The NovaCross CTO Microcatheter is intended to facilitate the intraluminal placement of conventional and steerable guidewires beyond the stenotic lesions, including chronic total occlusions (CTOs), and prior to PTCA or stent intervention. It is also intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.
Device Story
Sterile, single-use, over-the-wire percutaneous support catheter; used in coronary and peripheral vasculature. Input: manual manipulation by interventional cardiologist; guidewire. Device features stainless steel shaft and deployable/collapsible distal Nitinol scaffold; scaffold expands to artery width to provide anchoring/support near treatment site. Scaffold visible via fluoroscopy. Output: facilitated guidewire placement beyond stenotic lesions/CTOs. Device does not provide therapeutic benefit itself; enables subsequent use of atherectomy, PTCA, or stents. Used in cath lab/clinical setting. Benefits: improved guidewire support and access to difficult lesions.
Clinical Evidence
Multicenter, non-randomized, open-label, single-arm pivotal study (N=186). Primary safety endpoint: MACE rate (death, MI >3x ULN, urgent revascularization) was 12.3% (97.5% CI: 19.41%). Primary effectiveness endpoint: technical success (guidewire placement beyond CTO in true lumen) was 75.3%. Study concluded device is safe and effective.
Technological Characteristics
Single-lumen percutaneous catheter; stainless steel shaft; deployable/collapsible distal Nitinol scaffold for anchoring. Compatible with 0.014" guidewires. Sterilized via EO gas (ISO 11135:2014). Biocompatibility per ISO 10993-1. Mechanical testing: torque, tensile strength, pushability, retractability, scaffold durability.
Indications for Use
Indicated for patients requiring percutaneous cardiac intervention for chronic total occlusions (CTO) or access to discrete regions of coronary and peripheral vasculature to facilitate guidewire placement and exchange prior to PTCA or stent intervention.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
April 2, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.
Nitiloop Ltd. % Wanda Carpinella VP Regulatory Affairs Boston Biomedical Associates, LLC One Crowley Drive, Suite 216 Marlborough, Massachusetts 01752
# Re: K193322
Trade/Device Name: NovaCross CTO Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: March 3, 2020 Received: March 4, 2020
## Dear Wanda Carpinella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/ training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193322
Device Name NovaCross CTO Microcatheter
### Indications for Use (Describe)
The NovaCross CTO Microcatheter is intended to facilitate the intraluminal placement of conventional and steerable guidewires beyond the stenotic lesions, including chronic total occlusions (CTOs), and prior to PTCA or stent intervention. It is also intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional Premarket Notification Submission - 510(k) NovaCross™ CTO Microcatheter 510(k) Number K193322
Date Prepared: April 1, 2020
#### I. SUBMITTER
Nitiloop Ltd. 1 Tidhar Street P.O Box 1040, Pardes Hana 3711001 Israel Tel: +972-(0)4- 9874202
## Contact Person
Wanda Carpinella Company Regulatory Consultant Boston Biomedical Associates 100 Crowley Drive, Suite 216 Marlborough MA 01752 Cell: +1(781) 799-4572 wcarpinella@boston-biomedical.com
## Back-up Company Contact:
Orly Maor 25 Sirkin Street Kfar Saba 44421 Israel Tel: +972-9-7453607 oram.ma@gmail.com
#### II. DEVICE
Name of Device: NovaCross™ CTO Microcatheter Common or Usual Name: NovaCross™ CTO Microcatheter Classification Name: Percutaneous catheter (21 CFR 870.1250) Regulatory Class: II Product Code: DOY
#### PREDICATE DEVICE III.
Nitiloop Ltd. believes that the NovaCross™ CTO microcatheter, which is the subject of this notification, is substantially equivalent to the following predicate device:
- Boston Scientific; BridgePoint Medical system cleared under K102725
- ASAHI® Cosair and Cosair Pro Microcatheter cleared under K171933
In addition, the following reference device is used:
- . Nitiloop Ltd. NovaCross™ Microcatheter cleared under K143608 and K160389.
{4}------------------------------------------------
# IV. DEVICE DESCRIPTION
The NovaCross™ CTO Microcatheter is a sterile, single-use, single lumen, over-the-wire, disposable percutaneous support catheter designed to facilitate the intraluminal placement of conventional and steerable guidewires beyond stenotic lesions (including chronic total occlusions) prior to PTCA or stent intervention and to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature. The NovaCross™ CTO Microcatheter consists of stainless steel shaft, Over Tube, and a proximal Handle Body that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and collapsible distal Nitinol Scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide an anchoring to aid the user in establishing greater support near the treatment site.
Subsequent to conventional guidewire placement, therapeutic devices such as atherectomy devices, PTCA catheters, and/or stents may be used to provide therapeutic benefit. The NovaCross™ CTO Microcatheter by itself does not provide therapeutic benefit beyond simple facilitation of guidewire support. The NovaCross™ CTO Microcatheter is similar in its design and it achieves its intended use by means of the same mechanisms as the predicate devices.
