The Biosense Webster SOUNDSTAR® eco Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO® 3 EP Navigation Systems, the SOUNDSTAR® eco Catheter provides location information. Please refer to the Compatibility Matrix Insert for Compatible CARTO® 3 Systems as each catheter is compatible with a specific version of CARTO® 3 and is not backwards compatible with previous versions of CARTO® 3 EP Navigation Systems.
Device Story
The SOUNDSTAR eco 8F/8F G are 90cm, 8F diagnostic ultrasound catheters featuring an integrated acoustic phased array and electromagnetic (EM) location sensor. Used in clinical electrophysiology settings, the device connects to compatible ultrasound systems (Siemens/GE) and the CARTO 3 EP Navigation System. The EM sensor, containing three orthogonally assembled air-core coils, provides real-time location data for electro-anatomical mapping. The ultrasound transducer acquires real-time images. The device enables clinicians to visualize cardiac anatomy and device placement during procedures. Integration of ultrasound images with EM-acquired maps assists in navigation and clinical decision-making. The 8F design, featuring a smaller sensor than the 10F predicate, allows for use within the Stereotaxis NIOBE Remote Magnetic System. The device is single-use and sterilized via EtO.
Clinical Evidence
Bench testing only. Design verification testing, safety, and electromagnetic compliance testing were performed to validate the modified catheter design and compatibility with specified ultrasound systems.
Technological Characteristics
8F diameter, 90cm length, 64-channel acoustic phased array. EM location sensor uses three orthogonally assembled air-core coils. Connectivity via SWIFTLINK cable to ultrasound systems and CARTO 3 EP Navigation System. Sterilization: EtO. Single-use.
Indications for Use
Indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy, physiology, and other devices in the heart.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
SOUNDSTAR eco 10F (P/N:M-5276-17)
SOUNDSTAR eco 10FG (P/N:M-5276-17)
AcuNav 8F (P/N M-5723-09)
AcuNav 8FG (P/N M-5723-07)
Submission Summary (Full Text)
{0}------------------------------------------------
#### K140318 page 1 of 5
## Special 510K for SOUNDSTAR® eco 8F and SOUNDSTAR® eco 8F G Catheters
# 14 510(k) Summary
SPONSOR'S NAME & ADDRESS
Biosense Webster, Inc. 3333 Diamond Canyon Road Diamond Bar, CA 91765
# OFFICIAL CORRESPONDENT
Marina Guevrekian Senior Specialist, Regulatory Affairs Phone: 909-839-8714 Fax: 909-839-8804 Email: mguevrek@its.jnj.com
## SUBMISSION DATE
February 6, 2014
# TRADE NAME
# SOUNDSTAR® eco DIAGNOSTIC ULTRASOUND CATHETER
(Part Number: M-5723-17 / Catalogue Number: 10439011, Compatible with Siemens Ultrasound systems)
# SOUNDSTAR® eco DIAGNOSTIC ULTRASOUND CATHETER
(Part Number: M-5723-18 / Catalogue Number: 10439236, Compatible with GE Ultrasound systems)
#### Common Name
Electrophysiology Mapping/Ultrasound Catheter
# CLASSIFICATION NAME/PRODUCT CODE
Intravascular Ultrasound Catheter/OBJ
#### CLASSIFICATION
Class II, 21 CFR 870.1200
#### PREDICATE DEVICE
The modified SOUNDSTAR eco8F and SOUNDSTAR ecoF G Catheters are substantially equivalent to:
- SOUNDSTAR eco 10F P/N:M-5276-17 / Catalogue number 10439011 and SOUNDSTAR . eco 10FG P/N:M-5276-17 / Catalogue number 10439011.
page 79
{1}------------------------------------------------
# Special 510K for SOUNDSTAR® eco 8F and SOUNDSTAR® eco 8F G Catheters
- · AcuNav 8F P/N M-5723-09 / Catalogue number 10135936 and AcuNav 8FG P/N M-5723-07 / Catalogue number 10135910.
# DESCRIPTION OF MODIFIED DEVICE
The modified SOUNDSTAR eco 8F and SOUNDSTAR eco 8F G Catheters are both 90 cm long and 8F in diameter. They are IntraCardiac Echo (ICE) Catheters with acoustic array and magnetic location sensor equivalent to the currently cleared SQUNDSTAR eco 10F and SOUNDSTAR eco 10FG Catheters. The magnetic location sensor (providing location information to the CARTO 3 EP Navigation System Version 2.3 and 3.2) and an ultrasound transducer (acquiring real time ultrasound images) are embedded in the catheter tip.
