The ePM 10/ePM 12/ePM 15/ePM 10M/ePM 15M patient monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Carbon Dioxide (CO2), Oxygen (O2) and Anesthetic Gas (AG). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: · The PAWP monitoring is intended for adult and pediatric patients only; · C.O. monitoring is intended for adult patients only; The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The ePM 12/ePM 15/ePM 10M/ePM 12M/ePM 15M monitors are not intended for helicopter transport, hospital ambulance, or home use.
Device Story
Multi-parameter patient monitors for healthcare facilities; inputs include ECG, respiration, temperature, SpO2, pulse rate, NIBP, IBP, PAWP, cardiac output, CO2, O2, and anesthetic gas. Device processes physiological signals to display waveforms, numerics, and alarms; provides arrhythmia detection, ST-segment analysis, QT analysis, and resting 12-lead ECG interpretation. Operated by trained clinical professionals; integrates with central monitoring systems via wireless network. Output used for real-time patient status assessment and clinical decision-making; supports early warning scores (MEWS, NEWS, NEWS2) and Glasgow Coma Scale. Benefits include continuous patient surveillance and timely detection of physiological deterioration.
Clinical Evidence
Bench testing only. Validation included functional and system-level testing to verify performance against specifications and consensus standards (IEC 60601-2-27, AAMI/ANSI EC57). No clinical data provided.
Indicated for monitoring, displaying, reviewing, storing, alarming, and transferring physiological parameters (ECG, Resp, Temp, SpO2, PR, NIBP, IBP, PAWP, C.O., CO2, O2, AG) and resting 12-lead ECG interpretation in adult, pediatric, and neonatal patients. PAWP limited to adult/pediatric; C.O. limited to adult. For use in healthcare facilities by trained clinical professionals.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
June 8, 2020
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Yanhong Bai Manager Regulatory Affairs, Technical Regulation Department Mindray Building, Keji 12th Road South Hi-tech Industrial Park, Nanshan Shenzhen, 518057 China
Re: K200015
Trade/Device Name: ePM Series Patient Monitors Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DQA, DPZ, CCK, DXG, DSJ, CBQ, CBS, CBR. CCL Dated: May 9, 2020 Received: May 12, 2020
Dear Yanhong Bai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200015
Device Name ePM Series Patient Monitors
#### Indications for Use (Describe)
The ePM 10/ePM 12/ePM 15/ePM 10M/ePM 15M patient monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Carbon Dioxide (CO2), Oxygen (O2) and Anesthetic Gas (AG). The system also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: · The PAWP monitoring is intended for adult and pediatric patients only;
· C.O. monitoring is intended for adult patients only;
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The ePM 12/ePM 15/ePM 10M/ePM 12M/ePM 15M monitors are not intended for helicopter transport, hospital ambulance, or home use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the ePM Series Patient Monitors is provided below.
# 1. SUBMITTER
| Applicant: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81888998<br>Fax: +86 755 26582680 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Contact Person: Yanhong Bai<br>Title: Manager Regulatory Affairs<br>Phone: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared: | December 28, 2019 |
| DEVICE | |
| Device Trade Name: | ePM Series Patient Monitors |
| Device Common Name: | Patient Monitor |
| Classification Name: | 21 CFR 870.1025, Class II, Arrhythmia detector and<br>alarm (including ST-segment measurement and alarm) |
| Regulatory Class: | Class II |
| Panel: | Cardiovascular |
2.
