Technological Characteristics
10-lead/5-electrode ECG acquisition; Windows-based software; SQL database; proprietary digital transmission protocol. Hardware compliant with IEC 60601-1, 60601-1-2, and 60601-2-25:2011. Includes silicone strip on case. Connectivity via local or transtelephonic transmission. Software modules: VectraplexECG and Office Medic (ECG component).
Indications for Use
Indicated for adult patients presenting with chest pain or other presumed anginal equivalents requiring ECG monitoring or acquisition in hospitals or clinics.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
August 28, 2018
VectraCor, Inc. Andrew Schreck Manager of Regulatory Affairs 785 Totowa Rd. Suite 100 Totowa, New Jersey 07512
Re: K173952
Trade/Device Name: Universal SmartECG Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, MLD Dated: July 26, 2018 Received: July 27, 2018
Dear Andrew Schreck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Arielle Drummond -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K173952
Device Name Universal SmartECG
Indications for Use (Describe)
· The System is intended to derive, display and print a derived 12 lead ECG as well as the X, Y, Z leads from the acquisition of just 3 leads (5 electrodes).The System also has the capability to acquire the standard 12 lead ECG using the standard 10 electrodes and convert the ECG signal into a digital format.
· The 12 lead interpretive software is a windows-based program intended to interpret electrocardiograms. The software receives, displays and stores a single, three or standard 12 lead simultaneous ECG recording, which is tres- mitted either locally or transtelephonically from an ECG monitor using a proprietary digital data transmission protocol. The device contains proprietary software algorithms to receive, store, analyze and interpret the 12 lead ECG signal only.
· In 5 electrode mode, the System analyzes data from 3 leads (5 electrodes) and produces visual and audible alarms for ECG changes that may be consistent with 12-lead ECG signs of Acute Myocardial Ischemic Injury (AMII), including Acute Myocardial Infarction (AMI).This mode is not intended to be a sole means of diagnosis, but prompts, the user to acquire a standard 12 lead ECG (using 10 electrodes) for interpretation by a physician when the Cardian Electrical Biomarker (CEB®) is 95 or greater. Monitoring patients with a CEB® is only indicated for patients presentinewith chest pain or other presumed anginal equivalents.
· The interpretation software is only available for the standard 12 lead ECG utilizing the standard 10 electrodes.
· The System is intended to be used by healthcare professionals, or trained personnel where ECG monitoring /acquisition is indicated for hospitals and/or clinics.
· The System can be used within electro-surgical environments.
· Device is for Adult use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed announg the and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 1. 510k Summarv
Image /page/3/Picture/1 description: The image shows the VectraCor logo. The logo consists of the word "VECTRACOR" in blue, with a red line underneath. Below the line is the text "Diagnosing Better Care" in a smaller font. To the right of the text is a red heartbeat line.
#### Submitters Name and Address: 1.
VectraCor, Inc. 785 Totowa Rd. Suite 100 Totowa, NJ 07512 Tel: 973-904-0444 Fax: 973-904-0411 e-mail: ajschreck@vectracor.com Official Correspondent: Andrew Schreck, Manager or Regulatory Affairs
Contact Person for this Submission: Andrew Schreck
Date submission was prepared: November 30, 2017
#### 2. Trade Name, Common Name and Classification Name:
- a. Trade Name: Universal SmartECG
- b. Product Code: Z-7000-0500
- Common Name: Electrocardiogram, ECG, EKG C.
- Classification Name and Product Code: ರ.
| ECG | DPS Class II | 21 CFR 870.2340 |
|----------------------------|--------------|-----------------|
| ST Segment Monitor & Alarm | MLD Class II | 21 CFR 870.1025 |
#### 3. Predicate Device Identification:
- VectraplexECG System with VectraplexAMI (K102378) O
- Universal ECG (K990266) O
#### 4. Device Description:
The Universal SmartECG is a device with 10 lead wires attaching to electrodes on a patient's chest and one wire connecting to a computer. The system is able to display, store, and print a 12 lead ECG with 10 electrodes attached to the patient. In addition, The Universal SmartECG incorporates the VectraplexECG software (K102378), to derive up to a 15 lead (standard 12 leads and Frank vector X, Y, and Z leads) ECG from only three measured leads (5 electrodes). The device also monitors, in real-time, ECG changes that may be indicative of acute myocardial ischemic injury, including AMI*, via a displayed index number, the Cardiac Electrical Biomarker (CEB; VectraplexAMI). The system also provides the vector loops. The System's derivation algorithm is able to provide additional information based on geometric principals relating to the placement of electrodes on the body. The Cardiac Electrical Biomarker (CEB) is produced
{4}------------------------------------------------
based on an algorithm accounting for the multipolar contributions of the electrical field around the heart, which was FDA cleared via K102378.
*There will be times when the ECG shows signs consistent with AMI and the CEB shows signs consistent with AMI, but Troponin may not be completed or available at that time.
