The Globalcare GUS622 automatic Blood Pressure Monitor is indicated for home use for the non-invasive measurement of diastolic and systolic blood pressures and pulse rate of adults by means of an inflatable cuff which is wrapped around the upper arm. The cuff circumference is limited to 22 to 42 cm.
Device Story
Automatic upper arm blood pressure monitor; uses oscillometric method. Input: arterial oscillations captured by pressure sensors in inflatable cuff during deflation. Transformation: software algorithm analyzes pressure signals to compute systolic/diastolic blood pressure and pulse rate; detects irregular heartbeats (IHB) by comparing interval times to average; identifies hemodynamic rest condition. Output: LCD display shows blood pressure, pulse rate, IHB symbol, risk level, and hemodynamic status. Used in home environment by adult patients. Benefits: enables self-monitoring of blood pressure and pulse rate to support cardiovascular health management.
Clinical Evidence
Bench testing only. Compliance verified against ISO 81060-2:2013 and EN 1060-3:1997+A2:2009 for clinical performance. Safety/EMC testing per ANSI/AAMI ES60601-1, IEC 60601-1-11, IEC 60601-1-2, and IEC 80601-2-30. Biocompatibility testing per ISO 10993-5 and ISO 10993-10. Internal bench tests included life testing (500 hours), storage (12 months), and drop testing.
Technological Characteristics
Oscillometric measurement; electronic unit with LCD; powered by 4x AAA batteries or 100-240Vac mains adapter. Cuff circumference 22-42 cm. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 80601-2-30, ISO 10993-5/10. Type BF applied part. No USB/PC connectivity.
Indications for Use
Indicated for non-invasive measurement of systolic/diastolic blood pressure and pulse rate in adults (cuff circumference 22-42 cm) in a home environment.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Blood Pressure Monitors Models KD-927 and KD-928 (K141984)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 21, 2020
Globalcare Medical Technology Co., Ltd % Roger Grav VP Quality and Regulatory Donawa Lifescience Consulting Srl Piazza Albania 10 Rome, 00153 ITALY
Re: K192609
Trade/Device Name: Globalcare GUS622 Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 4, 2020 Received: March 9, 2020
Dear Roger Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K192609
Device Name
Globalcare GUS622 Blood Pressure Monitor
Indications for Use (Describe)
The Globalcare GUS622 automatic Blood Pressure Monitor is indicated for the non-invasive measurement of diastolic and systolic blood pressures and pulse rate of adults by means of an inflatable cuff which is wrapped around the upper arm. The cuff circumference is limited to 22 to 42 cm.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
| Device Name: | GUS622 Blood Pressure Monitor |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of 510(k) submission: | Traditional |
| Date of submission: | 4 March 2020 |
| Manufacturer: | Globalcare Medical Technology Co., Ltd<br>No.7 Factory, European Industrial Park<br>No. 39 Mid Industrial Main Road<br>Xiaolan Town, Zhongshan City<br>Guangdong, CHINA 528415 |
| Phone: | +86-760-22589901 |
| FDA Establishment Reg. Number: | 3010880718 |
| 510(k) Owner and Submitter: | Globalcare Medical Technology Co., Ltd<br>No.7 Factory, European Industrial Park<br>No. 39 Mid Industrial Main Road<br>Xiaolan Town, Zhongshan City<br>Guangdong, CHINA 528415 |
| Owner/Operator Reg. Number: | 10046812 |
| 510(k) Contact: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy |
| Phone:<br>Fax:<br>Email: | +39 06 578 2665<br>+39 06 574 3786<br>rgray@donawa.com |
| FDA Product Code: | DXN |
| FDA Regulation Number: | 21 CFR 870.1130 |
| FDA Classification Name: | Noninvasive blood pressure measurement system |
| Classification Panel: | Cardiovascular |
| Common Name: | Upper arm blood pressure monitor |
| FDA Classification: | Class II |
| Submission Type: | 510(k) |
| Indications for Use: | The Globalcare GUS622 automatic Blood Pressure Monitor is<br>indicated for home use for the non-invasive measurement of<br>diastolic and systolic blood pressures and pulse rate of adults by<br>means of an inflatable cuff which is wrapped around the upper<br>arm. The cuff circumference is limited to 22 to 42 cm. |
{4}------------------------------------------------
### Device Description:
The Globalcare GUS622 Blood Pressure Monitor is an automatic upper arm blood pressure monitor intended for non-invasive measurement or monitoring of adults' arterial blood pressure and pulse rate.
The device is powered by batteries or by a mains-connected power supply.
