CORDIS WEBSTER BRAIDED GUIDING SHEATH EXCHANGE SYSTEM

K970264 · Cordis Webster, Inc. · DYB · Oct 28, 1997 · Cardiovascular

Device Facts

Record IDK970264
Device NameCORDIS WEBSTER BRAIDED GUIDING SHEATH EXCHANGE SYSTEM
ApplicantCordis Webster, Inc.
Product CodeDYB · Cardiovascular
Decision DateOct 28, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The intended use of the percutaneous Braided Guiding Sheath Exchange System is for the introduction of intravascular electrophysiology catheters into any cardiac chamber.

Device Story

System consists of guiding sheaths with sideport/valve, pre-shaped distal tip with radiopaque brite tip, guidewire, conventional vessel dilator, and long tapered exchange vessel dilator with multipurpose curve. Sheaths feature nylon body reinforced with stainless steel braid wire; transition segments of varying durometer provide stiffness for cannulation and flexibility for atraumatic distal tip. Used by physicians in clinical settings to facilitate intravascular passage of electrophysiology catheters and infusion of heparinized saline or contrast medium into cardiac chambers.

Clinical Evidence

Bench testing only. Nonclinical performance testing compared the subject device to predicate devices, indicating no significant differences in outcomes affecting safety or effectiveness.

Technological Characteristics

Components: guiding sheath (nylon body, stainless steel braid reinforcement, polyurethane brite tip), guidewire, vessel dilators. Features: non-tapered lumen, variable durometer transition segments for stiffness/flexibility. Radiopaque filler in distal tip. Non-powered, mechanical device.

Indications for Use

Indicated for the introduction of intravascular electrophysiology catheters into any cardiac chamber.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Submission Summary (Full Text)

{0} K970264 510(k) Submission Cordis Webster Braided Guiding Sheath Exchange System OCT 28 1997 # 510(k) Summary ## Classification Name Common/Usual Name: Introducer Proprietary Name: Cordis Webster Braided Guiding Sheath Exchange System ## Name of Predicate Device Cordis 9 FR Guiding Catheter Daig Fast-Cath™ Hemostasis Introducer ## Device Description The Cordis Webster Braided Guiding Sheath Exchange System consists of guiding sheaths with a sideport and valve, pre-shaped distal tip section with radiopaque brite tip, a guidewire, a conventional vessel dilator, and a long tapered exchange vessel dilator with a multipurpose curve at the distal end. The guiding sheaths have large non-tapered lumens that allow for intravascular passage of catheters and infusion of heparinized normal saline and/or contrast medium. The guiding sheaths feature a nylon body reinforced with a tightly wound stainless steel braid wire. The braid wire extends from the proximal hub through several flexible transition segments to the “brite tip” which is a low durometer Polyurethane with radiopaque filler at the distal end. The transition segments are comprised of several durometers of material designed to vary the stiffness along the entire length of the guiding sheath shaft. The higher durometer is used on the proximal segment of the shaft to yield a stiffer segment. This stiffer segment provides the added support required to manipulate the guiding sheath during cannulation. The distal transition segments utilize an increasingly softer durometer which provides the flexible atraumatic distal tip segment. * CORDIS WEBSTER, INC. - STRICTLY PRIVATE * {1} 510(k) Submission Cordis Webster Braided Guiding Sheath Exchange System # 510(k) Summary (continued) ## Intended Use The intended use of the percutaneous Braided Guiding Sheath Exchange System is for the introduction of intravascular electrophysiology catheters into any cardiac chamber. ## Technological Characteristics The Cordis Webster Braided Guiding Sheath Exchange System and the Cordis Guiding Catheter, one of the predicate devices, consist of the same components, with exception of the multipurpose curve (i.e. shaped section of the vessel dilator) and a shorter sheath length to assist physician technique. ## Performance Data (Nonclinical Testing) The nonclinical performance testing conducted on the Cordis Webster Braided Guiding Sheath Exchange System compared to the predicate devices indicated that there were no significant differences in the outcome of the tests for each of the devices that would affect the safety and effectiveness of the device. ## Conclusions Drawn from the Nonclinical Tests The results of the nonclinical performance tests indicate that the Cordis Webster Braided Guiding Sheath Exchange System performs as well as the predicate devices and that the differences in testing outcome are not significant; therefore, Cordis Webster concludes that the Cordis Webster Braided Guiding Sheath Exchange System is substantially equivalent to the predicate devices, the 9F Cordis Guiding Catheter and the Daig Fast-Cath™ Hemostasis Introducer. * CORDIS WEBSTER, INC. - STRICTLY PRIVATE * {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 Ms. Mary Adams Regulatory Affairs Manager Cordis Webster, Inc. 4750 Littlejohn Street Baldwin Park, California 91706 OCT 28 1997 Re: K970264 Trade Name: Cordis Webster Braided Guiding Sheath Exchange System Regulatory Class: Two (II) Product Code: 74DYB Dated: July 22, 1997 Received: July 23, 1997 Dear Ms. Adams: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the {3} Page 2 - Ms. Mary Adams Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health {4} Page 1 of 510(K) Number (if known): K970264 Device Name: Cordis Webster Braided Guiding Sheath Exchange System Indications for Use: Introduction of intravascular electrophysiology catheters into any cardiac chamber. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K970264 (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...