ZYMED TELEMETRY SYSTEM: MODEL EASIVIEW

K990093 · Agilent Technologies · DSI · May 11, 1999 · Cardiovascular

Device Facts

Record IDK990093
Device NameZYMED TELEMETRY SYSTEM: MODEL EASIVIEW
ApplicantAgilent Technologies
Product CodeDSI · Cardiovascular
Decision DateMay 11, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Intended Use

Assessment of symptoms that may be related to rhythm disturbances of the heart: Patients with . palpitations; The evaluation of arrhythmia's in patients from pediatric to adult age. . Assessment of risk in patients with or without symptoms of arrhythmia. . . Assessment of efficacy of Antiarrhythmic therapy. . . Assessment of Pacemaker Function. . . Assessment of real time ST segment analysis . . Assessment of symptomatic or asymptomatic patients, to evaluate for, ischemic heart disease and arrhythmia analysis during exercise testing. . . Assessment of transferred SpO2 level date from an MDE Escort Bedside monitor to EasiView for display and alarm. . . Assessment of transferred Invasive and Non-Invasive blood pressure, both systolic and diastolic, data from an MDE Escort Bedside monitor to EasiView for display and alarm. . Assessment of transferred temperature parameter data from an MDE Escort Bedside monitor to ● EasiView for display and alarm. . Assessment of transferred respiration data from an MDE Escort Bedside monitor to Easi View for display. . Assessment is for single-hospital environment.

Device Story

Zymed EasiView Telemetry System provides real-time cardiac monitoring for up to eight patients; inputs include ECG signals from transmitters (VHF band 174-216 MHz) and multi-parameter data (SpO2, NIBP, IBP, temperature, respiration) transferred from MDE Escort bedside monitors. System performs real-time ECG waveform analysis to detect/classify abnormal beats and rhythm disturbances; features include automatic ST analysis, trend data, and full disclosure storage. Used in hospital settings by clinicians; output displayed on central station monitor with visual/audible alarms. Clinicians use data for diagnosis, treatment planning, and assessment of antiarrhythmic therapy efficacy. System benefits patients through continuous monitoring and automated detection of cardiac events.

Clinical Evidence

Bench testing only. Performance validated against industry-standard AHA, MIT, and European ST-T (EST) databases. Metrics included sensitivity (SE), positive predictivity (+P), and false positive rate (FPR) for QRS, ventricular beats, couplets, and runs. Episode sensitivity (ESE), episode positive predictivity (E+P), duration sensitivity (DSE), and durative predictivity (D+P) measured for ST analysis. System demonstrated performance equal to or better than the predicate in high heart rates (>300 bpm) and noisy environments (baseline, electrode, or muscle noise).

Technological Characteristics

IBM PC AT compatible platform; 200 MHz Pentium Pro CPU; 64 MB RAM; 1.6 GB hard disk; SVGA display. Transmitters operate on tunable VHF frequencies (174-216 MHz). Supports 4, 6, or 8 beds. Software runs on Windows NT/DOS. Oscillometric NIBP measurement. Connectivity via network/strip chart/laser printer. Sterilization not applicable (non-invasive).

