K103142 · Spacelabs Healthcare, Inc. · DSI · Mar 3, 2011 · Cardiovascular
Device Facts
Record ID
K103142
Device Name
SPACELABS MULTI-PARAMETER MODULE
Applicant
Spacelabs Healthcare, Inc.
Product Code
DSI · Cardiovascular
Decision Date
Mar 3, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use. Physiological parameters that may be monitored include ECG with arrhythmia detection, respiration, invasive and noninvasive blood pressure, temperature, oxygen saturation (SpO2) and cardiac output. Acquired data may then be communicated to an information network for display, recording, editing and analysis.
Device Story
Spacelabs Multi-parameter Module is a plug-in module for Patient Care Management System (PCMS) monitors (e.g., Model 91387). It collects physiological data via parameter-specific cables/sensors; processes ECG (including arrhythmia detection), respiration, blood pressure, temperature, SpO2, and cardiac output. Data is transmitted to the monitor for waveform/numeric display, alarm management, and printing. Used in clinical settings by healthcare providers; supports network communication for data recording/analysis. Provides real-time monitoring and alarm alerts to assist clinical decision-making and patient management.
Clinical Evidence
No clinical data. Bench testing performed in accordance with ANSI/AAMI EC57:1998/(R)2003 demonstrated performance equal to or better than the predicate device. Software verification and validation conducted per FDA guidance.
Technological Characteristics
Plug-in module for PCMS monitors. Parameters: ECG (arrhythmia detection), respiration, invasive/noninvasive BP, temperature, SpO2, cardiac output. Connectivity: Networked via PCMS. Software-based processing. No hardware modifications. Not sterilized by user.
Indications for Use
Indicated for adult and neonate/infant populations requiring monitoring of ECG (with arrhythmia detection), respiration, invasive/noninvasive blood pressure, temperature, SpO2, and cardiac output in clinical environments. Not for home use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Spacelabs Medical Multi-parameter Module 91496 (K050605)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
K103142
MAR - 3 2011
| Submission Date: | 17 December 2010 | | |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|---------------------------------------------------------------------------------|
| Submitter<br>Correspondent: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029<br>Mr. David J. Geraghty<br>Spacelabs Healthcare<br>Phone: +1 (425) 657-7200, ext 5889<br>Fax: +1 (425) 657-7210<br>Email: david.geraghty@spacelabs.com | | |
| Application<br>Correspondent: | Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466<br>Mr. Thomas Kroenke<br>Phone: +1 (303) 956-4232<br>Fax: +1 (303) 642-3141<br>Email: tkroenke@speedtomarket.net | | |
| Trade Name: | Spacelabs Multi-parameter Module | | |
| Common Name: | Arrhythmia Detector and Alarm | | |
| Classification Name: | Detector and Alarm, Arrhythmia | | |
| Classification<br>Regulation: | 21 CFR §870.1025, 21 CFR §870.2710, 21 CFR §870.2700,<br>21 CFR §870.1110, 21 CFR §870.1435, 21 CFR §870.1130,<br>21 CFR §880.2910, 21 CFR | | |
| Product Code: | DSI, DPZ, DQA,<br>DSK, DXG, DXN,<br>FLL, LOS | | |
| Substantially<br>Equivalent Devices: | New Spacelabs Model | Predicate 510(k)<br>Number | Predicate<br>Manufacturer / Model |
| | Spacelabs Multi-parameter<br>Module | K050605 | Spacelabs Medical Inc. /<br>Spacelabs Medical Multi-<br>parameter, Module 91496 |
{1}------------------------------------------------
The Spacelabs Multi-parameter Module provides monitoring capability for the following parameters: ECG with arrhythmia detection, respiration, invasive and noninvasive blood pressure, temperature, oxygen saturation (SpO2) and cardiac output. It is a "plug-in" module that is used in conjunction with a Spacelabs Monitoring Station such as the Model 91387.
The patient is connected to the Spacelabs Multi-parameter Module via parameter-specific cables/sensors which provide the monitor with the patient's physiological data. The physiological data for each active parameter is accumulated by the Spacelabs Multi-parameter Module, and sent to the monitor which performs any necessary analysis and provides both waveform data and numeric values for display on the monitor screen. The monitor provides the display and printing capabilities for the care provider.
Setting of alarm limits, enabling or disabling alarm monitoring, and definition of alarm responses are determined on a parameter-byparameter basis, and can be defined by the user for each parameter. In general, alarm limits can be set to provide the user with both audible and visible indications of the alarm condition. Each parameter provides a hierarchy of menu structures and keys that are activated through controls of the monitor by touching the monitor screen.
