K112124 · Navilyst Medical, Inc. · DQO · Aug 19, 2011 · Cardiovascular
Device Facts
Record ID
K112124
Device Name
MICROCATHETER
Applicant
Navilyst Medical, Inc.
Product Code
DQO · Cardiovascular
Decision Date
Aug 19, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The NMI Microcatheter is intended for use in small vessels and superselective anatomy during diagnostic and interventional procedures in the peripheral vasculature. Upon access of the desired region, the microcatheter can be used for the controlled and selective infusion of diagnostic, therapeutic and embolic materials. All products must be used in accordance with the original manufacturer's instructions for use. The catheter is not intended for use in coronary or neuro- vasculature.
Device Story
NMI Microcatheter is a diagnostic/interventional catheter for peripheral vasculature; used in small vessels and superselective anatomy. Device consists of a shaft, proximal hub, and radiopaque distal marker; provided with shaping mandrel, rotating hemostatic valve, guidewire, insertion tool, and torque device. Operates via fluoroscopic visualization to guide placement; allows controlled, selective infusion of diagnostic, therapeutic, or embolic materials. Used by physicians in clinical settings. Benefits include precise delivery of materials to target peripheral sites.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Radiopaque shaft with hydrophilic coating on distal end; 3.0F/2.8F or 3.0F/2.4F configurations. Includes shaping mandrel, rotating hemostatic valve, guidewire, insertion tool, and torque device. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for use in small vessels and superselective anatomy during diagnostic and interventional procedures in the peripheral vasculature for infusion of diagnostic, therapeutic, and embolic materials. Contraindicated for use in coronary or neuro-vasculature.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
Boston Scientific Renegade Hi-Flo Microcatheter (K000177)
Boston Scientific Renegade STC Microcatheter (K023861)
Related Devices
K024122 — NAUTICA MICRO CATHETER, MODEL 105-5094-153 · Micro Therapeutics, Inc. · Mar 7, 2003
K113260 — MODIFIED CONCENTIRC MICROCATHETER · Concentric Medical, Inc. · Mar 9, 2012
{0}------------------------------------------------
K112124 page 1 of 2
# AUG 1 9 2011
# 510(k) SUMMARY
Date Prepared: August 17, 2011
#### A. Sponsor
Navilyst Medical, Inc 26 Forest Street Marlborough, MA 01752
#### B. Contact
Wanda Carpinella Sr. Manager Global Regulatory Affairs 508-658-7929
#### C. Device Name
Trade Name: Common/Usual name: Classification Name:
Classification
#### D. Predicate Device(s)
Trade Name/510(k) Notification:
Vice President, Global Regulatory Affairs 508-658-7945
### NMI MC
Lorraine M. Hanley
Microcatheter Diagnostic Intravascular Catheter 21 CRF §870.1200, ProCode DQO Class II
Boston Scientific Renegade Hi-Flo Microcatheter
K000177 Boston Scientific Renegade STC Microcatheter
### E. Device Description
The proposed MC is offered in two French size configurations. The first configuration has an outside diameter of 3.0F/2.8F (tapers proximally to distally) and has a nominal inside diameter of 0.027" along the entire catheter length. The second configuration has an outside diameter of 3.0F/2.4F (tapers proximally to distally) and has a nominal insider diameter of 0.021" along the entire catheter length.
K023861
Both MC configurations consist of a shaft and proximal hub assembly. The radiopaque shaft transitions from a stiffer proximal region to a flexible 30cm distal end. A radiopaque marker is located 1mm from the catheter's distal end to identify the location of the distal tip under fluoroscopic visualization. The distal end of the catheter's outer surface is hydrophilic coated. The microcatheter is provided with a shaping mandrel and a rotating hemostatic valve. The microcatheter kit additionally includes a guidewire, a guidewire insertion tool and a torque device.
### F. Intended Use
The NMI Microcatheter is intended for use in small vessels and supreselective anatomy during diagnostic and interventional procedures in the peripheral vasculature. Upon access of the desired region, the microcatheter can be used for the controlled and selective infusion of diagnostic, therapeutic and embolic materials. All products delivered through the microcatheter must be used in accordance with the original manufacturer's instruction for use. The catheter is not intended for use in coronary or neuro- vasculature.
## G. Technology Characteristics
The proposed device has similar materials, design and components and technological characteristics as currently marketed microcatheters.
{1}------------------------------------------------
K1121241 page 2 of 2
# H. Performance Data
The NMI MC is substantially equivalent to the specified predicate device based on comparison of technological characteristics and the results of non-clinical tests, which included:
- Coating performance ●
- Trackability .
- Pressure testing
- Dimensional verification .
- Radiopacity .
- . Kink resistance
- Flow rate .
- Tip shapability .
- Compatibility tests �
- Particulate generation .
- Luer performance ●
- Biocompatibility per ISO 10993-1 �
#### Conclusion I.
The results of the non-clinical testing and a comparison of similarities and differences demonstrate that the proposed and predicate devices are substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The seal is black and white and appears to be a low-resolution image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Navylist Medical, Inc. c/o Ms. Wanda Carpinella 26 Forest St. Marlborough, MA 01752
AUG 1 9 2011
Re: K112124
Trade/Device Name: NMI Microcatheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II (two) Product Code: DQO Dated: 21 July 2011 Received: 25 July 2011
Dear Ms. Carpinella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
ff your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
\$\mu\$
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K 112124 page 1 of 1
# Indications for Use
510(k) Number (if known): K / / 2 12 4
Device Name: Microcatheter
Indication For Use:
The NMI Microcatheter is intended for use in small vessels and superselective anatomy during diagnostic and interventional procedures in the peripheral vasculature. Upon access of the desired region, the microcatheter can be used for the controlled and selective infusion of diagnostic, therapeutic and embolic materials. All products must be used in accordance with the original manufacturer's instructions for use. The catheter is not intended for use in coronary or neuro- vasculature.
| Prescription Use<br>(21 CFR Part 801 Subpart D) | X |
|-----------------------------------------------------|---|
| And/Or | |
| Over the Counter Use<br>(21 CFR Part 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K112124 |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.