The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
Device Story
AngioSculpt PTA Scoring Balloon Catheter; over-the-wire (OTW) angioplasty catheter; incorporates nitinol scoring element over conventional balloon. Input: guidewire for navigation to stenosis. Operation: balloon inflation with contrast medium expands balloon; nitinol scoring element focuses force to dilate lesions. Output: mechanical dilation of arterial or fistula stenosis. Used in clinical settings by physicians. Benefits: effective lesion dilation via focal force application. Single-use; sterile.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing including balloon profile, rated burst pressure, fatigue, bond/tip pull strength, inflation/deflation time, and biocompatibility (ISO 10993-5).
Technological Characteristics
Nitinol scoring element over conventional balloon; dual-lumen catheter; radiopaque markers; 6F sheath compatibility; 0.018" guidewire compatibility. Materials biocompatible per ISO 10993-5. Sterilization per AAMI TIR:28 2009.
Indications for Use
Indicated for dilatation of lesions in iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Contraindicated for coronary or neuro-vasculature use.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K081220 — MODIFICATION TO ANGIOSCULPT PTA SCORING BALLOON CATHETER · Angioscore, Inc. · May 28, 2008
K123358 — PACIFIC PLUS · Medtronic Vascular · Jan 29, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem depicts a stylized human figure with three faces in profile, representing the department's focus on health and human well-being. The logo is simple and recognizable, conveying the department's mission and purpose.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 31, 2016
The Spectranetics Corporation Ms. Kimberley Kline Senior Manager, Regulatory Affairs 9965 Federal Drive Colorado Springs, CO 80921
Re: K150634
Trade/Device Name: AngioSculpt PTA Scoring Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PNO Dated: March 10, 2015 Received: March 11, 2015
Dear Ms. Kline,
This letter corrects our substantially equivalent letter of April 8, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Misti L. Malone -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K150634
#### Device Name
AngioSculpt PTA Scoring Balloon Catheter, 7.0/8.0mm x 40mm
#### Indications for Use (Describe)
For dilatation of lesions in the iliac, femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
K150634 Page 1 of 3
Image /page/3/Picture/1 description: The image shows the Spectranetics logo. The logo consists of a blue geometric shape to the left of the word "Spectranetics" in blue, with the registered trademark symbol to the right. Below the word "Spectranetics" is the tagline "Always Reaching Farther" in a smaller font.
## 510(k) Summary
This 510(k) summary was prepared in accordance with 21 CFR 807.92
Prepared on March 10, 2015
| 510(k) Submitter / Holder: | Spectranetics<br>5055 Brandin Court<br>Fremont, CA 94538<br>Establishment Registration No: 3005462046 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Kimberley Kline<br>Senior Manager, Regulatory Affairs<br>Direct: 510.933.7989<br>Fax: 510.933.7994<br>Email: Kimberley.Kline@Spectranetics.com |
#### Subject Device
| Device Trade Name: | AngioSculpt® PTA Scoring Balloon Catheter |
|----------------------------|-------------------------------------------|
| Device Common Name: | Angioplasty catheter |
| Device Class: | II |
| Classification Regulation: | 21 CFR 870.1250 |
| Regulation Description: | Percutaneous catheter |
| Product Code: | LIT |
| 510(k) Type: | Special |
| Model Numbers: | 2332, 2333, 2334 |
#### Predicate Device
The AngioSculpt PTA Scoring Balloon Catheter was compared to the following legally marketed predicate device:
| Trade Name/510(k) Number: | AngioSculpt® PTA Scoring Balloon Catheters<br>K133998 4.0-6.0x200mm (137cm length) |
|---------------------------|------------------------------------------------------------------------------------|
| Manufacturer: | Spectranetics |
| Device Common Name: | Angioplasty catheter |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Spectranetics. The logo consists of a blue diamond shape with a white center, followed by the word "Spectranetics" in blue, sans-serif font. Below the word "Spectranetics" is the tagline "Always Reaching Farther" in a smaller, sans-serif font.
#### Device Description
The 7.0/8.0mm x 40mm AngioSculpt PTA Scoring Balloon Catheters are comprised of a conventional balloon catheter that incorporates a nitinol scoring element over the balloon. The balloon is designed to expand to a specified diameter at a specified pressure. The catheter has radiopaque markers to aid in positioning of the balloon in the stenosis. The catheter has two lumens; one lumen is used for inflation of the balloon with contrast medium; the other lumen permits the use of a quide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The product is provided in the over-the-wire (OTW) delivery platform. The catheters have a sheath compatibility of 6F and guidewire compatibility of 0.018". The catheter is single-use only and provided sterile to the user.
#### Intended and Indications for Use
The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
#### Technological Characteristics
The 7.0/8.0mm x 40mm AngioSculpt PTA Scoring Balloon catheters are equivalent to the currently cleared and marketed 4.0-6.0mm x 40mm and 200mm AngioSculpt family of devices because they share substantially the same; indications for use, device operation, material type, basic design confiquration and overall technical and functional capabilities.
#### Performance Data
The following testing was conducted to validate and verify that the subject device was substantially equivalent to the predicate devices
#### Design Verification and Validation Testing
- Catheter Diameter and Balloon Profile ●
- Balloon Preparation, Deployment and Retraction
- Rated Burst Pressure (RBP)
- Device Fatigue
- Balloon Compliance (Diameter vs. Pressure)
- Balloon Inflation/ Deflation Time ●
- Bond (Tensile) Strength
- Tip Pull Strength
- Catheter Diameter and Balloon Profile (post deflation) ●
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Spectranetics logo. The logo consists of a blue diamond shape with a white circle in the center, followed by the word "Spectranetics" in blue, bold font. Below the word "Spectranetics" is the tagline "Always Reaching Farther" in a smaller, regular font.
- Catheter Flexibility and Kink ●
- Torque Strength
- Radiopacity ●
- Pushability, Trackability and Secure Edges ●
- Freedom from Stent Interference
- Guide Wire Compatibility
- Catheter Effective Length
- Catheter Surface Appearance ●
- Luer Compatibility
- o Focal Force
#### Sterilization
- Product adoption equivalency per AAMI TIR:28 2009
#### Biocompatibility
The materials used in the subject device pose no greater significant biocompatibility challenge than the current AngioSculpt products. Biocompatibility test results of the currently marketed product have previously been performed and are biocompatible. Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Hemocompatibility, and Genotoxicity tests were conducted according to current standards (e.g. ISO 10993-5). Due to similarities in design, manufacturing processes, sterilization, intended use, and device categorization, the combined biocompatibility test results of predicate product can be used to confirm the biocompatibility of the subject device.
#### Preclinical and Clinical Data
Preclinical and clinical data were not required to demonstrate substantial equivalence. The design characteristics of the subject device are similar to the predicate. The design verification and validation test results demonstrated that the subject device is as safe and clinically effective as the predicate device.
#### Substantial Equivalence
Based on the similarities in design between the subject and predicate devices, and the performance testing performed, the subject AngioSculpt PTA Scoring Balloon Catheter, 7.0/8.0mm x 40mm, 50,90,137cm is substantially equivalent to the predicates, AngioSculpt 4.0-6.0x200mm (50, 90, 137cm length) devices cleared under K133998 and the 4.0-6.0 x 40mm AngioSculpt devices cleared under K091966.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.