N Series Patient Monitors

K213799 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · MHX · Jul 12, 2022 · Cardiovascular

Device Facts

Record IDK213799
Device NameN Series Patient Monitors
ApplicantShenzhen Mindray Bio-Medical Electronics Co., Ltd.
Product CodeMHX · Cardiovascular
Decision DateJul 12, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Indications for Use

The BeneVision N12/N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: - BIS, RM, CCO, SvO2/ScvO2, PAWP, NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only. CCO using FloTrac is intended for adult patients only; - C.O. monitoring and A-Fib are intended for adult patients only; - ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg; - rSO2 monitoring is intended for use in individuals greater than 2.5kg. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport, hospital ambulance, or home use. The BeneVision N1 Patient Monitor is intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, OT Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: - PAWP, PNP, and PNC are intended for adult and pediatric patients only; - A-Fib is intended for adult patients only; The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment, whereas N1 configured with WMTS technology can be used inside the hospital only. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.

Device Story

BeneVision N Series (N1, N12, N15, N17, N19, N22) are multiparameter patient monitors. They acquire physiological signals (ECG, SpO2, IBP, NIBP, CO2, etc.) via sensors and modules. The system processes these signals to display waveforms, numerics, and alarms. It includes arrhythmia detection, ST-segment analysis, and 12-lead ECG interpretation. Used in healthcare facilities by clinicians; N1 supports patient transport. Output is viewed on integrated touchscreens or external displays to assist clinical decision-making regarding patient status. Benefits include continuous monitoring and early detection of physiological changes.

Clinical Evidence

Bench testing only. Performance validated through functional and system-level testing against consensus standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-25/26/27/34, ISO 80601-2-55/56/61, and ANSI AAMI EC57. Testing confirmed accuracy specifications and equivalence to the predicate device.

Technological Characteristics

Multiparameter patient monitors with modular architecture. Features include ECG, SpO2, NIBP, IBP, CO2, AG, EEG, and hemodynamic monitoring. Connectivity via wired/wireless network (2.4G/5G, WMTS). Power via AC or Li-ion battery. Software-based signal processing for arrhythmia, ST-segment, and QT analysis. Compliance with IEC 60601-1, 60601-1-2, and specific collateral standards for monitoring equipment.