{4}------------------------------------------------
#### Primary Product Code: MHX - Monitor, Physiological, Patient (with arrhythmia detection or alarms)
| Table 1: | Secondary Product Codes |
|----------|-------------------------|
|----------|-------------------------|
| Regulation<br>Number/Class | Product<br>Code | Regulation description | Device Common Name |
|----------------------------|-----------------|----------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| 870.1025, II | DSI | Arrhythmia detector and alarm<br>(including ST-segment measurement<br>and alarm) | Detector and alarm,<br>arrhythmia |
| 870.1025, II | MLD | Arrhythmia detector and alarm<br>(including ST-segment measurement<br>and alarm) | Monitor, st segment with<br>alarm |
| 870.2300, II | DRT | Cardiac Monitor (including<br>cardiotachometer and rate alarm) | Monitor, cardiac (incl.<br>cardiotachometer & rate<br>alarm) |
| 870.1130, II | DXN | Noninvasive blood pressure<br>measurement system | System, measurement,<br>blood-pressure, non-invasive |
| 870.1110, II | DSK | Blood pressure computer | Computer, blood-pressure |
| 880.2910, II | FLL | Clinical electronic thermometer | Thermometer, electronic,<br>clinical |
| 870.2700, II | DQA | Oximeter | Oximeter |
| 870.2710, II | DPZ | Ear oximeter | Oximeter, ear |
| 868.1400, II | CCK | Carbon dioxide gas analyzer | Analyzer, gas,<br>carbon-dioxide,<br>gaseous-phase |
| 870.1435, II | DXG | Single-function, preprogrammed<br>diagnostic computer | Computer, diagnostic,<br>pre-programmed,<br>single-function |
| 870.1100, II | DSJ | Blood pressure alarm | Alarm, blood-pressure |
| 868.1500, II | CBQ | Enflurane gas analyzer | Analyzer, gas, enflurane,<br>gaseous-phase (anesthetic<br>concentration) |
| 868.1620, II | CBS | Halothane gas analyzer | Analyzer, gas, halothane,<br>gaseous-phase (anesthetic<br>conc.) |
| 868.1700, II | CBR | Nitrous oxide gas analyzer | Analyzer, gas, nitrous-oxide,<br>gaseous phase (anesthetic<br>conc.) |
| 868.1720, II | CCL | Oxygen gas analyzer | Analyzer, gas, oxygen,<br>gaseous-phase |
{5}------------------------------------------------
# 3. PREDICATE DEVICES
K191769 - ePM Series Patient Monitors (Including ePM 10/ePM 15/ePM 15/ePM 10M/ePM 12M/ePM 15M) (SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD)
# 4. REFERENCE DEVICE
K103142 - The spacelabs Multi-parameter Module (SPACELABS HEALTHCARE, INC.)
# 5. DEVICE DESCRIPTION
The subject ePM Series Patient Monitors includes six monitors:
- ePM 10 Patient Monitor
- ePM 12 Patient Monitor
- ePM 15 Patient Monitor
- . ePM 10M Patient Monitor
- ePM 12M Patient Monitor ●
- ePM 15M Patient Monitor ●
The ePM Series Patient Monitors are Mindray's new generation monitoring product family with ergonomic and flexible design in platform of both software and hardware to meet the clinical needs of monitoring.
# 6. INDICATIONS FOR USE
The ePM 10/12/15/10M/12M/15M patient monitors are intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead . 5-lead.6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, OT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Carbon Dioxide (CO2), Oxygen (O2) and Anesthetic Gas (AG). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
- The PAWP monitoring is intended for adult and pediatric patients only; ●
- C.O. monitoring is intended for adult patients only; .
- The monitors are to be used in healthcare facilities by clinical professionals or under their
{6}------------------------------------------------
guidance. They should only be used by persons who have received adequate training in their use. The ePM 10/12/15/10M/12M/15M monitors are not intended for helicopter transport, hospital ambulance, or home use.
# 7. SUBSTANTIAL EQUIVALENCE
## Comparison of Indications
Both the predicate devices and the subject devices are multi parameter patient monitors intended to be used under the direction of clinical professionals. The monitoring parameters supported by the subject ePM series are identical with those supported by the predicate ePM series monitors (K191769).
The indications for use statement of the subject devices include Arrhythmia Detection used on neonatal patients. Although this feature is not present in the primary predicate devices, it is present in the cleared spacelabs Multi-parameter Module (K103142) and does not constitute a new intended use.
In conclusion, the minor differences of the indications for use do not change the fundamental intended use of the ePM Series as multiparameter monitor.
