FULL AUTOMATIC (NIBP) BLOOD PRESSURE MONITOR (MODEL HL168KF)

K113239 · Health & Life Co., Ltd. · DXN · Mar 16, 2012 · Cardiovascular

Device Facts

Record IDK113239
Device NameFULL AUTOMATIC (NIBP) BLOOD PRESSURE MONITOR (MODEL HL168KF)
ApplicantHealth & Life Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateMar 16, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

HL168KF automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for use by people over the age of 18 with wrist circumference ranging from approx. 5.3 to 7.7 inches(13.5 cm to 19.5 cm) and for home use. When the device detects the appearance of irregular heartbeats during measurement, an indicated symbol will appear with measuring readings. Besides, HL168KF features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth device after measurement.

Device Story

Wrist-worn oscillometric blood pressure monitor; measures systolic/diastolic pressure and heart rate. Input: cuff pressure oscillations. Output: LCD display of vitals; irregular heartbeat indicator; Bluetooth transmission of results to paired devices. Used in home setting by adult patients. User sets personal target values; device alerts if measurement exceeds target. Benefits: enables self-monitoring of blood pressure and heart rate; facilitates data tracking via Bluetooth connectivity.

Clinical Evidence

Bench testing only. Device demonstrated compliance with ANSI/AAMI SP10:2002/(R)2008 standards for electronic sphygmomanometers. Safety and performance verified through IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), FCC Part 15, and ISO 10993 (biocompatibility) testing.

Technological Characteristics

Oscillometric measurement; automatic inflation/deflation. Materials: ABS housing, rubber keys. Power: 2x AAA alkaline batteries. Connectivity: Bluetooth for data transmission. Standards: ANSI/AAMI SP10, IEC 60601-1, IEC 60601-1-2, ISO 10993, ISO 14971.