Indications for Use

Indicated for monitoring physiological parameters (ECG, Resp, Temp, SpO2, PR, NIBP, IBP, PAWP, C.O., CCO, SvO2/ScvO2, CO2, O2, AG, ICG, BIS, RM, NMT, EEG, rSO2) in adult, pediatric, and neonatal patients. Specific exclusions apply for certain parameters (e.g., BIS, RM, CCO, PAWP, NMT, A-Fib, ICG, rSO2) based on patient age, weight, or clinical condition. Used in healthcare facilities by trained clinical professionals.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. July 12, 2022 Shenzhen Mindray Bio-medical Electronics Co., Ltd. Li Lei Manager Regulatory Affairs Mindray Building, Keji 12th Road South Hi-tech Industrial Park, Nanshan Shenzhen, Guangdong 518057 China Re: K213799 Trade/Device Name: N Series Patient Monitors Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DQA Dated: June 9, 2022 Received: June 10, 2022 Dear Li Lei: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K213799 #### Device Name BeneVision N Series Patient Monitors (including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19. BeneVision N22) #### Indications for Use (Describe) The BeneVision N12N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: - · BIS, RM, CCO, SvO2/ScvO2, PAWP, NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only. CCO using FloTrac is intended for adult patients only; - · C.O. monitoring and A-Fib are intended for adult patients only; - · ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg. - · rSO2 monitoring is intended for use in individuals greater than 2.5kg. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport, hospital ambulance, or home use. (Continue on next page N1 indications for use) | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |-----------------------------------------------------------------------------------| | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 510(k) Number (if known): Indications for Use (con't) #### BeneVision N1 Patient Monitor: The BeneVision N1 Patient Monitor is intended for monitoring, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO₂), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: - PAWP, PNP, and PNC are intended for adult and pediatric patients only; - A-Fib is intended for adult patients only; The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment, whereas N1 configured with WMTS technology can be used inside the hospital only. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use. {4}------------------------------------------------ In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Mindray BeneVision N Series Monitors is provided below. #### SUBMITTER 1. | Applicant: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81888998<br>Fax: +86 755 26582680 | |-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Contact Person: Li Lei<br>Title: Manager Regulatory Affairs<br>Phone: +86 755 81885953<br>Fax: +86 755 26582680<br>E-mail: lilei.js@mindray.com | | Date Prepared: | December 2, 2021 | | 2.<br>DEVICE | | | Device Trade Name: | BeneVision N Series Patient Monitors (Including<br>BeneVision N12, BeneVision N15, BeneVision N17,<br>BeneVision N19, BeneVision N22, BeneVision N1) | | Device Common Name: | Patient Monitor | | Classification Name: | 21 CFR 870.1025, Class II, Arrhythmia detector and<br>alarm (including ST-segment measurement and alarm) | | Regulatory Class: | Class II | | Primary Product Code: | MHX - Monitor, Physiological, Patient (with<br>arrhythmia detection or alarms) | | Panel: | Cardiovascular | {5}------------------------------------------------ | Regulation<br>Number/Class | Product<br>Code | Regulation description | Device Common Name | |----------------------------|-----------------|----------------------------------------------------------------------------------|--------------------------------------------------------------------------| | 870.1025, II | DSI | Arrhythmia detector and alarm<br>(including ST-segment measurement<br>and alarm) | Detector and alarm, arrhythmia | | 870.1025, II | MLD | Arrhythmia detector and alarm<br>(including ST-segment measurement<br>and alarm) | Monitor, st segment with alarm | | 870.2300, II | DRT | Cardiac Monitor (including<br>cardiotachometer and rate<br>alarm) | Monitor, cardiac (incl.<br>cardiotachometer & rate alarm) | | 870.1130, II | DXN | Noninvasive blood pressure<br>measurement system | System, measurement, blood-pressure,<br>non-invasive | | 870.1110, II | DSK | Blood pressure computer | Computer, blood-pressure | | 880.2770, II | DSB | Impedance plethysmograph. | Plethysmograph, impedance | | 880.2700, II | MUD | Oximeter | Oximeter, tissue saturation | | 870.2340, II | MLC | Electrocardiograph | Monitor, st segment | | 870.2370, II | KRC | Electrocardiograph surface electrode<br>tester. | Tester, electrode, surface,<br>electrocardiographic | | 880.2910, II | FLL | Clinical electronic thermometer | Thermometer, electronic, clinical | | 870.2700, II | DQA | Oximeter | Oximeter | | 870.2300, II | MSX | Cardiac monitor (including<br>cardiotachometer and rate alarm). | System, network and communication,<br>physiological monitors | | 870.2910, II | DRG | Radiofrequency physiological signal<br>transmitter and receiver. | Transmitters and receivers,<br>physiological signal, radiofrequency | | 868.1400, II | CCK | Carbon dioxide gas analyzer | Analyzer, gas,<br>carbon-dioxide, gaseous-phase | | 870.1435, II | DXG | Single-function, preprogrammed<br>diagnostic computer | Computer, diagnostic, pre-<br>programmed, single-function | | 870.1100, II | DSJ | Blood pressure alarm | Alarm, blood-pressure | | 870.2850, II | DRS | Transducer, blood-pressure,<br>extravascular | Extravascular blood pressure<br>transducer. | | 868.1500, II | CBQ | Enflurane gas analyzer. | Analyzer, gas, enflurane, gaseous-<br>phase (anesthetic concentration) | | 868.1500, II | NHO | Enflurane gas analyzer. | Analyzer, gas, desflurane, gaseous-<br>phase (anesthetic concentration) | | 868.1500, II | NHP | Enflurane gas analyzer. | Analyzer, gas, sevoflurane, gaseous-<br>phase (anesthetic concentration) | | 868.1500, II | NHQ | Enflurane gas analyzer. | Analyzer, gas, isoflurane, gaseous-<br>phase (anesthetic concentration) | | Regulation<br>Number/Class | Product<br>Code | Regulation description | Device Common Name | | 868.1620, II | CBS | Halothane gas analyzer. | Analyzer, gas, halothane, gaseous-<br>phase (anesthetic conc.) | | 868.1700, II | CBR | Nitrous oxide gas analyzer. | Analyzer, gas, nitrous-oxide, gaseous<br>phase (anesthetic conc.) | | 868.1720, II | CCL | Oxygen gas analyzer. | Analyzer, gas, oxygen, gaseous-phase | | 868.2775, II | KOI | Electrical peripheral