K050093 · Ge Medical Systems Information Technologies · DQK · May 13, 2005 · Cardiovascular
Device Facts
Record ID
K050093
Device Name
MAC-LAB/CARDIOLAB EP/COMBOLAB SYSTEM
Applicant
Ge Medical Systems Information Technologies
Product Code
DQK · Cardiovascular
Decision Date
May 13, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Mac-Lab System is intended for use in a catheterization and related cardiovascular specialty laboratories under the direct supervision of a licensed healthcare practitioner. It is intended to monitor, calculate and/or record cardiovascular data from adult and pediatric patients undergoing cardiac catheterization procedures. The data may be manually entered or acquired via interfaced devices. Data includes: ECG, heart rate, pulse oximetry (SpO2), respiration rate, EtCO2, temperature, valve gradients and areas, cardiac output, hemondynamic measurements, invasive and noninvasive blood pressure and procedural information and optional intracardiac electrocardiogram (IECG). Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, and generate reports on the data. Additionally, the system may acquire, amplify, display and record data received from other interfaced medical devices typically used during these procedures, such as imaging devices. The Mac-Lab System does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze data acquired during procedure. The Mac-Lab System does not transmit alarms or arrhythmias and does not have arrhythmia detection capabilities. The CardioLab EP System is intended for use in an electrophysiological laboratory and related specialty laboratories under the direct supervision of a licensed healthcare practitioner. It is intended to monitor, calculate and/or record electrophysiological data from adult and pediatric patients under going electrophysiological studies. Data includes: ECG, pressure, and intracardiac electrocardiogram (IECG) waveforms, heart rate, pulse oximetry (SpO2), respiration rate, EtCO2, temperature, invasive and noninvasive blood pressure, and procedural information. Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, and generate reports on the data. Additionally, the system may acquire, amplify, display and record data received from other interfaced medical devices typically used during these procedures, such as imaging devices and RF generators. The CardioLab EP System does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze data acquired during the procedure. The CardioLab EP System does not transmit alarms or arrhythmias and does not have arrhythmia detection capabilities. The ComboLab System is the combination of the both the Mac-Lab and CardioLab EP systems. The ComboLab System is intended for use in either a catheterization laboratory or electrophysiological laboratory and related speciality laboratories under the direct supervision of a licensed healthcare practitioner. The ComboLab System allows the user to run either the Mac-Lab System or the CardioLab EP System, although only one may be used at a time.
Device Story
Microprocessor-based data acquisition system for cath/EP labs; monitors, calculates, records physiological data. Inputs: ECG, pressure, respiration, IECG, SpO2, heart rate, EtCO2, temperature, valve gradients, cardiac output, hemodynamic measurements, blood pressure; data entered manually or via interfaced modules (TRAM, CardioLab Amplifier, DASH/Solar monitors, MUSE). System integrates acquisition devices, computer processors, software, high-resolution displays, printers, keyboard/mouse. Amplifier filters, amplifies, digitizes, transmits data via fiber optic cable. Software performs data acquisition, display, storage, reporting, and transmission to CardioLink INW server/networked hospital systems. Used by clinicians to monitor procedures, review data, generate reports. Does not control energy, administer drugs, perform life-support, or detect arrhythmias. Benefits: centralized physiological monitoring and documentation during invasive cardiovascular procedures.
Clinical Evidence
Bench testing only. Quality assurance measures included risk analysis, requirements/design reviews, module verification, system integration testing, and final validation. No clinical data presented.
Technological Characteristics
Microprocessor-based data acquisition system. Components: computer processors, high-resolution displays, power supply, printers, keyboard, mouse. Connectivity: fiber optic cable for amplifier-to-computer transmission; networked via CardioLink INW server. Software-based data processing. No life-supporting/sustaining functions. No arrhythmia detection/alarm transmission.
