THE CROSSER SYSTEM

K072776 · Flowcardia, Inc. · PDU · Dec 7, 2007 · Cardiovascular

Device Facts

Record IDK072776
Device NameTHE CROSSER SYSTEM
ApplicantFlowcardia, Inc.
Product CodePDU · Cardiovascular
Decision DateDec 7, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The CROSSER System is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions. The device is contraindicated for use in carotid arteries. The CROSSER Catheter is only intended for use with the CROSSER Electronics System. Refer to the CROSSER Electronics System Manual of Operations for proper use.

Device Story

The CROSSER System is a percutaneous catheter system used to navigate through peripheral artery chronic total occlusions. The system comprises a reusable electronic generator, a foot switch, a high-frequency transducer, and a single-use catheter. The generator and transducer convert AC power into high-frequency mechanical vibrations, which are transmitted through a Nitinol core wire to a Titanium tip on the catheter. The device is operated by a physician in a clinical setting. By applying mechanical vibration to the occlusion, the device facilitates the passage of a conventional guidewire, potentially improving procedural success in treating chronic total occlusions. The catheter features a Pebax body with a hydrophilic coating.

Clinical Evidence

Clinical performance was verified through a human clinical study designed to demonstrate the intended use of the device. No specific numerical metrics (e.g., success rates, p-values) were provided in the summary document.

Technological Characteristics

System components: electronic generator, foot switch, transducer, and catheter. Catheter materials: Nitinol core wire, Titanium tip, Pebax body, hydrophilic coating. Energy source: AC power converted to high-frequency mechanical vibrations. Connectivity: Catheter connects to generator via transducer. Sterilization: Not specified.

Indications for Use

Indicated for patients with peripheral artery chronic total occlusions to facilitate intra-luminal placement of conventional guidewires. Contraindicated for use in carotid arteries.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the text "FLOWCARDIA, INC." in a bold, sans-serif font. There are two horizontal lines below the text. To the right of the text, there is a sine wave pattern. ## 5) 510(k) Summary This summary of 510(k) information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21CFR807.92. **510(k) Number** K072776 | <b>Applicant Information</b> | | |------------------------------|--------------------| | Date Prepared: | September 27, 2007 | | Name and Address: | FlowCardia, Inc. | ・・ 2007 ្រីក | | 745 N. Pastoria Avenue | |-----------------|----------------------------------------| | | Sunnyvale, CA 94085 | | | Ph: (408) 617-0352 | | Contact Person: | Dustin Michaels, Sr. Director of RA/CF | | | Ph: (408) 617-0352 x302 | | | Fax: (408) 617-9198 | # Device Information | Classification: | DQY | |----------------------|-------------------------------------------------| | Trade Name: | The CROSSER System | | Common Name: | Percutaneous Catheter | | Classification Name: | Percutaneous Catheter, 74 DQY / 21 CFR 870.1250 | # Predicate Device The CROSSER System manufactured by FlowCardia, Inc. (K062868) # Indications for Use: The CROSSER System is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions. The device is contraindicated for use in carotid arteries. The CROSSER Catheter is only intended for use with the CROSSER Electronics System. Refer to the CROSSER Electronics System Manual of Operations for proper use. {1}------------------------------------------------ # Device Description The CROSSER System consists of a re-usable electronic Generator, Foot Switch, highfrequency Transducer, and single-use CROSSER Catheter. The CROSSER Catheter is available in three sizes: - CROSSER 14P (014 guidewire compatible), . - CROSSER 14S (014 guidewire compatible supportive), . - CROSSER 18 (018 guidewire compatible). . Each Catheter is connected to the electronic Generator through the Transducer. The Foot Switch is used to activate the system. The Generator and Transducer convert AC power into high-frequency mechanical vibrations which are propagated through a Nitinol core wire to the Titanium tip of the CROSSER Catheter. The main body of the catheter is constructed from Pebax and a hydrophilic coating which covers the distal end of the catheter. ## Technological Characteristics The predicate and CROSSER System catheters are substantially equivalent with respect to materials, design construction and performance. The CROSSER System catheters differ only slightly with respect to tip diameter, tip material and guidewire lumen length. These differences are due to the intended use of the CROSSER Catheters over the predicate device; the predicate device is indicated for use in coronary chronic occlusions and the CROSSER System is indicated for use in peripheral artery occlusions. ### Physical Test Data Design analysis, bench, and biocompatibility testing were conducted according to the relevant guidance documents to demonstrate that the FlowCardia CROSSER System met the acceptance criteria and performed similarly to the predicate device. In addition to dimensional verification, the following functional tests were performed: Tensile Strength, Torque Strength, Torqueability, Tip Flexibility, Coating Adherence/Integrity, Biocompatibility, Bench top Simulated Efficiency, Catheter Fatigue Testing, Shelf Life and Package Integrity Testing. #### Clinical Test Data The clinical performance of the CROSSER System was verified through a human clinical study designed to demonstrate the intended use of the device. #### Conclusion Based upon device physical and clinical comparisons the CROSSER System is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 FlowCardia, Inc. Mr. Dustin Michaels Senior Director of RA/CR 745 N. Pastoria Avenue Sunnyvale, CA 94085 SEP 1 8 2013 Re: K072776 Trade/Device Name: The CROSSER System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: September 27, 2007 Received: September 28, 2007 Dear Mr. Michaels: This letter corrects our substantially equivalent letter of December 7, 2007. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Zm.z- for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Statement of Indications for Use 510(k) Number (if known): K072776 Devicc Name: # The CROSSER System Indications for Use: The CROSSER System is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions. The device is contraindicated for use in carotid arteries. The CROSSER Catheter is only intended for use with the CROSSER Electronics System. Refer to the CROSSER Electronics System Manual of Operations for proper use. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Duna R. Vc Ames (Division Sir. .off) Division ci Cardiovascular Devices 510(k) number k072776
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...