The BeneVision N12/N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: · BIS, RM, CCO, SvO2/ScvO2, PAWP, and NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only: · C.O. monitoring and A-Fib are intended for adult patients only; • ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg. · rSO2 monitoring is intended for use in individuals greater than 2.5kg. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport. hospital ambulance, or home use. The BeneVision N1 Patient Monitor is intended for monitoring, displaying, storing , alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP) , Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: - PAWP, PNP, and PNC are intended for adult and pediatric patients only; - A-Fib is intended for adult patients only; The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.
Device Story
Multiparameter patient monitors (N1, N12, N15, N17, N19, N22) collect physiological signals via sensors (ECG leads, SpO2 probes, NIBP cuffs, IBP transducers, etc.). Data processed by internal modules (MPM 3.0, etc.) to calculate parameters, detect arrhythmias, and perform ST/QT analysis. Outputs displayed on integrated touchscreens; alarms triggered for abnormal values. Used in clinical settings by trained professionals; N1 supports transport. Provides real-time monitoring, trend analysis, and clinical decision support (EWS, SepsisSight, GCS). Benefits include continuous patient surveillance and early deterioration detection.
Clinical Evidence
Bench testing only. Functional and system-level testing validated performance against specifications and consensus standards (IEC 60601-1, IEC 60601-2-25/27/34, ISO 80601-2-61). Software verification and validation conducted per FDA guidance. No clinical data presented.
Indicated for monitoring, displaying, storing, alarming, and transferring physiological parameters (ECG, Resp, Temp, SpO2, PR, NIBP, IBP, PAWP, C.O., CCO, SvO2/ScvO2, CO2, O2, AG, ICG, BIS, RM, NMT, EEG, rSO2) in adult, pediatric, and neonatal patients in healthcare facilities. Specific parameter exclusions apply based on patient age/size (e.g., A-Fib/C.O. for adults only).
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized eagle, and the FDA name is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
January 07, 2021
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. Yanhong Bai Manager Regulatory Affairs, Technical Regulation Department Mindray Building, Keji 12th Road South High-tech Industrial Park, Nanshan Shenzhen, Guangdong 518057 China
Re: K202405
Trade/Device Name: BeneVision N Series Patient Monitors (Including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX Dated: December 10, 2020 Received: December 11, 2020
Dear Yanhong Bai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202405
## Device Name
BeneVision N Series Patient Monitors (Including BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1)
Indications for Use (Describe) BeneVision N12/N15/N17/N19/N22
The BeneVision N12N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
· BIS, RM, CCO, SvO2/ScvO2, PAWP, and NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only:
· C.O. monitoring and A-Fib are intended for adult patients only;
• ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
· rSO2 monitoring is intended for use in individuals greater than 2.5kg.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport. hospital ambulance, or home use.
(Continue on next page for N1 indications for Use)
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Number (if known): K202405 Indications for Use (con't)
BeneVision N1 Patient Monitor:
The BeneVision N1 Patient Monitor is intended for monitoring, displaying, storing , alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP) , Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
- PAWP, PNP, and PNC are intended for adult and pediatric patients only;
- A-Fib is intended for adult patients only;
The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.
{4}------------------------------------------------
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Mindray BeneVision N Series Monitors is provided below.
### 1. SUBMITTER
| Applicant: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81888998<br>Fax: +86 755 26582680 |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Contact Person: Yanhong Bai<br>Title: Manager Regulatory Affairs<br>Phone: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared: | December 08, 2020 |
| 2.<br>DEVICE | |
| Device Trade Name: | BeneVision N Series Patient Monitors (Including<br>BeneVision N12, BeneVision N15, BeneVision N17,<br>BeneVision N19, BeneVision N22, BeneVision N1) |
| Device Common Name: | Patient Monitor |
| Classification Name: | 21 CFR 870.1025, Class II, Arrhythmia detector and<br>alarm (including ST-segment measurement and alarm) |
| Regulatory Class: | Class II |
| Primary Product Code: | MHX - Monitor, Physiological, Patient (with<br>arrhythmia detection or alarms) |
| Panel | Cardiovascular |
{5}------------------------------------------------
| Regulation<br>Number/Class | Product<br>Code | Regulation description | Device Common Name |
