CORDIS WEBSTER PREFACE BRAIDED GUIDING SHEATH & PERRY EXCHANGE DILATOR, MODEL #'S 301-803A, 301-803P, 301-803M, 301-813S
K982740 · Cordis Webster, Inc. · DYB · Aug 28, 1998 · Cardiovascular
Device Facts
Record ID
K982740
Device Name
CORDIS WEBSTER PREFACE BRAIDED GUIDING SHEATH & PERRY EXCHANGE DILATOR, MODEL #'S 301-803A, 301-803P, 301-803M, 301-813S
Applicant
Cordis Webster, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Aug 28, 1998
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The intended use of the Modified Device is to provide stability and directional control during transcatheter ablation of endocardial tissue. PREFACE Braided Guiding Sheath - The intended use of the percutaneous Braided Guiding Sheath is for the introduction of intravascular electrophysiology catheters into any cardiac chamber. PERRY Exchange Dilator - The intended use of the PERRY Exchange Dilator is for the transseptal exchange of Cordis Webster PREFACE Braided Guiding Sheaths which are used for the introduction of intravascular electrophysiology catheters. The exchange dilator permits exchange of the guiding sheaths while maintaining transseptal left heart access and control of sheath location.
Device Story
System comprises PREFACE Braided Guiding Sheath and PERRY Exchange Dilator; used during transcatheter ablation of endocardial tissue. Guiding sheath provides stability/directional control for electrophysiology catheters in cardiac chambers. Exchange dilator facilitates sheath replacement while maintaining transseptal access/sheath position. Device inputs include manual manipulation by physician; outputs are physical access/stability for catheters. Radiopaque tungsten tips added to dilators for improved fluoroscopic visibility. Used in clinical/lab settings by physicians. Benefits include maintained access during sheath exchange, reducing procedural time/risk.
Clinical Evidence
Bench testing only. Verification testing compared stiffness and insertion force of the PERRY Exchange Dilator (with radiopaque tip) against the standard 8F AVANTI Sheath (K970392). No clinical data provided.
Technological Characteristics
System includes braided guiding sheath, vessel dilator, and exchange dilator. Materials identical to predicate with addition of tungsten for radiopaque tips. Features include posterior, anterior, or multipurpose distal tip shapes. Connectivity: None (mechanical device). Sterilization: Not specified.
Indications for Use
Indicated for patients requiring introduction of intravascular electrophysiology catheters into any cardiac chamber for transcatheter ablation of endocardial tissue. The PERRY Exchange Dilator is indicated for transseptal exchange of guiding sheaths while maintaining transseptal left heart access.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
AUG 28 1998
K982740
Special 510(k): Device Modificatio Cordis Webster Braided Guiding Sheath Exchange System
## Appendix A: 510(k) Summary of Safety and Effectiveness
| Statement | Information supporting claims of substantial equivalence, as defined under the<br>Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness is<br>summarized below. |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | For the convenience of the Reviewer, this summary is formatted in<br>accordance with the Agency's final rule "...510(k) Summaries and 510(k)<br>Statements..." (21 CFR §807) and can be used to provide a substantial<br>equivalence summary to anyone requesting it from the Agency. |
| Device<br>description | The components of the Predicate Device (K970264) consists of 1) a Guiding<br>Sheath with posterior, anterior or multipurpose shaped distal tip section, 2) a<br>conventional sheath vessel dilator ( <i>vessel dilator</i> ), 3) a dual-tapered long<br>vessel dilator ( <i>exchange dilator</i> ), 4) a .032", 260cm guidewire, and 5) a<br>guidewire funnel. The components of the Predicate Device are packaged in<br>four configurations: 1) Guiding Sheath w/ posterior curve and vessel dilator,<br>2) Guiding Sheath w/ anterior curve and vessel dilator, 3) Guiding Sheath w/<br>multipurpose curve, vessel dilator, exchange dilator, .032", 260cm guidewire<br>and guidewire funnel, and 4) Posterior, Anterior and Multipurpose Guiding<br>Sheaths, vessel dilator, exchange dilator, .032", 260cm guidewire and<br>guidewire funnel. |
