Full Automatic (NIBP) Blood Pressure Monitor

K153214 · Health & Life Co., Ltd. · DXN · Jan 7, 2016 · Cardiovascular

Device Facts

Record IDK153214
Device NameFull Automatic (NIBP) Blood Pressure Monitor
ApplicantHealth & Life Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateJan 7, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

HL858DI automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use. HL858DI detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

Device Story

Device is an automated, oscillometric, upper-arm blood pressure monitor for home use; operates during cuff inflation. Inputs: pressure signals from arm cuff. Processing: oscillometric algorithm calculates systolic/diastolic pressure and heart rate; detects irregular heartbeats (rhythm variation >25% from average interval); categorizes readings via Risk Category Indicator. Outputs: LCD display of blood pressure, heart rate, irregular heartbeat symbol, and risk category. Operated by patient; provides immediate visual feedback for self-tracking of blood pressure levels.

Clinical Evidence

Clinical validation performed per ISO 81060-2:2013. Study included 105 subjects. Results demonstrated that mean difference and standard deviation met required limits for automated non-invasive sphygmomanometers.

Technological Characteristics

Materials: ABS housing/keys. Sensing: Oscillometric method during inflation. Power: 6V 1A (4x AA batteries or AC adapter). Form factor: Upper arm cuff, LCD display. Connectivity: None. Software: IEC 62304 compliant. Standards: ISO 81060-2, IEC 60601-1, IEC 60601-1-11, IEC 80601-2-30, ISO 10993, ISO 14971.