|----------------------------|-----------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------|
| 870.1025, II | DSI | Arrhythmia detector and alarm<br>(including ST-segment measurement<br>and alarm) | Detector and alarm, arrhythmia |
| 870.1025, II | MLD | Arrhythmia detector and alarm<br>(including ST-segment measurement<br>and alarm) | Monitor, st segment with alarm |
| 870.2300, II | DRT | Cardiac Monitor (including<br>cardiotachometer and rate<br>alarm) | Monitor, cardiac (incl.<br>cardiotachometer & rate alarm) |
| 870.1130, II | DXN | Noninvasive blood pressure<br>measurement system | System, measurement, blood-pressure<br>non-invasive |
| 870.1110, II | DSK | Blood pressure computer | Computer, blood-pressure |
| 880.2770, II | DSB | Impedance plethysmograph. | Plethysmograph, impedance |
| 880.2700, II | MUD | Oximeter | Oximeter, tissue saturation |
| 870.2340, II | MLC | Electrocardiograph | Monitor, st segment |
| 870.2370, II | KRC | Electrocardiograph surface electrode<br>tester. | Tester, electrode, surface,<br>electrocardiographic |
| 880.2910, II | FLL | Clinical electronic thermometer | Thermometer, electronic, clinical |
| 870.2700, II | DQA | Oximeter | Oximeter |
| 870.2300, II | MSX | Cardiac monitor (including<br>cardiotachometer and rate alarm). | System, network and communication,<br>physiological monitors |
| 870.2910, II | DRG | Radiofrequency physiological signal<br>transmitter and receiver. | Transmitters and receivers,<br>physiological signal, radiofrequency |
| 868.1400, II | CCK | Carbon dioxide gas analyzer | Analyzer, gas,<br>carbon-dioxide, gaseous-phase |
| 870.1435, II | DXG | Single-function, preprogrammed | Computer, diagnostic, pre- |
Table 1:
| 870.2910, II | DRG | Radiofrequency physiological signal<br>transmitter and receiver. | Transmitters and receivers,<br>physiological signal, radiofrequency |
|--------------|-----|------------------------------------------------------------------|--------------------------------------------------------------------------|
| 868.1400, II | CCK | Carbon dioxide gas analyzer | Analyzer, gas,<br>carbon-dioxide, gaseous-phase |
| 870.1435, II | DXG | Single-function, preprogrammed<br>diagnostic computer | Computer, diagnostic, pre-<br>programmed, single-function |
| 870.1100, II | DSJ | Blood pressure alarm | Alarm, blood-pressure |
| 870.2850, II | DRS | Transducer, blood-pressure,<br>extravascular | Extravascular blood pressure<br>transducer. |
| 868.1500, II | CBQ | Enflurane gas analyzer. | Analyzer, gas, enflurane, gaseous-<br>phase (anesthetic concentration) |
| 868.1500, II | NHO | Enflurane gas analyzer. | Analyzer, gas, desflurane, gaseous-<br>phase (anesthetic concentration) |
| 868.1500, II | NHP | Enflurane gas analyzer. | Analyzer, gas, sevoflurane, gaseous-<br>phase (anesthetic concentration) |
| 868.1500, II | NHQ | Enflurane gas analyzer. | Analyzer, gas, isoflurane, gaseous-<br>phase (anesthetic concentration) |
{6}------------------------------------------------
| Regulation<br>Number/Class | Product<br>Code | Regulation description | Device Common Name |
|----------------------------|-----------------|-----------------------------------------|-------------------------------------------------------------------|
| 868.1620, II | CBS | Halothane gas analyzer. | Analyzer, gas, halothane, gaseous-<br>phase (anesthetic conc.) |
| 868.1700, II | CBR | Nitrous oxide gas analyzer. | Analyzer, gas, nitrous-oxide, gaseous<br>phase (anesthetic conc.) |
| 868.1720, II | CCL | Oxygen gas analyzer. | Analyzer, gas, oxygen, gaseous-phase |
| 868.2775, II | KOI | Electrical peripheral nerve stimulator. | Stimulator, nerve, peripheral, electric |
| 882.1400, II | OLW | Electroencephalograph. | Index-generating<br>electroencephalograph software |
| 882.1400, II | OLT | Electroencephalograph. | Non-normalizing quantitative<br>electroencephalograph software |
| 882.1400, II | OMC | Electroencephalograph. | Reduced-montage standard<br>electroencephalograph |
| 882.1400, II | ORT | Electroencephalograph | Burst suppression detection software<br>for electroencephalograph |
| 882.1320, II | GXY | Cutaneous electrode. | Electrode, cutaneous |
#### PREDICATE DEVICES 3.
- . Predicate: K192972 – BeneVision N Series Patient Monitors (Including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1)
#### REFERENCE DEVICES 4.
- K200015 - ePM series Patient Monitors (ePM 10/ePM 12/ePM 15/ePM 10M/ePM 12M/ePM 15M): provided a provided as reference devices for ECG algorithm supporting arrhythmia detection in neonate that has been added to the subject BeneVision N Series Patient Monitors.
#### DEVICE DESCRIPTION ನ.
The subject BeneVision N Series Patient Monitors includes six monitors:
- BeneVision N12 Patient Monitor
- BeneVision N15 Patient Monitor
- BeneVision N17 Patient Monitor ●
- . BeneVision N19 Patient Monitor
- . BeneVision N22 Patient Monitor
- BeneVision N1 Patient Monitor ●
{7}------------------------------------------------
Mindray's BeneVision N Series Patient Monitors provide a flexible software and hardware platform to meet the clinical needs of patient monitoring.