Indications for Use

Indicated for pediatric to adult patients requiring cardiac rhythm monitoring, arrhythmia evaluation, pacemaker function assessment, ST segment analysis, and monitoring of SpO2, blood pressure, temperature, and respiration in a single-hospital environment.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 1 1 1999 7 vmed K990093 ## 1 Zymed Inc. 20 North Aviador Street Camarillo, California 93010-8348 800.235.5941 - 805.987.9611 Fax 805.987.9532 ## 510(k) Summary Submitter: Dudley Harris, Director of Regulatory Affairs/QA Zymed Medical Instrumentation 20 North Aviador Street Camarillo, CA 93010 805/987-9532 Fax: 800/235-5941 (401) Phone: Date of Summary: 1-8-99 D. Harris - see above Contact: Trade Name: Common Name: Classification Name: (per 21 CFR 870.1025) Zymed Telemetry System: Model EasiView Telemetry Central Station Monitor Detection and Alarm, Arrhythmia Legally marketed device to which S.E. is claimed. Zymed Telemetry System: Model EasiView - 510(k) K 980186 Medical Data Electronics: Escort 100/300 Series B Patient Monitor System - 510(k) K930004/A Description: A Zymed Telemetry monitoring system consists of a series of interface devices to include ECG transmitters, multi-parameter transmitters, a central Telemetry monitoring computerized unit with a strip chart recorder, Easi 5 lead and laser printer. The Zymed central monitor supports up to eight patients for real time cardiac monitoring. The system displays each patient's ECG continuously on the screen while performing real time ECG waveform analysis for all eight patients. This analysis permits immediate detection and classification of abnormal beats, cardiac rhythm disturbances and variations. The system will also display each patient's Oxygen Saturation, Respiration, Non-Invasive Blood Pressure, Invasive Blood Pressure and Temperature. Each ECG transmitter's frequency can be programmed to operate at any frequency within the entire VHF band. For US domestic sites, the transmitters will comply with FCC band allocations (174-216 Mhz). In addition to ECG data, the transmitters also detect and transmit cardiac pacemaker information. Other information including transmitter status and individual lead impedance is also transmitted to the Zymed system for overall system safety and efficacy. The Zymed system presents the user with a number of clinical tools such as visual and audible alarms for the diagnosis of patients with various heart conditions. The system also provides tools to review a patient's {1}------------------------------------------------ cardiac performance. On-line review mechanisms as well as detailed analysis screens have been designed into the system to facilitate and to enhance the patient's diagnosis and treatment. Features such as individual ECG printouts, multi-channel automatic ST analysis, trend data analysis, and Full Disclosure data further enhance the system's qualities as a valuable and practical clinical tool. The Zymed system presents the user with a number of clinical tools such as visual and audible alarms for multi-parameter monitoring of the patient. The Zymed system provides a means to monitor multiple patients from a central monitoring station for ECG, Oxygen Saturation, Non-Invasive Blood Pressure, Invasive Blood Pressure, Temperature, and Respiration. The system has the following options available: Choice of 4, 6, or 8 bed central monitor Full disclosure screen and printout (full resolution programmable from 0 to 168 hours) Choice of 6 lead sets, based on transmitter capability Full arrhythmia analysis to include multi-channel automatic ST Analysis Laser Printer, print server options Strip Chart Networking Intended use: - Assessment of symptoms that may be related to rhythm disturbances of the heart: Patients with . palpitations; The evaluation of arrhythmia's in patients from pediatric to adult age. - . Assessment of risk in patients with or without symptoms of arrhythmia. - . Assessment of efficacy of Antiarrhythmic therapy. - . Assessment of Pacemaker Function. - . Assessment of real time ST segment analysis - . Assessment of symptomatic or asymptomatic patients, to evaluate for, ischemic heart disease and arrhythmia analysis during exercise testing. - . Assessment of transferred SpO2 level date from an MDE Escort Bedside monitor to EasiView for display and alarm. - . Assessment of transferred Invasive and Non-Invasive blood pressure, both systolic and diastolic, data from an MDE Escort Bedside monitor to EasiView for display and alarm. - Assessment of transferred temperature parameter data from an MDE Escort Bedside monitor to ● EasiView for display and alarm. - Assessment of transferred respiration data from an MDE Escort Bedside monitor to Easi View for display. - Assessment is for single-hospital environment. A review of the technological