## Comparison of Technological Characteristics
The table below compares the key technological feature of the subject device to the primary predicate device (ePM series Patient Monitors, K191769). The features in gray are features which are different between the predicate devices and the subject devices.
| | Predicate Devices (K191769) | | | | | Subject ePM Devices | | | | | | |
|------------------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|
| Featu<br>re | ePM 15 | ePM<br>12 | eP<br>M<br>10 | eP<br>M<br>15<br>M | eP<br>M<br>12<br>M | eP<br>M<br>1<br>0<br>M | eP<br>M<br>15 | eP<br>M<br>12 | ePM<br>10 | eP<br>M<br>15<br>M | eP<br>M<br>12<br>M | e<br>P<br>M<br>1<br>0<br>M |
| Displa<br>y and<br>touchs<br>creen | 15.6”<br>1366*768<br>pixels | 12.1”<br>1280*8<br>00pixels | 10.1”<br>1280*8<br>00pixels | 15.6”<br>1366*768<br>pixels | 12.1”<br>1280*8<br>00pixels | 10.1”<br>1280*8<br>00pixels | 15.6”<br>1366*768<br>pixels | 12.1”<br>1280*8<br>00pixels | 10.1”<br>1280*8<br>00pixels | 15.6”<br>1366*768<br>pixels | 12.1”<br>1280*8<br>00pixels | 10.1”<br>1280*8<br>00pixels |
| Secon | Mirrored display | | | | | | Same | | | | | |
| Table 2: | Device Comparison Table | |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| dary<br>display | | |
| Wireless | 2.4GHz/5GHz dual band module | Same |
| Power<br>supply | Battery or AC power | Same |
| Battery | Rechargeable Lithium-Ion, 10.8VDC, 5600 mAh.<br>Rechargeable Lithium-Ion, 10.95 VDC, 4500 mAh.<br>Rechargeable Lithium-Ion, 10.95 VDC, 2600 mA | Same |
| Data<br>storage | Embedded Multi Media Card(eMMC) | Same |
| Data<br>Recorder | Supports internal thermal recorder | Same |
| Device<br>integration | Use the RS-232 interface to integrate 3rd party devices.<br>Allow parameter values, waveforms and alarms from 3rd party devices to be displayed, stored and printed. | Same |
| Speaker | Give alarm tones (45 to 85 dB), key tones, QRS tones; support PITCH TONE and multi-level tone modulation | Same |
| Alarm<br>system | The alarm lamp is cyan, yellow, or red depending on alarm type | Same |
| ECG | 3-lead, 5-lead, 6-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, an interpretation of resting 12-lead ECG, J-point Auto detection, Dual Channel Pace detection, adjustable QRS threshold and heart rate (HR). | 3-lead, 5-lead, 6-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, an interpretation of resting 12-lead ECG, J-point Auto detection, Dual Channel Pace detection, adjustable QRS threshold and heart rate (HR). |
| | Arrhythmia detection is intended for adult and | Arrhythmia detection is intended for adult, |
| | pediatric. | pediatric and neonate. |
| | Supports intelligent arrhythmia alarm. | Supports intelligent arrhythmia alarm. |
| | ST segment analysis is intended for adult, pediatric | ST segment analysis is intended for adult, |
| | and neonate. | pediatric and neonate. |
| | | |
| | | Refer to Section 12.3. |
| | Asystole, VFib/Vtac, Vtac, Vent.Brady, Extreme | |
| Arrhy | Tachy, Extreme Brady, PVCs, Couplet, Bigeminy, | |
| thmia | Trigeminy, R on T, Run PVCs, PVCs/min, Tachy, | |
| Analy | Brady, Missed Beats, Vent Rhythm, Pacer Not | Same |
| | Pacing, Pacer Not Capture, Multif.PVC, | |
| sis | Nonsus.Vtac, Pause, Vent.Rhythm, Afib, | |
| | Pauses/min, Pauses/min | |
| Respir | Measurement range: | |
| ation | Adult: 0 to 120 rpm; | |
