SoundStar eco 8F 3D Diagnostic eco Ultrasound Catheters, SoundStar eco 8F G 3D Diagnostic eco Ultrasound Catheters

K161700 · Sterilmed, Inc. · OWQ · Nov 21, 2016 · Cardiovascular

Device Facts

Record IDK161700
Device NameSoundStar eco 8F 3D Diagnostic eco Ultrasound Catheters, SoundStar eco 8F G 3D Diagnostic eco Ultrasound Catheters
ApplicantSterilmed, Inc.
Product CodeOWQ · Cardiovascular
Decision DateNov 21, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1200
Device ClassClass 2

Indications for Use

The Reprocessed 3D Diagnostic Ultrasound eco Catheters are indicated for intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible Carto® 3 EP Navigation Systems (version 2.3 and higher) the reprocessed 3D Diagnostic Ultrasound eco Catheters provide location information.

Device Story

Reprocessed 3D Diagnostic Ultrasound eco Catheters are intra-cardiac echo (ICE) catheters; feature embedded ultrasound transducer and magnetic location sensor in catheter tip. Input: acoustic signals for real-time ultrasound imaging; magnetic field data for 3D navigation. Operation: connects to compatible ultrasound systems and Carto® 3 EP Navigation System; provides real-time integration of ultrasound images with electromagnetic maps. Used in clinical settings by physicians; assists in visualizing cardiac anatomy/physiology and other devices during procedures. Benefits: enables precise navigation and visualization within heart/great vessels. Reprocessed by Sterilmed, Inc. up to 3 times; includes cleaning, sterilization, and functional testing (joint strength, deflection, sensor calibration). Output: visual images and location data displayed on navigation system console; informs clinical decision-making during electrophysiology procedures.

Clinical Evidence

No clinical data. Bench testing only. Performance validated via cleaning/sterilization validation, biocompatibility (ISO 10993-1), packaging/shelf-life validation, and mechanical/functional testing (joint bond strength, torsional strength, flexation/deflection fatigue, tip buckling, sensor calibration, EEPROM reset, acoustic verification, and thrombogenicity). Electrical safety verified per IEC 60601-1 (defibrillation, leakage current) and IEC 61161 (acoustic power).

Technological Characteristics

8F, 90 cm diagnostic ultrasound catheter. Features integrated ultrasound transducer and magnetic location sensor. Materials identical to OEM. Connectivity: compatible with Carto® 3 EP Navigation System and specific Siemens/GE ultrasound systems. Sterilization: Ethylene Oxide (ISO 11135). Standards: ISO 10993 (biocompatibility), ASTM D4169/F88/F2096 (packaging), ASTM 1980-99 (shelf life), IEC 60601-1 (electrical safety), IEC 61161 (acoustic power).

Indications for Use

Indicated for intra-luminal visualization of cardiac and great vessel anatomy and physiology, and visualization of other devices in the heart. Compatible with Carto® 3 EP Navigation Systems (v2.3+) for location information.