# 5. Indications for use:
- · The System is intended to derive, display and print a derived 12 lead ECG as well as the X, Y, Z leads from the acquisition of just 3 leads (5 electrodes).The System also has the capability to acquire the standard 12 lead ECG using the standard 10 electrodes and convert the ECG signal into a digital format.
- · The 12 lead interpretive software is a windows-based program intended to interpret electrocardiograms. The software receives, displays and stores a single, three or standard 12 lead simultaneous ECG recording, which is transmitted either locally or transtelephonically from an ECG monitor using a proprietary digital data transmission protocol. The device contains proprietary software algorithms to receive, store, analyze and interpret the 12 lead ECG signal only.
- · In 5 electrode mode, the System analyzes data from 3 leads (5 electrodes) and produces visual and audible alarms for ECG changes that may be consistent with 12-lead ECG signs of Acute Myocardial Ischemic Injury, including Acute Myocardial Infarction . This mode is not intended to be a sole means of diagnosis, but prompts the user to acquire a standard 12 lead ECG (using 10 electrodes) for interpretation by a physician when the Cardiac Electrical Biomarker (CEBS) is 95 or greater. Monitoring patients with a CEB® is only indicated for patients presenting with chest pain or other presumed anginal equivalents.
- · The interpretation software is only available for the standard 12 lead ECG utilizing the standard 10 electrodes.
- · The System is intended to be used by healthcare professionals, or trained personnel where ECG monitoring /acquisition is indicated for hospitals and/or clinics.
- · The System can be used within electro-surgical environments.
- · Device is for Adult use.
Prescription Use YES Over-The-Counter Use Prescription Use______________________________________________________________________________________________________________________________________________________________ NO (21 CFR 801 Subpart C)
# 6. Comparison to Predicate Device:
The new Universal SmartECG System is the same principle device as the previous version of hardware, the Universal ECG (K990266). The databases for the VectraplexECG and Office Medic software modules, which are currently FDA-cleared, have been combined in this product. The technological characteristics (design, material, chemical composition, energy source) of the device are the same as the predicate.
The software, Office Medic and VectraplexECG, have been updated to be compatible with a new version of hardware, the Universal SmartECG (previously the Universal ECG). VectraCor has periodically released minor updates for the VectraplexECG System over the past few years.
{5}------------------------------------------------
The new Universal SmartECG System is the same principle device as the previous version of hardware, the Universal ECG (K990266). New hardware is designed to be compliant with IEC 60601-2-25:2011 standard. The Universal SmartECG system includes the hardware and 2 ECG software modules, VectraplexECG (K102378) and Office Medic. The Office Medic module has 2 components including ECG and spirometry functions. This 510k submission only affects the ECG component of Office Medic. The ECG component of Office Medic was previously known as CardioView 3000 (K083749) and was incorporated by Office Medic. The VectraplexECG and ECG component of Office Medic, which are FDA cleared and are currently 2 separate software programs, will share a database in the Universal SmartECG release. The technological characteristics (design, material, chemical composition, energy source) of the device are the same as the predicate.
Prior to this 510k submission, VectraCor has periodically released minor updates for the VectraplexECG System over the past few years. All updates have been considered in relation to the FDA memorandum "Deciding When to Submit a 510(k) for a Change to an Existing Device" and were determined that at the time a 510(k) was not necessary. These updates as well as enhancements to the new system include:
| Note to file enhancements | |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Operating System upgrades<br>(compatibility with new Windows versions) | Added reports to increase physician productivity with use of the VectraplexECG System |
| SQL Database upgrades | Added drivers and updates for better compatibility to the hardware device |
| Added a graph of the CEB over time | |
| Changes necessitated new 510k submission | |
|-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| • Updated software for compatibility to<br>new data acquisition hardware (new<br>revision of the previous hardware) | • The 2 ECG modules, VectraplexECG<br>and Office Medic, will share a<br>database in the new release |
| • Due to changes in the new hardware<br>(Universal SmartECG) the device can<br>now be used in electrosurgical<br>environments | • Updated the ECGlib.dll which<br>measures ECG waveform<br>components |
| • An outer silicone strip has been<br>added to the outside seams of the<br>case | |
### 7. Testing and Conclusions:
Verification, validation and testing procedures were performed to assure that the new device works and is substantially equivalent to the predicate. Testing supported the proposed label.
The new hardware device, the Universal SmartECG, has been tested to, and passed. 60601-1.-1-2. and 2-25 standards.
{6}------------------------------------------------
Since this update is not changing the CEB algorithm or the derivation algorithm from the VectraplexECG System it has been determined that no clinical data is necessary to support these functions.
VectraCor has determined, based on the performance testing, that the Universal SmartECG conforms to the design specifications and is substantially equivalent to the predicate devices, Universal ECG and VectraplexECG, for acquiring ECG leads and for monitoring a patient for ECG changes that may be consistent with 12 lead ECG signs of myocardial ischemic injury, including AMI. The device, as designed, is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.