The modified SOUNDSTAR eco 8F and SOUNDSTAR eco 8F G Catheters have a bifurcated 'tail' originating from the catheter handle which is identical to the bifurcated tail of the predicate device; the SOUNDSTAR eco 10F. One leg terminates in the SOUNDSTAR Flex Tab connector, which connects via the appropriate SWIFTLINK cable, to the appropriate Ultrasound system. Specifically, for the SOUNDSTAR eco Catheter 8F. the SWIFTLINK cable connects to the Acuson Cypress™, the Acuson Sequoia™ or the Acuson X300™ and Acuson SC2000™ Ultrasound systems. For the SOUNDSTAR eco 8F G Catheter, the SWIFTLINK cable connects to Vivid i and Vivid q Ultrasound systems.
Both versions of SOUNDSTAR eco 8F and SOUNDSTAR eco 8F G Catheters are based on the existing SOUNDSTAR eco 10F and SOUNDSTAR eco 10FG catheters and have the same intended use and clinical applications. The predicate and modified catheters share the majority of components and manufacturing process as explained previously.
The modified SOUNDSTAR eco 8F and SOUNDSTAR eco 8F G Catheters, when connected to the corresponding Ultrasound Systems, will provide real-time integration of ultrasound images with CARTO 3 Version 2.3 and Version 3.2 electromagnetic acquired maps.
## INDICATIONS FOR USE
The Biosense Webster SOUNDSTAR® eco Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO® 3 EP Navigation Systems, the SOUNDSTAR® eco Catheter provides location information.
Please refer to the Compatibility Matrix Insert for Compatible CARTO® 3 Systems as each catheter is compatible with a specific version of CARTO® 3 and is not backwards compatible with previous versions of CARTO® 3 EP Navigation Systems.
{2}------------------------------------------------
# DESCRIPTION OF MODIFICATION
The modifications to the predicate devices are:
] . A smaller diameter for the catheter body shaft. Consequently a smaller French size. Therefore change from 10 French to 8 French.
2. Change in the sensor design
The magnetic sensor in the predicate devices; SOUNDSTAR® eco 10F, SOUNDSTAR® eco 10FG catheters had a diameter of 1.3mm. The sensor used in the modified devices is a smaller sensor; magnetic sensor of 1.0mm, to allow the new sensor to fit in the smaller catheter body. In both catheters the Electromagnetic (EM) sensor enables accurate magnetic navigation for electro-anatomical mapping.
SOUNDSTAR eco 10F and SOUNDSTAR eco 10F G catheters sensor contains 3 coils in three X, Y, and Z axis. The coils have Ferrite core inside them and are assembled orthogonally with each other.
The coils have been repositioned for SOUNDSTAR® eco 8F catheter in comparison to the SOUNDSTAR® eco 10F catheters. Basically, the ferrites inside the coils of the SOUNDSTAR eco 8F and SOUNDSTAR eco 8F G catheters sensor have been replaced by air cores and assembled orthogonally.
The replacement of the ferrites by air cores in the sensor will allow the catheters to be used in the Stereotaxis Niobe® environment.
The main differences between the sensors are in the design and manufacturing process, while the end product is equivalent in functionality and performance
3. The SOUNDSTAR® eco 8F Catheters will be able to generate ultrasound images within the Stereotaxis NIOBE® Remote Magnetic System without impacting the location or image quality. For the catheter functionality, as mentioned above, the coils located inside of the sensor have been repositioned and the existing ferrites have been removed and replaced by air cores to allow the catheter to be used in the Sterotaxis Niobe® environment. The sensor material remains the same.
## SUMMARY OF NONCLINICAL TESTS
To support the design change of the catheter, Design Verification Testing was performed on the modified SOUNDSTAR eco8F and SOUNDSTAR eco 8F G Catheters.
Safety and Electromagnetic Compliance testing was also performed to support the design changes on the predicate device. The Safety and Electromagnetic compliance testing validates the use of the modified devices; the SOUNDSTAR eco 8F and SOUNDSTAR eco 8F G Catheters with all the validated Siemens and GE Ultrasound systems.
Image /page/2/Picture/14 description: This image shows the text 'page 81' vertically oriented. The word 'page' is stacked on top of the number '81'. The text is in a simple, sans-serif font and appears to be part of a page number within a larger document.
{3}------------------------------------------------
# Special 510K for SOUNDSTAR® eco 8F and SOUNDSTAR® eco 8F G Catheters
The modified catheters are compatible with SOUNDSTAR eco Extension Cable and the CARTO 3 version 2.3 and 3.2 EP Navigation system. This compatibility has been documented by a Letter to file, filed on December 10, 2013.
# SUBSTANTIAL EQUIVALENCE CONCLUSION
# SIMILARITIES TO THE PREDICATE DEVICES .
a) The modified SOUNDSTAR eco 8F and SOUNDSTAR eco8F G Catheters are identical to the currently cleared SOUNDSTAR eco 10F and SOUNDSTAR eco 10FG Catheters in the following aspects:
- Have the same intended use .
