K071818 · St Jude Medical · DQK · Dec 20, 2007 · Cardiovascular
Device Facts
Record ID
K071818
Device Name
ENSITE FUSION DYNAMIC REGISTRATION TOOL
Applicant
St Jude Medical
Product Code
DQK · Cardiovascular
Decision Date
Dec 20, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Intended Use
The EnSite System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated. • When used with the EnSite Catheter, the EnSite System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone. OR • When used with the EnSite NavX Surface Electrode Kit, the EnSite System is intended to display the position of conventional electrophysiology catheters in the heart. EnSite Fusion is indicated for registering the EnSite NavX navigation system to anatomic models, derived from CT scans, of the four individual cardiac chambers.
Device Story
EnSite System is a computerized storage/display system for cardiac electrophysiology studies; consists of console workstation, patient interface unit, and mapping catheter or surface electrode kit. EnSite Array catheter detects intracardiac electrograms without endocardial contact; system generates color-coded isopotential maps to identify arrhythmia sources. NavX surface patches provide non-fluoroscopic navigation/visualization of conventional EP catheters. EnSite Fusion tool registers NavX navigation system to CT-derived 3D anatomic models of cardiac chambers. Used in clinical settings by electrophysiologists to navigate to therapy sites; aids in treating complex, non-sustained, or poorly tolerated arrhythmias. Output includes reconstructed electrograms and 3D anatomical visualization, enabling precise catheter placement and arrhythmia identification.
Clinical Evidence
Bench testing only. Device validation testing conducted in accordance with in-house procedures to evaluate system changes.
Technological Characteristics
Computerized workstation and patient interface unit; utilizes electrophysiology mapping catheters or surface electrode kits. Software-based system for signal processing and 3D anatomical registration. Connectivity via patient interface unit. No specific materials or sterilization standards provided.
Indications for Use
Indicated for patients undergoing electrophysiology studies. EnSite System with EnSite Catheter: patients with complex arrhythmias in the right atrium. EnSite System with NavX Surface Electrode Kit: patients requiring visualization/navigation of conventional EP catheters. EnSite Fusion: registration of EnSite NavX navigation system to CT-derived anatomic models of the four cardiac chambers.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
K083328 — ENSITE SYSTEM, MODEL EE3000, AND ENSITE NAVX SURFACE ELECTRODE, MODEL EN0010 · St Jude Medical · May 22, 2009
K030129 — MODIFICATION TO ENSITE 3000 SYSTEM, MODEL EE3000 · Endocardial Solutions, Inc. · Apr 22, 2003
K060954 — MODIFICATION TO ENSITE SYSTEM, MODEL EE3000 · St. Jude Medical, Inc.-Endocardial Solutions · Apr 21, 2006
K070902 — MODIFICATION TO: ENSITE SYSTEM, MODEL, EE3000 · St Jude Medical · May 1, 2007
K042493 — ENSITE SYSTEM, MODEL EE3000 · Endocardial Solutions, Inc. · Oct 15, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
K071818
r 1/2
## 510(k) Summary for Public Disclosure
| Submitter: | St. Jude Medical<br>1350 Energy Lane, Suite 110<br>St. Paul, MN 55108 USA<br>Phone: 651-523-6900<br>Fax: 651-644-7897 | |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Contact: | Karen J. McKelvey<br>Principal Regulatory Compliance Engineer<br>DEC 20 2007 | |
| Date Prepared: | December 19, 2007 | |
| Trade Name: | EnSite® System (Model EE3000)<br>a) EnSite™ Array (Model EC1000)<br>b) EnSite NavX™ Surface Electrode Kit (Model EN0010)<br>d) EnSite Verismo™ Segmentation Tool (Model EV1000) | |
| Common name: | Electrophysiology cardiac mapping system | |
| Classification Name: | a) Electrode recording catheter or electrode recording probe (21CFR 870.1220)<br>b) Programmable diagnostic computer (21 CFR 870.1425) | |
| Predicate Device: | EnSite System<br>510(k) No. K070902 | |
| Device Description: | The EnSite System is a computerized storage and display system<br>for use in electrophysiology studies of the human heart. The<br>system consists of a console workstation, patient interface unit, and<br>an electrophysiology mapping catheter or surface electrode kit.<br><br>Unlike currently available electrode recording catheters, the EnSite<br>Array does not require direct contact with the endocardium for the<br>detection of intracardiac electrograms. The EnSite System is a<br>system that facilitates mapping and treatment of arrhythmias.<br>When used with the EnSite catheter, the system is useful for<br>treating patients with complex, non-sustained, or poorly tolerated | |
| | arrhythmias that are difficult, if not impossible, to map with<br>current mapping techniques. By visualizing the global activation<br>pattern seen on the color-coded isopotential maps in the EnSite<br>System, in conjunction with the reconstructed electrograms, the<br>electrophysiologist can identify the arrhythmia source and can<br>navigate to the defined area for therapy. When used with NavX<br>patches, the system is useful in treating patients with simpler<br>arrhythmias by providing non-fluoroscopic navigation and<br>visualization of conventional EP catheters | K071818<br>p 2 / 2 |
| EnSite System<br>V7.0 | | |
| Intended use: | The EnSite System is a suggested diagnostic tool in patients for<br>whom electrophysiology studies are indicated. | |
| | • When used with the EnSite Catheter, the EnSite System is<br>intended to be used in the right atrium of patients with complex<br>arrhythmias that may be difficult to identify using conventional<br>mapping systems alone.<br>OR<br>• When used with the EnSite NavX Surface Electrode Kit, the<br>EnSite System is intended to display the position of<br>conventional electrophysiology catheters in the heart. | |
| | | |
| EnSite Fusion<br>Dynamic<br>Registration Tool | | |
| Intended use: | EnSite Fusion is indicated for registering the EnSite NavX<br>navigation system to anatomic models, derived from CT scans, of<br>the four individual cardiac chambers. | |
| Technological<br>Characteristics: | The new device has the same technological characteristics as the<br>legally marketed predicate device. | |
| Non-clinical<br>Performance Data: | The changes made to the EnSite System underwent a battery of<br>bench and user tests. Device validation testing was conducted in<br>accordance with in-house procedures. | |
| Conclusion: | An evaluation of the device changes indicates that the device is as<br>safe and effective as the previously marketed device to which it is<br>being compared and does not raise any new issues of safety and<br>effectiveness. | |
:
{1}------------------------------------------------
:
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure, represented by three curved lines that form a profile. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
DEC 2 0 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
St. Jude Medical c/o Ms. Karen J. McKelvey Principal Regulatory Compliance Engineer 1350 Energy Lane. Suite 110 St. Paul. MN 55108
Re: K071818
Trade/Device Name:EnSite Fusion Dynamic Registration Tool Regulation Number: 21 CFR 870,1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: October 25, 2007 Received: October 26, 2007
Dear Ms. McKelvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Karen J. McKelvey
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
B. Bemmmer fr
Bram Ø. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 8. Indications for Use
510(k) Number (if known): _K071818_
Device Name: EnSite Fusion Dynamic Registration Tool
Indications for Use:
EnSite Fusion is indicated for registering the EnSite NavX navigation system to anatomic models, derived from CT scans, of the four individual cardiac chambers.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhumanna
Sian-Off of Cardiovascular Devices 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.