#### 6. INTENDED USE/INDICATIONS FOR USE
# BeneVision N12/N15/N17/N19/N22 Patient Monitors:
The BeneVision N12/N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, OT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
- BIS, RM, CCO, SvO2/ScvO2, PAWP, NMT monitoring, PNP, and PNC are intended ● for adult and pediatric patients only;
- C.O. monitoring and A-Fib are intended for adult patients only; ●
- ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
- . rSO2 monitoring is intended for use in individuals greater than 2.5kg.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport, hospital ambulance, or home use.
# BeneVision N1 Patient Monitor:
The BeneVision N1 Patient Monitor is intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP) , Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
{8}------------------------------------------------
- . PAWP, PNP and PNC are intended for adult and pediatric patients only;
- . A-Fib is intended for adult patients only;
The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.
### 7. SUBSTANTIAL EQUIVALENCE
# Comparison of Indications
Both the predicate devices and the subject devices are multiparameter patient monitors intended to be used under the direction of clinical professionals.
The indications for use statement of the subject BeneVision N1/N12/N15/N17/N19/N22 has been modified to include support for arrhythmia detection in neonates when used with the MPM 3.0 module. Although this feature is not present in the primary predicate devices, it is present in other cleared multiparameter patient monitors such as the Mindray ePM series Patient Monitors (K200015, ePM 10/ePM 12/ePM 15/ePM 10M/ePM 15M). The inclusion of supporting arrhythmia detection in neonates in the indications for use does not constitute a new fundamental intended use.
In conclusion, the minor difference in the indications for use statement does not change the fundamental intended use of the N Series monitors as multiparameter monitors.
# Technological Comparison
The tables below compare the key technological feature of the subject devices to the predicate device (N series Patient Monitors, K192972). The features in green are features which are different between the predicate devices and the subject devices.
{9}------------------------------------------------
| Feature | N22 | N19 | N17 | N15 | N12 | N22 | N19 | N17 | N15 | N12 |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|----------------|--------------|-----------|
| | Cleared in K192972 | | | | | Subject Device | | | | |
| Primary display<br>and touchscreen | 22"<br>1680*1050<br>pixels. | 19"<br>1680*1050<br>pixels. | 18.5"<br>1920*1080<br>pixels. | 15.6"<br>1920*1080<br>pixels. | 12.1"<br>1280*800<br>pixels. | No<br>change | No<br>change | No change | No<br>change | No change |
| Secondary<br>display | Independent control and<br>display. | Independent control and<br>display. | Independent<br>control and<br>display.<br>Size: 21.5";<br>Model:2203<br>L. | Mirrored display.<br>Size: 21.5";<br>Model: 2203L. | | No change | No change | No change | No change | No change |
| iView | Change the PC platform<br>Capacity of RAM: 8GB<br>Model of CPU: N4200<br>Windows 10 | | | Not supported. | | No change | No change | No change | No change | No change |
| Wireless radio<br>module | The 2.4G/5G module for connecting to a network wirelessly,<br>constructing a monitoring network with a central monitoring system<br>(CMS).<br>WiFi module: MSD45N | | | | | The 2.4G/5G module for connecting to a network wirelessly,<br>constructing a monitoring network with a central monitoring system<br>(CMS).<br>WiFi module: MSD45N and SX-SDMAC-2832S+ | | | | |
| Module rack | Must be connected to the<br>main unit to provide up to<br>8 standard module slots. | | Optional for the patient<br>monitors, adding 8 standard<br>module slots to extend the<br>measurement capabilities of<br>the system. | Not<br>supported. | | No change | No change | No change | No change | No change |
| Power supply | One rechargeable Lithium-ion battery or AC power supply. | | | | | No change | | | | |
| Battery | Chargeable Lithium-Ion,<br>10.8 VDC, 5600 mAh. | Chargeable Lithium-Ion, 10.95 VDC,<br>4500 mAh. | | | | No change | No change | No change | No change | No change |
| Feature | N22 | N19 | N17 | N15 | N12 | N22 | N19 | N17 | N15 | N12 |
| | Cleared in K192972 | | | | | Subject Device | | | | |
| Data storage | Solid State Hard Drive<br>(SDD) | | | | Embedded Multi Media Card (eMMC). | No change | | | | |
| Data recorder | Supports the thermal<br>recorder module, needs to<br>be plugged into the SMR. | | Supports the thermal recorder module and<br>the built-in thermal recorder, but they<br>cannot work at the same time. | | | No change | | | | |
| Speaker | Provides audible alarm tones (45 to 85 dB), key tones, QRS tones;<br>support PITCH TONE and multi-level tone modulation. | | | | | No change…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.