- . Use the same operating principle
- Are manufactured in the same manufacturing facility .
- Use the same fundamental scientific technology .
- Incorporate the same materials and construction .
- Have the same deflection mechanism .
- The magnetic sensor has the same material and location .
- Have the same Ultrasound connections (to both Siemens and GE . Ultrasound systems)
- Catheter body & strain relief have identical assembly process .
- Have the same 64-channel acoustic phased array .
- Have the same acoustic array location and connection .
- Have the same packaging and are packaged using the same materials and . processes
- Have the same Electronics; the PCB is inside the Hypertronic connector . and Extension Cable
- Identical magnetic sensor location (not connected to the Printed Circuit . Board (PCB))
- Identical handle, magnetic location sensor wires and connections .
- Have the same handle material .
- Have the same Steering mechanism .
- Have identical Hypertronic connector .
- Most of the manufacturing processes are identical .
- Have the same shelf life .
- Use the same sterilization method (EtO sterilization) .
- . Are single use devices
b) The modified SOUNDSTAR eco 8F and SOUNDSTAR eco 8F G Catheters are identical to the currently cleared AcuNav 8F and AcuNav 8FG catheters in the following aspects
- . Have the same shaft material, color and process with a 90cm useable length
{4}------------------------------------------------
- . The product functional specification (hi pot, acoustic, articulation) and the DVT criteria is identical
- . Have the same strain relief material. composition and mechanical design
# DIFFERENCES FROM THE PREDICATE DEVICES
a) The modified SOUNDSTAR eco8F and SOUNDSTAR eco8FG Catheters are different from the currently cleared SNDSTR10 and SNDSTR10G Catheters in the following aspects:
- · Have different Tip dimensions
- · Have a smaller French size for the catheter body shaft. Change from 10French to 8French diameter
- · A smaller sensor to fit in a smaller diameter of shaft.
- · The SOUNDSTAR eco 8F catheters will be able to generate ultrasound images within the Stereotaxis NIOBE® Remote Magnetic System, without impacting the location or image quality.
- The labels, IFUs and Matrices are slightly modified .
- The handle color is blue .
c) The modified SOUNDSTAR eco 8F and SOUNDSTAR eco 8F G Catheters are different to the currently cleared AcuNav 8F and AcuNav 8FG catheters in the following aspects
- . The last .5 cm of the shaft distal end out of 90 cm of the entire shaft length has additional internal grove to accommodate the sensor. The product functional specifications ( hi pot, acoustic, articulation, length, OD, ID, Shaft Material) and the DVT criteria do not change despite of this change.
- 트 The strain relief color is black
- . Has the new Heat Shrink Tube shown at figure 5 on page 19 as the item F.
page
83
{5}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60S Silver Spring, MD 20993-0002
Image /page/5/Picture/3 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized symbol that resembles an abstract human figure or a caduceus-like design.
March 12, 2014
Biosense Webster Marina Guevrekian Senior Specialist Regulatory Affairs 3333 Diamond Canyon Rd. Diamond Bar, CA 91765 US
K140318 Re:
Trade/Device Name: Soundstar ECO 8FG Ultrasound Catheter, Soundstar ECO 8F Ultrasound Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Intravascular Ultrasound Catheter Regulatory Class: II Product Code: OBJ Dated: February 6, 2014 Received: February 10, 2014
Dear Marina Guevrekian,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems
{6}------------------------------------------------
Page 2 - Marina Guevrekian
(QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDR/'s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDeviccs/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Linda J. Ricci - S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# K140318 Special 510K for SOUNDSTAR® eco 8F and SOUNDSTAR® eco 8F G Catheters
#### 6 INDICATIONS FOR USE
510(k) No (if known): TBI)
# Device Name:
SOUNDSTAR eco 8F Catheter SOUNDSTAR eco 8F G Catheter
# Indications for Use:
The Biosense Webster SOUNDSTAR® eco Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO" 3 EP Navigation Systems, the SOUNDSTAR" eco Catheter provides location information.
Please refer to the Compatibility Matrix Insert for Compatible CARTO® 3 Systems as each catheter is compatible with a specific version of CARTO® 3 and is not backwards compatible with previous versions of CARTO® 3 EP Navigation Systems.
Prescription Use ---------イ — AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
の日、2008年には、2005年には、2007年には、2006年には、2006年には、2008年の100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
Concurrence of CDRH, Office of Device Evaluation (ODE)
Linda J. Ricci - S
Page 1 of 1
Page 5 (
BIOSENSE WEBSTER, INC.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.