Indications for Use
Indicated for adult and pediatric patients undergoing cardiac catheterization or electrophysiological studies in a clinical laboratory setting under direct supervision of a licensed healthcare practitioner.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Mac-Lab/CardioLab EP/ComboLab System Version 6.0 (K032577)
K021366 — MAC-LAB/CARDIOLAB EP/COMBOLAB · General Electric Medical Systems Information Techn · Jul 3, 2002
K032577 — MODIFICATION TO: MAC-LAB/CARDIOLAB EP/COMBOLAB SYSTEM · Ge Medical Systems Information Technologies · Oct 10, 2003
K061741 — MAC-LAB/CARDIOLAB /COMBOLAB/SPECIALSLAB SYSTEM · GE Healthcare · Sep 15, 2006
K993414 — PRUCKA ENGINEERING INC. CARDIOLAB EP SYSTEM,VERSION 4.2 · Prucka Engineering, Inc. · Apr 7, 2000
K111200 — MAC-LAB SYSTEM, CARDIOLAB SYSTEM, COMBOLAB SYSTEM, SPECIALSLAB SYSTEM · GE Healthcare · Oct 12, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
# MAY 1 3 2005
# Section 2 Summary
.
K050093
.
14
## 510(k) Summary of Safety and Effectiveness
| Date: | January 7, 2005 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems Information Technologies<br>8200 West Tower Avenue<br>Milwaukee, WI 53223 USA |
| Contact Person: | Ronald N. Blaski<br>Regulatory Affairs Specialist<br>GE Medical Systems Information Technologies<br>Phone: (414) 362-2348<br>Fax: (262) 293-2585 |
| Device: | |
| Trade Name: | Mac-Lab/CardioLab EP/ComboLab System |
| Common/Usual Name: | Cardiac Catheterization Laboratory System |
| Classification Names: | 21 CFR 870.1425 Programmable Diagnostic Computer |
| Predicate Device: | Mac-Lab/CardioLab EP/ComboLab System Version 6.0 (K032577) |
{1}------------------------------------------------
050093 2/4
### The Mac-Lab System:
The Mac-Lab System is a microprocessor based data acquisition system used during cath procedures to monitor, calculate and record physiological data from pediatric or adult patients. Data may be entered manually or acquired via an interfaced GE Medical Systems Information Technologies acquisition device, such as: TRAM module (K011000), EtCO2 module (K904789), CardioLab Amplifier Module (K910307), DASH 3000/4000 Monitor (K001359), Solar 8000M Monitor (K993757), MUSE cardiovascular system (K992637) or other peripheral interface. Data includes: ECG, pressure, respiration intracardiac electrocardiograms (IECG), and SpO2 waveforms, heart rate, pulse oximetry (SpO2), respiration rate, valve gradients and areas, cardiac output, EtCO2, hemodynamic measurements, invasive and noninvasive blood pressure, temperature, and procedural information. The Mac-Lab joins together the acquisition devices with computer processors, software, highresolution display monitors, power supply, printers, keyboard and mouse. Digital data is transmitted, via cable, from the acquisition devices to the computer for processing. Major functions of the software include data acquisition and display, data storage, reporting of data, and transmission of data to the CardioLink INW server and other networked hospital information systems.
### The CardioLab EP System:
The CardioLab EP System is a microprocessor based data acquisition system used during electrophysiology procedures to monitor, calculate and record physiological data from pediatric or adult patients. Data may be entered manually or acquired via an interfaced GE Medical Systems Information Technologies acquisition device, such as: CardioLab Amplifier Module (K910307), TRAM module (K011000), EtCO2 module (K904789), DASH 3000/4000 Monitor (K001359), Solar 8000M Monitor (K993757), MUSE cardiovascular system (K992637) or other peripheral interface, such as RF generators and fluoro video Data includes: ECG. intracardiac pressure waveforms, heart rate, pulse electrocardiograms (IECG), and oximetry (SpO2), respiration rate, EtCO2, invasive and noninvasive blood pressure, temperature, and procedural information. The ECG, intracardiac and pressure data are acquired by an amplifier that is connected to the patient by third-party devices such as ECG leadwires and catheters. The amplifier filters, amplifies, digitizes and transmits the data to the computer via fiber optic cable. The CardioLab joins together the acquisition devices with computer processors, software, high-resolution display monitors, power supply, printers, Digital data is transmitted, via cable, from the keyboard and mouse. acquisition devices to the computer for processing. Major functions of the software include data acquisition and display, data storage, reporting of data, and transmission of data to the CardioLink INW server and other networked hospital information systems.