The Globalcare GUS622 Blood Pressure Monitor comprises the following parts:
| • Blood pressure monitor main device | x1 |
|----------------------------------------------|----|
| • Upper arm cuff | x1 |
| • 1.5 V LR03 AAA batteries | x4 |
| • Storage bag | x1 |
| • Instructions for use | x1 |
| • 100-240Vac 50/60Hz power supply (optional) | x1 |
None of the parts included with the device are supplied in a sterile condition.
The Globalcare GUS622 Blood Pressure Monit is an electronic unit incorporating a screen that displays results and other information relevant to device operation. The device is powered by batteries or by a mains-connected power adapter, and is designed for home ('OTC') use.
The clinical outputs from the GUS622 Blood Pressure Monitor are:
- . Systolic pressure (mm Hg)
- Diastolic pressure (mm Hg)
- Pulse rate (1/min) ●
- Cardiac arrhythmia (Irregular heart beat (IHB)) symbol .
- Level of risk ●
- Hemodynamic Rest Condition (HSD) indication
The device operates on the oscillometric method: pressure sensors on the cuff are used to capture pulse waves during constricted blood flow and the device then computes the results for display by means of an algorithm which analyses the pressures transmitted during arterial oscillations that occur during cuff deflation. See Section 16 of this submission for further details of the software utilized in the device.
Irregular heart beat (IHB) is detected by measuring the interval time between each heart beat and comparing these values with the average heart beat interval (heart beat per minutes shown on the display). If the difference between any heart beat interval and the average is higher or lower than 25 %, the device shows the cardiac arrhythmia (IHB) symbol on the display.
The pulse rate is determined by calculating the frequency of the cuff, which are cardiac synchronous.
The GUS622 is controlled by software which calculates the blood pressure (diastolic and pulse rate by the oscillometric method. To do this, it also controls the pneumatic components to inflate the cuff (with a pump), holds the cuff pressure, and then deflates the cuff (by means of a solenoid valve). The software collects the relevant data from the pressure sensor during deflation.
The software also drives the LCD display, from which the user can read the result of the blood pressure and pulse rate measurements, and undertakes the Hemodynamic Rest Condition (HSD) detection calculations, and reports whether any irregular heart beat is occurring.
### Performance data:
The GUS622 has been tested and found to be in compliance with the following standards:
Safetv and EMC:
- ANSI/AAMI ES60601-1:2005 / A2:2010
{5}------------------------------------------------
- IEC 60601-1-11:2015 .
- IEC 60601-1-2:2014 ●
- IEC 80601-2-30:2009/AMD1:2013 .
Clinical performance:
- ISO 81060-2:2013 ●
- EN 1060-3:1997+A2:2009 ●
Cuff biocompatibility:
- ISO 10993-5:2009 ●
- ISO 10993-10:2010 .
### Bench tests:
- Internal tests to verify performance ●
- Mains and Battery comparison o
- After 500 hour Life Test has been completed o
- After 12 months storage o
- o After drop test
The results of the above testing assist in the demonstration of substantial equivalence of the subject device with the predicate device.
### Substantial equivalence
The predicate device selected for comparison with the GUS622 Blood Pressure Monitor is:
| Predicate Device: | Blood Pressure Monitors Models KD-927 and KD-928 |
|----------------------|--------------------------------------------------|
| Sponsor: | Andon Health Co., Ltd., China |
| 510(k) Number: | K141984 |
| Clearance Date: | 29 April 2015 |
| FDA Product Code: | DXN |
| Classification Name: | Noninvasive blood pressure measurement system |
| Regulation No: | 21 CFR 870.1130 |
| Class: | II |
### Predicate device comparison table:
The following Table 1 provides evidence of substantial equivalence of the subject device with the selected predicate device.