Each parameter provides the capability to output recordings of selected information to a variety of recording devices. Recordings are available using the Spacelabs Monitor internal printer, network printers, and the Intesys Clinical Suite Print Manager Model 91881. Recordings can be automatically generated by each of the parameter's alarm managers or manually generated.
The Spacelabs Multi-parameter Module utilizes the same technology for each parameter as utilized by the predicate device.
The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use.
Physiological parameters that may be monitored include ECG with arrhythmia detection, respiration, invasive and noninvasive blood pressure, temperature, oxygen saturation (SpO2) and cardiac output. Acquired data may then be communicated to an information network for display, recording, editing and analysis.
> K103142 Page 2 of 4
Technology Comparison:
Intended Use:
{2}------------------------------------------------
## Performance Testing:
| Sterilization and<br>Shelf-Life | The Spacelabs Multi-parameter Module is not sterilized or sterilizable<br>by the user.<br>Accessories which may be sterilized or sterilizable by the user were not<br>affected by the software device modifications made to the Spacelabs<br>Multi-parameter Module. Therefore, this section is not applicable. |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | The Spacelabs Multi-parameter Module is not intended to directly or<br>indirectly contact the patient.<br>Accessories which may directly or indirectly contact the patient were<br>not affected by the software device modifications made to the<br>Spacelabs Multi-parameter Module. Therefore, this section is not<br>applicable. |
| Software Testing | The software device modifications made to the Spacelabs Multi-<br>parameter Module were designed and developed according to a robust<br>software development process, and were rigorously verified and<br>validated.<br>Software information is provided in accordance with:<br>FDA guidance: The content of premarket submissions for software<br>contained in medical devices, 11 May 05; and FDA guidance: General principles of software validation; Final<br>guidance for industry and FDA staff, 11 Jan 02. Test results indicate that the Spacelabs Multi-parameter Module<br>complies with its predetermined specification. |
| Electrical Safety and<br>Electromagnetic<br>Compatibility Testing | The software device modifications made to the Spacelabs Multi-<br>parameter Module did not affect the electrical safety or electromagnetic<br>compatibility performance of the device. No hardware modifications<br>were made to the device; therefore this section is not applicable. |
| Performance Testing<br>- Bench | In addition to the software verification and validation performed, the<br>software device modifications made to the Spacelabs Multi-parameter<br>Module were testing in accordance with ANSI/AAMI EC57: 1998 / (R)<br>2003. Test results demonstrate performance equal to or better than the<br>predicate device. |
| Performance Testing<br>– Animal: | Animal performance testing is not necessary to demonstrate safety and<br>effectiveness of the Spacelabs Multi-parameter Module. Therefore, this<br>section is not applicable. |
| Performance Testing<br>- Clinical: | Clinical performance testing is not necessary to demonstrate safety and<br>effectiveness of the Spacelabs Multi-parameter Module. Therefore, this<br>section is not applicable. |
K103142 Page 3 of 4
. . . . .
{3}------------------------------------------------
## Conclusion
Verification and validation activities were conducted to establish the performance and safety characteristics of the software device modifications made to the Spacelabs Multi-parameter Module. The results of these activities demonstrate that the Spacelabs Multiparameter Module is safe and effective when used in accordance with its intended use and labeling.
Therefore, the Spacelabs Multi-parameter Module is considered substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the circumference. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Spacelabs Healthcare Inc. c/o Mr. Thomas Kroenke Speed To Market, Inc PO Box 3018 Nederland, CO 80466
MAR - 3 2011
Re: K103142
Trade/Device Name: Spacelabs Multi-parameter Module Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II (two) Product Code: DSI, DQA, DSK, DXG, DXG, DXN, FLL Dated: February 7, 2011 Received: February 10, 2011
Dear Mr. Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Mr. Thomas Kroenke
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
27 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name:
Indications for Use:
Spacelabs Multi-parameter Module
K 103142
The Spacelabs Multi-parameter Module is intended for use with
the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use.
Physiological parameters that may be monitored include ECG with arrhythmia detection, respiration, invasive and noninvasive blood pressure, temperature, oxygen saturation (SpO2) and cardiac output. Acquired data may then be communicated to an information network for display, recording, editing and analysis.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
510(k) Number V
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) 3/3/204
Division of Cardiovascular Devices
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.