characteristics compared to the predicate devices are: {2}------------------------------------------------ | Platform | EasiView (New)<br>telemetry system | EasiView (Old)<br>telemetry system | |-------------------------------|------------------------------------|------------------------------------------------------| | Type | IBM PC AT Compatible | Same | | CPU | 200 Mhz Pentium Pro | Same | | RAM | 64 M Bytes | Same | | Hard Disk | 1.6 G Bytes | Same | | Display | SVGA | Same | | Transmitters: | | | | 3 channel | yes | yes | | Tunable frequencies | 174-216 Mhz | Same | | Software: | | | | Number of patients | 4, 6, or 8 | Same | | Operating System | Windows NT<br>DOS Compatible | Same | | Data Storage | 24 hrs of EKG/Channel | Same | | Number of leads for Analysis: | 3 | Same | | Automatic ST Analysis | yes | Same | | Manual ST Analysis | no | no | | Platform | EasiView (New)<br>telemetry system | MDE Escort Monitor System (Old)<br>monitoring system | | Temperature: | | | | Range | 20-50 degrees C | Same | | Alarms | 0-150 F (0-65 C) | 0.1-44.0 C | | Non-Invasive BP | | | | Measurement Method | Oscillometric | Same | | Alarms | | | | Systolic | 0-300 mmHg | 25-240 mmHg | | Mean | NA | 15-200 mmHg | | Diastolic | 0-200 mmHg | 10-180 mmHg | | Invasive BP | | | | Number of Channels | 1 or 2 | Same | | Alarms | | | | Systolic | 0-300 mmHg | 0-240 mmHg | | Mean | NA | 0-200 mmHg | | Diastolic | 0-200 mmHg | 0-180 mmHg | | SpO2 | | | | Range | 20- 100% | Same | | Alarms | 0-100% by 1%<br>0-100% by 1 % | 70-100% high limit<br>50-99% low limit | | Respiration | | | | Range | 4-200 | Same | Performance was measured against industry accepted AHA (AHA), MIT (MIT) and European ST-T (EST) databases. Results were typical for the real time monitoring environment for the Easi View as targeted. Separate sensitivities (SE), positive predictivity (+P), and false positive rate (FPR) were examined for each database and measured for QRS, Ventricular, Couplets, Short runs and Long runs. Separate Episode {3}------------------------------------------------ Sensitivities (ESE), Episode Positive Predictivity (E+P), Duration Sensitivity (DSE) and Durative Predictivity (D+P) were examined for the European ST-T(EST) database and measured for ST analysis. High heart rates to include pediatric patients were demonstrated to be within recommended guidelines in excess of 300 bpm, and performance in the presence of noise indicates the new system is equal to or better than the old system when looking at baseline, electrode or muscle as the cause of noise. In summary, performance data between the two systems shows nearly identical data, and therefore, supports a claim of Substantial Equivalence. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 1 1 1999 Mr. Dudley Harris Director of RA/QA Zymed, Inc. 20 North Aviador Street Camarillo, CA 93010-8348 Re: K990093 Easiview Telemetry Central Station Monitor Regulatory Class: III (three) Product Code: DSI Dated: April 20, 1999 Received: April 21, 1999 Dear Mr. Harris: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food Drug and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this {5}------------------------------------------------ response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, 2Misbranding by reference to premarket notification² (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J.Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## 510(k) Number: K990093 Device Name: EasiView Telemetry Monitoring System Indications for Use: - . Assessment of symptoms that may be related to rhythm disturbances of the heart; Patients with palpitations: The evaluations of arrhythmia's in patients from pediatric to adult age. - Assessment of risk in patients with or without symptoms of arrhythmia. . - Assessment of efficacy of Antiarrhythmia therapy. . - Assessment of Pacemaker Function. - . Assessment of real time ST segment analysis. - Assessment of symptomatic or asymptomatic patients, to evaluate for, ischemic heart disease . and arrhythmias analysis during exercise testing. - Assessment of transferred SpO2 level data from an MDE Escort Bedside monitor to ● EasiView for display and alarm. - Assessment of transferred invasive and Non-invasive blood pressure, both systolic and . diastolic, data from an MDE Escort Bedside monitor to EasiView for display and alarm. - Assessment of transferred temperature parameter data from an MDE Escort Bedside monitor . to EasiView for display and alarm. - Assessment of transferred respiration data from an MDE Escort Bedside monitor to EasiView . for display. - Assessment is indicated for single-hospital environment. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) F.K. Jones for the Prescription Use (CFR21 CFR 801.109) Over-The-Counter Use
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...