| rate | Pediatric, neonate: 0 to 150 rpm. | Same |
| (Resp | Accuracy: | |
| ) | 7 to 150 rpm: ±2 rpm or ±2%, whichever is greater; | |
| | 0 to 6 rpm: Not specified. | |
| Temp | | |
| eratur | | |
| e | Measurement range: 0 to 50 °C (32 to 122 F) | Same |
| (Temp | Accuracy: ±0.1 °C or ±0.2 F (without probe). | |
| ) | | |
| | Supports Mindray SpO2 function, Masimo SpO2 | |
| | function and Nellcor SpO2 function from multi | |
| Pulse | parameter module. | |
| oxyge | Mindray SpO2 function | |
| n | Measurement range: 0 to 100%; | |
| | Accuracy: | Same |
| satura | 70 to 100%: ±2% (adult/pediatric mode); | |
| tion | 70 to 100%: ±3% (neonate mode); | |
| (SpO2 | 0% to 69%: Not specified. | |
| ) | Masimo SpO2 function | |
| | Measurement range: 1 to 100%; | |
| | Accuracy: | |
| | 70 to 100%: ±2% (measured without | |
| | motion in adult/pediatric mode). | |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| | 70 to 100%: ±3% (measured without motion in<br>neonate mode);<br>70 to 100%: ±3% (measured with motion);<br>1% to 69%: Not specified.<br><br>Nellcor SpO2 function<br>Measurement range: 0 to 100%;<br>Accuracy:<br>70 to 100%: ±2% (adult/pediatric);<br>70 to 100%: ±3% (neonate);<br>0% to 69%: Not specified.<br><br>Note: The specifictions of the various SpO2<br>functions provided by each manufacturer are the<br>same across platforms. | | | |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|------|
| | | PR from built-in<br>Mindray SpO2<br>Module<br>Measurement<br>range:<br>20 to 254 bpm;<br>Accuracy: ±3 bpm. | | |
| | PR from built-in<br>Mindray SpO2 Module<br>Measurement range: 20<br>to 254 bpm;<br>Accuracy: ±3 bpm. | PR from built-in<br>Masimo SpO2<br>Module<br>Measurement<br>range:<br>25 to 240 bpm;<br>Accuracy: ±3 bpm<br>(measured<br>without<br>motion);<br>±5 bpm<br>(measured with<br>motion). | Same | Same |
| Pulse<br>rate<br>(PR) | PR from built-in Nellcor<br>SpO2 Module<br>Measurement range: 20<br>to 300 bpm;<br>Accuracy:<br>20 to 250 bpm: ±3 bpm;<br>251 to 300 bpm, not<br>specified. | PR from built-in<br>Nellcor SpO2<br>Module<br>Measurement<br>range:<br>20 to 300 bpm;<br>Accuracy:<br>20 to 250 bpm:<br>±3 bpm;<br>251 to 300 bpm,<br>not specified.<br><br>PR from<br>external IBP<br>Module<br>Measurement | | |
{10}------------------------------------------------
| | range:<br>25 to 350 bpm;<br>Accuracy:<br>$\pm$ 1 bpm or $\pm$ 1%,<br>whichever is<br>greater. | | | | | |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|------|
| Non-<br>invasi<br>ve<br>blood<br>pressu<br>re<br>(NIBP<br>) | Measurement range: | Adult | Pediatric | Neonate | Same | |
| | Systolic | 25-290 | 25-240 | 25-140 | | |
| | Diastolic | 10-250 | 10-200 | 10-115 | | |
| | Mean | 15-260 | 15-215 | 15-125 | | |
| | Accuracy: | | | | | |
| | Max mean error: $\pm$ 5 mmHg; | | | | | |
| | Max standard deviation: 8 mmHg. | | | | | |
| Invasi<br>ve<br>blood<br>pressu<br>re<br>(IBP) | Uses an internal IBP module<br>to measure invasive blood<br>pressure. The monitor can<br>monitor up to 8 invasive<br>blood pressures and displays<br>systolic, diastolic and mean<br>pressures and a waveform<br>for each pressure. Support<br>PPV function. Measurement<br>range: -50 to 300 mmHg;<br>Accuracy: $\pm$ 2% or $\pm$ 1<br>mmHg, whichever is greater<br>(without sensor). | | | Uses an internal IBP module or stand-alone IBP<br>Module to measure invasive blood<br>pressure. The<br>monitor can<br>monitor up to 8<br>invasive blood<br>pressures and<br>displays systolic,<br>diastolic and mean<br>pressures and a<br>waveform for each<br>pressure. Support<br>PPV function.<br>Measurement<br>range: -50 to 300<br>mmHg;<br>Accuracy: $\pm$ 2% or<br>$\pm$ 1<br>mmHg,<br>whichever is greater<br>(without sensor). | | Same |
