VOYAGER 12

K062469 · Del Mar Reynolds Medical, Ltd. · DPS · Nov 3, 2006 · Cardiovascular

Device Facts

Record IDK062469
Device NameVOYAGER 12
ApplicantDel Mar Reynolds Medical, Ltd.
Product CodeDPS · Cardiovascular
Decision DateNov 3, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2340
Device ClassClass 2

Indications for Use

The use of Voyager 12 is indicated when it is desired to record and interpret an electrocardiogram of a patient. This includes a resting electrocardiogram and a rhythm electrocardiogram.

Device Story

Voyager 12 is a portable, self-contained 12-lead digital electrocardiograph featuring an internal printer. It captures ECG signals from patients to provide both manual and automatic interpretation of resting and rhythm electrocardiograms. Designed for clinical use, the device allows healthcare providers to record, analyze, and print ECG data to assist in cardiac assessment and clinical decision-making.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including compliance with IEC 60601-1, IEC 60601-2-25, IEC 60601-2-51, and IEC 60601-1-2, as the device utilizes the same technology as the predicate devices.

Technological Characteristics

Portable 12-lead digital electrocardiograph; internal printer; electrical safety and performance per IEC 60601-1, IEC 60601-2-25, IEC 60601-2-51, and IEC 60601-1-2.

Indications for Use

Indicated for patients requiring resting or rhythm electrocardiogram recording and interpretation.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K062469 Del Mar Reynolds Medical Voyager 12 Added Information Page 19 # Del Mar Reynolds Medical Ltd. Special 510(k) 510(k) Summary Voyager 12 Revised October 18, 2006 #### 1. Submitter Information Name: Del Mar Reynolds Medical Ltd. Address: 1 Harforde Court John Tate Road Hertford, Herts SG13 7NW ENGLAND Telephone Number: +44-1992-507700 Contact Person: Dr. George Myers (Official Correspondent) Medsys Inc. 377 Route 17 S Hasbrouck Heights, NJ 07604 Telephone 201-727-1703 Fax 201-727-1708 Date Prepared: 19 October 2006 #### 2. Name of Device Trade Namc: Voyager 12 Common Name: 12-Lead Digital Electrocardiograph Classification name: Electrocardiograph #### 3. Equivalent legally-marketed devices. - Del Mar Reynolds Medical CardioDirect 121, K024283 - Burdick Eclipse 4, K946281 #### 4. Description NOV - 3 2006 P.1/2- {1}------------------------------------------------ Voyager 12 is a 12-lead digital electrocardiograph with interpretation of the electrocardiogram. Voyager 12 is a portable, self-contained unit with an internal printer. It permits either manual or automatic interpretation of electrocardiograms. ત્ત્વનું નિર્માનના P. 4-12. # న. Intended Use Voyager 12 is intended to be used record and interpret an electrocardiogram of a patient. This includes a resting electrocardiogram and a rhythm electrocardiogram. # 6. Performance Data - (a) Non-clinical tests - I. IEC 60601-1 - 2. IEC 60601-2-25 - 3. IEC 60601-2-51 - 4. Validation tests - 5. IEC 60601-1-2 - (b) Clinical tests Clinical tests are not necessary since Voyager 12 uses the same technology as the predicate device. # Conclusions (c) Voyager 12 is equivalent in safety and efficacy to the legally marketed predicate devices. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle's head is facing left. Surrounding the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV - 3 2006 Medsys, Inc. c/o Mr. George Myers President 377 Route 17 S Hasbrouck Heights, NJ Rc: K062469 Trade Name: Voyage 12 ECG Interpretation Cart Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: October 24, 2006 Received: October 25, 2006 Dear Mr. Myers: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the cnclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ # Page 2 - Mr. George Myers or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1000. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logally marketed predicate device results in a classification for your device and thus, permitts vor device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may althus, other general information on your responsibilities under the Act from the Division of Vouall Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Blymmmer for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K062469 Del Mar Reynolds Medical Voyager 12 Added Information Page 39 Indications for Use 510(k) Number (if known): K062469 Device Name: Voyager 12 Indications for Use: The use of Voyager 12 is indicated when it is desired to record and interpret an electrocardiogram of a patient. This includes a resting electrocardiogram and a rhythm electrocardiogram, Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page / of / (Posted November 13, 2003) Back to the Indications for Use Page CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer EDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page Center for Devices and Radiological B. Hemmerman Sign-Of on of Cardlovascular Devices 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...