Regulatory Classification

Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Public Health Service November 21, 2016 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Sterilmed. Inc. Neelu Gibson Senior Director, Regulatory Affairs 5010 Cheshire Parkway, Suite 2 Plymouth, Minnesota 55446 Re: K161700 Trade/Device Name: Reprocessed 3D Diagnostic Ultrasound eco Catheters (See attached list of models) Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OWO Dated: October 7, 2016 Received: October 11, 2016 Dear Neelu Gibson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Mucha Jellman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # K161700 ## Models intended for reprocessing | Original<br>Manufacturer | OEM<br>Model<br>Number | Reprocessed<br>Product Code | Description | |---------------------------|------------------------|-----------------------------|--------------------------------------------------------------------------------------------------------------------| | Biosense Webster,<br>Inc. | 10439011 | R10439011 | SoundStar® eco Diagnostic<br>Ultrasound Catheter, 8F, 90 cm<br>length. For use with Siemens<br>ultrasound systems. | | Biosense Webster,<br>Inc. | 10439236 | R10439236 | SoundStar® eco Diagnostic<br>Ultrasound Catheter, 8F, 90 cm<br>length. For use with GE ultrasound<br>systems. | {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161700 Device Name Reprocessed 3D Diagnostic Ultrasound eco Catheters Indications for Use (Describe) The Reprocessed 3D Diagnostic Ultrasound eco Catheters are indicated for intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible Carto® 3 EP Navigation Systems (version 2.3 and higher) the reprocessed 3D Diagnostic Ultrasound eco Catheters provide location information. Type of Use (Select one or both, as applicable) | <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Sterilmed. The logo consists of a stylized, partially-formed circle on the left, with the word "STERILMED" in all caps to the right of the circle. The circle is silver on the top and green on the bottom. The word "STERILMED" is in a simple, sans-serif font. | Submitter and Manufacturer: | Neelu Gibson<br>Sterilmed, Inc.<br>5010 Cheshire Parkway, Suite 2<br>Plymouth, MN 55446 | |---------------------------------|----------------------------------------------------------------------------------------------------------| | Manufacturing Facility Address: | 11400 73rd Avenue North<br>Maple Grove, MN 55369 | | Primary Contact: | Neelu Gibson<br>Sterilmed, Inc.<br>Tel: 908-705-3160<br>Fax: 763-488-3350<br>Email: ngibson9@its.jnj.com | | Date of Submission: | 21 May, 2016 | | Trade Name: | Reprocessed 3D Diagnostic Ultrasound<br>eco Catheters | | Regulation Name: | Diagnostic Intravascular Catheter | | Device Classification: | Class II | | Product Code: | OWQ | # 510(k) Summary - K161700 | Predicate<br>Device: | Biosense Webster, Inc.: SoundStar eco 8F G Ultrasound Catheter, and SoundStar<br>eco 8F Ultrasound Catheter (510(k) K140318) | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Secondary /<br>Reference<br>Predicates | Sterilmed Inc: Reprocessed 3D Ultrasound Diagnostic Catheter (510(k) K110076)<br>Sterilmed Inc: Reprocessed Ivus Imaging Catheter (510(k) K043453) | | Device<br>Description: | Sterilmed Reprocessed 3D Diagnostic Ultrasound eco Catheters are specially<br>designed catheters that provide two-dimensional imaging using an ultrasound<br>transducer and three-dimensional location sensor. | | | They are IntraCardiac Echo (ICE) Catheters with acoustic array and magnetic<br>location sensor equivalent to the currently cleared SoundStar eco 8F G Ultrasound<br>Catheter, and SoundStar eco 8F Ultrasound Catheters 510(k) K140318. The<br>magnetic location sensor (providing location information to the CARTO® 3 EP<br>Navigation System) and an ultrasound transducer (acquiring real time ultrasound<br>images) are embedded in the catheter tip. | | | The Reprocessed 3D Diagnostic Ultrasound Catheters, when connected to the<br>corresponding Ultrasound Systems, will provide real-time integration of the<br>ultrasound images with CARTO® 3 electromagnetic acquired maps. | | | Note: Only the catheter is the subject of this submission. Accessories and/or any<br>other related equipment are not included in the scope of this submission. The table<br>below lists the specific model numbers that are the subject of this submission. | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for STERILMED. The logo features a stylized, partially circular graphic on the left, with the word "STERILMED" in a sans-serif font to the right of the graphic. The graphic is two-toned, with the upper portion in silver and the lower portion in green. | Original<br>Manufacturer | OEM<br>Model<br>Number | Reprocessed<br>Product<br>Code | Description | |---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|--------------------------------------------------------------------------------------------------------------------| | Biosense Webster,<br>Inc. | 10439011 | R10439011 | SoundStar® eco Diagnostic<br>Ultrasound Catheter, 8F, 90 cm<br>length. For use with Siemens<br>ultrasound systems. | | Biosense Webster,<br>Inc. | 10439236 | R10439236 | SoundStar® eco Diagnostic<br>Ultrasound Catheter, 8F, 90 cm<br>length. For use with GE ultrasound<br>systems. | | Indications for<br>Use: | The Reprocessed 3D Diagnostic Ultrasound eco Catheters are indicated for intra-<br>cardiac and intra-luminal visualization of cardiac and great vessel anatomy and<br>physiology as well as visualization of other devices in the heart. When used with<br>compatible Carto® 3 EP Navigation Systems (version 2.3 and higher) the<br>reprocessed 3D Diagnostic Ultrasound eco Catheters provide location information. | | | | Technological<br>Characteristics: | The Reprocessed 3D Diagnostic Ultrasound eco Catheters are identical to the<br>predicate devices in design, materials of construction, and intended use. There are<br>no changes to the clinical applications, patient population, performance<br>specifications, or method of operation. | | | | Functional and<br>Safety Testing: | Representative samples of reprocessed devices were tested to demonstrate<br>appropriate functional characteristics.<br>Process validation testing was performed to validate the cleaning and sterilization<br>procedures as well as device packaging. In addition, the manufacturing process<br>includes visual and validated functional testing of all products produced.