#### V. INDICATIONS FOR USE
The NovaCross™ CTO Microcatheter device is intended to facilitate the intraluminal placement of conventional and steerable guidewires beyond stenotic lesions, including chronic total occlusions (CTO), prior to PTCA or stent intervention. It is also intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE AND REFERENCE DEVICE
The NovaCross™ CTO and the BridgePoint medical system and the ASAHI® Cosair / Cosair Pro Microcatheter have the same intended use. The NovaCross™ CTO and the BridgePoint Medical (the CrossBoss element), are both manufactured with stainless steel shafts with a lumen which is compatible for 0.014 guidwires. Also, both have stabilization systems - the BridgePoint system has an inflated balloon (Stingray) and the NovaCross™ CTO uses a nitinol scaffold. The ASAHI Corsair / Corsair Pro Microcatheters are also composed of a stainless steel shaft that supports placement of the guidewire beyond the chronic total occlusion.
The reference device, the cleared NovaCross™ MicroCatheter, and the NovaCross™ CTO have the same operating principles. Both use a Nitinol element (the superelasticity of the Nitinol is used), as the material to provide support to the guidewire and both include the same delivery system although the NovaCross™ CTO includes minor design modifications compared to the NovaCross™ device.
Questions related to clinical performance have been evaluated for the NovaCross™ CTO Microcatheter through extensive design verification and validation testing. including an IDE Clinical Study.
{5}------------------------------------------------
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
### Biocompatibility testing
An evaluation of biocompatibility for the NovaCross™ CTO microcatheter was conducted in accordance with FDA guidance, Use of International Standard ISO 10993-1. Testing included cytotoxicity, sensitization, intracutaneous study, pyrogenicity, SC5b-9 complement activation assay and hemolysis. All materials and processes of the NovaCross™ CTO device were shown to be biocompatible.
## Sterilization, Packaging and Shelf Life Testing
The NovaCross™ CTO Microcatheter will be sterilized by EO gas. The sterilization cycle has been validated in compliance with ISO 11135:2014. In addition, shelf life and packaging testing were performed to support the labeled shelf life. All tests were successfully completed.
### Mechanical Testing
Mechanical bench testing that has been performed on the NovaCross™ Microcatheter included the following:
- Dimensional .
- Simulated use ●
- . Tip flexibility
- . Torque response
- Tensile strength .
- . Fluid leak
- Air leakage into hub assembly during aspiration ●
- Pushability and retractability
- . Scaffold durability
All tests met the predefined acceptance criteria and demonstrate substantial equivalence to the predicate NovaCross device and support a one year shelf life for the NovaCross™ Microcatheter.
Particulate and lubricity and corrosion testing were leveraged from the predicate NovaCross™ Microcatheter.
## VIII. CLINICAL DATA
A multicenter, non-randomized, open label, single arm pivotal study was conducted to assess the safety and efficacy of the NovaCross™ CTO Microcatheter device in subjects undergoing Percutaneous Cardiac Intervention for Chronic Total Occlusion. The study was performed in two cohorts, and included 186 subjects that were recruited in 10 investigational sites.
## Primary Safety Endpoint
The primary safety endpoint of the study was the rate of post-procedure MI rate, defined as > X3 upper normal limit of CK-MB cardiac biomarker. the overall MACE rate observed in the combined analysis of cohort A and cohort B was 12.3% for the ITT population, with an upper confidence interval of 19.41% (97.5% Cl). The following tables summarize the MACE rate according to each definition and for the ITT population for both device- related and not related events.
{6}------------------------------------------------
| Category | Statistics | Cohort A<br>(N=145) | Cohort B<br>(N=41) | Cohort A + B<br>(N=186) | 97.5% Upper CI1 | P-Value1 |
|---------------------------|------------|---------------------|--------------------|-------------------------|-----------------|----------|
| MACE | % (n/N) | 14.0% (16/114) | 7.3% (3/41) | 12.3% (19/155) | 19.41% | 0.0020 |
| Death | N | 0 | 0 | 0 | | |
| Protocol definition of MI | N | 16 | 3 | 19 | | |
| Urgent revascularization | N | 0 | 0 | 0 | | |
### MACE events composite for ITT population
MACE defined as the composite of death, protocol definition of MI (CK-MB>3X ULN), or urgent revascularization (target vessel revascularization (TVR) or urgent coronary artery bypass surgery (CABG)).
## Primary Effectiveness Analysis
Primary effectiveness endpoint was set as a technical success, which was defined as the ability of the NovaCross™ CTO Microcatheter to successfully facilitate placement of a guidewire beyond a native coronary chronic total occlusion (CTO) in the true vessel lumen. The results of the study showed a technical success of 75.3%.
### Summary of Primary Effectiveness Endpoint for ITT Population:
| Category | Statistics | Cohort A<br>(N=145) | Cohort B<br>(N=41) | Cohort A + B<br>(N=186) |
|-------------------|------------|---------------------|--------------------|-------------------------|
| Technical success | % (n/N) | 75.9% (110/145) | 73.2% (30/41) | 75.3% (140/186) |
The results of this pivotal study show that the device has met both the primary safety endpoint and the primary effectiveness endpoint. It also met the secondary endpoints, associated with both the effectiveness and safety of the device. It was therefore concluded that the device is safe, effective and easily used and handled by interventional cardiologists.
## IX. CONCLUSIONS
The NovaCross™ CTO Microcatheter was determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.