| Cardi<br>ac<br>output | Use internal C.O. module.<br>The cardiac output (C.O.)<br>measurement invasively | | Use internal or<br>external C.O.<br>module. | Same | Same | |
{11}------------------------------------------------
| (C.O.) | measures | The cardiac output<br>(C.O.)<br>measurement | | |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|------|
| | cardiac output and other<br>hemodynamic parameters<br>using the right heart (atria)<br>thermodilution method. The<br>temperature change is<br>displayed as a curve in the<br>C.O. split screen, and the<br>monitor calculates the<br>C.O. value from this curve.<br>The monitor is capable of<br>storing 6 measurements.<br>Measurement range: C.O:<br>0.1 to 20 L/min; TB: 23 to<br>43 °C;<br>TI: 0 to 27 °C;<br>Accuracy:<br>C.O: ±5% or ±0.1 L /min,<br>whichever is greater;<br>TB, TI: ±0.1 °C (without<br>sensor). | invasively<br>measures<br>cardiac output and<br>other hemodynamic<br>parameters using<br>the right heart<br>(atria)<br>thermodilution<br>method. The<br>temperature change<br>is displayed as a<br>curve in the C.O.<br>split screen, and the<br>monitor calculates<br>the<br>C.O. value from<br>this curve. The<br>monitor is capable<br>of storing 6<br>measurements.<br>Measurement<br>range: C.O: 0.1 to<br>20 L/min; TB: 23 to<br>43 °C;<br>TI: 0 to 27 °C;<br>Accuracy:<br>C.O: ±5% or ±0.1 L<br>/min, whichever is<br>greater;<br>TB, TI: ±0.1 °C<br>(without sensor). | | |
| Carbon<br>dioxide(CO2) | Compatible with 3 internal<br>CO2 modules:<br>Sidestream CO2 2.0 module<br>Mainstream CO2<br>Module<br>MicroStream CO2<br>module | Compatible with 4<br>internal or external<br>CO2 modules:<br>Sidestream CO2<br>1.0 Module<br>Sidestream CO2<br>2.0 module<br>Mainstream CO2 | Same | Same |
ePM Series Patient Monitors
{12}------------------------------------------------
| | module<br>MicroStream CO2<br>module<br>*10M does not<br>support external<br>CO2 modules | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|------|
| Type: Sidestream CO2 Module (CO2 2.0):<br>Measurement range: 0~150mmHg<br>Accuracy:<br>0~40 mmHg: ±2mmHg,<br>41~76 mmHg: ±5% of reading,<br>77~99 mmHg: ±10% of reading,<br>100~150mmHg: ±(3mmHg + 8% of reading),<br>ISO accuracy mode:<br>Add ±2mmHg to the full accuracy mode<br>AwRR measurement:<br>awRR measurement range: 0 to 150rpm;<br>awRR:<br><60rpm, ±1rpm,<br>60~150rpm, ±2rpm. | | Same |
| Type: Microstream CO2 Module<br>Measurement range:<br>CO2: 0~99mmHg;<br>awRR: 0~150rpm;<br>Accuracy:<br>CO2:<br>0~38mmHg: ±2mmHg;<br>39~99mmHg: ±5% of the reading+0.08% of (the reading-38).<br>awRR:<br>0~70rpm: ±1rpm,<br>71~120rpm: ±2rpm,121~150rpm: ±3rpm | | Same |
| Type: Mainstream CO2<br>Module Measurement range:<br>CO2: 0~150mmHg;<br>awRR: 0~150rpm;<br>Accuracy:<br>CO2:<br>0~40mmHg: ±2mmHg,<br>41~70mmHg: ±5% of the reading,<br>71~100mmHg: ±8% of the reading | | Same |
{13}------------------------------------------------
| 101~150mmHg: ±10% of the reading; | | | | |
