AuST CSP Introducer

K260163 · CenterPoint Systems, LLC · DYB · Feb 19, 2026 · Cardiovascular

Device Facts

Record IDK260163
Device NameAuST CSP Introducer
ApplicantCenterPoint Systems, LLC
Product CodeDYB · Cardiovascular
Decision DateFeb 19, 2026
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The AuST CSP Introducer is intended for the introduction of diagnostic and therapeutic devices into the human vasculature for intracardiac placements. Do not use this device for neural placements.

Device Story

Single-use percutaneous catheter; provides pathway for diagnostic/therapeutic devices into human vasculature. Features braid-reinforced shaft, deflectable segment, radiopaque distal tip, handle with 3-way stopcock for flushing/aspiration, and hemostasis valve. Used in clinical settings by physicians to facilitate intracardiac procedures. Operator controls deflection via handle; hemostasis valve prevents fluid/air leakage. Device benefits patient by enabling safe, controlled access to cardiac anatomy for interventional procedures.

Clinical Evidence

Bench testing only. No clinical data. Testing included sterilization validation (AAMI TIR28), visual inspection, simulated use, deflection verification, valve leak testing, tensile testing, dimensional verification, flush testing, torque testing, and particulate characterization.

Technological Characteristics

Braid-reinforced catheter shaft; materials include PEBAX, PTFE, Stainless Steel, Vestamid, Polyvinylpyrrolidone. Features deflectable segment, radiopaque distal tip, hemostasis valve, and 3-way stopcock. 25F ID/30F OD; 39cm active length. Sterilized via Ethylene Oxide. Biocompatibility per ISO 10993-1.