<br>Where appropriate prior validation testing of previously cleared Reprocessed 3D<br>Diagnostic Ultrasound eco Catheter K110076 and Reprocessed AcuNav Imaging<br>Catheter K043453 was utilized for the same reprocessing procedures, equipment,<br>sterile barrier and sterilization processing. The Reprocessed 3D Diagnostic<br>Ultrasound eco Catheters are reprocessed no more than three (3) times. Each device<br>is marked and tracked through each reprocessing cycle. After the device has<br>reached the maximum number of reprocessing cycles (3), the device is rejected<br>from further reprocessing. Reprocessing is performed only by the manufacturer<br>Sterilmed, Inc. | | | | Summary of<br>Non-Clinical<br>Tests<br>Conducted: | Specific non-clinical tests performed included: cleaning validation, sterilization<br>validation (ISO 11135, USP <71>), biocompatibility testing (ISO 10993-1),<br>ethylene oxide residual testing (ISO 10993-7), packaging validation (ASTM<br>D4169, ASTM F88, ASTM F2096), and shelf life validation (ASTM 1980-99). In<br>addition, functional performance (including both the ultrasound and 3D location<br>sensor components) and mechanical reliability were validated using bench and<br>laboratory testing. | | | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for STERILMED. The logo features a stylized, circular graphic on the left, with the word "STERILMED" in a simple, sans-serif font to the right of the graphic. The graphic is two-toned, with the top portion in silver and the bottom portion in green. | | Catheters perform as originally intended. | |-------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Conclusion: | Sterilmed concludes that the Reprocessed 3D Diagnostic Ultrasound eco<br>Catheters are safe, effective, and substantially equivalent to the predicate<br>devices, Biosense Webster, Inc. SoundStar eco 8F and eco 8F G Ultrasound<br>Catheters (K140318), as described in this premarket notification submission.<br>The Reprocessed 3D Diagnostic Ultrasound eco Catheters are substantially<br>equivalent to the listed predicate devices with respect to their indications for<br>use (intended use) and technical characteristics. The information and data<br>provided in this 510(k) submission identify no new safety or effectiveness<br>issues. | | Verification<br>Testing | Description | Pass/<br>Fail | |------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------| | Joint Bond<br>Strength<br>(SM5653/5840) | Test measured tensile strength of device joints to<br>confirm reprocessing has no effect on device. | Pass | | Torsional Strength<br>(SM5652/5842) | Test measured the torsional strength of the flexible tip<br>and shaft sections to confirm reprocessing has no effect<br>on device. | Pass | | Flexation Fatigue<br>(SM5604/5827) | Test measured flexation through maximum range of<br>deflection 100 times to confirm that reprocessing has no<br>effect on device. | Pass | | Deflection Fatigue<br>(SM5605/5823) | Testing measured deflection in a simulated use<br>model through full articulation multiple times to<br>confirm that reprocessing has no effect on device. | Pass | | Tip buckling<br>(SM4091/5923) | Testing measured force required to cause tip<br>buckling, to ensure reprocessing has effect on<br>device. | Pass | | CCS check<br>(SM5523/5822) | Calibration check testing confirmed that<br>reprocessing has no effect on device's ability to<br>maintain sensor location. | Pass | | Auto ID<br>(SM5800/5830) | Testing confirmed that reprocessed devices continue<br>to be identified as Soundstar eco 8F catheters | Pass | | EEPROM reset<br>verification<br>(SM5802/5841) | Testing confirmed that EEPROM can be effectively<br>reset and continue to function as intended on<br>reprocessed devices. | Pass | | AMCS Acoustic<br>Verification<br>(SM5522/5828) | Testing confirmed that Acoustic Manual Calibration<br>System functions for reprocessed devices. | Pass | | Thrombogenicity<br>testing<br>SM6020/6828 | Testing confirmed that reprocessed devices do not<br>cause any new risk of thrombus formation. | Pass | # Summary of Performance Testing {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for STERILMED. The logo features a stylized, circular graphic on the left, with the word "STERILMED" in a combination of bold and light gray text to the right. The circular graphic appears to be a stylized recycling symbol, with the top half in silver and the bottom half in green. #### Biocompatibility testing . Biocompatibility data demonstrates that there is no risk presented by Reprocessed SoundStar eco 8F and eco 8F G Ultrasound Catheters. - Electromagnetic compatibility (EMC) testing was performed by Sterilmed per IEC 60601-1:2005 AI:2012 3nd Edition Medical electrical equipment - Part 1 General requirements for basic safety and essential performance were the defibrillation, leakage current and functional testing. ### Defibrillation Test report SM 5839 documents the defibrillation protection compliance to the IEC 60601-1:2005 Clause 8.5.5.1 for defibrillation protection testing. IEC 60601-1:2005 Clause 8.5.5.1 for defibrillation protection testing defines a peak value of the differential signal (Yl -Y2) to be less than 1 volt (passing the test). Voltage reading over 1 volt is considered a fail. ### Current Leakage and Functional Test To mitigate the concerns of possible damage to the device due to reprocessing the devices are: - 100% in-process inspected for electrical insulation via Hi-pot (high potential) . testing, in which each device is subjected to 1500VDC, and electrical leakage is monitored. Reference test report SM5831 - 100% in-process inspected for functionality. ● ### IEC 60601-1-2:2007 No physical changes or parts replacement is conducted by Sterilmed during the reprocessing. The devices are unchanged and are identical to the OEM devices as evidenced in functional tests SM5828 - AMCS Check - SoundStar eco 8F within Error! Reference source not found.. ### IEC 61161:2013 Total emitted acoustic power of ultrasonic transducers was verified and documented in SM5826. ### Summary The Reprocessed 8F Catheters and their predicate devices, the OEM 8F Catheters, have the same intended use and indications for use and fundamental scientific technology. Based on performance testing for the Reprocessed 8F Catheters against the OEM devices discussed above, Sterilmed concludes that the Reprocessed 8F catheters are safe, effective, and substantially equivalent to the predicate OEM 8F catheters.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...