CARDIMA PATHFINDER 1.5F MAPPING MICROCATHETER

K971975 · Cardima, Inc. · DRF · Jul 1, 1998 · Cardiovascular

Device Facts

Record IDK971975
Device NameCARDIMA PATHFINDER 1.5F MAPPING MICROCATHETER
ApplicantCardima, Inc.
Product CodeDRF · Cardiovascular
Decision DateJul 1, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1220
Device ClassClass 2

Intended Use

The Cardima Pathfinder® 1.5F Mapping Catheter is a diagnostic medical device designed to provide localized unipolar or bipolar sensing of electrical potentials as measured from within the venous system of the heart. The catheter is available with 4, 1mm long electrodes spaced either 2mm or 2-6-2mm apart providing for a total electrode recording length of 10-14 mm. The working outer diameter is 1.5F and will be available with working lengths of 135cm and 170cm. The Pathfinder is not indicated for pacing or ablation. The outer surface is treated with a hydrophilic coating to enhance lubriciousness.

Device Story

Diagnostic mapping microcatheter; 1.5F diameter; 135cm/170cm lengths. Inputs: electrical potentials from coronary venous system via 4 electrodes. Operation: catheter inserted into coronary veins; electrodes sense unipolar/bipolar electrical signals; signals transmitted to external recording equipment. Used by physicians in clinical electrophysiology labs. Output: electrical signals displayed for clinician interpretation. Benefit: enables localized cardiac electrical potential mapping to assist in diagnostic electrophysiology studies.

Clinical Evidence

Animal studies conducted to evaluate maneuverability and signal quality in coronary veins. Catheter successfully positioned in anterior interventricular, middle cardiac, posterolateral, lateral, and anterolateral veins. Signals reported as good quality and adequate for clinical applications. Minor complications noted (slight endothelial injury, small subepicardial hemorrhage) attributed to guiding catheter, not considered clinically significant. No human clinical data provided.

Technological Characteristics

1.5F mapping microcatheter; 4 electrodes (1mm length); hydrophilic coating. Sensing: unipolar/bipolar electrical potentials. Sterilization: gamma radiation. Packaging: polyethylene protective coil, Tyvek/polyester-polyethylene pouch. Connectivity: wired to external recording equipment.

Indications for Use

Indicated for electrogram recording from within the coronary venous system during diagnostic electrophysiology studies. Not indicated for pacing or ablation.

Regulatory Classification

Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Trade Name: Cardima Pathfinder 1.5F Mapping Microcatheter ### Manufacturer: Cardima, Inc. 47266 Benicia Street Fremont, CA 94538-7330 Contact: Jack P. Douglas, Ph.D. Establishment Registration Number: 9007594 #### Classification Name: Electrode Recording Catheter (21 CFR 870.1220) ## Device Classification: Class II (21 CFR 870.1220) Panel: Circulatory System Devices Panel, DCRND ## Intended Use and Product Description: The Cardima Pathfinder® 1.5F Mapping Catheter is a diagnostic medical device designed to provide localized unipolar or bipolar sensing of electrical potentials as measured from within the venous system of the heart. The catheter is available with 4, 1mm long electrodes spaced either 2mm or 2-6-2mm apart providing for a total electrode recording length of 10-14 mm. The working outer diameter is 1.5F and will be available with working lengths of 135cm and 170cm. The Pathfinder is not indicated for pacing or ablation. The outer surface is treated with a hydrophilic coating to enhance lubriciousness. ## Sterilization, Packaging and Pyrogenicity: The Pathfinder 1.5F Mapping Catheter is packaged individually in a polyethylene protective coil. Each coll is in turn placed in a clear, blue tinted tray and a translucent-white Tyvek lid that is heat sealed to the outside of the tray . This sealed tray is placed within a standard Tyvek/polyester-polyethylene pouch which is also heat sealed. All product is sterilized using gamma radiation. #### Substantial Equivalence: The Pathfinder 1.5F Mapping Catheter is a modified version of Cardima's 2.5F Pathfinder electrode recording catheter. Establishment of equivalence was based on similarities of labeling, design, materials, and physical characteristics as evaluated by physical bench testing and biocompatibility. The primary design modification involves a change of outer diameter from 2.5F to 1.5F. #### Summary of Safety and Effectiveness: Safety and effectiveness were evaluated through biocompatibility testing and animal studies. Biocompatibility testing was conducted on final sterilized Pathfinder product per ISO 10993-1 and shown to be acceptable for all categories. Reliability testing was conducted on sterilized product per "Electrode Recording Catheter Preliminary Guidance, March 1995". The tests were used to assess the mechanical and electrical properties of both the catheter and the connector cable and found to be acceptable for use. Animal studies were conducted to evaluate the maneuverability of the Pathfinder 1.5F catheter positioned in the coronary sinus and veins. The catheter was shown to be easily positioned into selected coronary veins including the anterior interventricular, middle cardiac, posterolateral, lateral and anterolateral veins without difficulty, and signals were of good quality and adequate for clinical applications. Effusion of dye outside the coronary venous system was not evident, nor was there pericardial effusion found on gross examination of the heart. Some procedural complications noted most likely attributable to the guiding catheter included slight endothelial injury with consequent thrombus formation in one case. Some small subepicardial hemorrhage was found around the proximal middle cardiac vein in 2 dogs, and the coronary sinus area in 2 dogs. These complications are not considered to be clinically significant. {1}------------------------------------------------ Image /page/1/Picture/10 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name arranged in a circular pattern around the perimeter. At the center of the seal is a stylized caduceus symbol, which is often associated with healthcare and medicine. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 1 1998 JUL - Jack P. Douglas, Ph.D. Director of Requlatory Affairs Cardima, Inc. 47266 Benicia Street P.O. Box 14172 Fremont, CA 94538-7330 Re: K971975 Cardima Pathfinder™ 1.5F Mapping Microcatheter Requlatory Class: II (two) Product Code: DRF Dated: April 7, 1998 Received: April 8, 1998 Dear Dr. Douglas: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {2}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radioloqical Health Enclosure {3}------------------------------------------------ Page _________________________________________________________________________________________________________________________________________________________________________ of K971975 / 510(k) Number (if known): Device Name: Cardima Pathfinder 1.5F Mapping Microcatheter TM Indications For Use: The Cardima Pathfinder 1.5F Microcatheter is intended for electrogram recording from within the coronary venous system during diagnostic electrophysiology studies. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off) Division of Cardi voscular, Respiratory, and Neurological Dev **510(k) Number** K971975 Image /page/3/Picture/9 description: The image shows the words "Prescription Use" and "(Per 21 CFR 801.109)". The words are written in a clear, legible font. There is a line going through the words "Prescription Use". OR Over-The-Counter Use (Optional Format 1-2-96) 000007
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...