Indications for Use

Indicated for introduction of diagnostic and therapeutic devices into human vasculature for intracardiac placements. Contraindicated for neural placements.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION February 19, 2026 CenterPoint Systems, LLC Conner Johnson Director of Regulatory 3338 Pkwy. Blvd. West Valley City, Utah 84041 Re: K260163 Trade/Device Name: AuST CSP Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: January 16, 2026 Received: January 20, 2026 Dear Conner Johnson: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K260163 - Conner Johnson Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K260163 - Conner Johnson Page 3 Sincerely, FINN E. DONALDSON -S Digitally signed by FINN E. DONALDSON -S Date: 2026.02.19 15:53:02 -05'00' For Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K260163 | | | Device Name AuST CSP Introducer | | | Indications for Use (Describe) The AuST CSP Introducer is intended for the introduction of diagnostic and therapeutic devices into the human vasculature for intracardiac placements. Do not use this device for neural placements. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K260163 Page 1 of 4 510(k) Summary 1.1 Submitter Name: CenterPoint Systems LLC Address: 3338 Parkway Blvd West Valley City UT Phone: 801-602-1923 Contact Person: Conner Johnson, Director of Regulatory Date Prepared: 16 January 2026 1.2 Device Name of Device: AuST CSP Introducer Classification Name: Catheter Introducer Regulatory Class: Class II per 21 CFR 870.1340 Product Code: DYB 1.3 Predicate Device Predicate Name and 510(k) Number: AuST Steerable Sheath, K251051 This predicate has not been subject to a design-related recall. Reference Device Name and 510(k) Number: Sentrant Introducer Sheath with Hydrophilic Coating, K171866 This predicate has not been subject to a design-related recall. 1.4 Device Description The AuST CSP Introducer is a single-use percutaneous catheter intended to provide a pathway through which diagnostic and therapeutic devices are introduced in the human vasculature. The catheter is not intended for neural placements. The AuST CSP Introducer product family consists of a variety of configurations to accommodate different anatomies and/or devices being introduced into the human vasculature. Each of the sheaths in the product family is comprised of a braid-reinforced catheter shaft, deflectable segment, radiopaque distal tip, a handle with a 3-way stopcock for flushing and aspiration, and a hemostasis valve to prevent air and fluid leakage. {5} K260163 Page 2 of 4 ## 1.5 Indications for Use The AuST CSP Introducer is intended for the introduction of diagnostic and therapeutic devices into the human vasculature for intracardiac placements. Do not use this device for neural placements. ## 1.6 Comparison of Technological Characteristics with the Predicate Devices The Proposed Device and Predicate Device are similar in indications for use, intended use, technological characteristics, and principles of operation. The differences between the Proposed Device and the Predicate Device are minor and raise no different questions of safety and effectiveness. In accordance with 21CFR807.92(a)(6) a summary of how the technological characteristics of the Proposed Device compares to the Predicate Device is provided below. | Feature | AuST CSP Introducer (proposed device) | Predicate: AuST Steerable Sheath (K251051) | Reference: Sentrant Introducer Sheath (K171866) | Same / Different between Proposed & Predicate | | --- | --- | --- | --- | --- | | Intended Use/Indications for Use | The AuST CSP Introducer is intended for the introduction of diagnostic and therapeutic devices into the human vasculature for intracardiac placements. Do not use this device for neural placements. | The currently cleared AuST Steerable Sheath (K251051) is intended for the introduction of diagnostic and therapeutic devices into the human vasculature, including but not limited to intracardiac, renal and other peripheral placements. Do not use this device for neural placements. | The Sentrant Introducer Sheath with Hydrophilic Coating is intended to provide a conduit for the insertion of diagnostic or endovascular devices into the vasculature and minimize blood loss associated with such insertions. | Substantially equivalent | | Device Class | II | II | II | Same | | Product Code | DYB | DYB | DYB | Same | | Regulation number | 21 CFR 870.1340 | 21 CFR 870.1340 | 21 CFR 870.1340 | Same | | Duration of use | Single-use, Transient | Single-use, Transient | Single-use, Transient | Same | | Sterilization | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Same | {6} K260163 Page 3 of 4 | Feature | AuST CSP Introducer (proposed device) | Predicate: AuST Steerable Sheath (K251051) | Reference: Sentrant Introducer Sheath (K171866) | Same / Different between Proposed & Predicate | | --- | --- | --- | --- | --- | | Prescription Device | Yes | Yes | Yes | Same | | Inner Diameter / Outer Diameter | 25F (30F) | 4 available: 8.5F (OD: 12.2F), 10F (OD: 14.1F), 12F (OD: 15.9F), 13.2F (OD: 17F) | Inner Diameter: 12F – 26F (2F Increments) | Substantially Equivalent to Predicate and Reference | | Active Length (Sheath) | 39cm | 44cm, 65cm, 74cm, 84cm | 28cm, 64cm | Substantially equivalent | | Deflection | Yes | Yes | No | Same | | Guidewire Compatibility | Max outside diameter 0.89mm (0.035”) | Max outside diameter 0.89mm (0.035”) | Max outside diameter 0.89mm (0.035”) | Same | | Tip with Radiopaque Materials | Yes | Yes | Market band at tip | Same | | Biocompatibility | Standard medical device materials, including PEBAX, PTFE, Stainless Steel, Vestamid, Polyvinylpyrrolidone The biocompatibility tests demonstrate that there are no adverse biocompatibility risks associated with use of this material. All test results met the requirements of ISO 10993-1. | Standard medical device materials, including PEBAX, PTFE, Stainless Steel, Vestamid, Polyvinylpyrrolidone The biocompatibility tests demonstrate that there are no adverse biocompatibility risks associated with use of this material. All test results met the requirements of ISO 10993-1. | Standard medical device materials including HDPE, PEBAX and Stainless Steel The biocompatibility tests demonstrate that there are no adverse biocompatibility risks associated with use of this material. All test results met the requirements of ISO 10993-1. | Same | | Features | Deflectable Sheath, Soft Atraumatic Tip with ventilation holes, Handle with Deflection Control, Hemostasis Valve, Side Port (Irrigation/Aspiration), Dilator | Deflectable Sheath, Soft Atraumatic Tip with ventilation holes, Handle with Deflection Control, Hemostasis Valve, Side Port (Irrigation/Aspiration), Dilator | Flexible sheath, dilator, hemostatic valve, side port. | Same | {7} K260163 Page 4 of 4 | Feature | AuST CSP Introducer (proposed device) | Predicate: AuST Steerable Sheath (K251051) | Reference: Sentrant Introducer Sheath (K171866) | Same / Different between Proposed & Predicate | | --- | --- | --- | --- | --- | | Hydrophilic Coating | Yes | Yes | Yes | Substantially Equivalent | The AuST CSP Introducer is used for the same intended use in the same anatomical location using the same principles of operation as the predicate device. ## 1.7 Performance Data All necessary performance testing has been conducted on the AuST Steerable Sheath to assure substantial equivalence to the predicate device and to demonstrate the device performs as intended. All testing was performed on test units representative of finished devices. The device passed the following tests, which were conducted in accordance with noted standards: - Sterilization validation adoption per AAMI TIR28 - Visual Inspection - Simulated use testing, including use/compatibility with ancillary devices - Deflection verifications - Valve liquid leak test - Tensile tests - Sheath and Dilator Dimensional verification, including OD/ID, working length, and curl diameter - Flush test - Torque test - Particulate characterization (coating integrity and durability) ## 1.8 Conclusions Based on the similarity of the subject and predicate device in terms of the intended use, principle of operation and overall technological characteristics, the modified AuST Steerable Sheath is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...