HERO, VERSION 2.0

K081473 · Medical Predictive Science Corp. · DPS · Jun 27, 2008 · Cardiovascular

Device Facts

Record IDK081473
Device NameHERO, VERSION 2.0
ApplicantMedical Predictive Science Corp.
Product CodeDPS · Cardiovascular
Decision DateJun 27, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2340
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Indications for Use

HeRO is intended to acquire, store, analyze, and report on ECG data collected from infants. HeRO is intended to be used under the direct supervision of a licensed health care practitioner in a hospital neonatal or pediatric ICU environment. HeRO is intended to be used for the analysis of the variability in RR intervals (heart rate) and to report measurements of the variability of heart rate data (HRV). The HRV measurements reported by HeRO are specialized in nature, and intended to identify periods of transient decelerations and/or reduced baseline variability in the heart rate. HeRO is intended to provide only specialized HRV measurements and is not intended to produce any interpretation of those measurements or any kind of diagnosis. The specialized HRV measurements produced by HeRO have not been approved by the FDA for any specific clinical diagnosis. HeRO acquires data from a user-supplied ECG monitor, and requires a user-supplied local area network.

Device Story

HeRO Version 2.0 is a microprocessor-based system for neonatal/pediatric ICU monitoring. It acquires ECG data from user-supplied bedside monitors via a local area network. Data is stored and processed on a central server (CMS). The system analyzes RR intervals to calculate HRV metrics, specifically identifying transient decelerations and reduced baseline variability. Results are displayed on a remote monitoring station (RMS) for clinician review. The device provides specialized measurements only; it does not perform automated diagnosis or clinical interpretation. It serves as an adjunct tool for clinicians to monitor heart rate patterns in infants.

Clinical Evidence

Bench testing only. The system was tested in accordance with UL 60601-1 and IEC 60601-1-2 standards. No clinical study data provided.

Technological Characteristics

Microprocessor-based system; off-the-shelf PC hardware; data acquisition device (DAD); central monitoring station (CMS); remote monitoring station (RMS). Connectivity via user-supplied local area network. Compliant with UL 60601-1 and IEC 60601-1-2.

Indications for Use

Indicated for infants in neonatal or pediatric ICU environments for the analysis of heart rate variability (HRV) from ECG data to identify transient decelerations or reduced baseline variability. Not for diagnostic use.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K081473 ps10f2 Special 510(k) Application #### 510(k) Summary 3 | JUN | 27 2008 | |-----|---------| |-----|---------| Medical Predictive Science Corporation Date: 5/23/08 Submitter: Contact: William E King CEO (800) 394 1625 (434) 220 0714 (Direct) (240) 220 6098 (FAX) wking@mpsc.biz 21CFR 870.2340 74DPS, Class II 2246 Ivy Rd, Suite 17 Charlottesville, VA 22903 Trade Name: Common Name: Classification Name: Classification Number: Predicate Device: K021230 (HeRO) HeRO Version 2.0 HRV Analysis System Electrocardiograph Device Description: HeRO Version 2.0 is comprised of off-the-shelf Personal Computers (PC's) and special-purpose hardware capable of acquiring, storing, analyzing, and reporting ECG data from the cardiac monitoring devices. Data is acquired on a special-purpose Data Acquisition Device (DAD). It is stored and analyzed on a server, also called the Central Monitoring Station (CMS), located in a data closet or other suitable location. Results of the analyses are reported on a viewing station, or Remote Monitoring Station (RMS). The analysis algorithms identify Heart Rate Variability (HRV) patterns that reflect transient decelerations and/or reduced baseline variability. {1}------------------------------------------------ K081473 pg 2 of 2 Special 510(k) Application | Intended Use: | HeRO is intended to acquire, store, analyze, and report<br>on ECG data collected from infants. HeRO is intended<br>to be used under the direct supervision of a licensed<br>health care practitioner in a hospital neonatal or<br>pediatric ICU environment. | |---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | HeRO is intended to be used for the analysis of the<br>variability in RR intervals (heart rate) and to report<br>measurements of the variability of heart rate data<br>(HRV). The HRV measurements reported by HeRO<br>are specialized in nature, and intended to identify<br>periods of transient decelerations and/or reduced<br>baseline variability in the heart rate. | | | HeRO is intended to provide only specialized HRV<br>measurements and is not intended to produce any<br>interpretation of those measurements or any kind of<br>diagnosis. | | | The specialized HRV measurements produced by<br>HeRO have not been approved by the FDA for any<br>specific clinical diagnosis. | | | HeRO acquires data from a user-supplied ECG monitor,<br>and requires a user-supplied local area network. | | Technology: | Microprocessor based system. | | Test Summary: | HeRO Version 2.0 was developed and is manufactured<br>in accordance with 21CFR820 Quality System<br>Regulations. The HeRO system has been third-party<br>tested in accordance with UL 60601-1 and IEC 60601-<br>1-2. | | Conclusion: | HeRO Version 2.0 has an identical intended use as the<br>predicate (unmodified) device. HeRO Version 2.0<br>raises no new questions of safety or efficacy when<br>compared with the predicate device. Therefore, HeRO<br>Version 2.0 is judged substantially equivalent to HeRO. | {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is positioned to the right of a circular inscription that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 7 2008 Medical Predictive Science Corporation c/o Mr. William E. King CEO 2246 Ivy Rd. Suite 17 Charlottesville, VA 22903 Rc: K081473 Trade/Device Name: HeRO Version 2.0 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: May 23, 2008 Received: May 27, 2008 Dear Mr. King: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Mr. William E. King CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use Statement 2 | 510(k) Number (if known): | K08 1473 | |---------------------------|----------| | Device Name: | HeRO | ### Indications for Use: HeRO is intended to acquire, store, analyze, and report on ECG data collected from infants. HeRO is intended to be used under the direct supervision of a licensed health care practitioner in a hospital neonatal or pediatric ICU environment. HeRO is intended to be used for the analysis of the variability in RR intervals (heart rate) and to report measurements of the variability of heart rate data (HRV). The HRV measurements reported by HeRO are specialized in nature, and intended to identify periods of transient decelerations and/or reduced baseline variability in the heart rate. HeRO is intended to provide only specialized HRV measurements and is not intended to produce any interpretation of those measurements or any kind of diagnosis. The specialized HRV measurements produced by HeRO have not been approved by the FDA for any specific clinical diagnosis. HeRO acquires data from a user-supplied ECG monitor, and requires a usersupplied local area network. ## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ﺎ (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) (Division Sigh-Off) Division of Cardiovascular Devices K08 413 510(k) Number - 2.1 -
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...