K112959 · Philips Health Care · DPS · Jan 26, 2012 · Cardiovascular
Device Facts
Record ID
K112959
Device Name
CALG-STR EXERCISE ECG ANALYSIS ALGORITHM
Applicant
Philips Health Care
Product Code
DPS · Cardiovascular
Decision Date
Jan 26, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The intended use of the CAlg-STR exercise ECG analysis algorithm is to analyze multi-channel ECG waveforms acquired from a patient and produce measurements such as heart rate, detect ventricular arrhythmias, form representative beats, and calculate ST segment deviation (elevation or depression) and ST slope for review by a trained physician or clinician in determining a diagnosis. The measurements should not be used as a sole means for determining a patient's diagnosis.
Device Story
Software module integrated into stress exercise systems; analyzes up to 16-lead ECG waveforms. Inputs: multi-channel ECG signals (1000 samples/sec). Processing: beat classification, heart rate calculation, arrhythmia detection, and averaging of similar morphology beats to form representative beats; calculates ST segment deviation and slope. Output: measurements displayed on host device for clinician review. Used in hospitals and clinics by trained medical personnel. Does not provide diagnostic interpretation; output supports physician decision-making regarding cardiac diagnosis and treatment. Benefits: provides automated, objective measurements of ST segments and rhythm during stress testing to assist clinical assessment.
Clinical Evidence
Bench testing only. Performance validated against ANSI/AAMI/ISO EC57:1988/(R)2008 requirements for cardiac rhythm and ST-segment measurement algorithms. Results demonstrated consistent performance meeting all design specifications.
Technological Characteristics
Software-based ECG analysis algorithm; 1000 samples/sec sampling rate; 0.050-150 Hz bandwidth; 5µV/LSB resolution. Uses Cubic Spline algorithm for baseline correction. Configurable notch filter. No hardware/biocompatibility requirements as it is a software module.
Indications for Use
Indicated for use in exercise ECG testing for patients 10 years or older to evaluate ECGs for diagnosis, treatment planning, treatment effectiveness, or symptom evaluation. Not for use as a physiological monitor.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
Predicate Devices
Case Cardiac Testing System/CS Cardiac Testing System
Submission Summary (Full Text)
{0}------------------------------------------------
K112959 pg 1 of 5
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
JAN 2 6 2012
### Appendix I
## 510(k) Summary
As required by 21 CFR 807.87(h), a 510(k) Summary for this Premarket Notification submission is provided below.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
#### CAlg-STR Exercise ECG Analysis Algorithm
#### Submitter's Name and Address
| Submitter's Name: | Philips Healthcare |
|-----------------------|-----------------------------------------------|
| Division: | Diagnostic ECG |
| Address: | 1525 Rancho Conejo Boulevard Suite 100 |
| City, State, and Zip: | Thousand Oaks, CA 91320 |
| Contact Name: | Gretel Lumley/Quality and Regulatory Engineer |
| Telephone / Fax: | ( 805 ) 214-5101 / (805) 214-5129 |
#### Manufacturers' Information: Establishment Registration Number Establishment name:
Address:
Establishment Registration No.
Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 1218950
#### Device Details
| | New Product | Predicate |
|-------------------------------|------------------------------------------------|------------------------------------------------------------|
| Proprietary or<br>Trade Name: | CAlg-STR Exercise<br>ECG Analysis<br>Algorithm | Case Cardiac Testing<br>Sytem/CS Cardiac<br>Testing System |
| Common<br>Name: | ECG Analysis System | ECG Analysis Computer |
| Device Class: | II | II |
| Device<br>ProCodes: | DPS | DQK |
| Device CFR: | 870.2340 | 870.1245 |
| Classification<br>Panel: | 74 Cardiovascular | 74 Cardiovascular |
| Classification<br>Name: | ECG Analysis System | Programmable<br>diagnostic computer |
Page 53 of 57
{1}------------------------------------------------
## K112959 pg 20f5
#### Intended Use
The intended use of the CAlg-STR Exercise ECG Analysis Algorithm is to analyze multi-channel ECG waveforms acquired from a patient and produce measurements such as heart rate, detect ventricular arrhythmias, form representative beats, and calculate ST segment deviation (elevation or depression) and ST slope for review by a trained physician or clinician in determining a diagnosis. The measurements should not be used as a sole means for determining a patient's diagnosis.
#### Device Description
The CAlg-STR Exercise ECG Analysis Algorithm is a software module which may be integrated into a stress exercise system. It is used to analyze ECG waveforms to find and classify heart beats to calculate heart rate, detect arrhythmias and calculate ST deviation and ST slope when an adult or pediatric patient is undergoing cardiac stress testing.
The exercise ECG Analysis Algorithm takes up to 16 ECG leads as input and provides analysis and measurements. The Exercise ECG Analysis Algorithm analyzes up to 3 leads of ECG waveforms to find and classify the heart beats and calculate heart rate. The beat classifications are then used to calculate arrhythmias. The beats with similar morphology are averaged to from representative beats for up to 16 leads. The representative beats are used to measure ST deviation and slope for up to 16 leads. Those measurements are used for display and report in the host device. The algorithm does not supply an interpretation of the data.
