Hansen Medical Magellan Robotic Catheter 9Fr

K160919 · Hansen Medical, Inc. · DRA · Sep 20, 2016 · Cardiovascular

Device Facts

Record IDK160919
Device NameHansen Medical Magellan Robotic Catheter 9Fr
ApplicantHansen Medical, Inc.
Product CodeDRA · Cardiovascular
Decision DateSep 20, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1280
Device ClassClass 2

Intended Use

The Hansen Medical Magellan™ Robotic Catheter 9Fr is intended to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices. The Magellan Robotic Catheter 9Fr Long Length (80 cm) is also indicated for use as a conduit for manual placement of therapeutic devices in the neuro vasculature. The Magellan™ Robotic Catheter 9Fr is intended to be used with the Hansen Medical Magellan™ M Robotic System and accessories.

Device Story

The Magellan Robotic Catheter (MRC) 9Fr is a steerable, telescoping catheter system designed for use with the Hansen Medical Magellan Robotic System. It consists of a steerable Leader catheter nested within a steerable Guide catheter. The system is operated by a physician via the robotic interface to navigate to targets in peripheral or neuro vasculature. The distal sections of both components articulate independently; the Leader can extend up to 21 cm beyond the Guide. Once the catheter is positioned at the target site, it serves as a conduit for the manual delivery of therapeutic devices. The device is intended for use in clinical settings where robotic-assisted vascular access is required. By providing precise, robotic-controlled articulation, the device facilitates navigation through complex vascular anatomy, potentially improving access to target sites compared to manual-only techniques.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including design verification and validation activities comparing performance attributes (diameters, lengths, tip shape) to the predicate device.

Technological Characteristics

Telescoping steerable catheter system (Leader and Guide). Leader: 6Fr OD, 3Fr ID, 3cm articulating section. Guide: 9Fr OD, 6Fr ID, 2.8cm articulating section. Materials include radiopaque marker bands. Articulation via four pull wires per component. Sterilized via EtO. Single-use. Designed for use with the Hansen Medical Magellan Robotic System.

Indications for Use

Indicated for navigation to anatomical targets in peripheral vasculature and as a conduit for manual placement of therapeutic devices. The 80 cm Long Length version is also indicated for use as a conduit for manual placement of therapeutic devices in the neuro vasculature.

