K955802 · Cardima, Inc. · DRF · Jan 29, 1997 · Cardiovascular
Device Facts
Record ID
K955802
Device Name
PATHFINDER
Applicant
Cardima, Inc.
Product Code
DRF · Cardiovascular
Decision Date
Jan 29, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
Both devices are made to aid physicians specializing in cardiac electrophysiology, cardiac surgery and interventional cardiology in the management of abnormal rhythms of the heart. The main function of both devices is diagnostic sensing of electrical impulses in the heart muscles to provide precise, localized data.
Device Story
Diagnostic electrode recording catheter; 2.5 Fr diameter; multiple electrodes at distal end; conductor wires connect to proximal connector; links to standard ECG recording/monitoring system. Used by cardiac electrophysiologists/surgeons to sense electrical signals in heart muscle; provides localized data for arrhythmia management. Accessed via peripheral vasculature using Seldinger technique; proximal torque control for navigation; soft, shapeable tip. Output is electrical signal displayed on external ECG monitor; assists physician in mapping cardiac electrical activity; enables precise diagnosis of abnormal rhythms. Benefit: high-quality diagnostic data via smaller diameter catheter compared to predicate.
Clinical Evidence
Controlled clinical investigation, 20 patients undergoing electrophysiology study for arrhythmia. Within-patient control design: signal recordings obtained with Pathfinder followed by Orthogonal device in coronary sinus. Endpoints: signal quality, maneuverability/placement, and complications. Results: Pathfinder demonstrated equivalent safety and effectiveness to Orthogonal device.
Technological Characteristics
2.5 Fr diagnostic catheter; 4, 8, 16 electrode configurations; distal recording length up to 7 cm; lengths 100-200 cm. Biocompatible materials. Passive sensing of cardiac electrical impulses. Connects to standard ECG recording/monitoring systems. Sterilized product.
Indications for Use
Indicated for physicians specializing in cardiac electrophysiology, cardiac surgery, and interventional cardiology for diagnostic sensing of cardiac electrical impulses in patients with abnormal heart rhythms.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}
CARDIMAP
JAN 29 1997
Pathfinder 510 (k)
K955802
12/22/95
# Section 2 - Summary of Safety and Effectiveness
This submission is prepared to show equivalence between the Cordis Webster Orthogonal (predicate) and the Cardima Pathfinder (new) electrode recording devices, according to the Safe Medical Device Act of 1990.
The criteria for equivalence included comparing labeling, physical characteristics and clinical performances of the new and predicate devices as well as physical and animal testing of the new device with biocompatibility and sterilization information.
Both devices are made to aid physicians specializing in cardiac electrophysiology, cardiac surgery and interventional cardiology in the management of abnormal rhythms of the heart.
The main function of both devices is diagnostic sensing of electrical impulses in the heart muscles to provide precise, localized data. While the predicate device is also indicated for pacing, the new device is not intended for pacing and it is labeled so.
Longitudinal characteristics of both catheters are the same, both were designed to access the heart through the peripheral vasculature using the standard Seldinger technique.
Both devices are capable to change direction during advancement using proximal torque control. Leading the change in direction of the relatively stiff predicate device is a deflectable tip, while the new device is designed with a soft, shapeable tip, much like coronary guidewires.
While the 7 Fr. predicate device exposes larger electrodes to the heart, the signal from the 2.5 Fr. smaller diameter new device yields the same quality diagnostic information to the physician. Included with this submission is the clinical data and a conclusive statement from the independent reviewers supporting the equivalence claim.
## Device Description
The 2.5 Fr. Pathfinder Electrode Recording Catheter is a diagnostic medical device designed to sense electrical signals by use of multiple electrodes placed on the distal end of the catheter, each of which is connected to a connector at the proximal end of the catheter by means of a conductor wire. The connector is matched with a connector cable which functions to link the catheter to a standard ECG recording/monitoring system. The device is a composite construction of biocompatible materials commonly used in other (FDA) approved catheters.
{1}
CARDIMA
Pathfinder 510 (k)
2/2
12/22/95
The catheter is available in 4, 8 and 16 electrode configurations and provides for a distal electrode recording length of up to 7 cm. The working outer diameter of the catheter is 2.3F or 2.5F depending upon the number of electrodes. The useable lengths are 100, 135, 150 and 200cm.
## Nonclinical Tests
**Bench Studies:** A series of tests were performed using the FDA “Electrode Recording Catheter Preliminary Guidance, June 1994” as a guide. The tests were used to assess the mechanical and electrical properties of both the catheter and the connector cable. Mechanical testing included tensile and torsional strength, angular rotation, coil/tip stiffness and flexural fatigue. Electrical testing included electrical phase shift and dielectric voltage breakdown. All test samples met the defined acceptance criteria for each performance test. In addition to mechanical and electrical testing, a full series of biocompatibility testing was performed on final, sterilized product. Testing included cytotoxicity, hemolysis, mutagenicity, systemic/ intracutaneous/intravenous toxicity, muscle implantation, and delayed contact sensitization.
**Animal Studies:** A pre-clinical GLP animal study was conducted to evaluate the performance of the Pathfinder device in recording cardiac signal recording from the coronary sinus when compared to the Orthogonal device. The study demonstrated that the Pathfinder could be successfully placed in the coronary sinus, could record cardiac signals of adequate quality from the coronary sinus, and did not present any increase in device associated complications as compared to the Orthogonal catheter.
## Clinical Tests
A controlled clinical investigation was conducted at two clinical sites consisting of a total of 20 patients undergoing an electrophysiology study for arrhythmia. The purpose of the investigation was to demonstrate equivalent safety and effectiveness of the Pathfinder device when compared to the Orthogonal device. The study design was to use patients as their own control, in which, in the same patient, signal recordings were obtained with the Pathfinder device immediately followed by signal recordings obtained with the Orthogonal device in the coronary sinus. To demonstrate safety and effectiveness, signal recordings, device maneuverability/placement, and complications were compared. The results indicated that the Pathfinder device is substantially equivalent to the Orthogonal device with respect to safety and effectiveness.
## Conclusions Drawn from Nonclinical and Clinical Tests
The results of the nonclinical and clinical tests summarized in sections (b)(1) and (b)(2) above, indicate that the Cardima Pathfinder Electrode Recording Catheter exhibits substantially equivalent safety and effectiveness when compared to the Cordis Webster Orthogonal Electrode Recording Catheter.