{2}------------------------------------------------
#### Product Comparison:
| Specification/<br>Feature | CAlg-STR Exercise<br>ECG Analysis<br>Algorithm | CASE Cardiac Testing System/CS Cardiac<br>Testing System | Comparison |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | Intended Use:<br>The intended use of<br>the CAlg-STR<br>exercise ECG analysis<br>algorithm is to<br>analyze mult-channel<br>ECG waveforms<br>acquired from a<br>patient and produce<br>measurements such as<br>heart rate, detect<br>ventricular<br>arrhythmias, form<br>representative beats,<br>and calculate ST<br>segment deviation<br>(elevation or<br>depression) and ST<br>slope for review by a<br>trained physician or<br>clinician in<br>determining a<br>diagnosis. The<br>measurements should<br>not be used as a sole<br>means for<br>determining a<br>patient's diagnosis. | Intended Use:<br>CASE Cardiac Testing System and the CS<br>Cardiac Testing System are intended to be used by<br>trained operators under direct supervision of a<br>licensed health care practitioner on adult and<br>pediatric patients. The CASE Cardiac Testing<br>System and the CS Cardiac Testing System are<br>designed to acquire, process, record, archive,<br>analyze and output (12 and 15 lead) ECG data<br>during a period of physiologic stress or during a<br>resting ECG test, acquire data from ancillary<br>devices (such as Spirometery and Ambulatory<br>Blood Pressure), provide median morphology<br>recordings and record ECG in real-time with and<br>without arrhythmia detection. <b>The arrhythmia<br/>detection portion of CASE Care provided to the<br/>user for the convenience of automatic detection<br/>of arrhythmias but does not provide alarms.</b><br>CASE Cardiac Testing System and the CS Cardiac<br>Testing System provide the control of external<br>devices (typically treadmill or Ergometer) and<br>communicate with centralized electronic/digital<br>storage system via network. CASE Cardiac<br>Testing System and the CS Cardiac Testing<br>System provide a user selectable option for<br>printouts of prognostic scores on select reports.<br>Vector loops are also available.<br>Case Cardiac Testing System and the CS<br>Cardiac Testing System can be configured in a<br>network environment for multiple CASE or CS<br>stations allowing the user to create a central<br>database of patient demographics and collected<br>patient physiological data. <b>CASE Cardiac<br/>Testing System and the CS Cardiac Testing<br/>System are intended to be used primarily in the<br/>hospital but can be used in clinics, physician<br/>offices, outreach centers or wherever exercise,<br/>stress testing ECG, Spirometry or ambulatory<br/>blood pressure testing is performed.</b><br><b>CASE Cardiac Testing System and the CS<br/>Cardiac Testing System offer no diagnostic<br/>opinion to the user. Instead it provides<br/>interpretive statements of morphology, rhythm,<br/>and conduction for with the physician renders<br/>his/her own medical opinion.</b> CASE/CS Cardiac<br>Testing System is not intended to be used as a<br>transport device or for home use. CASE Cardiac<br>Testing System and the CS Cardiac Testing<br>System are not interceded for the use as a vital<br>signs physiological monitor or for intracardiac<br>use. | Similar for<br>algorithm and<br>use<br>environment. |
| Indications for<br>Use | Indications for Use:<br>The analysis<br>algorithm is indicated<br>for use in those<br>situations where the<br>clinician decides to<br>evaluate the<br>electrocardiogram of<br>patients at 10 years or<br>older, as part of<br>decisions regarding<br>possible diagnosis,<br>potential treatment,<br>effectiveness of<br>treatment or to rule<br>out causes for<br>symptoms | | |
| Target<br>Population | Adult and Pediatric | Adult and Pediatric | Same |
| Specification/<br>Feature | CAlg-STR Exercise<br>ECG Analysis<br>Algorithm | CASE Cardiac Testing System/CS Cardiac<br>Testing System | Comparison |
| Where Used | Used in hospital and<br>clinical settings, by<br>qualified medical<br>personnel trained in<br>stress exercise testing. | Used primarily in the hospital but can be<br>used in clinics, physician offices, outreach<br>centers or wherever exercise, stress testing<br>ECG, Spirometry or ambulatory blood<br>pressure testing is performed. | Similar |
| Display | Displayed on host<br>device | Displayed on host device. | Same |
| Number of ECG<br>Signal Leads | Up to 16 leads | Up to 16 leads | Same |
| Sampling Rate | 1000 samples per<br>second | 500 samples per second | Different |
| Bandwidth | .050 - 150 Hz | .050 - 150 Hz | Similar |
| Input Signal<br>Resolution | 5µV/LSB at 1000 Hz | 4.88µV/LSB at 500 Hz | Similar, Hz<br>difference is<br>due to sampling<br>rate |
| Notch Filter | Yes, configurable | Yes, Configurable | Same |
| Baseline<br>Correction | Cubic Spline<br>Algorithm | Cubic Spline Algorithm | Same |
| Artifact/Baseline<br>Correction | No | Finite Residual Filter (FRF) Analysis | Different, did<br>not implement<br>FRF due to<br>potential<br>distortion of ST<br>segment |
