The NovaCross Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.
Device Story
NovaCross Microcatheter is a sterile, single-use, over-the-wire percutaneous support catheter. It consists of a telescopic shaft, an over-tube, and a proximal handle for manual manipulation and flushing. A key feature is a deployable/retractable distal Nitinol scaffold, visible under fluoroscopy, which expands to the artery width to provide distal anchoring and guidewire support. Used in coronary and peripheral vasculature by physicians; facilitates access for therapeutic devices like atherectomy catheters, PTCA catheters, or stents. The device does not provide therapeutic benefit itself; it serves as a support tool to improve guidewire positioning. Benefits include enhanced support near the treatment site, facilitating subsequent interventional procedures.
Clinical Evidence
Bench testing included biocompatibility (ISO 10993-1), sterilization (ISO 11135-1), and mechanical testing (torque, tensile, pushability, scaffold durability, radial force). A GLP animal study (n=6 pigs) evaluated safety in coronary and peripheral arteries. Endpoints included vessel integrity, gross pathology, and histopathology at 1 and 14 days post-procedure. Results showed no discernible pathological adverse effects and successful achievement of all performance criteria.
Technological Characteristics
Single-lumen, over-the-wire percutaneous catheter. Features a deployable Nitinol scaffold for anchoring. Compatible with 7F guiding catheters and 0.014" guidewires. Working length 135cm. Materials evaluated per ISO 10993-1. Sterilization per ISO 11135-1. No software or electronic components.
Indications for Use
Indicated for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange in patients requiring such access.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three human profiles facing to the right, stacked on top of each other, with a flowing fabric-like element below them.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 4, 2015
Nitiloop Ltd. % Michael Daniel President, Daniel & Daniel Consulting, LLC Daniel & Daniel Consulting 340 Jones Lane Gardnerville. Nevada 89460
Re: K143608
> Trade/Device Name: NovaCross Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: April 10, 2015 Received: April 10, 2015
Dear Michael Daniel,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrand
for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143608
Device Name NovaCross Microcatheter, 1
Indications for Use (Describe)
The NovaCross Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| <div> <span style="display:inline-block;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="display:inline-block;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services | |
| Food and Drug Administration | |
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional Premarket Notification Submission - 510(k) NovaCross™ Microcatheter 510(k) Number K143608
#### I. SUBMITTER
Nitiloop Ltd. 8 Hamada St. P.O.Box 12892 Herzliya, Israel 46733 Tel: +972-52-396-8925 Fax: +972-72-2501631
#### Contact Person
Michael A. Daniel Daniel & Daniel Consulting, LLC 340 Jones Lane Gardnerville, NV 89460 Tel: 775.392.2970 Cell: 415.407.0223 Fax: 610.545.0799 madaniel@clinregconsult.com
#### II. DEVICE
Name of Device: NovaCross™ Microcatheter Common or Usual Name: NovaCross™ Microcatheter Classification Name: Percutaneous catheter (21 CFR 870.1250) Regulatory Class: II Product Code: DQY
#### III. PREDICATE DEVICE
The primary predicate device is the Roxwood Medical, Inc. MultiCross Support Catheter cleared under K121763, (product code DQY, Regulation No. 870.1250).
A reference device was used in this submission: Corsair Microcatheter (Asahi Intecc Co., Ltd. K083127).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for "Nitiloop CROSSING THE LIMITS". The logo features a red square with a white design on the left side. To the right of the square is the word "Nitiloop" in a bold, sans-serif font. Below "Nitiloop" is the phrase "CROSSING THE LIMITS" in a smaller, lighter font.
## IV. DEVICE DESCRIPTION
The NovaCross™ Microcatheter is a sterile, single-use, single lumen, over-the-wire, disposable percutaneous support catheter designed for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature for guidewire exchange.
The NovaCross™ Microcatheter consists of telescopic shaft. Over Tube, and a proximal Handle Body that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and retractable distal Nitinol Scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide an anchoring to aid the user in establishing greater support near the treatment site.
Subsequent to conventional guidewire placement, therapeutic devices such as atherectomy devices, PTCA catheters, and/or stents may be used to provide therapeutic benefit. The NovaCross™ Microcatheter by itself does not provide therapeutic benefit beyond simple facilitation of guidewire support. The NovaCross™ Microcatheter is similar in its design and it achieves its intended use by means of the same mechanisms as the predicate devices.
### V. INDICATIONS FOR USE
The NovaCross™ Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Both the MultiCross, and the NovaCross™ Microcatheters have a working length of 135cm and are both compatible with a 7F guiding catheter, as well as a 0.014" guidewire. Like the MultiCross Nitinol scaffold, the NovaCross™ Nitinol scaffold is also expanded to the diameter of the artery when deployed.
The two devices use similar technology including a delivery system. The guidewire support with the two devices is accomplished by extending a Nitinol scaffold. The Nitinol element for the NovaCross™, like that of the MultiCross predicate, provides distal anchoring and support, thus exhibiting similar safety questions. These safety questions have been evaluated for the NovaCross™ Microcatheter through extensive design verification and validation testing, including a GLP Animal Study.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for "Nitiloop". The logo consists of a red square with a white curved line inside, followed by the word "Nitiloop" in a bold, sans-serif font. Below the word "Nitiloop" is the phrase "CROSSING THE LIMITS" in a smaller, sans-serif font.
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
#### Biocompatibility testing
An evaluation of biocompatibility was performed in compliance with ISO 10993-1. Biocompatibility testing included cytotoxicity, irritation, acute systemic toxicity, sensitization, and hemocompatibility testing. All tests were successfully completed.
### Sterilization, Packaging and Shelf Life Testing
Sterilization validation testing was performed to demonstrate compliance with ISO 11135-1. In addition, shelf life and packaging testing were performed to support the labeled shelf life. All tests were successfully completed.
#### Mechanical Testing
Mechanical bench testing included the following:
- Dimensional
- Simulated use ●
- Tip flexibility ●
- Tip and shaft durability
- Torque response
- Torque strength
- Tensile strength
- Hydrophilic coating integrity
- Corrosion
- Fluid leak
- Air leakage
- Pushability and retractability ●
- Scaffold durability
- Radial force
All tests met the predefined acceptance criteria.
### Animal Study
A GLP animal study, consisting of 6 pigs, was used to test whether the NovaCross™ device could be used safely and effectively as a tool to provide support to the guidewire. The system was used on three arteries (two coronary and one peripheral) in each animal. The safety of the system was evaluated 1 and 14 days post procedure in terms of the integrity of the treated vessels (compared to intact control segments), gross pathology and histopathology. The results demonstrated that no discernible pathological adverse effects were detected, in terms of local and systemic effects
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "Nitiloop". The logo consists of a red square with a white design inside, followed by the word "Nitiloop" in black letters. Below the word "Nitiloop" is the phrase "CROSSING THE LIMITS" in smaller, gray letters.
following the deployment of the Test Item NovaCross™ Microcatheter in coronary and peripheral arteries. The Test Item successfully achieved all performance criteria defined in the study protocol. Pathology and histopathology analyses confirmed the results of the study.
## VIII. CONCLUSIONS
The NovaCross™ Microcatheter was determined to be substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.