### The ComboLab System:
The product will be available in three configurations: CardioLab EP application only, Mac-Lab application only, or a combination of both CardioLab EP and Mac-Lab applications. The 'CardioLab EP only's configuration only allows the user to run the CardioLab EP mode. The 'Mac-Lab only' configuration only allows the user to run the Mac-Lab mode. The ComboLab configuration is the combination of both CardioLab EP and Mac-Lab modes, though only one mode may be used at a time (CardioLab EP for electrophysiological lab cases and Mac-Lab for catheterization lab cases).
{2}------------------------------------------------
Intended Use:
050093 3/4
### Mac-Lab System:
The Mac-Lab System is intended for use in a catheterization and related cardiovascular specialty laboratories under the direct supervision of a licensed healthcare practitioner. It is intended to monitor, calculate and/or record cardiovascular data from adult and pediatric patients undergoing cardiac catheterization procedures. The data may be manually entered or acquired via interfaced devices. Data includes: ECG, heart rate, pulse oximetry (SpO2), respiration rate, EtCO2, temperature, valve gradients and areas, cardiac output, hemondynamic measurements, invasive and noninvasive blood pressure and procedural information and optional intracardiac electrocardiogram (IECG). Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, and generate reports on the data. Additionally, the system may acquire, amplify, display and record data received from other interfaced medical devices typically used during these procedures, such as imaging devices. The Mac-Lab System does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze data acquired during procedure. The Mac-Lab System does not transmit alarms or arrhythmias and does not have arrhythmia detection capabilities.
### CardioLab EP System
The CardioLab EP System is intended for use in an electrophysiological laboratory and related specialty laboratories under the direct supervision of a licensed healthcare practitioner. It is intended to monitor, calculate and/or record electrophysiological data from adult and pediatric patients under going electrophysiological studies. Data includes: ECG, pressure, and intracardiac electrocardiogram (IECG) waveforms, heart rate, pulse oximetry (SpO2), respiration rate, EtCO2, temperature, invasive and noninvasive blood pressure, and procedural information. Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, and generate reports on the data. Additionally, the system may acquire, amplify, display and record data received from other interfaced medical devices typically used during these procedures, such as imaging devices and RF generators. The CardioLab EP System does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze data acquired during the procedure. The CardioLab EP System does not transmit alarms or arrhythmias and does not have arrhythmia detection capabilities.
### ComboLab System
The ComboLab System is the combination of the both the Mac-Lab and CardioLab EP systems. The ComboLab System is intended for use in either a catheterization laboratory or electrophysiological laboratory and related speciality laboratories under the direct supervision of a licensed healthcare practitioner. The ComboLab System allows the user to run either the Mac-Lab System or the CardioLab EP System, although only one may be used at a time.
The proposed Mac-Lab/CardioLab EP/ComboLab System employs the same Technology: functional scientific technology as the predicate device Mac-Lab/CardioLab EP/ComboLab System (K032577).
{3}------------------------------------------------
K050093 4
tem /
The Mac-Lab/CardioLab EP/ComboLab System Test Summary:
complies with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the device:
- Risk Analysis .
- . Requirements Reviews
- Design Reviews .
- Testing on unit level (Module verification) .
- . Clinical Use Validation
- Integration Testing (System verification) .
- Final acceptance testing (Validation) .
- . Performance testing
- . Safety testing
- Environmental Testing .
- The results of these measurements demonstrated that the Mac-Lab/CardioLab Conclusion: EP/ComboLab System is as safe, as effective, and performs as well as the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the top half of the circle. In the center of the seal is a stylized image of what appears to be an eagle or other bird-like figure, represented by three curved lines that suggest wings or feathers.
