The Nihon Kohden Vital Sign Telemeter GZ-140P is intended to monitor and transmit physiological data from a patient to a Nihon Kohden monitor via radiofrequency in the 802.11 band for continuous monitoring. GZ-140P transmits electrocardiogram (ECG), respiration data, blood oxygen saturation (SpO2) and non-invasive blood pressure (NIBP). The device may generate an audible and or visible alarm when an arrhythmia exists, when a measured physiological rate falls outside preset limits, or when a technical error is detected. Furthermore, the device can be configured for use as a temporary simple monitor to display the patient's vital signs on the device screen and generate alarms without transmitting the data to other Nihon Kohden monitor. The device is intended to be used by qualified medical facility, such as hospital or clinic, on all patient populations including adult, neonate, infant, child, and adolescent subgroups, with the exception of NIBP measurement which is not intended for use on neonates.
Device Story
Telemetry system for medical facilities; monitors and transmits ECG, respiration, SpO2, and NIBP data via 802.11 wireless LAN to central monitoring station. Operates as temporary simple monitor with local display and alarms. Used by qualified medical personnel. Inputs: physiological signals via sensors. Outputs: vital sign data, arrhythmia alerts, technical alarms. Clinical decision-making supported by real-time monitoring and alarm generation for physiological rate deviations or arrhythmias. Benefits: continuous patient surveillance and timely intervention.
Clinical Evidence
Bench testing only. Includes software verification/validation, unit/integration/system testing, wireless coexistence/design validation, and electromagnetic safety compliance. NIBP inflation mode validated via clinical testing per ISO 81060-2. Compliance with ANSI/AAMI EC57 for cardiac rhythm/ST-segment algorithms and various IEC/ISO standards for safety, alarm systems, and physiological monitoring.
Technological Characteristics
Wireless telemeter (802.11). NIBP measurement via oscillometric method (deflation and inflation modes). ECG, SpO2, respiration monitoring. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 60601-2-49, IEC 80601-2-30, ISO 80601-2-61, ISO 81060-2, ANSI/AAMI EC57, IEC 62304, ISO 10993-5/10, ISO 14971.
Indications for Use
Indicated for continuous monitoring of ECG, respiration, SpO2, and NIBP in all patient populations (adult, neonate, infant, child, adolescent) in hospitals or clinics. NIBP measurement is contraindicated for neonates.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Nihon Kohden BSM-6000 Series Bedside Monitor (K080342)
Nihon Kohden Vital Sign Telemeter GZ-120P and GZ-130P (K153707)
Nihon Kohden Afib Detection Program QP-039P (K152305)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a family of three people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 4, 2017
Nihon Kohden Corporation % Tom Bento Sr. Vice President, Quality And Regulatory Nihon Kohden America, Inc. 15353 Barranca Parkway Irvine, California 92618
Re: K163459 Trade/Device Name: Nihon Kohden Vital Sign Telemeter GZ-140P Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX Dated: March 7, 2017 Received: March 8, 2017
Dear Tom Bento:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Minde Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163459
Device Name
Nihon Kohden Vital Sign Telemeter GZ-140P
#### Indications for Use (Describe)
The Nihon Kohden Vital Sign Telemeter GZ-140P is intended to monitor and transmit physiological data from a patient to a Nihon Kohden monitor via radiofrequency in the 802.11 band for continuous monitoring. GZ-140P transmits electrocardiogram (ECG), respiration data, blood oxygen saturation (SpO2) and non-invasive blood pressure (NIBP). The device may generate an audible and or visible alarm when an arrhythmia exists, when a measured physiological rate falls outside preset limits, or when a technical error is detected. Furthermore, the device can be configured for use as a temporary simple monitor to display the patient's vital signs on the device screen and generate alarms without transmitting the data to other Nihon Kohden monitor.
The device is intended to be used by qualified medical facility, such as hospital or clinic, on all patient populations including adult, neonate, infant, child, and adolescent subgroups, with the exception of NIBP measurement which is not intended for use on neonates.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# Nihon Kohden Vital Sign Telemeter GZ-140P
| Submitter:<br>Address: | Nihon Kohden Corporation<br>1-31-4 Nishiochiai, 1-Chome, Shinjuku-Ku<br>Tokyo, Japan 161-8560 |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone number: | 81-2-59968020 |
| Contact person:<br>Phone number:<br>Fax number: | Tom Bento<br>(949) 680-9048<br>(913) 273-0732 |
| Date prepared: | December 06, 2016 |
| Device name:<br>Common name:<br>Device classification: | Nihon Kohden Vital Sign Telemeter GZ-140P<br>Patient monitor<br>Monitor, physiological, patient (with arrhythmia detection or alarms),<br>Transmitters and Receivers, Physiological Signal, Radio Frequency |
| Primary Product code:<br>Regulation: | MHX<br>21 CFR 870.1025 |
| Secondary Product Code:<br>Requlation: | DRG<br>21 CFR 870.2910 |
| Substantial equivalent<br>claimed to: | Nihon Kohden BSM-6000 Series Bedside Monitor (K080342) and<br>Nihon Kohden Vital Sign Telemeter GZ-120P and GZ-130P<br>(K153707) |
## Description:
The Nihon Kohden Vital Sign Telemeter GZ-140P is mainly used as a telemetry system within a medical facility. The device transmits patient's vital signs (ECG, SpO2, respiration, and blood pressure) and alarm information via wireless LAN connection to the central monitoring station. The device can be configured to display the patient's vital signs on the screen and generate alarms and used as a temporary simple monitor.