Indications for Use

Indicated for adults aged 18+ with wrist circumference 13.5–19.5 cm for home-based, non-invasive measurement of systolic/diastolic blood pressure and heart rate via oscillometric method.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # PREMARKET NOTIFICATION MAR 1 6 2012 # 510(k) SUMMARY #### (As Required By 21 CFR 807.92) This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR \$807.92. The assigned 510(k) number is: Date: 2011.11.1 #### 1. Submitter: Health & Life Co., Ltd. 9F, No.186, Jian Yi Road, Zhonghe District, New Taipei City, Taiwan, R.O.C. TEL : +886-2-8227-1300 FAX : +886-2-8227-1301 Contact person: Sarah Su/ Regulatory Affairs Dept. E-mail: sarah.su@hlmt.com.tw Tel: 886-2-8227-1300 ext.1201 Fax: 886-2-8227-1301 #### 2. Name of the Device: Trade Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL 168KF Common Name: Blood Pressure Monitor Classification Name: Non-invasive Blood Pressure Measurement System Classification: Class II, 21CFR 870.1130 Product Code: DXN Panel: Cardiovascular #### 3. Information for the 510(k) Cleared Device (Predicate Device): A. Full Automatic (NIBP) Blood Pressure Monitor, Model HL168KB, K092163 B. A&D Digital Blood Pressure Monitor, Model UA-767PBT, K043217 #### 4. Device Description: HL168KF uses the oscillometric method to automatically measure systolic and diastolic blood pressure as well as heart rate. The measurement position is at human being's wrist. All values can be read out in one LCD panel. The device is designed for home use and recommended for use by adults aged 18 years and older with wrist circumference ranging approx. 5.3 ~ 7.7 inch (13.5 ~ 19.5 cm). {1}------------------------------------------------ The user is able to set the personal target value and the device will flash the value when the measured blood pressure value exceeds the target one. Further, HL168KF features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth device after measurement. Additionally, the device will display a symbol of wor or mo. to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat during the measurement. #### 5. Intended Use HL168KF automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for use by people over the age of 18 with wrist circumference ranging from approx. 5.3 to 7.7 inches(13.5 cm to 19.5 cm) and for home use. #### 6. Comparison of device to predicate device: ### Product Specification Comparison Table of HL168KF and HL168KB (K092163) | Item | Predicate<br>HL168KB (K092163) | Subject device<br>HL168KF | |--------------------------|------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------| | Method of<br>measurement | Oscillometric | Same as left | | Measurement<br>range of | Pressure 0- 300mmHg<br>Pulse 40-199 beats/minute | Rated range of cuff pressure: 0- 300mmHg<br>Rated range of determination: 40- 280mmHg<br>Pulse 40-199 beats/minute | | Accuracy | Pressure +/- 3mmHg<br>Pulse +/- 5% | Same as left | | Inflation | Automatic inflation (Air pump) | Same as left | | Deflation of<br>pressure | Automatic air release<br>control valve | Same as left | | Display | Liquid Crystal Display | Same as left | | Memory | 40 memory * 3 users<br>(120 memory total) | 99 memory * 1 user<br>(99 memory total) | | Cuff size | Wrist circumference<br>approx. 135 ~ 195 mm<br>(approx.5.3 ~ 7.7 inch) | Same as left | | Operation<br>environment | 50°F ~ 104°F (10°C ~ 40°C),<br>15%~90%R.H. | Same as left | {2}------------------------------------------------ | Storage/<br>Transportation<br>environment | -4°F ~ +158°F (-20°C ~ +70°C),<br>≤ 90% R.H. | Same as left | |-------------------------------------------|----------------------------------------------|-----------------------------------------------------| | Power Supply | 2 × "AAA" (1.5V)<br>Alkaline battery | Same as left | | Material | ABS housing and<br>rubber keys | Same as left | | Number of<br>Push Bottom | 5 | Same as left | | Storage case | Yes | Same as left | | Unit Weight | Approx. 148g<br>(Including batteries) | Approx. 119 ± 5 g<br>(Excluding cuff and batteries) | #### Changes from the predicate devices HL168KB (K092163): *Changing of exterior casing design and sets of memory *Additional product feature of Bluetooth Data Transmission Function For the product feature of Bluetooth data transmission, was compared with the other predicate device A&D UA-767PBT (K043217). ### 7. Discussion of Clinical Tests Performed: HL168KF is compliant to the ANSI/AAM1 SP10:2002/(R) 2008 & ANSI/AAM1 SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 Standard for Manual, electronic, or automated sphygmomanometers. All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met. # 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: The subject device was tested to evaluate its safety and effectiveness, including the followings: - a. Safety Test: IEC 60601-1:2005, Medical electrical equipment Part 1: General requirements for basic safety and essential performance. - b. EMC Test: IEC 60601-1-2 Edition 3:2007-03, Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests. - c. FCC Test: FCC 47 CFR Part 15, Subpart B - d. Biocompatibility Test: - ISO 10993-1:2009, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. - ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity. {3}------------------------------------------------ - ISO 10993-10:2002/Amd. 1:2006(E), Biological evaluation of medical devices -- Part 10: Tests for irritation and delayed-type hypersensitivity amendment 1. - e. Reliability Test: ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008, Manual, electronic or automated sphygmomanometers. - ﯾﻨﮯ Risk Assessment: ISO 14971:2007 Medical devices - Application of usability engineering to medical device. #### 9. Conclusions: The subject device was tested and fulfilled the requirements from those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate devices. {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing strength and protection. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is simple and recognizable, representing the department's role in promoting the health and well-being of Americans. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAR 1 6 2012 Health & Life Co., Ltd c/o Ms. Sarah Su Manager 9F, No. 186, Jian Yi Road Zhonghe District 23553, New Taipei City Taiwan Re: K113239/S001 Trade/Device Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL168KF Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: November 1, 2011 Received: November 2, 2011 Dear Ms. Su: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ## Page 2 – Ms. Sarah Su Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indication for Use 510(k) Number (if known): Device Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL 168KF #### Indications for Use: HL168KF automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for use by people over the age of 18 with wrist circumference ranging from approx. 5.3 to 7.7 inches(13.5 cm to 19.5 cm) and for home use. When the device detects the appearance of irregular heartbeats during measurement, an indicated symbol will appear with measuring readings. Besides, HL168KF features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth device after measurement. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ V (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Devices Evaluation (ODE) Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...