MAY 1 3 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
GE Medical Systems Information Technologies c/o Mr. Ronald N. Blaski Regulatory Affairs Specialist 8200 West Tower Avenue Milwaukee WI 53223
Re: K050093
K050095
Trade Name: Mac-Lab / CardioLab EP / ComboLab Systems
Trade Name: Mac-Lab / CardioLab 1945 Regulation Number: 21 CFR 870.1245 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: April 11, 2005 Received: April 12, 2005
### Dear Mr. Blaski:
We have reviewed your Section 510(k) premarket notification of intent to market the device to We have reviewed your Section 3 IQ(x) premantially equivalent (for the indications
referenced above and have determined the devices marketed in interstate referenced above and have deemined the ucvices ated in interstate
for use stated in the enclosure) to legally marketed prediced Device Amendments, for use stated in the enclosure) to the enactment date of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food, Drug, commerce prior to May 28, 1976, the enactment and of the Federal Food, Drug, devices that have been reclassified in accordance who meetires and onlines of PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act. The and Cosmetic Act (Act) that do not require approval or approval on the Act. The Act. The
You may, therefore, market the device, subject to the general on the Act. The You may, therefore, market the device, subject to the general cregistration, listing of
general controls provisions of the Act 1 heline and recisibitions anainst misbranding general controls provisions of the Act Include requirements to the many of the many adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it
em and the can and the caller and a Fristing major regulations affecti If your device is classified (see above) mio entrely in the states of the major can
may be subject to such additional controls. Existing major regulations affective your dev may be subject to such additional controis: Existing major regal to 898. In addion, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. be found in the Code of Federal Kegulations, Priro 24, 2017
publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Mr. Ronald N. Blaski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be advised that I Driv issualles over device complies with other requirements of the Act that I Dri has made a aved regulations administered by other Federal agencies. You must or any I edolar bathates the requirements, including, but not limited to: registration and listing comply with an the Hec b requirement, at 801); good manufacturing practice requirements as set (21 CFR Part 807), acoming (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control pro resin marketing your device as described in your Section 510(k) I mis icher witi anow you to ogin maing of substantial equivalence of your device to a legally premarket notification: "The PDF Intellegsification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrieves y at (240) 276-0120.. Also, please note the regulation entitled, Colliati the Office of Compulline in (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bummer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 510(k) Number (if known): Кобосорз
### Device Name:
### Mac-Lab/CardioLab EP/ComboLab System
### Indications For Use:
#### Mac-Lab System:
The Mac-Lab System is intended for use in a catheterization and related cardiovascular specialty laboratories under the direct supervision of a licensed healtheare practitioner. It is intended to monitor, calculate and/or record cardiovascular data from adult super raiser or a neens on and ac catheterization procedures. The data may be manually entered or acquired via interfaced devices. Data includes: ECG, heart rate, pulse oximetry (SpO2), respiration rate, ExCO2, temperature, valve gradients million on the output, herrondynamic measurements, invasive and noninvasive blood pressure and procedural information and optional intracardiac electrocardiogram (IECG). Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, and generate reports on the data. Additionally, the system may acquire, amplify, display and received from other interfaced medical devices typically used during these procedures, such as imaging devices. The Mac-Lab System does not control the delivery of energy, administer drugs, perform any life-supporting or lifesustaining functions, or analyze data acquired during procedure. The Mac-Lab System does not transmit alarms or arritythmias and does not have arrhythmia detection capabilities.
#### CardioLab EP System:
The CardioLab EP System is intended for use in an electrophysiological laboratory and related specially laboratories under the I in Caroletao en Cystom is mended to monitor, it is intended to monitor, calculate and/or record electrophysiological data from adult and petiative patients under going electrophysiological studies. ECG, pressure, and intracadiac tual tron and pecificile patients and gate, pulse oximetry (SpO2), respiration rate, ElCO2, temperature, invasive and enouraliations and procedural information. Physiological parameters such as diastolic, systolic, and mean blood pressure, heartere, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to pressure, its acquisition of data, review the data. Additionally, the system may acquire, amplify, display and record data received from other interfaced medical devices typically used during these procedures, such as imaging devices and RF generators. The CardioLab EP System does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze data acquired during the procedure. The CardioLab EP System does not transmit alarms or arrhythmias and does not have arrhythmia detection capabilities.
#### ComboLab System:
The Combol ab System is the combination of the both the Mac-Lab and CardioLab EP systems. The ComboLab System is intended for use in either a catherization laboratory or clectrophysiological laboratory and related speciality laboratories under the direct supervision of a licensed health are practitioner. The ComboLab System allows the user to run either the Mac-Lab System or the CardioLab EP System, although only one may be used at a time.
Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhummar
Divisi
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.