Regulatory Classification

Identification

A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, layered on top of each other to create a sense of depth. The profiles are black and are positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the graphic. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 20, 2016 Hansen Medical, Inc. Joy Sacmar VP. OA & RA 800 E Middlefield Road Mountain View, California 94043 Re: K160919 Trade/Device Name: Hansen Medical Magellan Robotic Catheter 9Fr Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable Catheter Regulatory Class: Class II Product Code: DRA Dated: August 15, 2016 Received: August 16, 2016 Dear Joy Sacmar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Mude Yellman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160919 Device Name Hansen Medical Magellan™ Robotic Catheter 9Fr Indications for Use (Describe) The Hansen Medical Magellan™ Robotic Catheter 9Fr is intended to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices. The Magellan Robotic Catheter 9Fr Long Length (80 cm) is also indicated for use as a conduit for manual placement of therapeutic devices in the neuro vasculature. The Magellan™ Robotic Catheter 9Fr is intended to be used with the Hansen Medical Magellan™ M Robotic System and accessories. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### SECTION 5.0 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. #### K160919 510(k) Number ## Applicant Information | Owner Name: | Hansen Medical, Inc. | |---------------------------------------|-------------------------------------------------------| | Address: | 800 East Middlefield Road<br>Mountain View, CA. 94043 | | Office Phone: | 650-404-5800 | | Establishment<br>Registration Number: | 3006026430 | | Contact Person: | Marysa Loustalot | | Company Contact: | Joy M. Sacmar, Vice-President RA/QA | | Phone Number: | 650 404 2777 | | Facsimile Number: | 650 404 5901 | | Date Prepared: | September 13, 2016 | #### De vice Information | Regulatory Class: | Class II | |----------------------|----------------------------------------------| | Trade/Device Name: | Hansen Medical Magellan Robotic Catheter 9Fr | | Common Name: | Robotic Control Catheter | | Classification Name: | Steerable catheter | | Regulation Number: | 21 CFR 870.1280 | | Product Code: | DRA | ## Predicate Device The Hansen Medical Magellan Robotic Catheter 9Fr is substantially equivalent in intended use and method of operation to the MicroVention Chaperon 6Fr Guiding Catheter (K082385). {4}------------------------------------------------ ## Reference Device The Hansen Medical Magellan Robotic Catheter 9Fr (K132369). ## De vice De scription The proposed change to the Hansen Medical Magellan™ Robotic Catheter (MRC) 9Fr is to the Indications for Use only; there is no change to the fit, form, or function of the device or to any of its components and accessories cleared under K132369. The MRC 9Fr is a telescoping design comprised of a steerable Leader within a steerable Guide and is designed to be used with the Hansen Medical Magellan™ Robotic System (Magellan System). The distal sections of each Hansen Catheter component articulate individually and the distal end of the Leader can extend up to 21cm beyond the distal end of the Guide. The Leader is a tubular catheter with four pull wires articulating the distal 3cm. The Leader has a 6Fr outer diameter and 3Fr inner diameter, and is available in 125cm, 141cm, and 158cm lengths. The Leader fits inside the Guide and is only used in conjunction with the Guide. The distal end of the Leader has a 3cm flexible articulating section. The Leader has a radiopaque marker band between the articulating section and the atraumatic tip. The tip extends 4mm distal to the marker band. The Guide is a tubular catheter with four pull wires articulating the distal 2.8cm. The Guide has a 6Fr inner diameter and 9Fr outer diameter. The Guide is available in 50cm, 65cm, or 80cm lengths. The Guide fits around the Leader and may be used with or without the Leader. When the Leader is removed from the Guide, a 6Fr compatible percutaneous catheter may be delivered through the Guide. Figure 5-1 below is a diagram taken from the current commercially available MRC 9Fr Guide (80cm) Box Label providing additional details on tip dimensions. Image /page/4/Figure/8 description: The image shows two diagrams of medical instruments with measurements. The first diagram on the left shows a blue instrument with an inner diameter (ID) of 0.039 inches (1.00 mm) and an outer diameter (OD) of 0.084 inches (2.13 mm). The instrument is shown bending 180 degrees over a length of 30 mm, with a height of 23 mm. The second diagram on the right shows a gray instrument with an ID.6 Fr of 0.084 inches (2.13 mm) and an OD.9 Fr of 0.128 inches (3.25 mm). This instrument is shown bending 90 degrees over a length of 27 mm, with a height of 23 mm. Figure 5-1 MRC 9Fr diagram taken from label PN 20995 Rev. C {5}------------------------------------------------ The Hansen Catheter family consists of three different lengths of Leader and Guide pairs, and all sizes are sterilized via EtO for single use only. However, the MRC 9Fr Short (50cm) and MRC 9Fr Medium (65cm) catheters are not included for the expanded indication as they are not long enough to access the target anatomy. Table 5-1 below lists all sizes the MRC 9Fr is currently available in. | Catalog<br>Number | 1. Guide<br>Effective<br>Length<br>(cm) | 2. Guide<br>Overall<br>Length<br>(cm) | 3. Leader<br>Overall<br>Length<br>(cm) | Total<br>Effective<br>Length<br>(cm) | Minimum<br>Guide Wire<br>Length for<br>Exchange (cm) | |-------------------|-----------------------------------------|---------------------------------------|----------------------------------------|--------------------------------------|------------------------------------------------------| | MC9F50<br>(11142) | 50 | 74 | 125 | 71 | 260 | | MC9F65<br>(11141) | 65 | 90 | 141 | 86 | 260 | | MC9F80<br>(11140) | 80 | 107 | 158 | 101 | 300 | Table 5-1 Magellan™ Robotic Catheter 9Fr - Lengths Image /page/5/Figure/3 description: The image shows a diagram with three measurements labeled 1, 2, and 3. Measurement 1 spans from the second square block to the start of the filtration system. Measurement 2 spans from the second square block to the end of the second filtration unit. Measurement 3 spans from the first square