| Table 1: Predicate device comparison table | | | |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Feature | Subject device | Predicate device | Similarity |
| Device name | GUS622 | KD-927/928 | N/A |
| Device Manufacturer | Globalcare, China | Andon Health, China | N/A |
| 510(k) Reference | This submission | K141984 | N/A |
| FDA Product Code | DXN | DXN | Same |
| FDA Classification Name | Noninvasive blood pressure measurement system | Noninvasive blood pressure measurement system | Same |
| FDA Regulation Number | 21 CFR 870.1130 | 21 CFR 870.1130 | Same |
| Device description | Automatic upper arm blood pressure monitor intended for non-invasive measurement or monitoring of adults' arterial blood pressure and pulse rate via an upper arm cuff. | Automatic upper arm blood pressure monitor intended for non-invasive measurement or monitoring of adults' arterial blood pressure and pulse rate via an upper arm cuff. | Same |
| Feature | Subject device | Predicate device | Similarity |
| Indications for<br>use | The Globalcare GUS622<br>automatic Blood Pressure<br>Monitor is indicated for home use<br>for the non-invasive<br>measurement of diastolic and<br>systolic blood pressures and<br>pulse rate of adults by means of<br>an inflatable cuff which is<br>wrapped around the upper arm.<br>The cuff circumference is limited<br>to 22 to 42 cm. | KD-927 and KD-928 Fully<br>Automatic Electronic Blood<br>Pressure Monitor are for use by<br>medical professionals or at home<br>and are non-invasive blood<br>pressure measurement system<br>intended to measure the diastolic<br>and systolic blood pressures and<br>pulse rate of an adult individual<br>by using a non-invasive<br>technique in which an inflatable<br>cuff is wrapped around the upper<br>arm. The cuff circumference is<br>limited to 22 cm-48 cm. | Substantially equivalent |
| Use environment | Indoor use, home environment | Indoor use, home and healthcare<br>facilities environments | Difference has no effect on<br>safety or effectiveness |
| Measuring<br>method | Oscillometric method, automatic<br>inflation and measurement | Oscillometric method, automatic<br>inflation and measurement | Same |
| Device<br>measurements<br>(output<br>parameters) | Diastolic and systolic blood<br>pressure<br>Pulse rate | Diastolic and systolic blood<br>pressure<br>Pulse rate | Same |
| Additional output<br>indications | Irregular heart beat<br>hemodynamic instability | Irregular heart beat<br>hemodynamic instability | Same |
| Output result<br>calculation | Software algorithm | Software algorithm | Same |
| Measurement<br>range | Cuff pressure: 0-300 mmHg<br>Systolic: 50-280 mmHg<br>Diastolic: 30-200 mmHg<br>Pulse: 40-199 beats/min | Cuff pressure: 0-300 mmHg<br>Systolic: 60-260 mmHg<br>Diastolic: 40-199 mmHg<br>Pulse: 40-180 beats/minute | Substantially equivalent:<br>Differences have no effect<br>on safety or effectiveness |
| Accuracy | Pressure: ±3 mmHg<br>Pulse rate: ±5% | Pressure: ±3 mmHg<br>Pulse rate: ±5% | Same |
| External Power<br>Supply | Input 100-240Vac 50/60Hz 0.5A<br>max; Output 6VDC<br>(+/-5%) / 600mA (optional) | Input 100-240Vac 50/60Hz 0.5A<br>max; Output 6VDC<br>(+/-5%) / 600 mA | Same |
| Unit voltage | 6V dc | 6V dc | Same |
| Batteries | 4 x AAA alkaline 1.5V dc | 4 x AAA alkaline 1.5 V dc | Same |
| Battery life | Approx 500 cycles | Approx 200 cycles | Difference has no effect on<br>safety or effectiveness |
| Standards<br>compliance | ANSI/AAMI ES60601-1:2005 /<br>A2:2010<br>IEC 60601-1-11:2015<br>IEC 60601-1-2:2014<br>IEC 80601-2-30:2009/<br>AMD1:2013<br>ISO 81060-1:2012<br>EN 1060-3:1997+A2:2009 | IEC 60601-1:2005/AC:2010<br>IEC60601-1-2:2007/AC:2010<br>IEC 80601-2-30:2009 +<br>Cor.2010 | Subject device complies<br>with more recent standards |
| Device Protection | IEC 60601-1: Class 2 | IEC 60601-1: Class 2 | Same |
| Application Part | IEC 60601-1: Type BF | IEC 60601-1: Type BF | Same |
| Memory | 60 x 2 user | 60 x 2 user | Same |
| Dimensions | L 98 mm x W 140 mm x H 53<br>mm | L 95 mm x W 186 mm x H 56<br>mm | Differences have no effect<br>on safety or effectiveness |
| Weight | 207 g device without cuff and<br>batteries | 330 g without cuff and batteries | Difference has no effect on<br>safety or effectiveness |
| USB Port? | No | Yes | Different |
| Compatible PC<br>software? | No | Yes | Different |
{6}------------------------------------------------
{7}------------------------------------------------
The subject device and the predicate device have many identical or similar properties or features. The differences that exist and are identified in the above table include:
- Indications for use
- . No USB port on subject device
- . Subject device has no compatible PC software
None of the identified differences introduce new aspects of safety or effectiveness.
### Conclusion
The subject and predicate devices have very similar intended uses and fundamental technological characteristics. Any differences in technological characteristics between subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.