THE PHILIPS MP20, MP30, MP40, MP50, MP60, MP70 AND MP90 INTELLIVUE PATIENT MONITORS
Applicant
Philips Medizin Systeme Boeblingen GmbH
Product Code
MHX · Cardiovascular
Decision Date
Feb 11, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are intended for use in transport situations within additionally hospital environments.
Device Story
IntelliVue patient monitors (MP20-MP90) collect physiological data via various sensors/probes (ECG, blood pressure, SpO2, gas, etc.). System processes signals to provide real-time monitoring, recording, and alarming. Used by healthcare professionals in hospital environments and during intra-hospital transport. Release B.1.2 introduces software updates and an ITS wireless interface. Output displayed on integrated screens for clinical decision-making; enables continuous patient surveillance and timely intervention. Benefits include improved patient safety through reliable physiological tracking and alarm management.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level tests, performance tests, and safety testing based on hazard analysis. Pass/Fail criteria were established against predicate device specifications; results demonstrated substantial equivalence.
Technological Characteristics
Multi-parameter patient monitoring system. Includes wireless interface (ITS). Supports various physiological inputs (ECG, blood pressure, gas, etc.). Software-driven processing. Class II device. Connectivity via wireless interface. Sterilization not applicable (non-invasive).
Indications for Use
Indicated for monitoring physiological parameters in adult, pediatric, and neonatal patients in hospital settings, including transport for MP20-MP50 models. ST segment monitoring restricted to adults. Transcutaneous gas measurement (tcpO2/tcpCO2) restricted to neonates.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
## 510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Sate Medical Devices Act of 1990 and 21 C.F.R. §807.92.
1. The submitter of this premarket notification is: Hauke Schik Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-2031 Fax: ++49 7031 463-2442 e-mail: hauke.schik@philips.com
This summary was prepared on January 21, 2005.
- 2. The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitors. Classification names are as follows:
| Device Panel | Classification | ProCode | Description |
|------------------------------------------|----------------|---------|-----------------------------------------------------------------------------|
| Circulatory<br>System Devices<br>(12625) | \$870.1025, II | DSI | Detector and alarm, arrhythmia |
| | \$870.1025, II | MLD | Monitor, ST Segment with Alarm |
| | \$870.1025, II | MHX | Monitor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms) |
| | \$870.1100, II | DSJ | Alarm, Blood Pressure |
| | \$870.1110, II | DSK | Computer, Blood Pressure |
| | \$870.1130, II | DXN | System, Measurement, Blood-<br>Pressure, Non-Invasive |
| | \$870.1435, II | DXG | Computer, Diagnostic, Pre-<br>Programmed, Single-Function |
| | \$870.1915, II | KRB | Probe, Thermodilution |
| | \$870.2060, II | DRQ | Amplifier and Signal<br>Conditioner, Transducer Signal |
| | \$870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm) |
| | \$870.2300, II | MSX | System, Network and<br>Communication, Physiological<br>Monitors |
| | \$870.2340, II | DPS | Electrocardiograph |
| | \$870.2340, II | MLC | Monitor, ST Segment |
| | \$870.2350, II | DRW | Electrocardiograph, Lead<br>Switching Adapter |
| | \$870.2370, II | KRC | Tester, Electrode, Surface,<br>Electrocardiograph |
| | \$870.2450, II | DXJ | Display, Cathode-Ray Tube,<br>Medical |
| | \$870.2600, I | DRJ | System, Signal Isolation |
| | \$870.2700, II | DQA | Oximeter |
| | \$870.2770, II | DSB | Plethysmograph, Impedance |
| | \$870.2800, II | DSH | Recorder, Magnetic tape,<br>Medical |
| | \$870.2810, I | DSF | Recorder, Paper Chart |
| | \$870.2850, II | DRS | Extravascular Blood Pressure<br>Transducer |
{1}------------------------------------------------
| | \$870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector |
|------------------------------------------------------|----------------|-----|------------------------------------------------------------------------------|
| | \$870.2910, II | DRG | Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency |
| Anesthesiology<br>and Respiratory<br>Therapy (12624) | \$868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase |
| | \$868.1500, II | CBQ | Analyzer, Gas, Enflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1500, II | NHO | Analyzer, Gas, Desflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1500, II | NHP | Analyzer, Gas, Sevoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1500, II | NHQ | Analyzer, Gas, Isoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1620, II | CBS | Analyzer, Gas, Halothane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | \$868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-<br>Phase |
| | \$868.2375, II | BZQ | Monitor, Breathing Frequency |
| | \$868.2480, II | LKD | Monitor, Carbon Dioxide,<br>Cutaneous |
| | \$868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia |
| General Hospital<br>and Personal Use<br>(12520) | \$880.2910, II | FLL | Thermometer, Electronic,<br>Clinical |
| Neurological<br>(12513) | \$882.1400, II | GWR | Electroencephalograph |
| | \$882.1420, I | GWS | Analyzer, Spectrum |
- 3. The modified devices are substantially equivalent to previously cleared Philips devices marketed pursuant to K014159, K021778, K030038, K032858, K040304, K040183 and K040357, K041235, K41741 and K042845.
- 4. The modification is the introduction of Release B.1.2 software for the IntelliVue patient monitor devices and incorporating a ITS wireless interface.
- 5. The modified devices have the same intended use as the legally marketed predicate devices. They are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in hospital environment and during transport within hospital environments.
- 6. The modified devices have the same technological characteristics as the legally marketed predicate devices.
Electroencephalogram Signal
{2}------------------------------------------------
K050141
P³/₃
- 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is an abstract image of an eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 1 2005
Phillips Medizin Systeme Boeblingen GmbH c/o Mr. Hauke Schik Sr. Regulatory Affairs Engineer Hewlett-Packard Str. 2 D-71034 Boeblingen GERMANY
Re: K050141
Trade Name: The Phillips MP20, MP30, MP40, MP50, MP60, MP70 and MP90 Intellivue Patient Monitors, Release B.1.2. Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-segment measurement and alarm) Regulatory Class: II (two) Product Code: MHX Dated: January 21, 2005 Received: January 24 2005
Dear Mr. Hauke Schik:
We have reviewed your Section 510(k) premarket notification of intent to market the device w & no re review a your we determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, de nees mat have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Hauke Schik
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Brimmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):_ Ko5o14 l
Device Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70 and MP90 IntelliVue Patient Monitors, Release B.1.2.
Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The MP20, MP30, MP40 and MP50 are intended for use in transport situations within additionally hospital environments.
ST Segment monitoring is restricted to adult patients only.
The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.
Prescription Use yes___________________________________________________________________________________________________________________________________________________________________________ Over-The-Counter Use _ No (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blinums
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.