Pirouette 035

K161427 · Arravasc Limited · LIT · Jul 30, 2016 · Cardiovascular

Device Facts

Record IDK161427
Device NamePirouette 035
ApplicantArravasc Limited
Product CodeLIT · Cardiovascular
Decision DateJul 30, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

Piroutte 035 PTA Catheter is intended for balloon of the iliac, femoral, popliteal, infra-popliteal and renal arteries.

Device Story

Pirouette 035 is an over-the-wire (OTW), semi-compliant, coaxial PTA catheter; features balloon mounted on distal tip; hydrophilic coating; dual-lumen shaft (guidewire and inflation); radiopaque markers for fluoroscopic visualization. Used by physicians in clinical settings for peripheral artery dilation. Device tracks over 0.035-inch guidewire; balloon inflated/deflated via manifold. Enables vessel dilation to improve blood flow; physician monitors placement via fluoroscopy. Benefits include minimally invasive treatment of arterial stenosis.

Clinical Evidence

No clinical studies performed. Evidence based on bench testing including dimensional verification, balloon fatigue, burst pressure, compliance, inflation/deflation time, bond strength, flexibility, torque, radiopacity, coating integrity, and biocompatibility (ISO 10993-1).

Technological Characteristics

OTW, semi-compliant, coaxial catheter; hydrophilic coating; radiopaque marker bands. Compatible with 0.035" guidewires. Balloon diameters 3-12mm; lengths 20-200mm. Sterilized via Ethylene Oxide. Complies with ISO 10555-1 and ISO 10555-4.