block to the end of the second filtration unit. Figure 5-2 Magellan™ Robotic Catheter 9Fr {6}------------------------------------------------ #### Intended Use The Hansen Medical Magellan™ Robotic Catheter 9Fr is intended to be used to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices. The Magellan Robotic Catheter 9Fr Long Length (80 cm) is also indicated for use as a conduit for manual placement of therapeutic devices in the neuro vasculature. The Magellan™ Robotic Catheter 9Fr is intended to be used with the Hansen Medical Magellan™ Robotic System and accessories." #### Comparison to Predicate Device(s) The Hansen Medical Magellan Robotic (MRC) 9Fr Long Length (80cm) Guide and Leader is substantially equivalent to the MicroVention 6Fr Chaperon Guiding Catheter System. The MRC 9Fr remains as a telescoping design, comprised of a steerable Leader within a steerable Guide and is designed to be used with the Hansen Medical Magellan™ Robotic System (Magellan System). The distal sections of each Hansen Catheter component articulate individually, and the distal end of the Leader can extend up to 21 cm beyond the distal end of the Guide. The MRC 9Fr Long Length is the only size available in the MRC 9Fr family that can access the neurovasculature at a total insertion length (Guide and Leader Catheter) of 101 cm. #### Technological Characteristics/Performance Data The MicroVention 6Fr Chaperon Guide catheter was chosen as the predicate device as it has a similar intended use and target anatomy as the MRC 9Fr Long Length (80cm) expanded indication. The Chaperon and Magellan catheters are both access catheters intended to provide a conduit for manual delivery of therapy. Diameters, lengths, and tip shape are the critical characteristics that determine the performance ability of an access catheter. The MRC 9Fr Long Length (80cm) with the expanded indication and the predicate MRC 9Fr device cleared under K132369 have the same technological characteristics. In comparison to the MicroVention 6Fr Chaperon guide and inner catheters (the predicate device), the MRC 9Fr Long Length (8cm) has similar characteristics that determine the performance of accessing the neuro vasculature. The minor differences in the characteristics between the two products do not alter the performance and do not present any new issues of safety or effectiveness. Performance of the MRC 9Fr {7}------------------------------------------------ Long Length (80cm) to access the neuro vasculature was further demonstrated in the design validation and verification activities described in Section 18.0 Performance Testing. Therefore, the Hansen Medical Magellan Robotic 9Fr Long Length (80cm) is substantially equivalent with respect to intended use and technological characteristics as the predicate device, and is expected to perform equivalently in accessing the neuro vasculature. # Clinical Testing Clinical evaluation is not required for this device. #### Substantial Equivalence | Product | Micro Vention Inc.<br>6 Fr Chaperon Guiding Catheter<br>System | Hansen Medical<br>Magellan Robotic Catheter 9Fr<br>w/ Expanded Indication | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Code | DQY (Percuraneous Catheter) | DRA (Steerable Catheter) | | Class | Class II | Class II | | 21CFR | 870.1250 | 870.1280 | | K Number | K082385 | K160919 | | Indications<br>for Use | Chaperon Guiding Catheter is<br>intended for general intravascular<br>use, including the neuro and<br>peripheral vasculature. Chaperon<br>Guiding Catheter can be used to<br>facilitate introduction of diagnostic<br>or therapeutic devices. Chaperon<br>Guiding Catheter is not intended for<br>use in coronary arteries. | The Hansen Medical Magellan™<br>Robotic Catheter 9Fr is intended to<br>be used to facilitate navigation to<br>anatomical targets in the peripheral<br>vasculature and subsequently<br>provide a conduit for manual<br>placement of therapeutic devices.<br><br>The Magellan Robotic Catheter 9Fr<br>Long Length (80 cm) is also<br>indicated for use in the U.S. as a<br>conduit for manual placement of<br>therapeutic devices in the neuro<br>vasculature.<br><br>The Magellan™ Robotic Catheter<br>9Fr is intended to be used with the<br>Hansen Medical Magellan™<br>Robotic System and accessories. | {8}------------------------------------------------ | MRC 9Fr and the Microvention Chaperon 6Fr | | | | | | |-------------------------------------------|------------------------------------------------------------------------|----------------------------------|-------------------------------------------------------------------------|--------------------|----------------------| | Performance<br>Specification | MRC 9Fr<br>Guide<br>(Long) | Chaperon<br>Guide | MRC 9Fr<br>Leader | Chaperon<br>Leader | Rationale<br>Section | | OD | 9Fr<br>0.128 in | 6F<br>0.084 in | 6Fr<br>0.084 in | 5F<br>0.071 in | 12.3.3.1 | | ID | 6Fr<br>0.084 in | 5Fr<br>0.071 in | 3Fr<br>0.039 in | 3Fr<br>0.048 in | 12.3.3.2 | | Guidewire<br>Compatibility | 0.035 in<br>0.018 in<br>0.014 in | 0.035 in<br>0.018 in<br>0.014 in | 0.035 in<br>0.018 in<br>0.014 in | 0.035 in | 12.3.3.3 | | Total<br>Insertion<br>Length | 80 cm | 95 cm | 101 cm | 117 cm | 12.3.3.4 | | Distal Flex<br>Length | 2.7 cm | 7 cm | 3.0 cm | 7 cm | 12.3.3.5 | | Hydrophilic<br>Coating<br>Length | 45 cm | N/A | 85 cm | 15 cm | 12.3.3.6 | | Tip<br>Configuration | Configurable<br>bend to any<br>angle up to<br>90° in all<br>directions | STR, MP2,<br>BUR | Configurable<br>bend to any<br>angle up to<br>180° in all<br>directions | VTR,<br>SIM2, JB2 | 12.3.3.7 | Performance Attribute and Characteristic Comparison of the Hansen Medical MRC 9Fr and the MicroVention Chaperon 6Fr ## Conclusion In comparison to the MicroVention 6Fr Chaperon guide and inner catheters (the predicate device), the MRC 9Fr Long Length (8cm) has similar characteristics that determine the performance of accessing the neuro vasculature. The minor differences in the characteristics between the two products do not alter the performance and do not present any new issues of safety or effectiveness. Therefore, the Hansen Medical Magellan Robotic 9Fr Long Length (80cm) is substantially equivalent with respect to intended use and technological characteristics as the predicate device, and is expected to perform equivalently in accessing the neuro vasculature as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...