|-----------------------------------|-------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|------|
| awRR: ±1rpm. | | | | |
| Oxyg<br>en<br>(02) | Not support | Oxygen values are<br>measured by the<br>Sidestream CO2 2.0<br>or the AG module<br>using<br>a<br>paramagnetic<br>method.<br>Measurement<br>range: 0~100%<br>(CO2 and AG),<br>Accuracy:<br>0~25%, ±1%;<br>26~80%,±2%;<br>81~100%, ±3%<br>(CO2 and AG). | Same | Same |
| Anest<br>hetic<br>gas<br>(AG) | Not support | The AG module<br>analyzes gas<br>samples from the<br>patient and<br>calculates CO2, O2,<br>N2O, and AA<br>waveforms and<br>related numerics<br>that include airway<br>respiratory rate and<br>MAC (minimum<br>alveolar<br>concentration).<br>Measurement<br>range:<br>CO2, HAL, ENF,<br>ISO, SEV, DES:<br>0~30 %;<br>O2, N2O: 0~100 %;<br>awRR: 2~100 rpm.<br>Accuracy: | Same | Same |
{14}------------------------------------------------
| | mode: | |
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | CO2:0%≤CO2≤1% | |
| | , ±0.1%ABS, 1%< | |
| | CO2<5%, = =0.2% | |
| | ABS, ABS, ABS, ABS,<br>5%< | |
| | CO2≤7%, | |
| | ±0.3% ABS, 7%< | |
| | CO2<10%, ±0.5% | |
| | ABS, 10%< CO2 | |
| | not specified; | |
| | N2O: 0〜20%REL: | |
| | ±2% ABS, | |
| | 20~100%REL: | |
| | ±3% ABS; | |
| | 02 0~25%, ±1%; | |
| | 26~80%, ±2%; | |
| | 81~100%, ±3%; | |
| | HAL, ENF, ISO: | |
| | 0~1%REL: ±0.15% | |
| | ABS, 1~5%REL: | |
| | ±0.2% | |
| | ABS, >5%REL, not | |
| | specified; | |
| | SEV: 0~1%REL: | |
| | =0.15%<br>ABS, | |
| | 1~5%REL: | |
| | #0.2% = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =<br>ABS, | |
| | 5~8%REL: ±0.4% | |
| | ABS, >8%REL, not | |
| | specified; | |
| | DES: 0~1%REL: | |
| | 40.15%<br>ABS, | |
| | 1~5%REL: =0.2% | |
| | ABS, 5~10%REL: | |
| | ABS,<br>±0.4% ======================================================================================================================================================================== | |
| | 10~15%REL: | |
| | =0.6% | |
| | ABS ,15~18%REL: | |
| | ±1% | |
| | ABS ,>18%REL, | |
| | not specified; | |
| | awRR:2~60rpm, | |
{15}------------------------------------------------
| | ±1rpm, >60rpm, not specified. | |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Wirel<br>ess<br>functi<br>on | The Wireless radio module (Laird) is used for<br>connecting to a wireless monitoring network with a<br>central monitoring system (CMS). | Same |
| ECG<br>24h<br>Summ<br>ary | Provides the function to statistical results of heart<br>rate changes and cardiac arrhythmia of patients<br>within 24 hours, including HR statistics, ARR<br>statistics, ST statistics, QT/QTc statistics and Pace<br>statistics. | Same |
| EWS | The EWS is a set of early warning scores that are<br>intended to assist clinicians in recognizing the early<br>signs of deterioration in patients based on vital<br>signs and clinical observations. The three types of<br>EWS provided are Modified Early Warning Score<br>(MEWS), National Early Warning Score (NEWS<br>and NEWS2) and user configurable Custom Score. | Same |
| Glasg<br>OW<br>Coma<br>Scale<br>(GCS) | The GCS a well-established scoring system used to<br>assess the state of consciousness based three sub-<br>components: eye-opening response, verbal<br>response, and limb movement. | Same |
## Substantial Equivalence Conclusion
In conclusion, the differences in technological charact…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.