Indications for Use

Indicated for adults aged 18+ with arm circumference 9-17 inches (23-43 cm) for home-use blood pressure and heart rate monitoring via oscillometric method.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 7, 2016 Health & Life Co., Ltd. Sarah Su Director 9F, No. 186, Jian Yi Road Zhonghe District, New Taipei City, TW 23553 Taiwan Re: K153214 Trade/Device Name: Full Automatic (NIBP) Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 3, 2015 Received: November 5, 2015 Dear Sarah Su: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mitchell Stein for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) # K153214 Device Name Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DI ### Indications for Use (Describe) HL858DI automatically measures human's Systolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use. HL858DI detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator with the readings on the screen for the user tracking their blood pressure level. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### K153214 page 1 of 6 # PREMARKET NOTIFICATION # 510(k) SUMMARY (As Required By 21 CFR 807.92) This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. NOV 0 3 2015 The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ Date: ### 1. Submitter: Health & Life Co., Ltd. 9F. No.186, Jian Yi Road, Zhonghe District, New Taipei City, Taiwan, R.O.C TEL: +886-2-8227-1300 FAX: +886-2-8227-1301 Contact person: Sarah Su/ Regulatory Affairs Dept. E-mail: sarah.su@hlmt.com.tw Tel: 886-2-8227-1300 ext.1201 Fax: 886-2-8227-1301 #### 2. Name of the Device: Trade Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DI Common Name: Blood Pressure Monitor Classification Name: Non-invasive Blood Pressure Measurement System Classification: Class II, 21 CFR 870.1130 Classification Panel: 74 Cardiovascular Product Code: DXN #### 3. Information for the 510(k) Cleared Device (Predicate Device): Full Automatic (NIBP) Blood Pressure Monitor, Model: HL858CB (K142968) {4}------------------------------------------------ ### 4. Device Description: HL858DI automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use. The device will display a symbol ( ** or ma), to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level. #### 5. Intended Use HL858DI automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use. HL858DI detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level. ### 6. Comparison of device to predicate device: # Product Specification Comparison Table of Subject Device HL858DL, and Predicate Device HL858CB (K142968) | Item | Predicate Device<br>HL858CB (K142968) | Subject Device<br>HL858DI | |--------------------------|---------------------------------------|---------------------------| | Method of<br>measurement | Oscillimetric | Same as left | | Measurement | During deflation | During inflation | {5}------------------------------------------------ | Type | | | |-------------------------------------------|-----------------------------------------------------------------------------|------------------------------------------------------| | Range of<br>measurement | Pressure 0- 300mmHg,<br>Pulse 40-199 Beats/minute | Same as left | | Accuracy | Pressure ± 3mmHg<br>Pulse ± 5% | Same as left | | Pressure<br>Changed Rate | 2~5mmHg/sec. | Same as left | | Exhaust | Automatic exhaust valve | Same as left | | Display | Liquid Crystal Digital | Same as left | | Power Supply | 6V 1A, 4 × AA/1.5V (LR6)<br>Alkaline batteries, or<br>AC adapter (optional) | Same as left | | Storage/<br>Transportation<br>Environment | - 25℃ ~ +70℃ (approx. -13°F ~<br>approx. +158°F),<br>≤ 93% R.H. | Same as left | | Operating<br>Environment | 5℃ ~ 40℃ (approx. 41°F~approx. 104°F),<br>15% ~ 93% R.H. | Same as left | | Material | ABS housing and<br>ABS keys | Same as left | | Sets of<br>memory | 2*120, total 240 | 2*60,total 120 | | Number of<br>Push Button | 7 +<br>2 switch control<br>(Triple-Check (Multi-Read),<br>Rest Assure) | 5 key control | | Storage pouch | Yes | Same as left | | Cuff size | Arm circumference approx.<br>23~43cm / 9~17 inch<br>(Universal cuff) | Same as left | | Unit Weight | Approx. 393 ± 10g<br>(Excluding cuff and Batteries ) | Approx. 310 ± 10g<br>(Excluding cuff and Batteries ) | | Risk Category<br>Indicator | Yes | Same as left | | Irregular<br>Heartbeat<br>Detector | Yes | Same as left | | Triple-Check<br>(Multi-Read)<br>Function | Yes | No | | Rest Assure<br>Function | Yes | No | | Data Link<br>function | Yes<br>(Via Bluetooth) | No | {6}------------------------------------------------ #### Changes from the predicate devices HL858CB (K142968): - * Modifying the measurement type from "during deflation" into "during inflation". - * Changing the memory of 240 (2*120, total 240) to 120 (2*60, total 120). - * Changing the 7 key & 2 switch control to the 5 key control. - * Changing the Unit Weight from Approx. 393 ± 10g to Approx. 310 ± 10g. - * Removing the Triple-Check (Multi-Read) Function, Rest Assure Function and Bluetooth Data Transmission Function. These features has been verified and validated and do not affect the safety and effectiveness of subject device HL858DI.Please refer to Section 12. Substantial Equivalence Discussion for detail information. #### 7. Discussion of Clinical Tests Performed: HL858DI is compliant to the standard of ISO 81060-2: Second Edition 2013-05-01 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device HL858DI in the group of 105 subjects with qualified distribution. Thus, all the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met. ## 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows: {7}------------------------------------------------ #### K153214 page 5 of 6 The subject device was tested to evaluate its safety and effectiveness, including the followings: - a. EMC Test: IEC 60601-1-2 Edition 3:2007-03, Medical Electrical Equipment -Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Tests - b. Safety Test: -IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance -IEC 60601-1-11:2010, Medical electrical equipment-Part 1-11: General Requirement for basic safety and essential performance- Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment - c. FCC Test: FCC 47 CFR Part 15, Subpart B - d. Biocompatibility Test: -ISO 10993-1:2009, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process -ISO 10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for In Vitro cytotoxicity -ISO 10993-10:2010, Third Edition Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization - e. Reliability Test: IEC 80601-2-30 Edition1.1 2013-07 Medical electricalequipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. - f. Risk Assessment: ISO 14971:2007 Second Edition, Medical devices -Application of risk management to medical devices - g. Software Verification and Validation: -IEC 62304 Ed.1.0 (2006), Medical device software - Software life cycle processes. -IEC 60601-1-4 Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems, edition 1.1 {8}------------------------------------------------ ### h. Usability Validation: -IEC 62366-1:2015 Medical devices - Application of usability engineering to medical devices -IEC 60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability ## 9. Conclusions: The subject device was tested and fulfilled the requirements of those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...