## Indications for Use:
The Nihon Kohden Vital Sign Telemeter GZ-140P is intended to monitor and transmit physiological data from a patient to a Nihon Kohden monitor via radiofrequency in the 802.11 band for continuous monitoring. GZ-140P transmits electrocardiogram (ECG), respiration data, blood oxygen saturation (SpO2) and non-invasive blood pressure (NIBP). The device may generate an audible and/or visible alarm when an arrhythmia exists, when a measured physiological rate falls outside preset limits, or when a technical error is detected. Furthermore, the device can be configured for use as a temporary simple monitor to display the patient's vital signs on the device screen and generate alarms without transmitting the data to other Nihon Kohden monitor.
{5}------------------------------------------------
The device is intended to be used by qualified medical personnel within a medical facility, such as hospital or clinic, on all patient populations including adult, neonate, infant, child, and adolescent subgroups, with the exception of NIBP measurement which is not intended for use on neonates.
## Technological Characteristics - Substantial Equivalence Discussion:
The Nihon Kohden Vital Sign Telemeter GZ-140P is substantially equivalent to the predicate devices, Nihon Kohden BSM-6000 Series Bedside Monitor (K080342) and Nihon Kohden Vital Sign Telemeter GZ-120P and GZ-130P. Differences between the devices are minor and do not raise questions reqarding safety or efficacy.
These differences include:
- . The GZ-140P is substantially equivalent to the GZ-120P/GZ-130P Vital Sign Telemeter other than the additional functions of NIBP measurement and atrial fibrillation detection.
- . The GZ-140P has two NIBP measurement modes: Conventional Mode (Deflation Mode) where NIBP measurements are made during deflation of the cuff and iNIBP (Inflation Mode) where NIBP measurements are made during inflation of the cuff. The Conventional Mode (Deflation Mode) is the same as that of the predicate device, BSM-6000 series bedside monitor. The iNIBP (Inflation Mode) measurement uses the same oscillometric method as the Conventional Mode (Deflation Mode) measurement and was validated in the clinical testing (Attachment 015). The similar technology with iNIBP is acquired by Welch Allyn Spot Ultra Vital Signs (K040490), where the blood pressure is measured while the device is inflating a cuff.
- . The atrial fibrillation detection algorithm is exactly same as that of Nihon Kohden Afib Detection Program QP-039P (K152305).
# Test Summary:
Performance testing for the Nihon Kohden Vital Sign Telemeter GZ-140P includes software verification and validation test, software unit test, integration test, and testing to compliance standards for electromagnetic safety. Wireless coexistence testing and evaluation was performed following FDA Guidance, "Radio Frequency Wireless Technology in Medical Device", and the device's immunity to proximity fields from radio frequency wireless communications equipment was validated. Traceability has been documented between all system specifications to validation test results.
# Standards compliance testing includes:
- ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012: General requirements for basic safety and ● essential performance
- IEC 60601-1-2: 2007 Medical Electrical Equipment Part 1-2: General requirements for Basic . Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
- . IEC 60601-1-8: 2006 & A1:2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- . IEC 60601-2-27:2011 Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
{6}------------------------------------------------
- IEC 60601-2-49:2011 Medical electrical equipment Part 2-49: Particular requirements for the . basic safety and essential performance of multifunction patient monitoring equipment
- . IEC 80601-2-30:2013 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for the basic safety and essential performance of pulse oximeter equipment
- . ISO 81060-2:2013 Non-invasive sphygmomanometers – Part 2: Clinical validation of automated measurement type
- ANSI/AAMI EC57:2012 Testing and reporting performance results of cardiac rhythm and STsegment measurement algorithms
- . IEC 62304:2006 Medical device software - Software life-cycle processes
- . ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity
- . ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
- . ISO 14971:2007 Medical devices – Application of risk management to medical devices
Verification and validation testing includes:
- . Software Verification and Validation Test
- Software Unit Test
- . Integration Test
- System Test
- . Wireless Coexistence Test
- . Wireless Design Validation and Verification Test
# Conclusion:
The performance of the Nihon Kohden Vital Sign Telemeter GZ-140P is substantially equivalent to the predicate devices, the Nihon Kohden BSM-6000 Series Bedside Monitor (K080342) and the Nihon Kohden Vital Sign Telemeter GZ-120P and GZ-130P (K153707), and raises no safety or effectiveness issues.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.