| ST<br>Measurements<br>and Calculations | ST amplitude, ST<br>Slope | ST amplitude, ST Slope, integral, index,<br>ST/HR slope, ST/HR loops, ST/HR index up<br>to 15 leads | Similar, only<br>ST amplitude<br>and ST slope<br>are calculated<br>other<br>measurements<br>are provided by<br>the host device |
| Heart Rate | Automatic<br>Arrhythmia detection,<br>documentation and<br>annotation | Automatic Arrhythmia detection,<br>documentation and annotation | Similar |
| Post J-Point<br>selection | Configurable | Configurable | Same |
| Biocompatibility | Not applicable | Not applicable | Identical |
| Sterility | Not applicable | Not applicable | Identical |
Philips Medical Systems
K112959 CAIg-STR Exercise ECG Analysis Algorithm Response
-
Page 55 of 57
{3}------------------------------------------------
#### KII 2959 pg 4of 5
The CAIg-STR Exercise ECG Analysis Algorithm has same technological characteristics except for the sampling rate and Artifact/Baseline correction as shown in the table above. The difference between the predicate's sampling rate and the CAIg-STR Exercise ECG Analysis Algorithm is not significant because a higher sampling rate creates a better fidelity of the ECG signal that is analyzed. The predicate device chose to include Artifact/Baseline correction which is not included in the CAlg-STR Exercise ECG Analysis Algorithm. Philips Medical Systems chose not to include Artifact/Baseline Correction due to potential distortion of the ST segment which would have potentially
{4}------------------------------------------------
K112959 pg 5 of 5
caused decreased accuracy of the algorithm and additional design changes to the algorithm would have been necessary to address distortion of the ST segment.
#### Performance Data
The performance data demonstrated that the CAIg-STR Exercise ECG Analysis Algorithm meet with the design specifications of the algorithm. The CAIg-STR Exercise ECG Analysis Algorithm was tested to the requirements of ANSVAAMI/ISO EC57:1988/(R)2008: Testing and reporting performance results of cardiac rhythm and ST-sgement measurement algorithms. The performance results from this testing demonstrated that CAlg-STR Exercise ECG Analysis Algorithm performed consistentlyand meets the requirements of ANSI/AAMI/ISO EC57:1988/(R)2008: Testing and reporting performance results of cardiac rhythm and ST-sgement measurement algorithms.
#### Conclusions
In conclusion, the performance and non-clinical testing demonstrated that the CAIg-STR Exercise ECG Analysis Algorithm is as safe and effective as the analysis software of the predicate device. This is supported by the adherence to the FDA recognized consenus standard (ANSI/AAMI/ISO EC57:1988/(R) 2008: Testing and reporting performance results of cardiac rhythm and ST-sgement measurement algorithms) and comparison testing to other algorithms already approved for market.
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle, with its wings spread. The eagle is facing to the right.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 2 6 2012
Philips Healthcare c/o Ms. Gretel Lumley 1525 Rancho Conejo Blvd, Suite 100 Thousand Oaks, CA 91320
Re: K112959
Trade/Device Name: CAlg-STR Exercise ECG Analysis Algorithm Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: January 10, 2012 Received: January 11, 2012
Dear Ms. Lumley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Gretel Lumley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bran D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K112959 pg 1 of 1
#### Appendix H
#### Indications for Use Statement
510(k) Number: Device Name: Philips CAlg-STR Exercise Analysis Algorithm
Indications for Use:
#### Intended Use:
The intended use of the CAIg-STR exercise ECG analysis algorithm is to analyze multi-channel ECG waveforms acquired from a patient and produce measurements such as heart rate, detect ventricular arrhythmias, form representative beats, and calculate ST segment deviation (elevation or depression) and ST slope for review by a trained physician or clinician in determining a diagnosis. The measurements should not be used as a sole means for determining a patient's diagnosis.
#### Indications for Use:
The analysis algorithm is indicated for use in exercise ECG testing where the clinician decides to evaluate the electrocardiogram of patients at 10 years or older, as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to rule out causes for symptoms. The analysis algorithm is not intended to be used as a physiological monitor.
| Prescription Use (21 CFR 801 Subpart D) | X |
|---------------------------------------------|---|
| AND/OR | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
Philips Medical Systems
K112959 CAlg-STR Exercise ECG Analysis Algorithm Response
(Division Sign-Off)
Division of Cardiovascular Devices
| Page | 52 of 57 |
|---------------|----------|
| 510(k) Number | K112959 |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.