| | The packaging configuration of the Predicate Device has been modified to<br>offer the Guiding Sheath and Exchange System in two separate packages.<br>The Guiding Sheath ( <i>PREFACE Braided Guiding Sheath</i> ) with posterior,<br>anterior or multipurpose shaped distal tip sections is packaged with a vessel<br>dilator and a short .032", 150cm guidewire. The Exchange System includes<br>the exchange dilator ( <i>PERRY Exchange Dilator</i> ), .032", 260cm guidewire,<br>and guidewire funnel. The distal end of the vessel dilator and both ends of the<br><i>PERRY Exchange Dilator</i> were modified by adding radiopaque tips with<br>tungsten to improve visibility under fluoroscopy. The function and intended<br>use of the Modified Device as compared to the Predicate Device are identical,<br>as are all component materials with the exception of the tungsten dilator<br>material. The addition of the tungsten material does not affect safety or<br>effectiveness as demonstrated in the Verification Test Data provided in<br>Appendix I. |
| Device<br>Description -<br>Continued | The PREFACE Braided Guiding Sheaths are intended to provide stability and<br>directional control during catheter ablation. The PREFACE Braided Guiding<br>Sheath is packaged with a vessel dilator with a radiopaque distal tip, and a<br>.032", 150cm guidewire. The PERRY Exchange Dilator will be utilized by<br>the physician to exchange the existing PREFACE Braided Guiding Sheath for<br>an alternate PREFACE Braided Guiding Sheath without losing the established<br>intracardiac position. The PERRY Exchange Dilator with radiopaque ends is<br>packaged with a .032", 260cm guidewire and guidewire funnel. |
| Intended use | The intended use of the Modified Device is to provide stability and directional<br>control during transcatheter ablation of endocardial tissue. |
| Indications<br>statement | PREFACE Braided Guiding Sheath - The intended use of the percutaneous<br>Braided Guiding Sheath is for the introduction of intravascular<br>electrophysiology catheters into any cardiac chamber.<br> |
| | PERRY Exchange Dilator - The intended use of the PERRY Exchange<br>Dilator is for the transseptal exchange of Cordis Webster PREFACE Braided Guiding Sheaths which are used for the introduction of intravascular<br>electrophysiology catheters. The exchange dilator permits exchange of the<br>guiding sheaths while maintaining transseptal left heart access and control of<br>sheath location. |
| Technological<br>characteristics | The technological characteristics of the Modified Device are the same as the<br>Predicate Device. |
| | Continued on next page |
| Performance<br>data | Verification testing of the Modified Device was conducted to compare the<br>stiffness and the insertion force of the PERRY Exchange Dilator with<br>radiopaque tip and the dilator (of the PREFACE Braided Guiding Sheath)<br>with the standard 8F AVANTI Sheath (K970392). |
| Conclusion | Based on the 510(k) summaries and the 510(k) statements (21 CFR §807) and<br>the information provided herein, we conclude that the Modified Device is<br>substantially equivalent to the Predicate Device under the Federal Food, Drug<br>and Cosmetic Act. |
| Contact | Mary Adams<br>Regulatory Affairs Manager<br>Cordis Webster, Inc.<br>4750 Littlejohn Street<br>Baldwin Park, CA 91706 |
| Date | August 5, 1998 |
Continued on next page
{1}------------------------------------------------
## Appendix A: 510(k) Summary of Safety and Effectiveness, Continued
{2}------------------------------------------------
## Appendix A: 510(k) Summary of Safety and Effectiveness,
Continued
----
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right. The profiles are stacked on top of each other and are connected by a flowing line.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 28 1998
Ms. Mary Adams Requlatory Affairs Manager Cordis Webster, Inc. 4750 Littlejohn Street Baldwin Park, CA 91706
Re: K982740 Cordis Webster PREFACE Braided Guiding Sheath and PERRY Exchange Dilator Regulatory Class: II (two) Product Code: 74 DYB Dated: Auqust 5, 1998 Received: Auqust 6, 1998
Dear Ms. Adams:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturinq Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this
{4}------------------------------------------------
obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, ² Misbranding by reference to premarket notification² (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Appendix B: Indications for Use Statement
Statement
Indications for Use Statement:
510(k) Number: K_982740
Device Name: PREFACE Braided Guiding Sheath and PERRY Exchange Dilator
Indications for Use: The PREFACE Braided Guiding Sheath is intended for the introduction of intravascular electrophysiology catheters into any cardiac chamber.
Indications for Use: The intended use of the PERRY Exchange Dilator is for the transseptal exchange of Cordis Webster PREFACE Braided Guiding Sheaths which are used for the introduction of intravascular electrophysiology catheters. The exchange dilator permits exchange of the guiding sheaths while maintaining transseptal left heart access and control of sheath location.
Qarly Tillh
**Prescription Use**
(Per 21 CFR 801.109)