nerve stimulator. | Stimulator, nerve, peripheral, electric | | 882.1400, II | OLW | Electroencephalograph. | Index-generating<br>electroencephalograph software | | 882.1400, II | OLT | Electroencephalograph | Non-normalizing quantitative<br>electroencephalograph software | | 882.1400, II | OMC | Electroencephalograph. | Reduced- montage standard<br>electroencephalograph | | 882.1400, II | ORT | Electroencephalograph | Burst suppression detection software<br>for electroencephalograph | | 882.1320, II | GXY | Cutaneous electrode. | Electrode, cutaneous | #### Table 1: Secondary Product Codes {6}------------------------------------------------ #### PREDICATE DEVICES 3. - Predicate: K202405 BeneVision N Series Patient Monitors (Including BeneVision N12, . BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1) #### 4. REFERENCE DEVICES - Philips InterlliVue MX600, MX800 patient monitors (K161531): provided as reference devices for SVT and SVCs/min High, Artifact flag of ABP shields alarms function, and combined alarm function. - . GE Carescape B850 patient monitor (K131414) is provided as reference device for SVT and SVCs/min High and SVCs/min value. - . Edwards HemoSphere Advanced Monitor (K180881) is provided as reference device for the Flotrac CCO cable. - . Philips InterlliVue MX850 patient monitors (K210906) are provided as reference devices for the aEEG module. - Mindray BeneVision central monitoring system (K193391, including TM70 telemetry ● system) is provided as reference device for the WMTS module of TM70. - Philips InterlliVue MX series patient monitors (K182979) are provided as reference ● devices for the Alarm Threshold. {7}------------------------------------------------ #### DEVICE DESCRIPTION 5. The subject BeneVision N Series Patient Monitors includes six monitors: - . BeneVision N12 Patient Monitor - BeneVision N15 Patient Monitor - BeneVision N17 Patient Monitor ● - BeneVision N19 Patient Monitor ● - BeneVision N22 Patient Monitor ● - BeneVision N1 Patient Monitor Mindray's BeneVision N Series Patient Monitors provide a flexible software and hardware platform to meet the clinical needs of patient monitoring. #### INTENDED USE/INDICATIONS FOR USE 6. ### BeneVision N12/N15/N17/N19/N22 Patient Monitors: The BeneVision N12/N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: - BIS, RM, CCO, SvO2/ScvO2, PAWP, NMT monitoring, PNP, and PNC are intended ● for adult and pediatric patients only. CCO using FloTrac is intended for adult patients only: - C.O. monitoring and A-Fib are intended for adult patients only; ● - ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg; - . rSO2 monitoring is intended for use in individuals greater than 2.5kg. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport, hospital ambulance, or home use. {8}------------------------------------------------ ### BeneVision N1 Patient Monitor: The BeneVision N1 Patient Monitor is intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, OT Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: - PAWP, PNP, and PNC are intended for adult and pediatric patients only; - A-Fib is intended for adult patients only; ● The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment, whereas N1 configured with WMTS technology can be used inside the hospital only. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use. #### 7. SUBSTANTIAL EQUIVALENCE ### Comparison of Indications Both the predicate devices and the subject devices are multiparameter patient monitors intended to be used under the direction of clinical professionals. The indications for use statement of the subject BeneVision N12/N15/N17/N19/N22 has been modified to include support for CCO using FloTrac in adult patients. Although this feature is not present in the primary predicate devices, it is present in HemoSphere Advanced Monitor (K180881, including Pressure Cable). The indications for use statement of the subject BeneVision N1 has been modified to include WMTS technology for use inside the hospital only. In conclusion, the minor difference in the indications for use statements do not change the fundamental intended use of the N Series monitors as multiparameter monitors. ### Technological Comparison The tables below compare the key technological feature of the subject devices to the predicate device (N series Patient Monitors, K202405). The features in grey are features which are different between the predicate devices and the subject devices. {9}------------------------------------------------ | Table 2: | Device Comparison Table (BeneVision 22/N19/N17/N15/N12) | | |----------|---------------------------------------------------------|--| |----------|---------------------------------------------------------|--| | Feature | N22 | N19 | N17 | N15 | N12 | N22 | N19 | N17 | N15 | N12 | | | |-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|-----------|-----------|-----------|--| | | Cleared in K202405 | | | | Subject Device | | | | | | | | | Primary display and<br>touchscreen | 22"<br>1680*1050<br>pixels. | 19"<br>1680*1050<br>pixels. | 18.5"<br>1920*1080<br>pixels. | 15.6"<br>1920*1080<br>pixels. | 12.1"<br>1280*800<br>pixels. | No change | No change | No change | No change | No change | | | | Secondary display | Independent control and<br>display. | | Independent<br>control and<br>display.<br>Size: 21.5"<br>Model:<br>ET2203LM. | Mirrored display.<br>Size: 21.5"<br>Model: ET2203LM. | | No change | No change | No change | No change | No change | | | | iView | iView is an isolated PC platform that<br>allows the user to run 3rd party<br>applications using a limited set of<br>functions. The iView system is not<br>intended to be used as a primary alarm<br>device.<br>Capacity of RAM: 8GB<br>Model of CPU: N4200<br>Windows 10 | | | Not supported. | | No change | | | No change | | | | | Wireless radio module | The 2.4G/5G module for connecting to a network wirelessly,<br>constructing a monitoring network with a central monitoring system<br>(CMS). | | | | | No change | | | | | | | | Module rack | Must be connected to the<br>main unit to provide up<br>to 8 standard module<br>slots. | | Optional for the patient<br>monitor, adding 8 standard<br>module slots to extend the<br>measurement capabilities of<br>the system. | | | Not<br>supported. | No change | | | | | | | Power supply | One rechargeable Lithium-ion battery or AC power supply. | | | | | No change | | | | | | | | Battery | Chargeable Lithium-Ion,<br>10.8 VDC, 5600 mAh. | | Chargeable Lithium-Ion, 10.95 VDC,<br>4500 mAh. | | | No change | | | | No change | | | | Feature | N22…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...