Indications for Use

Indicated for balloon dilation of iliac, femoral, popliteal, infra-popliteal, and renal arteries in patients requiring percutaneous transluminal angioplasty.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 30, 2016 ArraVasc Limited % Patsy J. Trisler, JD, RAC Qserve Group US, Inc. 5600 Wisconsin Avenue Chevy Chase, MD 20815 Re: K161427 Trade/Device Name: Pirouette 035 Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: July 6, 2016 Received: July 6, 2016 Dear Ms. Trisler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Brian D. Pullin -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K161427 Device Name Pirouette 035 Indications for Use (Describe) Piroutte 035 PTA Catheter is intended for balloon of the iliac, femoral, popliteal, infra-popliteal and renal arteries. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary #### 5.1 Submitter | Submitter Address: | ArraVasc Limited<br>2 Ballybrit Business Park,<br>Galway<br>Ireland | |--------------------|---------------------------------------------------------------------| | Phone Number: | +353-91-758939 | | Fax Number: | +353-91-758930 | | Contact Person: | Aoife Donoghue | | Date Prepared: | 13 June 2016 | ### 5.2 Device | Device Trade Name: | Pirouette 035 | |------------------------|----------------------------------------------| | Common Name: | OTW PTA catheter | | Classification Name: | Peripheral Transluminal Angioplasty Catheter | | Classification number: | 21 CFR 870.1250 | | Product code: | LIT | | Class: | II | | Classification Panel: | Cardiovascular | #### 5.3 Predicate Device | Primary predicate: | Pirouette 018 OTW PTA Catheter | |--------------------|--------------------------------------------------------| | | ArraVasc Ltd. | | | 510(K) number: K151153 | | Reference Device | Advance 35LP Low Profile PTA balloon Dilation Catheter | | | Cook Inc. | | | 510(k) number K132020 | 5.4 Device Description Device Description: The Pirouette 035 Percutaneous Transluminal Angioplasty (PTA) Catheter Family are standard over the wire (OTW), semi-compliant, coaxial design catheters with a balloon mounted on the distal tip. The distal portion of the catheter has a hydrophilic coating. The manifold connector and shaft design consists of a quidewire lumen allowing the catheter to track over a quidewire and an inflation lumen, used to inflate and deflate the balloon. Radiopaque markers are positioned on the shaft within the balloon to enable visualization of the catheter/balloon under fluoroscopy. The catheter is compatible with .035 inch (0.89 mm) wire guides. The Pirouette 035 PTA Catheter Family includes multiple balloon sizes ranging from 3 to 12 mm in diameter and 20 to 200mm in length. The nominal balloon diameter (mm) and the balloon length (mm) are inscribed on the quidewire hub of the manifold. The effective lengths of the catheter are 75cm and 130cm. Physical Description: Percutaneous Transluminal Angioplasty Catheter (PTA Catheter). Indication for use Indications for Use Pirouette 035 is intended for balloon dilation of the iliac, femoral, Statement: popliteal, infra-popliteal and renal arteries. {4}------------------------------------------------ ### 5.5 Comparison with the predicate device The Pirouette 035 is substantially equivalent to the predicate device in design, materials, packaging, fundamental scientific technology, manufacturing, sterilization and intended use. The following table provides a summary of general and technical characteristics as compared to the predicate device. ### Table 1 Comparison of Pirouette 035 to Pirouette 018 (predicate) | Parameter | Characteristics | |---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Classification | Class II, 21 CFR 870.1250<br>Same Classification as predicate devices. | | Intended Use | Same intended use: balloon dilatation of the iliac, femoral, popliteal<br>and infra-popliteal arteries and renal arteries. | | Balloon material | Same material | | Balloon diameter | Comparable range of balloon diameters: 3 - 12 mm versus 2 - 9 mm<br>for the predicate device | | Balloon length | Comparable range of balloon lengths: 20 – 200 versus 20 - 300mm for<br>the predicate device | | Nominal pressure (atm) | Same nominal pressure: 8 atm. | | Rated burst pressure<br>(atm) | Comparable rated burst pressure: 16atm (2.0-5.0mm balloon<br>diameter) / 14atm (6.0mm balloon diameter) / 12atm (7.0-10.0mm<br>balloon diameter) / 10atm (12.0mm) | | Radiopaque | Two Markerbands, one at distal and one at proximal side of the<br>balloon, with the same function. | | Marker bands | Same material | | Outer shaft | Same material | | Catheter shaft<br>outer diameter | Comparable catheter diameter: 5.2 Fr - 5.4 Fr versus: 3.6 – 4.2 Fr for<br>the Predicate | | Catheter Usable<br>Lengths (cm) | Comparable range of usable lengths: 75 cm and 130 cm versus 45 -<br>150cm for the Predicate device. | | Recommended<br>Introducer Sheath<br>compatibility | Comparable size: 5 - 7 Fr versus 4 – 5 Fr for the Predicate device. | | Recommended<br>Guidewire diameter | Maximum 0.035"" versus; maximum 0.018" | | Catheter strain relief &<br>manifold material | Same material, with the same function. | | Catheter manifold<br>design | Same design, dual lumen Y design, with the same function. | | Catheter coating | Same coating. | | Sterilization Method | Same method; Ethylene Oxide. | | Single Use / Reusable | Single Use | ### Performance Data 5.6 As per ArraVasc Risk Analysis procedures, the Pirouette 035 was thoroughly tested on the bench to evaluate and verify that it meets the required performance specifications and to support a determination of substantial equivalence. The bench testing plan was developed with the consideration of the recommendations outlined in the applicable FDA guidance documents, tests recommended in ISO 1055-1, Intravascular catheters- Sterile and single-use catheters - Part 1: General Requirements, and 10555-4, Intravascular catheters- Sterile and single-use catheters - Part 4: Balloon dilatation catheters. Testing performed on Pirouette included the following: {5}------------------------------------------------ ### Non-Clinical Tests: - Dimensional verification ● - Balloon preparation, pushability, trackability, deployment, withdrawal and balloon reinsertion ● - Balloon rated burst pressure (RBP) . - . Balloon fatique (repeated balloon inflations) - Balloon Compliance at nominal and rated burst inflation pressure . - Balloon Inflation/Deflation Time ● - Catheter Bond Strength (Tip to balloon, balloon to proximal shaft, shaft to manifold) - Flexibility and Kink test ● - Torque Strength ● - Radiopacity ● - Coating Integrity ● - . Particulate evaluation - Guide wire compatibility ● - Introducer sheath compatibility ● The results showed that the Pirouette 035 met the pre-determined acceptance criteria. No new safety or effectiveness issues were raised during this testing. ### Biocompatibility Tests: Per ISO10993-1:2009, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process, the Pirouette 035 is classified as an externally communicating device, which contacts circulating blood during a limited contact duration (≤24hours). For reason that the Pirouette 035 has the same indication, general design, materials, and packaging, the Biocompatibility tests with the Pirouette 018 were considered valid for the Pirouette 035. Biocompatibility testing with the Pirouette 018 was conducted per ISO 10993-1:2009 were completed and passed: - . MTT Cytotoxicity Study - ISO Maximization Sensitization Study ● - ISO Intracutaneous Study ● - ISO Systemic Toxicity Study ● - ASTM Hemolysis - C3a Complement Activation Assay - SC5b-9 Complement Activation Assay - ASTM Partial Thromboplastin Time ● - In Vivo Thromboresistance Study Juqular Vein ● - Pyrogenicity ● ### Sterilization, Shelf life tests and Packaging validation: - EQ sterilzation validation by an adoption validation using a sub-lethal cycle approach; ● - . EO/ECH residues of Pirouette 035 were tested; - Shelf life testing was leveraged from the data on the Pirouette 018; ● - Packaging validation was leveraged on the data of the Pirouette 018. The results show that the Pirouette 035 is sterility is maintained by the packaging during the entire shelf life of the device. From the results it was also concluded that the device meets the criteria for Residual Testing, i.e. EO/ECH residues. ### Clinical Performance Data: - . No clinical studies were performed for the purpose of obtaining safety and effectiveness data. - . The Pirouette 035 has been approved for marketing in the European Union (CE certified) in 2015, and has been marketed in the EU thereafter. ### 5.7 Conclusions Based upon the intended use, fundamental scientific technology, performance characteristics, nonclinical performance testing, and comparison to legally marketed devices, it is concluded that the Pirouette 035 is appropriate for its intended use, and is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...